Submitted:
27 February 2026
Posted:
04 March 2026
You are already at the latest version
Abstract
Keywords:
1. Introduction
2. Implementing Relative Primary Radiometry in Practice
2.1. HTFP cells
2.2. High temperature furnace
2.3. The radiometer
2.4. Characterizing of the LP5
2.4.1. Size-of-source (SSE)
2.4.2. The Radiometer Linearity
2.5. Determination of Thermodynamic Temperature by the RPRT Method
2.6. Software Workflow Diagram
3. Experiments
3.1. Plateau realization
3.2. The radiance temperature distribution across the furnace
4. Software Implementation
4.1. Plateau Analysis and Data Preparation
4.2. Point of Inflection (POI)

| HTFPs | POI, K | |||
| CCT WG | Cubic interpolation | Average | (Max-min)/2 | |
| Fe-C (T) | 1426.741 | 1426.728 | 1426.735 | 0.007 |
| Co-C (T) | 1597.374 | 1597.390 | 1597.382 | 0.008 |
| Pd-C (T) | 1764.970 | 1764.963 | 1764.967 | 0.004 |
| Ru-C (T) | 2226.724 | 2226.699 | 2226.712 | 0.013 |
| WC-C (T) | 3020.360 | 3020.365 | 3020.363 | 0.003 |
4.4. Preliminary Scale Realization
4.5. Corrections
4.5.1. SSE correction
4.5.2. Non- Linearity correction
4.5.3. Emissivity correction
4.5.4. Temperature drop correction
4.6. Scale Realization
4.7. ITS-90 routines
5. Uncertainty Evaluation
6. Case Study: Optimization of the Calibration Scheme
7. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Preston-Thomas, H. The International Temperature Scale of 1990 (ITS-90). Metrologia 1990, 27, 3–10. [Google Scholar] [CrossRef]
- Stock, M.; Davis, R.; de Mirandés, E.; Milton, M.J.T. The revision of the SI—the result of three decades of progress in metrology. Metrologia 2019, 56, 022001. [Google Scholar] [CrossRef]
- Machin, G. The kelvin redefined. Meas. Sci. Technol. 2018, 29, 022001. [Google Scholar] [CrossRef]
- Pearce, J.; Machin, G.; Todd, A.; Martin, M.J.; Pan, Y.; Zhou, K.; Olsen, Å.; Kowal, A.; Gaiser, C.; Kawamura, Y.; Nakano, T. Future traceability of practical primary thermometry and self-validating thermometry. Philos. Trans. R. Soc. A 2026, 384, 20240453. [Google Scholar] [CrossRef] [PubMed]
- Fischer, J. The Boltzmann constant for the definition and realization of the kelvin. Ann. Phys. 2019, 531, 1800304. [Google Scholar] [CrossRef]
- Consultative Committee for Thermometry (CCT). Mise en pratique for the definition of the kelvin in the SI; Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2019. [Google Scholar]
- Fellmuth, B.; Fischer, J.; Machin, G.; Picard, S.; Steur, P.P.M.; Tamura, O.; Yoon, H. The kelvin redefinition and its mise en pratique. Philos. Trans. R. Soc. A 2016, 374, 20150037. [Google Scholar] [CrossRef]
- Machin, G.; Anhalt, K.; Bloembergen, P.; Sadli, M.; Lowe, D.; Saunders, P.; Yamada, Y.; Yoon, H. MeP-K Relative Primary Radiometric Thermometry (Edition 2017); Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2017. [Google Scholar]
- Machin, G.; Anhalt, K.; Bloembergen, P.; Yamada, Y.; Yoon, H. MeP-K Absolute Primary Radiometric Thermometry, Section 1: Absolute (Spectral-Band) Radiometry (Radiation Thermometry); Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2012. [Google Scholar]
- Machin, G.; Bloembergen, P.; Anhalt, K.; Hartmann, J.; Sadli, M.; Saunders, P.; Woolliams, E.; Yamada, Y.; Yoon, H. Practical implementation of the mise en pratique for the definition of the kelvin above the silver point. Int. J. Thermophys. 2010, 31, 1779–1788. [Google Scholar] [CrossRef]
- Machin, G.; Anhalt, K.; Bloembergen, P.; Sadli, M.; Saunders, P.; Woolliams, E.; Yamada, Y.; Yoon, H. MeP-K Absolute Primary Radiometric Thermometry; Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2020. [Google Scholar]
- Machin, G.; Anhalt, K.; Bloembergen, P.; Hartmann, J.; Saunders, P.; Woolliams, E.; Yoon, H. Realisation and dissemination of thermodynamic temperature above the silver point (1234.93 K). CCT 25th Session, Doc. CCT/2010-12, 2010.
