Preprint
Article

This version is not peer-reviewed.

Numerical Investigation of Welding Parameter Effects on Dynamic Stability of Medical Equipment Frames

Submitted:

11 January 2026

Posted:

12 January 2026

You are already at the latest version

Abstract
This study numerically investigates how welding parameters influence the dynamic stability of medical equipment frames. A set of 30 frame models with varying weld bead sizes and heat inputs was analyzed using finite element modal and harmonic response analysis. The first natural frequency varied between 62 and 91 Hz depending on welding conditions, leading to up to 2.1-fold differences in resonance amplification factors. The findings indicate that welding design plays a critical role not only in static strength but also in dynamic performance and service life of medical equipment.
Keywords: 
;  ;  ;  ;  
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.