Submitted:
31 May 2024
Posted:
03 June 2024
You are already at the latest version
Abstract
Keywords:
1. Introduction
2. Experimental Test Method
2.1. Test Method
2.2. Geometry
2.3. Setup and Test Equipment Recommendations
2.4. Test Conditions and Preparation
2.5. Test Protocol
- Prepare all equipment to be used.
- Connect all test elements together and secure the test setup, making sure that all fittings and interconnections are correctly and tightly connected.
- If necessary, fill the reservoir with the test fluid medium at least 30 min before beginning the tests.
- Connect all required electrical power supplies to the pressure controllers and sensors, and turn on all instrumentation at least 30 min before beginning the tests.
- Prime the test setup using the test media at the maximum admissible pressure, limited by the component (pressure controller, pressure sensors, microfluidic device) which has the lowest maximum labeled pressure.
- Zero the pressure and flow sensors to minimize measurement offsets.
- Set a target pressure for the pressure controller and observe the rise and stabilization of the pressure reading using the pressure sensors.
- After a stable pressure is reached (not drifting upwards of downwards within the stability stated in the pressure controller datasheet), record the measurement of pressure, differential pressure, or flow rate, if possible digitally, to allow higher sampling rate, reducing the effect of collection time impact on a time-averaged pressure value.
- Repeat step 7 and 8 for all the required target pressure, differential pressure or flow rate, encompassing all the operating conditions at which the microfluidic system will be used.
- Additional measurements should be equally carried out at each operating condition without the microfluidic device in the test system. These control tests will serve to measure the pressure differential associated with the location of the pressure taps and the overall setup supporting the experiment.
- Calculate the hydrodynamic resistance associated with each test condition according to eq. 2. The pressure drop associated solely with the experimental setup without the microfluidic device (control) should be subtracted from the test conditions with the microfluidic device (total) to determine the pressure drop within the microfluidic device itself (eq. 3).
2.6. Implementation of the Test Protocol
3. Theoretical Calculation of Hydrodynamic Resistance
4. CFD Simulations of Hydrodynamic Resistance
5. Results and Discussion
6. Conclusions
Author Contributions
Funding
Data Availability Statement
Conflicts of Interest
References
- Copeland, M., Ogheard, F., Batista, E., & Silverio, V. MFMET A2.2.2: Development of test protocols for microfluidic devices. 2023. [CrossRef]
- Copeland, M., Ogheard, F., Batista, E., & Heeren, H. van. Whitepaper flow resistivity testing. 2023. [CrossRef]
- Bruus, H. 2008, Theoretical Microfluidics. Publisher: Oxford University Press.
- van Heeren H. (2015) Results survey on microfluidics flow control https://mfmet.eu/wp-content/uploads/2022/02/2015-Microfluidics-Flow-Control-results-from-survey-enablingMNT.pdf [Accessed 9 November 2023].
- https://MFMET.EU.
- https://www.microfluidic-chipshop.com/catalogue/microfluidic-chips/polymer-chips/straight-channel-chips-microscopy-slide-format/straight-channel-chips-with-eight-parallel-channels-fluidic-157/ [Accessed 22/12/2023].
- BIPM JCGM WG1, GUM - JCGM 100:2008(E) – Evaluation of measurement data — Guide to the expression of uncertainty in measurement.
- White F. M. (1991) Viscous fluid flow, 2nd edition, McGraw-Hill, USA, pp 120, equation 3-48, ISBN 10:0070697132.
- Batista E., Almeida N., Godinho I., Filipe E. (2015) Uncertainty calculation in gravimetric microflow measurements. In Advanced Mathematical and Computational Tools in Metrology and Testing X. 98-104. [CrossRef]
- Silverio, V., Cardoso, S. (2021). Lab-on-a-chip: Systems integration at the microscale. In Drug Delivery Devices and Therapeutic Systems (pp. 63–87). Elsevier. [CrossRef]




| Measurements | Calculation (eq. 2) | |||
|---|---|---|---|---|
| 1 | 1 | 2 | 2 | |
| [kPa] | [µL/min] | [kPa] | [µL/min] | [1012 Pa∙s/m3] |
| 3.90 | 18.0 | 3.20 | 14.0 | 10.5 |
| 3.20 | 14.0 | 2.50 | 9.0 | 8.40 |
| 2.50 | 9.0 | 2.05 | 6.0 | 9.00 |
| 2.05 | 6.0 | 1.50 | 2.5 | 9.43 |
| Average | 9.33 | |||
| U (k = 2) | 1.79 | |||
| Measurements | Calculation (eq. 2) | |||
|---|---|---|---|---|
| 1 | 1 | 2 | 2 | |
| [kPa] | [µL/min] | [kPa] | [µL/min] | [1012 Pa∙s/m3] |
| 3.90 | 57.0 | 3.20 | 48.0 | 4.67 |
| 3.20 | 48.0 | 2.50 | 37.5 | 4.00 |
| 2.50 | 37.5 | 2.05 | 28.0 | 2.84 |
| 2.05 | 28.0 | 1.50 | 18.0 | 3.30 |
| Average | 3.70 | |||
| U (k = 2) | 1.61 | |||
| Setup | Method | RH | Uncertainty |
|---|---|---|---|
| [1012 Pa∙s/m3] | [1012 Pa∙s/m3] | ||
| Circuit Configuration B with chip | experimental measurement | 9.33 | 1.79 |
| Circuit Configuration B without chip – control setup | experimental measurement | 3.70 | 1.61 |
| Chip (Figure 3i, a, b, c, d and e) | calculated from measurements | 5.63 | 2.41 |
| Straight square cross section channel (Figure 3i, c) | eq. 5 | 4.98 | |
| Straight square cross section channel (Figure 3i, c) | eq. 6 | 4.97 | |
| Straight square cross section channel (Figure 3i, c) | COMSOL Multiphysics, $OpenFoam | 5.00 | |
| Chip (Figure 3i, a, b, c, d, e) | COMSOL Multiphysics | 5.27 | |
| Chip (Figure 3i, a, b, c, d, e) | OpenFoam | 5.18 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).