- Saunders, P. The non-uniqueness of ITS-90 above the silver point and its impact on values of T − T90. Metrologia 2020, 57, 045007. [Google Scholar] [CrossRef]
- Anhalt, K.; Machin, G. Thermodynamic temperature by primary radiometry. Philos. Trans. R. Soc. A 2016, 374, 20150041. [Google Scholar] [CrossRef]
- Machin, G.; Engert, J.; Gavioso, R.M.; Sadli, M.; Woolliams, E.R. The EURAMET metrology research programme project implementing the new kelvin (InK). Int. J. Thermophys. 2014, 35, 405–416. [Google Scholar] [CrossRef]
- Machin, G.; Engert, J.; Gianfrani, L.; McEvoy, H.; Sparasci, F. The European Metrology Programme for Innovation and Research project: implementing the new kelvin 2 (InK2). J. Phys. Conf. Ser. 2018, 1065, 122002. [Google Scholar] [CrossRef]
- Machin, G.; Sadli, M.; Engert, J.; Kirste, A.; Pearce, J.; Gavioso, R.M. Progress with realizing the redefined kelvin. AIP Conf. Proc. 2024, 3230, 020001. [Google Scholar] [CrossRef]
- Woolliams, E.R.; Anhalt, K.; Ballico, M.; Bloembergen, P.; Bourson, F.; Briaudeau, S.; Campos, J.; Cox, M.G.; del Campo, D.; Dong, W.; Dury, M.R.; Gavrilov, V.; Grigoryeva, I.; Hernanz, M.L.; Jahan, F.; Khlevnoy, B.; Khromchenko, V.; Lowe, D.H.; Lu, X.; Yuan, Z. Thermodynamic temperature assignment to the point of inflection of the melting curve of high-temperature fixed points. Philos. Trans. R. Soc. A 2016, 374, 20150044. [Google Scholar] [CrossRef]
- Martin, M.J.; Mantilla, J.M. The transition from ITS-90 to primary thermometry above 1235 K. Philos. Trans. R. Soc. A 2026, 384, 20240448. [Google Scholar] [CrossRef]
- CCT-WG5 on Radiation Thermometry. Uncertainty Budgets for Realisation of Scales; Bureau International des Poids et Mesures (BIPM): Sèvres, France.
- Yamada, Y.; Sakate, H.; Sakuma, F.; Ono, A. Radiometric observation of melting and freezing plateaus for a series of metal–carbon eutectic points in the range 1330 °C to 1950 °C. Metrologia 1999, 36, 207–209. [Google Scholar] [CrossRef]
- Machin, G. Twelve years of high temperature fixed point research: A review. AIP Conf. Proc. 2013, 1552, 305–316. [Google Scholar] [CrossRef]
- Battuello, M.; Florio, M.; Sadli, M.; Bourson, F. Investigations on two Co–C fixed-point cells prepared at INRIM and LNE-Cnam. Int. J. Thermophys. 2011, 32, 1732–1743. [Google Scholar] [CrossRef]
- Bourson, F.; Briaudeau, S.; Rougié, B.; Sadli, M. Developments around the Co–C eutectic point at LNE-INM/Cnam. Acta Metrol. Sin. 2008, 29. [Google Scholar]
- Khlevnoy, B.B.; Grigoryeva, I.A.; Otryaskin, D.A. Development and investigation of WC–C fixed-point cells. Metrologia 2012, 49, S59–S67. [Google Scholar] [CrossRef]
- Yamada, Y.; Wang, Y.; Sasajima, N. Metal carbide–carbon peritectic systems as high-temperature fixed points in thermometry. Metrologia 2006, 43, L23–L27. [Google Scholar] [CrossRef]
- Yamada, Y.; Duan, Y.; Ballico, M.; Park, S.N.; Sakuma, F.; Ono, A. Measurement of Rh–C, Pt–C and Ru–C eutectic points by four national metrology institutes. Metrologia 2001, 38, 203–211. [Google Scholar] [CrossRef]
- Yamada, Y.; Anhalt, K.; Battuello, M.; Bloembergen, P.; Khlevnoy, B.; Machin, G.; Matveyev, M.; Sadli, M.; Wang, T. Construction of high-temperature fixed-point cells for thermodynamic temperature assignment. AIP Conf. Proc. 2013, 1552, 335–339. [Google Scholar] [CrossRef]
- Bloembergen, P.; Yamada, Y. Measurement of thermodynamic temperature above the silver point on the basis of the scheme n = 2. Int. J. Thermophys. 2011, 32, 45–67. [Google Scholar] [CrossRef]
- Lowe, D.; Machin, G. Low uncertainty thermodynamic temperature above the silver point using relative primary radiometry. AIP Conf. Proc. 2024, 3230, 100002. [Google Scholar] [CrossRef]
- Lowe, D. A pyrometer for calibration traceable to a future thermodynamic temperature scale. Meas. Sci. Technol. 2013, 24, 015901. [Google Scholar] [CrossRef]
- MultiFixRad—Improving the realisation of the kelvin by multiple fixed-point radiation thermometry. Available online: https://projects.lne.eu/jrp-multifixrad/the-project/.
