Introduction
Dosage accuracy is essential for medical devices based on microemulsions, particularly when the active ingredient is present in micellar form. The pharmaceutical literature indicates that drop formation is influenced by multiple physical and rheological factors (Lachman et al., 1986; Washington, 1996).
Furthermore, European regulations require medical devices to ensure stable and reproducible delivery throughout their entire period of use (European Commission, 2017).
To verify the hypothesis that the pipette dropper was intrinsically less precise than a gravity dropper, a two-phase experimental study was conducted.
Subjects, Materials and Methods
Twenty subjects (10 males and 10 females), young adults with a mean age of 34 ± 6 years, were enrolled and divided into two groups of 10 participants each.
Each subject was instructed to:
Dispense 10 drops onto a laboratory precision scale
Record 10 individual weight measurements
Repeat the procedure using the assigned dropper (pipette or gravity)
In the second phase, 10 volunteers used a bottle equipped with a gravity dropper for 30 days, taking five drops twice daily and recording drop weight at three time points (beginning, middle, and end of the month).
The methodology follows EMA standards for the evaluation of oral solutions (European Medicines Agency, 2020).
Results
Phase 1
The gravity dropper showed significantly lower variability (SD ±2 mg) compared to the pipette dropper (SD ±5 mg), with statistical significance (p < 0.05).
These findings are consistent with the literature on microemulsions and surfactant-based delivery systems (Torchilin, 2006).
Phase 2
Subjects who previously showed high variability with the pipette dropper displayed values similar to those of the gravity dropper group. The drop weight remained stable across all 30 days, indicating that:
bottle opening
daily use
progressive emptying
did not affect the amount delivered.
Discussion
The pipette dropper proved less precise and less reproducible than the gravity dropper.
The greater variability observed with the pipette dropper is consistent with Washington’s (1996) findings on the sensitivity of drop formation to operational conditions.
Conclusions
The gravity dropper ensures more accurate and reproducible dosing, in accordance with European regulatory requirements (European Commission, 2017).
For this reason, the gravity dropper has been adopted as the standard dispensing system for NegEnt Micellar Drops, CE Medical Device.
Author Contributions
Tullio Scrimali: study design, scientific supervision, data interpretation, manuscript drafting. Advanced artificial intelligence tools: support in text structuring, linguistic optimisation, formal revision, and translation.
Funding
Herbal Neurocare funded this research.
Acknowledgments
The author also wishes to emphasise, for the purpose of developing new scientific methodologies, that advanced artificial intelligence tools were employed in conducting the research presented, in drafting the present article, and in translating it from Italian into English. Scientific, methodological, and interpretative responsibility for all content remains entirely with the human author.
Conflicts of Interest
The author declares no conflicts of interest other than the institutional relationship with the funding company.
References
- European Commission. (2017). Medical Device Regulation (EU) 2017/745.
- European Medicines Agency. (2020). Guideline on quality of oral solutions and suspensions.
- Lachman, L., Lieberman, H. A., & Kanig, J. L. (1986). The theory and practice of industrial pharmacy (3rd ed.). Lea & Febiger.
- Torchilin, V. P. (2006). Structure and design of polymeric surfactant-based drug delivery systems. Journal of Controlled Release, 111(1–2), 1–33.
|
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).