Redesign of Electrical Stapler Design: An Approach to Reduce Muscle Fatigue of Surgeons

Version 1 : Received: 12 December 2023 / Approved: 13 December 2023 / Online: 13 December 2023 (10:15:40 CET)

How to cite: Qingmin, F.; Yufeng, X.; Mengya, H.; Qiang, Z.; Chuanqing, W.; Shenglin, L.; Zhengke, G.; Yan, F. Redesign of Electrical Stapler Design: An Approach to Reduce Muscle Fatigue of Surgeons. Preprints 2023, 2023120936. https://doi.org/10.20944/preprints202312.0936.v1 Qingmin, F.; Yufeng, X.; Mengya, H.; Qiang, Z.; Chuanqing, W.; Shenglin, L.; Zhengke, G.; Yan, F. Redesign of Electrical Stapler Design: An Approach to Reduce Muscle Fatigue of Surgeons. Preprints 2023, 2023120936. https://doi.org/10.20944/preprints202312.0936.v1

Abstract

Muscle fatigue caused by using staplers affects the surgical procedures of surgeons. Compared to manual staplers, electric staplers decrease the muscle fatigue by replacing repeated gripping and releasing with automatic model. Still powered by the battery, the electric staplers are confronted with new operation fatigue due to the extra weight of the battery. The weight and manual operation of the electric stapler can cause user's muscle fatigue. To tackle the unsettled fatigue issue, the study is to propose a new design method to offset the influence of extra weight of the battery attached to the electrical stapler by changing the opening and closing of the stapler end clamp to electric control. The purpose of this study is to verify whether the design of this electric stapler can offset the impact of weight. Experiments are set up to test the feasibility of the design on 12 male Chinese surgeons. The muscle fatigue is indicated by measuring the EMG (electromyography) signal, and the stapler's working effect is indicated by measuring the end movement of the stapler. The result shows the new-designed stapler reduce the operation fatigue, counteracting the impact of the battery weight.

Keywords

laparoscopic surgery; endoscopic stapler; jitter; surface EMG signal; muscle fatigue

Subject

Engineering, Bioengineering

Comments (0)

We encourage comments and feedback from a broad range of readers. See criteria for comments and our Diversity statement.

Leave a public comment
Send a private comment to the author(s)
* All users must log in before leaving a comment
Views 0
Downloads 0
Comments 0
Metrics 0


×
Alerts
Notify me about updates to this article or when a peer-reviewed version is published.
We use cookies on our website to ensure you get the best experience.
Read more about our cookies here.