- Sadli, M.; Falnes Olsen, Å.A.; Kňazovická, L.; Pavlásek, P.; Nasibli, H.; Bourson, F.; Bojkovski, J.; Broberg, P.; Clausen, S.; Gözönünde, C.; Holmsten, M.; Kadı, O.F.; Králová, M.; Kozlova, O.; Maniur, M.I.; Mlačnik, V.; Næsby Rasmussen, A.; Opel, K.; Zuzek, V. MULTIFIXRAD: A European project for the mise-en-pratique of the kelvin at high temperature using HTFPs. 2025.
- Sadli, M.; Kňazovická, L.; Pavlásek, P.; Nasibli, H.; Bourson, F.; Falnes Olsen, Å.A. New capabilities for the realisation and the dissemination of the kelvin at high temperature in Europe. Meas. Sens. 2025, 38, 101624. [Google Scholar] [CrossRef]
- Saunders, P.; White, D.R. Physical basis of interpolation equations for radiation thermometry. Metrologia 2003, 40, 195–203. [Google Scholar] [CrossRef]
- Saunders, P.; White, D.R. Interpolation errors for radiation thermometry. Metrologia 2004, 41, 41–46. [Google Scholar] [CrossRef]
- Saunders, P. Uncertainties in the realization of thermodynamic temperature above the silver point. Int. J. Thermophys. 2011, 32, 26–44. [Google Scholar] [CrossRef]
- Saunders, P. Propagation of uncertainty for an interpolated radiation thermometry scale; CCT Working Document ID-326; Bureau International des Poids et Mesures (BIPM): Sèvres, France.
- White, D.R.; Saunders, P. The propagation of uncertainty with calibration equations. Meas. Sci. Technol. 2007, 18, 2157–2164. [Google Scholar] [CrossRef]
- Joint Committee for Guides in Metrology (JCGM). Evaluation of Measurement Data—Guide to the Expression of Uncertainty in Measurement (GUM); JCGM 100:2008; Joint Committee for Guides in Metrology: 2008.
- Can, M.; Gözönünde, C.; Arifoviç, N.; Yıldız, F.; Nasibov, H. Large-area Fe–C eutectic fixed-points for radiation and contact thermometry. Meas. Sci. Technol. 2023, 34, 124001. [Google Scholar] [CrossRef]
- Can, M.; Kadı, Ö.F.; Korkmaz, M.; Gözönünde, C.; Nasibli, H. Realizing Fe–C, Pd–C, Ru–C and WC–C eutectic fixed-points at UME. AIP Conf. Proc. 2024, 3230, 070002. [Google Scholar] [CrossRef]
- Saunders, P.; Woolliams, E.; Yoon, H.; Todd, A.; van der Ham, E.; Anhalt, K.; Khlevnoy, B. Uncertainty Estimation in Primary Radiometric Temperature Measurement; Mise en pratique for the definition of the kelvin (MeP-K); Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2018. [Google Scholar]
- Yoon, H.W.; Allen, D.W.; Saunders, R.D. Methods to reduce the size-of-source effect in radiometers. Metrologia 2005, 42, 89–96. [Google Scholar] [CrossRef]
- Sakuma, F.; Hattori, S. Temperature: Its measurement and control in science and industry. In Temperature: Its Measurement and Control in Science and Industry; Schooley, J.F., Ed.; American Institute of Physics: New York, NY, USA, 1982; Vol. 5, pp. 421–427. [Google Scholar]
- Todd, A.D.W.; Anhalt, K.; Bloembergen, P.; Khlevnoy, B.B.; Lowe, D.H.; Machin, G.; Saunders, P. On the uncertainties in the realization of the kelvin based on thermodynamic temperatures of high-temperature fixed-point cells. Metrologia 2021, 58, 035007. [Google Scholar] [CrossRef]
- Bloembergen, P.; Yamada, Y.; Sasajima, N.; Wang, Y.; Wang, T. The effect of the eutectic structure and the residual effect of impurities on the uncertainty in the eutectic temperatures of Fe–C and Co–C. Metrologia 2007, 44, 279–293. [Google Scholar] [CrossRef]
- Lowe, D.; Mingard, K.; Malik, Z.; Quested, P. The dependence of the melting temperature of cobalt–carbon eutectic on the morphology of its microstructure. Int. J. Thermophys. 2007, 28, 2019–2027. [Google Scholar] [CrossRef]
- Bloembergen, P. On the uncertainty in the correction for the size-of-source effect. Metrologia 2009, 46, 544–553. [Google Scholar] [CrossRef]
- Yamada, Y.; Anhalt, K.; Battuello, M.; et al. Evaluation and selection of high-temperature fixed-point cells for thermodynamic temperature assignment. Int. J. Thermophys. 2015, 36, 1834–1847. [Google Scholar] [CrossRef]
- Xie, Y.; Dai, C.; Wang, Y.; Wu, Z.; Li, L.; He, S. A selective multiple fit method of the point of inflection for the large-area HTFP melting plateau curve. Measurement 2021, 173, 108576. [Google Scholar] [CrossRef]
- Bloembergen, P. On the correction for the size-of-source effect corrupted by background radiation. In Proceedings of TEMPMEKO 1999, 1999; pp. 607–612.
- Yoon, H.W.; Butler, J.J.; Larason, T.C.; Eppeldauer, G.P. Linearity of InGaAs photodiodes. Metrologia 2003, 40, S154–S158. [Google Scholar] [CrossRef]
- Sadli, M.; Bourson, F.; Lowe, D.; Anhalt, K.; Taubert, D.; Martin, M.J.; Mantilla, J.M.; Girard, F.; Florio, M.; Gözönünde, C.; Nasibli, H.; Kňazovická, L.; Sasajima, N.; Lu, X.; Kozlova, O.; Briaudeau, S.; Machin, G. Thermodynamic temperatures of Fe–C, Pd–C, Ru–C and WC–C for the mise-en-pratique of the kelvin up to 3020 K. AIP Conf. Proc. 2024, 3230, 020004. [Google Scholar] [CrossRef]
- Castro, P.; Machin, G.; Villamañan, M.A.; Lowe, D. Calculation of the temperature drop for high-temperature fixed points for different furnace conditions. Int. J. Thermophys. 2011, 32, 1773–1785. [Google Scholar] [CrossRef]
- Bureau International des Poids et Mesures (BIPM). Guide to the Realization of the ITS-90—Radiation Thermometry; BIPM: Sèvres, France, 2018. [Google Scholar]
- Saunders, P. (Measurement Standards Laboratory, New Zealand). Personal communication.
- Khlevnoy, B.B.; Sapritsky, V.I.; Ogarev, S.A.; Sakharov, M.K.; Samoylov, M.L.; Pikalev, Y.A. Development of fixed points above 2700 K based on MC and MC–C eutectics at VNIIOFI for radiation thermometry and radiometry. In Proceedings of the 9th International Symposium on Temperature and Thermal Measurements in Industry and Science (TEMPMEKO), Dubrovnik, Croatia, June 2004; pp. 203–208. [Google Scholar]
- Khlevnoy, B.; Grigoryeva, I.; Ivashin, E.; Otryaskin, D.; Solodilov, M. The candela realisation using molybdenum carbon fixed-point blackbody. Metrologia 2024, 61, 055011. [Google Scholar] [CrossRef]
- Nasibov, H.; Diril, A.; Pehlivan, O.; Kalemci, M. Comparative study of two InGaAs-based reference radiation thermometers. Int. J. Thermophys. 2017, 38, 112. [Google Scholar] [CrossRef]
- Battuello, M.; Girard, F.; Florio, M. Extrapolation of radiation thermometry scales for determining the transition temperature of metal–carbon points: Experiments with the Co–C. Metrologia 2009, 46, 26–32. [Google Scholar] [CrossRef]














| Reference Point | Melting temperature , K | ||||
| POI | UPOI | LP | ULP | ||
| Fe-C (T) | 1426.92 | 0.14 | 1427.02 | 0.16 | |
| Co-C (T) | 1597.39 | 0.13 | 1597.48 | 0.14 | |
| Pd-C (T) | 1765.05 | 0.16 | 1765.18 | 0.18 | |
| Pt-C (T) | 2011.43 | 0.18 | 2011.50 | 0.22 | |
| Ru-C (T) | 2226.99 | 0.24 | 2227.08 | 0.24 | |
| Re-C (T) | 2747.84 | 0.35 | 2747.91 | 0.44 | |
| WC-C (T) | 3020.85 | 0.40 | 3020.92 | 0.40 | |
| Freezing point, K | |||||
| Cu (T) | 1357.802 | U=0.08 | |||
| HTFPs | Liquidus, K | |||
| Fraction | Intersection | Average | (Max-min)/2 | |
| Fe-C (T) | 1426.848 | 1426.838 | 1426.843 | 0.005 |
| Co-C (T) | 1597.545 | 1597.537 | 1597.441 | 0.004 |
| Pd-C (T) | 1765.076 | 1765.053 | 1765.065 | 0.007 |
| Ru-C (T) | 2226.783 | 2226.773 | 2226.778 | 0.005 |
| WC-C (T) | 3020.404 | 3020.394 | 3020.399 | 0.005 |
| HTFPs | Corrections, mK | |||
| SSE Correction |
Linearity Correction |
Emissivity Correction |
Temperature Drop Correction | |
| Cu | 9 | 0.1 | 25 | 1.3 |
| Fe-C | 10 | 4 | 28 | 1.5 |
| Co-C | 12 | 5 | 35 | 2.4 |
| Pd-C | 14 | 10 | 42 | 3.6 |
| Ru-C | 22 | 16 | 67 | 9 |
| WC-C | 39 | 29 | 124 | 31 |
| Uncertainty component | Fe–C | Co-C | Pd–C | Ru–C | WC–C |
|---|---|---|---|---|---|
| Gain Ratio | 0.009 | 0.012 | 0.014 | 0.022 | 0.041 |
| Calibration of the radiometer | 0.066 | 0.110 | 0.167 | 0.233 | 0.562 |
| Linearity of radiometer | 0.014 | 0.017 | 0.021 | 0.034 | 0.062 |
| Ambient Temperature | 0.001 | 0.001 | 0.001 | 0.002 | 0.004 |
| The drift of the radiometer since calibration | 0.006 | 0.007 | 0.009 | 0.014 | 0.025 |
| SSE | 0.018 | 0.022 | 0.028 | 0.045 | 0.083 |
| Alignment of radiometer | 0.018 | 0.022 | 0.028 | 0.045 | 0.083 |
| Determination of POI | 0.035 | 0.010 | 0.006 | 0.003 | 0.009 |
| Stability of source | 0.031 | 0.008 | 0.006 | 0.005 | 0.004 |
| Combined Uncertainty, K, (k = 1) | 0.09 | 0.12 | 0.18 | 0.25 | 0.58 |
| Expanded Uncertainty, K, (k = 2) | 0.17 | 0.23 | 0.35 | 0.49 | 1.16 |
| Source of uncertainty | Uncertainty Contribution, K |
|---|---|
| Impurity | 0.005 |
| Emissivity | 0.009 |
| Temperature drop | 0.003 |
| Cu–Plateau Determination | 0.011 |
| Repeatability | 0.025 |
| Wavelength | 0.000 |
| Repeatability | 0.000 |
| Short time Drift | 0.006 |
| Out-of-band transmission | 0.000 |
| Size-of-source | 0.005 |
| Non-linearity | 0.015 |
| Drift | 0.026 |
| Combined Uncertainty (k=1), U | 0.05 |
| Expanded Uncertainty (k=2), U | 0.09 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |