Abstract
Purpose: Bronchoscopy has matured from a diagnostic channel into a therapeutic platform, yet its guidance still runs on static pre-procedural CT maps while the target moves. This review defines what digital twins currently are in bronchoscopic intervention, maps the components that already behave like twins, and proposes an agenda for the field’s dynamic era.Methods: A structured PubMed search (22 November 2026) identified all publications explicitly adopting the digital-twin concept in bronchoscopy; enabling technologies were systematically mapped along the bronchoscopic workflow and synthesized as a four-layer technical stack.Results: Nine publications explicitly adopting the digital-twin term were identified; the only randomized evidence to date—two pioneering trials—addresses training and pre-procedural risk communication rather than intra-procedural guidance, while motion modeling and predictive targeting remain open opportunities. Meanwhile every layer of a dynamic procedure twin has validated components in clinical use: airway-tree segmentation and pathway planning, four-dimensional motion mathematics from radiation oncology, scope-tracking and cone-beam CT confirmation, and a workflow that already generates quantized ground-truth labels (radial ultrasound views, tool-in-lesion confirmation). We formalize the field’s error structure as a dual registration chain, propose a tier-by-correction-cadence classification, and stage a five-rung validation ladder with five falsifiable research propositions.Conclusion: The static-map era of bronchoscopic guidance is complete and clinically validated; the dynamic era is an assembly problem. The first group to close the loop between a patient-specific replica and the bronchoscope will define the next standard of navigational and therapeutic bronchoscopy.
Keywords:
bronchoscopy
; digital twin
; navigation bronchoscopy
; cone-beam CT
; respiratory motion
; transbronchial ablation
2. Clinical Maturity of Bronchoscopic Intervention: The Application Soil for a Digital Twin
2.1. Diagnostic Bronchoscopy: Navigation Mature, Yield Plateaued
Peripheral bronchoscopy is the most comprehensively engineered access route in interventional pulmonology. Virtual bronchoscopic navigation precomputes a pathway through the segmented airway tree; electromagnetic navigation bronchoscopy (ENB) tracks a locatable guide within that tree; shape-sensing robotic platforms hold the scope tip under continuous positional control; and intra-procedural imaging—radial endobronchial ultrasound (rEBUS), augmented fluoroscopy, and mobile CBCT—confirms whether the instrument has truly arrived [2,3,21]. Robotic platforms have reported navigation success exceeding 95% in prospective multicenter cohorts [22,23], and mobile CBCT demonstrably improves targeting when added to robotic guidance [24].
Two features of this maturity deserve emphasis because a digital twin will be built on top of them. First, the field’s endpoints are already quantized: navigation success, radial-probe view (concentric, eccentric, or adjacent to the lesion), tool-in-lesion confirmation on CBCT, and final diagnostic yield are distinct, countable events rather than gestalt judgments [13,24]. Second, the ceiling is equally well measured: pooled diagnostic yield across navigation technologies has remained near 70% for years [13], and the residual failures concentrate precisely where the static map fails—small, mobile, lower-lobe targets without a bronchus sign. The field knows exactly how good its maps are, and exactly where they break.
2.2. Therapeutic Bronchoscopy: Ablation Reaching Clinical Maturity
The therapeutic conversion of the bronchoscope has been rapid. After preclinical feasibility work with bronchoscopic vapor and microwave systems in porcine lung [25,26], Chan and colleagues reported the initial 30-case experience of ENB-guided transbronchial microwave ablation [5], followed by prospective evaluation of a flexible ablation probe [6] and the multicenter NAVABLATE study combining transbronchial ablation with CBCT and augmented fluoroscopy [27]. Electromagnetic-navigation-guided three-dimensional conformal ablation [7] and transbronchial cryoablation with thin cryoprobes [8,28] have extended the modality range; single-session treatment of multiple nodules [29] and a head-to-head comparison against CT-guided percutaneous ablation [30] frame the route’s competitive position; and dedicated reviews now treat bronchoscopic ablation of peripheral lung cancer as an emerging frontier rather than an experiment [9,10,31,32,33]. A multicenter treat-and-resect protocol coupling bronchoscopic ablation to surgical histopathology (NCT05281237) will provide an unprecedented validation substrate [34]. The energy–tissue modeling that therapeutic delivery demands—dielectric characterization, heat-sink physics, thermal-dose integration—is the subject of our companion review and is not duplicated here [19].
The transbronchial route intensifies rather than relaxes the guidance problem. The applicator approaches through a branching, deformable conduit, so achievable orientation is constrained by bronchial anatomy; the target shifts between planning CT, registration, and energy delivery; and CBCT confirmation captures a single respiratory state [11,12]. Achieved ablation zones deviate measurably from planned ones even under percutaneous guidance [14]; under the airway route, with its additional degrees of freedom, patient-specific prediction becomes both harder and more valuable.
2.3. The Shared Unmet Need: A Dual Registration Chain
We propose that the guidance problem of bronchoscopic intervention has a precise structure that distinguishes it from every percutaneous route. Percutaneous targeting solves one registration: image to patient, with the needle passing through static air. Bronchoscopic targeting solves two in series: scope-to-map (the registration of the bronchoscope tip into the pre-procedural CT, maintained by electromagnetic or shape-sensing tracking) and map-to-patient (the correspondence between that CT and the breathing patient in front of the operator, degraded by CT-to-body divergence). Error propagates through this dual registration chain multiplicatively, not additively: a perfect scope registration cannot rescue a stale map, and a refreshed map cannot rescue a drifting scope. Every contemporary correction—registration refresh, tomosynthesis update, CBCT confirmation—repairs one link at one moment, episodically [12,24]. What no system does is maintain the chain continuously and predictively.
The same structure explains why bronchoscopy, of all ablation and biopsy routes, is uniquely ready for a dynamic solution. The workflow already emits quantized ground truth at every step: the radial-probe view classifies instrument–lesion geometry, tool-in-lesion imaging confirms arrival, and navigation success is scored separately from diagnostic yield [13,24]. A self-labeling workflow of this kind is precisely what a predictive model requires for calibration and validation; percutaneous routes have nothing comparable. The thesis of this review is that bronchoscopic intervention possesses the most structured guidance problem and the richest validation substrate in procedural lung medicine—and that the two facts together make it the natural first home of a procedure-level digital twin.
3. What Counts as a Digital Twin: A Hierarchy for Bronchoscopic Intervention
The term “digital twin” dilutes in proportion to its popularity, and procedural medicine is not immune. We adopt the three-tier hierarchy introduced in our companion review [19], itself adapted from manufacturing and medical frameworks [15,16,35], classified by the direction and continuity of data flow between patient and replica. A Tier 1 static patient-specific model is built once from pre-procedural data and simulated without updating; most computational models, however sophisticated, live here. A Tier 2 digital shadow adds one-directional flow from physical to virtual—the replica is updated with intra-procedural data and informs the operator without steering therapy; augmented-reality localization registered to the operative field is the lung’s clearest example [36]. A Tier 3 digital twin closes the loop bidirectionally: the replica is repeatedly synchronized with the patient and its predictions feed back to modulate the intervention itself, as approached by virtual-heart systems in cardiology [17].
Three extensions adapt this hierarchy to the bronchoscopy suite, and each names something the field currently discusses without a term. First, the replica-role taxonomy: a replica may serve as a display (a better picture), a decision object (a thing the operator interrogates before acting), or a prediction object (a thing that forecasts what the intervention will do). Today’s bronchoscopic replicas are overwhelmingly displays; the dynamic era begins when the replica becomes a prediction object—a performance twin that forecasts targeting success, achievable margins, and complication risk for this patient, this lesion, this plan. Second, the procedure twin: the object being replicated in bronchoscopy is not merely an organ but an event—the scope, the airway corridor, the moving target, and the operator, bound together for thirty minutes. Section 4 shows that every digital twin published in this field so far twins the procedure for training or communication; none yet twins it for guidance. Third, correction cadence: the frequency and mode with which reality is allowed to update the replica. Current systems correct episodically—a registration refresh here, a CBCT there; a dynamic twin corrects continuously and, ultimately, predictively, compensating motion before divergence occurs. Correction cadence converts the static–shadow–twin hierarchy from a philosophical axis into a measurable engineering specification, and Figure 1 maps contemporary bronchoscopic systems onto the resulting two-axis grid.
Applied with this shared terminology, the current picture is clear. Bronchoscopic intervention possesses abundant Tier 1 models (planning reconstructions, computational prototypes [37]) and genuine Tier 2 shadows (AR localization [36], CBCT-corrected navigation [24]), but no Tier 3 system. The nearest digital-twin framework in pulmonary medicine—a CT-derived thoracic replica for light-dosimetry personalization [38]—likewise qualifies only as Tier 1 and never enters the procedure room. The first closed-loop procedure twin for bronchoscopy remains to be built, and, as in lung ablation [19], its construction is an assembly problem more than an invention problem.
4. The Current Digital-Twin Literature in Bronchoscopy and Airway Medicine: An Inventory
Which bronchoscopy publications have explicitly adopted the digital-twin concept? Our structured search returned nine records (Table 1), and their distribution is instructive. Encouragingly, the field’s first randomized evidence is already here: a single-center randomized study in which a CT-derived digital-twin bronchoscopy simulator improved novice performance and three-month skill retention, with training on anatomically diverse patient-specific models outperforming training on a single standard model [39]; and a multicenter randomized trial in which patient-specific simulator visualization during pre-procedural risk disclosure reduced anxiety and improved satisfaction in adults undergoing elective bronchoscopy [40]. A third randomized study of digital-twin simulation-based training is ongoing [41]. The remaining records extend the digital-twin concept to adjacent pulmonary problems—ventilator-protocol testing [42], aerosol deposition modeling [43,44], acute-respiratory-failure decision rules [45], airway-remodeling prediction in asthma [46], inhalable biomaterials [47], and breathing-gas logistics [48]—rather than to bronchoscopic guidance.
Three observations follow. First, these pioneering twins have established their first clinical foothold exactly where replica technology matures fastest: outside the procedure room, on reconstructed anatomy, for training and communication. We read this as confirmation that the underlying replica technology is real and randomized-trial-grade, and as the natural beachhead from which guidance applications can build. Second, respiratory-motion modeling and predictive intra-procedural outputs have not yet been incorporated into these works; under the hierarchy of Section 3 they correspond to the Tier 1–2 foundation stages on which closed-loop systems will build. Third, the concentration is no accident of history: bronchoscopy already possesses a mature simulation culture, with systematic-review-grade evidence that simulation-based training works and codified best practices for delivering it [49,50]. The specialty that built that culture is the specialty most likely to build the guidance twin—and the two randomized trials are best read as its proof of concept.
5. A Four-Layer Technical Stack for the Bronchoscopy Digital Twin
We organize the assembly problem into four layers, each expressed in the field’s own vocabulary (Figure 2). For every layer we identify the validated components already in clinical or prototype use, the bronchoscopy-specific barrier, and the concrete gap; Table 2 summarizes the mapping with measurable success criteria.
5.1. Layer 1: Airway-Tree and Target Modeling
The geometric substrate is a patient-specific reconstruction of the bronchial tree—generation by generation—together with the target, its feeding bronchus, and the surrounding vessels and fissures. This layer is the most mature: airway segmentation from thin-section CT is routine in every navigation platform, deep-learning methods have substantially closed the discontinuity problem that once fragmented peripheral branches [51], and virtual bronchoscopic planning already computes pathways to the target that operators follow in daily practice [2]. The CT bronchus sign—a bronchus leading into rather than merely adjacent to the lesion—remains the field’s strongest single predictor of biopsy success and is, in twin terms, a topological property of the patient-specific model [13,21]. Two bronchoscopy-specific caveats remain. First, ground-glass targets have indistinct margins whose segmentation uncertainty propagates into every downstream layer. Second, the model must encode not only anatomy but passability: scope diameter, flexion limits, and working-channel geometry constrain which computed paths are physically traversable, a constraint no planning system yet exploits explicitly.
5.2. Layer 2: Motion and CT-to-Body Divergence Modeling
This is the layer where bronchoscopy’s problem is unique and where the enabling mathematics is, fortunately, most mature—just not in bronchoscopy. Respiratory target motion of one to two centimeters, with lower-lobe predominance, hysteresis, and day-to-day variability, is the dominant error source of the map-to-patient link, and its clinical signature is CT-to-body divergence [11]. Radiation oncology has spent two decades industrializing exactly this mathematics: four-dimensional CT acquisition, deformable image registration with quantified nonlinearity [52], deep-learning registration of four-dimensional CT and cone-beam images [53], and surface-guided surrogate monitoring [54,55] all carry known error bounds [56]. Within bronchoscopy itself, a personalized image-guided intervention system built on a patient-specific respiratory-motion model has demonstrated feasibility for peripheral biopsy guidance in prototype [37], and ventilation strategy—jet ventilation, apnea windows, controlled atelectasis—offers a pragmatic intra-procedural handle on the same physics [57]. The gap is integration: no navigation platform carries a motion model into the procedure, no system predicts where the target will be at the moment of sampling, and CBCT confirmation captures one respiratory phase of a moving object [12,24]. This is simultaneously the largest recoverable error source and the clearest opportunity for measurable improvement.
5.3. Layer 3: Scope–Twin Coupling and Confirmation
The third layer binds the physical procedure to the replica: continuous localization of the scope and instruments within the patient-specific model, and confirmation of instrument–target geometry. Its components are the field’s daily tools—electromagnetic and shape-sensing tracking [22,23], vision-based pose optimization that corrects registration from the scope’s own camera [58], radial-probe ultrasound classification of instrument–lesion geometry, and CBCT or augmented-fluoroscopy confirmation of tool-in-lesion [24,27]. What distinguishes bronchoscopy here is the quantized vocabulary of confirmation noted in Section 2.3: a concentric rEBUS view and a tool-in-lesion CBCT are not impressions but labels, and a twin coupled to this layer inherits calibration data with every procedure. The barrier is that these channels currently terminate at the operator’s eyes rather than in the model: confirmation events are consumed episodically (correct the registration, repeat the pass) instead of being assimilated as continuous observations that update the replica’s state. Closing that channel is an engineering task, not a conceptual one.
5.4. Layer 4: Closed-Loop Recalibration
The final layer is what promotes a shadow to a twin: the replica absorbs the confirmation stream, re-estimates the procedure state, and issues concrete corrective or predictive guidance—redirect two generations distally, wait for end-expiration, the margin is insufficient at this orientation. Its prerequisites are speed and assimilation logic. Fast re-solution is achievable: model-order-reduction methods already generate patient-specific mechanical surrogates that run orders of magnitude faster than full simulation [59], and the confirmation channels of Layer 3 provide the observations. The barrier is that nothing in the current stack performs state estimation at all; correction cadence (Section 3) is the honest metric of how far the field has to travel—from episodic, operator-initiated correction to continuous, model-initiated compensation. The trajectory of adjacent fields suggests the endpoint is realistic: cardiac electrophysiology already rehearses ablation strategies patient-specifically before delivering them [17], and radiation oncology corrects motion continuously during delivery [55,56]. Bronchoscopy’s twin will be declared operational the day the map stops going stale.
6. A Staged Validation Pathway: From Phantom to Prospective Trial
We propose a five-rung validation ladder (Table 3), ordered by evidentiary weight and exploiting the self-labeling workflow that distinguishes this field. Rung 1, phantom and simulation: performance twins are exercised against benchtop airway models and the existing simulator infrastructure, with targeting error under programmed motion as the endpoint; the simulation culture that produced the field’s first digital-twin randomized trials provides ready methodology [39,50]. Rung 2, retrospective digital rehearsal: the twin replays archived procedures—navigation CT, intra-procedural imaging, recorded confirmation events—and must reproduce the observed outcomes (radial-probe view, tool-in-lesion, diagnostic yield) better than chance and better than the static plan; planning-model precedents in percutaneous ablation demonstrate both the feasibility and the pitfall of this stage, and a twin should demonstrate superiority over well-built regression models on retrospective data before its guidance value is claimed [60,61]. Rung 3, the accuracy gate: a prospective, pre-specified demonstration that twin-informed guidance reduces targeting error—divergence between predicted and confirmed instrument–target geometry—below the plateau documented in contemporary cohorts [13,24]. Rung 4, prospective utility: a randomized comparison of twin-guided versus conventional navigation, using the noninferiority-then-superiority logic already established for augmented-reality localization [62], with diagnostic yield, procedure time, and complications as endpoints. Rung 5, therapeutic extension and histological validation: for transbronchial ablation, the treat-and-resect design offers what no other organ offers—direct histopathological scoring of predicted-versus-achieved treatment effect [34], with three-dimensional margin assessment as the quantitative endpoint [63,64].
7. Research Agenda: Five Propositions
The agenda (Figure 3) is deliberately falsifiable; each proposition carries a pre-specifiable endpoint, and Figure 4 depicts the end-to-end proof-of-concept workflow we consider achievable within existing technology. P1: motion-envelope overlay. A four-dimensional target-motion envelope, computed from pre-procedural four-dimensional CT and deformably registered to the navigation map, displayed on the console as a probabilistic target volume; the mathematics is imported, not invented [52,53], and the endpoint is reduced scope-to-target divergence at sampling. P2: registration-error dashboard. Instrument the dual registration chain: log scope-to-map and map-to-patient error continuously through routine procedures, publish the error budget of contemporary navigation, and convert CT-to-body divergence from an anecdote into a monitored quantity [11,58]. P3: scaled procedure twins for training and credentialing. Extend the two randomized training trials into multi-anatomy, multi-operator programs with skill-retention endpoints, using the existing simulation evidence base [39,40,49]. P4: predictive targeting and digital margin rehearsal. For biopsy and transbronchial ablation alike, rehearse the intervention in the twin—trajectory, expected radial-probe geometry, predicted margin at the achievable applicator orientation—and report pre-specified prediction accuracy before claiming guidance value; the energy–tissue modeling this requires for ablation is specified in our companion review [7,19]. P5: the longitudinal twin. Extend the replica beyond the procedure, predicting ablation-zone involution and integrating surveillance imaging to separate expected evolution from recurrence, aligned with consensus follow-up frameworks [65].
Underneath the propositions sits the architecture that a dynamic procedure twin implies—five layers from data ingestion through modeling, twinning, decision support, and clinical integration (Figure 2)—and a division of labor with the ablation-focused companion review: the present manuscript carries the guidance and targeting stack, the companion carries the energy–tissue stack [19]. Between them, the two manuscripts define the full assembly problem for procedural lung medicine.
8. Challenges and Limitations
Several challenges are structural. Motion modeling in aerated lung must contend with deformation, hysteresis, and day-to-day variability that phantom studies systematically understate. The randomized digital-twin evidence in this field is currently concentrated in one pioneering research group and one simulator platform; broader, independent replication is therefore an explicit agenda item (P3) and a natural next step for the community. Our inventory relies on PubMed-indexed literature and on authors’ self-identification of digital-twin status; enabling technologies were mapped by structured supplementary searches rather than by a systematic review protocol, and some relevant engineering literature outside biomedical indexing may be missed. The tier-by-cadence classification, like any taxonomy, imposes boundaries on a continuum, and the assignment of individual commercial systems to cells of Figure 1 is descriptive rather than evaluative; commercial platforms are mentioned solely to describe the technological state of the art, without endorsement. Finally, the agenda’s most important risk is comparative: statistical prediction from clinical covariates already performs respectably in adjacent ablation planning [61], and a performance twin establishes its value by outperforming such simpler models on pre-specified quantitative endpoints.
9. Conclusions
Bronchoscopic intervention has completed its static era. Navigation platforms, robotic control, and intra-procedural confirmation have converted the peripheral lung into a routine destination, and the field’s own measurements—quantized confirmation vocabulary, a well-characterized yield plateau, a named divergence problem—define exactly what remains unsolved. What is missing is not any single component but the closed-loop assembly: a patient-specific procedure twin that carries motion, maintains the dual registration chain continuously, and earns clinical trust through a staged, falsifiable validation pathway. The components exist; the workflow already labels its own outcomes; the adjacent disciplines have solved the mathematics. The first group to assemble a working procedure twin will not merely improve navigation bronchoscopy—it will define the standard against which the next generation of bronchoscopic intervention is judged.
Author Contributions
Conceptualization, Y.Z.; methodology, Y.Z.; writing—original draft preparation, Y.Z.; writing—review and editing, C.M. and Y.Z.; supervision, Y.Z. All authors have read and agreed to the published version of the manuscript.
Funding
The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.
Data availability
No new data were created or analysed in this study. The structured search strategy, full search terms, and hit counts (22 November 2026) are available from the corresponding author on reasonable request.
Conflicts of Interest
The authors declare no conflicts of interest. As a non-financial disclosure, the authors are developing a research program related to the agenda proposed in this review (Section 8).
Use of AI tools
During the preparation of this manuscript, the authors used AI-based tools for language editing only. The authors have reviewed and edited all content and take full responsibility for the integrity of this work.
Ethics approval
Not applicable; this article is a review of previously published literature and does not contain any studies with human participants or animals performed by any of the authors.
Consent to participate
Not applicable.
Consent to publish
Not applicable.
References
- Bray, F.; Laversanne, M.; Sung, H.; Ferlay, J.; Siegel, R.L.; Soerjomataram, I.; Jemal, A. Global cancer statistics 2022: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J. Clin. 2024, 74, 229–263. [Google Scholar] [CrossRef] [PubMed]
- Kemp, S.V. Navigation Bronchoscopy. Respiration 2020, 99, 277–286. [Google Scholar] [CrossRef] [PubMed]
- Zhang, X.B.; Hogarth, D.K. Multimodal innovations and clinical applications of Robotic-assisted bronchoscopy in pulmonary nodule diagnosis: a review of recent advances. BMC Pulm. Med. 2025, 26, 56. [Google Scholar] [CrossRef] [PubMed]
- Fernandez-Bussy, S.; Chandra, N.C.; Koratala, A.; Yu Lee-Mateus, A.; Barrios-Ruiz, A.; Garza-Salas, A.; Koirala, T.; Funes-Ferrada, R.; Balasubramanian, P.; Patel, N.M.; Chadha, R.; Hazelett, B.N.; Robertson, K.S.; Reisenauer, J.; Abia-Trujillo, D. Robotic-assisted bronchoscopy: a narrative review of systems. J. Thorac. Dis. 2024, 16, 5422–5434. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.W.Y.; Lau, R.W.H.; Ngai, J.C.L.; Tsoi, C.; Chu, C.M.; Mok, T.S.K.; Ng, C.S.H. Transbronchial microwave ablation of lung nodules with electromagnetic navigation bronchoscopy guidance-a novel technique and initial experience with 30 cases. Transl. Lung Cancer Res. 2021, 10, 1608–1622. [Google Scholar] [CrossRef] [PubMed]
- Pritchett, M.A.; Reisenauer, J.S.; Kern, R.; Wilson, D.S.; Meyers, E.E.; Szapary, P.O.; Laeseke, P.F. Novel Image-Guided Flexible-Probe Transbronchial Microwave Ablation for Stage 1 Lung Cancer. Respiration 2023, 102, 182–193. [Google Scholar] [CrossRef] [PubMed]
- Liu, Y.; Huang, Y.; Lin, X.; Rong, Y.; Mao, L.; Lu, S.; Chen, J.; Chen, J.; Xie, X.; Luo, K. Electromagnetic navigation bronchoscopy-guided three-dimensional conformal microwave ablation for unresectable peripheral high-risk pulmonary nodules: a mid-term assessment with no real-time intraoperative CT/cone-beam computed tomography guidance. Transl. Lung Cancer Res. 2026, 15, 274. [Google Scholar] [CrossRef] [PubMed]
- Yang, R.; Chen, J.; Xie, F.; Gu, C.; Hong, S.; Zhu, H.; Jiang, Y.; Sun, J. Mid-term efficacy and safety of transbronchial cryoablation for stage IA peripheral lung cancer: a retrospective analysis. Transl. Lung Cancer Res. 2026, 15, 79. [Google Scholar] [CrossRef] [PubMed]
- Steinfort, D.P.; Herth, F.J.F. Bronchoscopic treatments for early-stage peripheral lung cancer: Are we ready for prime time? Respirology 2020, 25, 944–952. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.W.Y.; Ng, C.S.H. Transbronchial ablation for lung cancers: Ready for prime time? Respirology 2025, 30, 16–17. [Google Scholar] [CrossRef] [PubMed]
- Ho, A.T.N.; Ma, X.; Christos, P.J.; Pedoto, A.; Lee, R.P.; Oberg, C.L.; Rakočević, R.; Bott, M.J.; Chawla, M.; Husta, B.C.; Kalchiem-Dekel, O. Virtual vs. reality in cone-beam CT-guided robotic-assisted bronchoscopy. J. Thorac. Dis. 2025, 17, 9051–9062. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.W.Y.; Chang, A.T.C.; Siu, I.C.H.; Lau, R.W.H.; Ng, C.S.H. Electromagnetic navigation bronchoscopy transbronchial lung nodule ablation with IllumisiteTM platform corrects CT-to-body divergence with tomosynthesis and improves ablation workflow: a case report. AME Case Rep. 2023, 7, 13. [Google Scholar] [CrossRef] [PubMed]
- Kops, S.E.P.; Heus, P.; Korevaar, D.A.; Damen, J.A.A.; Idema, D.L.; Verhoeven, R.L.J.; Annema, J.T.; Hooft, L.; van der Heijden, E.H.F.M. Diagnostic yield and safety of navigation bronchoscopy: A systematic review and meta-analysis. Lung Cancer 2023, 180, 107196. [Google Scholar] [CrossRef] [PubMed]
- Han, X.; Zhang, T.; Meng, M.; Wang, J.; Huang, G.; Dai, J.; Zhao, Z.; Yang, X. Comparison of the difference between the actual ablation zone in lung tumors after computed tomography-guided percutaneous microwave ablation and the predicted value in the planning system: A retrospective study. J. Cancer Res. Ther. 2026, 22, 651–658. [Google Scholar] [CrossRef] [PubMed]
- Laubenbacher, R.; Mehrad, B.; Shmulevich, I.; Trayanova, N. Digital twins in medicine. Nat. Comput Sci. 2024, 4, 184–191. [Google Scholar] [CrossRef] [PubMed]
- Sadée, C.; Testa, S.; Barba, T.; Hartmann, K.; Schuessler, M.; Thieme, A.; Church, G.M.; Okoye, I.; Hernandez-Boussard, T.; Hood, L.; Shmulevich, I.; Kuhl, E.; Gevaert, O. Medical digital twins: enabling precision medicine and medical artificial intelligence. Lancet Digit Health 2025, 7, 100864. [Google Scholar] [CrossRef] [PubMed]
- Hwang, T.; Lim, B.; Kwon, O.S.; Kim, M.H.; Kim, D.; Park, J.W.; Yu, H.T.; Kim, T.H.; Uhm, J.S.; Joung, B.; Lee, M.H.; Hwang, C.; Pak, H.N. Clinical usefulness of digital twin guided virtual amiodarone test in patients with atrial fibrillation ablation. npj Digit Med. 2024, 7, 297. [Google Scholar] [CrossRef] [PubMed]
- Olawade, D.B.; Oisakede, E.O.; Bello, O.J.; Analikwu, C.C.; Egbon, E.; Ojo, A. Digital twins in oncology: From predictive modelling to personalised treatment strategies. Crit. Rev. Oncol. Hematol. 2026, 220, 105171. [Google Scholar] [CrossRef] [PubMed]
- Ma, C.; Zhao, Y. Digital Twin Technology for Image-Guided Thermal Ablation of Pulmonary Nodules: From Computational Proof-of-Concept to Clinical Translation; Preprints, 2026. [Google Scholar] [CrossRef]
- Baethge, C.; Goldbeck-Wood, S.; Mertens, S. SANRA-a scale for the quality assessment of narrative review articles. Res. Integr. Peer Rev. 2019, 4, 5. [Google Scholar] [CrossRef] [PubMed]
- Quah, P.; Lee, P. Peripheral bronchoscopy: back to basics. Curr. Opin. Pulm. Med. 2026, 32, 8–16. [Google Scholar] [CrossRef] [PubMed]
- Ost, D.E.; Folch, E.E.; Reisenauer, J.S.; Majid, A.; Casal, R.F.; Keyes, C.; Parikh, M.S.; Diaz-Mendoza, J.; Fernandez-Bussy, S.; Simoff, M.J. The PRECIsE study: a prospective, multicenter study of shape-sensing robotic-assisted bronchoscopy with 2 years of follow-up. Ann. Am. Thorac. Soc. 2026, 23, 1360–1368. [Google Scholar] [CrossRef] [PubMed]
- Husta, B.C.; Cheng, G.Z.; Batra, H.; Reisenauer, J.S.; Bartek, W.M.; Kalchiem-Dekel, O.; Zouk, A.; Patel, N.; Chawla, M.; Eapen, G.A.; Jimenez, C.A.; Lee, R.P.; Bott, M.J.; Oh, S.; Casal, R.F. Shape-sensing robotic-assisted bronchoscopy with integrated mobile cone-beam CT for small nodules: results from the prospective multicentre CONFIRM study. Thorax 2026, 81, 267–275. [Google Scholar] [CrossRef] [PubMed]
- Chen, J.; McNierney, D.; Zein, J.G.; Vaszar, L.T.; Swanson, K.L.; Azadeh, N.; Sakata, K.K. Diagnostic performance of monarch robotic bronchoscopy with and without mobile cone-beam CT support. Lung Cancer 2026, 213, 108910. [Google Scholar] [CrossRef] [PubMed]
- Steinfort, D.P.; Christie, M.; Antippa, P.; Rangamuwa, K.; Padera, R.; Müller, M.R.; Irving, L.B.; Valipour, A. Bronchoscopic Thermal Vapour Ablation for Localized Cancer Lesions of the Lung: A Clinical Feasibility Treat-and-Resect Study. Respiration 2021, 100, 432–442. [Google Scholar] [CrossRef] [PubMed]
- De Leon, H.; Royalty, K.; Mingione, L.; Jaekel, D.; Periyasamy, S.; Wilson, D.; Laeseke, P.; Stoffregen, W.C.; Muench, T.; Matonick, J.P.; Kaluza, G.L.; Cipolla, G. Device safety assessment of bronchoscopic microwave ablation of normal swine peripheral lung using robotic-assisted bronchoscopy. Int. J. Hyperth. 2023, 40, 2187743. [Google Scholar] [CrossRef] [PubMed]
- Lau, K.K.W.; Lau, R.W.H.; Baranowski, R.; Krzykowski, J.; Ng, C.S.H. Transbronchial Microwave Ablation of Peripheral Lung Tumors: The NAVABLATE Study. J. Bronchol. Interv. Pulmonol. 2024, 31, 165–174. [Google Scholar] [CrossRef] [PubMed]
- Gu, C.; Yuan, H.; Yang, C.; Xie, F.; Chen, J.; Zhu, L.; Jiang, Y.; Sun, J. Transbronchial cryoablation in peripheral lung parenchyma with a novel thin cryoprobe and initial clinical testing. Thorax 2024, 79, 633–643. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.W.Y.; Lau, R.W.H.; Chang, A.T.C.; Siu, I.C.H.; Chu, C.M.; Mok, T.S.K.; Ng, C.S.H. Concomitant electromagnetic navigation transbronchial microwave ablation of multiple lung nodules is safe, time-saving, and cost-effective. JTCVS Tech 2023, 22, 265–272. [Google Scholar] [CrossRef] [PubMed]
- Wang, F.; Yang, B.; Zhang, X.; Subinuer, K.; Li, W.; Rong, H.; Tong, Z. Comparative study of bronchoscopic and CT-guided percutaneous microwave ablation for inoperable non-small cell lung cancer. Transl. Lung Cancer Res. 2025, 14, 3529–3541. [Google Scholar] [CrossRef] [PubMed]
- Chaddha, U.; Hogarth, D.K.; Murgu, S. Bronchoscopic Ablative Therapies for Malignant Central Airway Obstruction and Peripheral Lung Tumors. Ann. Am. Thorac. Soc. 2019, 16, 1220–1229. [Google Scholar] [CrossRef] [PubMed]
- Villalba, A.J.A.; Ost, D.E. Bronchoscopic treatment of early-stage peripheral lung cancer. Curr. Opin. Pulm. Med. 2024, 30, 337–345. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.W.Y.; Siu, I.C.H.; Chang, A.T.C.; Li, M.S.C.; Lau, R.W.H.; Mok, T.S.K.; Ng, C.S.H. Transbronchial Techniques for Lung Cancer Treatment: Where Are We Now? Cancers 2023, 15. [Google Scholar] [CrossRef] [PubMed]
- Beattie, J.; Parrish, R.; Ayasa, L.; Aranguren, P.; Uribe-Buritica, F.L.; Lopez, M.N.; Pineda, C.M.; Cheng, G.; Senitko, M.; Abdelghani, R.; Magge, A.; Avendano, C.A.; Kheir, F.; Swenson, K.; Parikh, M.; Wilson, J.; VanderLaan, P.A.; Gangadharan, S.; Majid, A. Safety and feasibility of bronchoscopic microwave ablation technology for peripheral lung cancer: a multi-center, prospective, single-arm study protocol. BMC Surg. 2025, 25, 497. [Google Scholar] [CrossRef] [PubMed]
- Sun, T.; He, X.; Song, X.; Shu, L.; Li, Z. The Digital Twin in Medicine: A Key to the Future of Healthcare? Front Med. 2022, 9, 907066. [Google Scholar] [CrossRef] [PubMed]
- Li, C.; Ji, A.; Jian, Z.; Zheng, Y.; Feng, X.; Guo, W.; Lerut, T.; Lin, J.; Li, H. Augmented reality navigation-guided intraoperative pulmonary nodule localization: a pilot study. Transl. Lung Cancer Res. 2023, 12, 1728–1737. [Google Scholar] [CrossRef] [PubMed]
- Wang, T.; He, T.; Zhang, Z.; Chen, Q.; Zhang, L.; Xia, G.; Yang, L.; Wang, H.; Wong, S.T.C.; Li, H. A personalized image-guided intervention system for peripheral lung cancer on patient-specific respiratory motion model. Int. J. Comput Assist Radiol. Surg. 2022, 17, 1751–1764. [Google Scholar] [CrossRef] [PubMed]
- Diaz Tovar, J.S.; Dos Reis, F.P.; Moriyama, L.T.; Pêgo-Fernandes, P.M.; Nomura, C.H.; Yakovlev, V.; Bagnato, V.S.; Doronin, A.; Kurachi, C. Anatomy-resolved digital twin framework for personalized light dosimetry in lung therapies. Proc. Natl. Acad. Sci. U S A 2026, 123, e2612370123. [Google Scholar] [CrossRef] [PubMed]
- Deng, M.; Li, F.; Tang, F.; Chen, W.; Wang, F.; Tang, C.L.; Tong, R.; Yang, Z.; Xu, W.; Zhang, N.; Xia, Y.; Li, S.; Herth, F.J.F.; Hou, G. Digital twin-based bronchoscopy simulator improves training performance and skill retention of novices: a randomised controlled study. Thorax 2026, 81, 483–491. [Google Scholar] [CrossRef] [PubMed]
- Deng, M.; Xu, W.; Tang, F.; Chen, H.; Yang, Z.; Wang, F.; Zhang, N.; Wu, H.; Li, J.; Zheng, Z.; Wu, S.; Hou, G. Digital Twin-Assisted Risk Disclosure in Adults Undergoing Elective Bronchoscopy: Multicenter Randomized Controlled Trial. J. Med. Internet Res. 2026, 28, e81382. [Google Scholar] [CrossRef] [PubMed]
- ClinicalTrials.gov. Learning curves in digital-twin simulation–based bronchoscopy training (first 30 real-patient cases), NCT07718451. 2026. Available online: https://clinicaltrials.gov/study/NCT07718451 (accessed on 22 November 2026).
- Ang, C.Y.S.; Lee, J.W.W.; Chiew, Y.S.; Wang, X.; Tan, C.P.; Cove, M.E.; Nor, M.B.M.; Zhou, C.; Desaive, T.; Chase, J.G. Virtual patient framework for the testing of mechanical ventilation airway pressure and flow settings protocol. Comput Methods Programs BioMed 2022, 226, 107146. [Google Scholar] [CrossRef] [PubMed]
- Taki, A.; Murphy, B.; Richter, J.; Schwenck, N.; Grill, M.J.; Joyce, M.; Leime, C.O.; McNea, N.; MacLoughlin, R. Evaluation of aerosol delivery in ventilated patients via vibrating mesh nebuliser versus other delivery devices using a patient-specific in silico whole-lung model. Respir. Res. 2026, 27. [Google Scholar] [CrossRef] [PubMed]
- Kolanjiyil, A.V.; Dutta, R.; Maradiaga, A.; Grey, C.; Aladwani, G.; Dalton, C.; Momin, M.A.M.; Hindle, M.; Worth Longest, P. Validating complete-airway computational fluid dynamics (CA-CFD) predictions of regional aerosol deposition in human subjects: comparisons with in vivo data of pharmaceutical aerosol delivery during HFNC therapy. Int. J. Pharm. 2026, 707, 127605. [Google Scholar] [CrossRef] [PubMed]
- Montgomery, A.J.; Litell, J.; Dang, J.; Flurin, L.; Gajic, O.; Lal, A. Digital Twin Platform for education, research, and healthcare delivery investigator group (2023) Gaining consensus on expert rule statements for acute respiratory failure digital twin patient model in intensive care unit using a Delphi method. Biomol. BioMed 23, 1108–1117. [CrossRef] [PubMed]
- Gangemi, S.; Manti, S.; Virchow, J.C.; Canonica, G.W. Targeting Airway Remodeling in Severe Asthma: Is There a Window of Opportunity for Biologic Therapy Predicting Effects Using Causal Artificial Intelligence? Allergy 2026, 81, 3157–3170. [Google Scholar] [CrossRef] [PubMed]
- Fan, Q.; Li, Z.; Sun, Y.; Liu, Y.; Chen, H. Intelligence-enabled smart inhalable biomaterials for chronic inflammatory airway diseases: integrating responsive carrier engineering, lung-on-chip validation, and patient-specific pulmonary modeling. Front Bioeng. Biotechnol. 2026, 14, 1964825. [Google Scholar] [CrossRef] [PubMed]
- Rohren, F.; von Platen, P.; Leonhardt, S.; Walter, M. Modelling and Validation of a Decentralized Breathing Gas Source. Annu Int. Conf. IEEE Eng. Med. Biol. Soc. 2022, 2022, 625–628. [Google Scholar] [CrossRef] [PubMed]
- Gerretsen, E.C.F.; Chen, A.; Annema, J.T.; Groenier, M.; van der Heijden, E.H.F.M.; van Mook, W.N.K.A.; Smeenk, F.W.J.M. Effectiveness of Flexible Bronchoscopy Simulation-Based Training: A Systematic Review. Chest 2023, 164, 952–962. [Google Scholar] [CrossRef] [PubMed]
- Cold, K.M.; Konge, L. Simulation-Based Training in Flexible Bronchoscopy: Best Practices and Future Directions. Chest 2023, 164, 820–821. [Google Scholar] [CrossRef] [PubMed]
- Nan, Y.; Ser, J.D.; Tang, Z.; Tang, P.; Xing, X.; Fang, Y.; Herrera, F.; Pedrycz, W.; Walsh, S.; Yang, G. Fuzzy Attention Neural Network to Tackle Discontinuity in Airway Segmentation. IEEE Trans. Neural Netw. Learn Syst. 2024, 35, 7391–7404. [Google Scholar] [CrossRef] [PubMed]
- Matkovic, L.; Lei, Y.; Fu, Y.; Wang, T.; Kesarwala, A.H.; Axente, M.; Roper, J.; Higgins, K.; Bradley, J.D.; Liu, T.; Yang, X. Deformable lung 4DCT image registration via landmark-driven cycle network. Med. Phys. 2024, 51, 1974–1984. [Google Scholar] [CrossRef] [PubMed]
- Teng, X.; Chen, Y.; Zhang, Y.; Ren, L. Respiratory deformation registration in 4D-CT/cone beam CT using deep learning. Quant. Imaging Med. Surg. 2021, 11, 737–748. [Google Scholar] [CrossRef] [PubMed]
- McKivitt, O.; Roche, M.; Lawless, K.; Collins, P.; Walker, C.; O’Connell, A.; O’Donovan, T.; O’Cathail, S.M.; Codd, L.; McGrath, E.; Nobre Gois, I.; Murphy, J.; Barry, A. Implementation of surface-guided radiotherapy for motion management in liver SBRT: A scoping review and clinical workflow development. Tech Innov. Patient Support Radiat. Oncol. 2026, 38, 100400. [Google Scholar] [CrossRef] [PubMed]
- Grama, D.; Dahele, M.; van Rooij, W.; Slotman, B.; Gupta, D.K.; Verbakel, W.F.A.R. Deep learning-based markerless lung tumor tracking in stereotactic radiotherapy using Siamese networks. Med. Phys. 2023, 50, 6881–6893. [Google Scholar] [CrossRef] [PubMed]
- Keall, P.J.; Mageras, G.S.; Balter, J.M.; Emery, R.S.; Forster, K.M.; Jiang, S.B.; Kapatoes, J.M.; Low, D.A.; Murphy, M.J.; Murray, B.R.; Ramsey, C.R.; Van Herk, M.B.; Vedam, S.S.; Wong, J.W.; Yorke, E. The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med. Phys. 2006, 33, 3874–900. [Google Scholar] [CrossRef] [PubMed]
- Carrero-Cardenal, E.; Vollmer-Torrubiano, I.; Torres-López, M.; Martín-Barrera, G.; Casanovas-Mateu, G.; Tercero-Machin, F.J.; Paez-Carpio, A.; Fábregas-Julià, N.; Valero-Castell, R. Continuous positive airway pressure is unsafe for radiofrequency ablation of lung cancer under sedation: a randomised controlled trial. Insights Imaging 2024, 15, 153. [Google Scholar] [CrossRef] [PubMed]
- Shu, H.; Soberanis-Mukul, R.D.; Xu, J.; Ding, H.; Ringel, M.; Shen, M.; Sayed, S.I.; Rafii-Tari, H.; Unberath, M. BronchOpt: vision-based pose optimization with fine-tuned foundation models for accurate bronchoscopy navigation. Int. J. Comput Assist Radiol. Surg. 2026, 21, 1033–1041. [Google Scholar] [CrossRef] [PubMed]
- Lauzeral, N.; Borzacchiello, D.; Kugler, M.; George, D.; Rémond, Y.; Hostettler, A.; Chinesta, F. A model order reduction approach to create patient-specific mechanical models of human liver in computational medicine applications. Comput Methods Programs BioMed 2019, 170, 95–106. [Google Scholar] [CrossRef] [PubMed]
- Sebek, J.; Taeprasartsit, P.; Pathomparai, C.; Dupuy, D.E.; Wibowo, H.; Prakash, P. Computational modeling of microwave ablation of lung tumors: Assessment of model-predictions against post-treatment imaging. Med. Phys. 2025, 52, e17897. [Google Scholar] [CrossRef] [PubMed]
- Kim, D.H.; Chiang, J.; Chen, L.; Lamba, A.; Abtin, F.; Genshaft, S.; Quirk, M.; Suh, R. Patient-Specific Prediction of Immediate Phase Lung Microwave Ablation Zone Size. J. Vasc. Interv. Radiol. 2024, 35, 1635–1643. [Google Scholar] [CrossRef] [PubMed]
- Song, Z.; Li, Y.; Tian, Q.; Sun, C.; Liu, H.; Chong, K.; Zhang, Q.; Chen, J.; Li, P.; Song, L.; Tosi, D.; Kim, M.P.; Lin, Z.; Luo, Q.; Yu, L.; Cheng, X. Augmented reality guided versus computed tomography guided percutaneous lung nodule localization: a noninferiority randomized clinical trial. Int. J. Surg. 2025, 111, 2933–2941. [Google Scholar] [CrossRef] [PubMed]
- Yan, P.; Tong, A.N.; Nie, X.L.; Ma, M.G. Assessment of safety margin after microwave ablation of stage I NSCLC with three-dimensional reconstruction technique using CT imaging. BMC Med. Imaging 2021, 21, 96. [Google Scholar] [CrossRef] [PubMed]
- Mohn, F.; Pal, K.; O’Connor, C.; Paolucci, I.; Castelo, A.; Ahrar, K.; Yevich, S.; Kuban, J.D.; Odisio, B.C.; Tam, A.L.; Sheth, R.A.; Brock, K.K. Association between Three-dimensional Margin Assessment and Local Tumor Control in Percutaneous Lung Cryoablation of Colorectal Metastases. Radiol. Imaging Cancer 2026, 8, e250113. [Google Scholar] [CrossRef] [PubMed]
- Maas, M.; Beets-Tan, R.; Gaubert, J.Y.; Gomez Munoz, F.; Habert, P.; Klompenhouwer, L.G.; Vilares Morgado, P.; Schaefer, N.; Cornelis, F.H.; Solomon, S.B.; van der Reijd, D.; Bilbao, J.I. Follow-up after radiological intervention in oncology: ECIO-ESOI evidence and consensus-based recommendations for clinical practice. Insights Imaging 2020, 11, 83. [Google Scholar] [CrossRef] [PubMed]
Figure 1.
The tier-by-cadence classification of digital replicas in bronchoscopic intervention. Vertical axis: data-flow tier (Tier 1 static model; Tier 2 digital shadow; Tier 3 digital twin, per the hierarchy of the companion review). Horizontal axis: correction cadence (none; episodic, operator-initiated; continuous; predictive). Contemporary navigation platforms, confirmation tools, and the two randomized simulator trials are placed descriptively; the empty upper-right region is the agenda of this review.
Figure 1.
The tier-by-cadence classification of digital replicas in bronchoscopic intervention. Vertical axis: data-flow tier (Tier 1 static model; Tier 2 digital shadow; Tier 3 digital twin, per the hierarchy of the companion review). Horizontal axis: correction cadence (none; episodic, operator-initiated; continuous; predictive). Contemporary navigation platforms, confirmation tools, and the two randomized simulator trials are placed descriptively; the empty upper-right region is the agenda of this review.

Figure 2.
The four-layer technical stack of the bronchoscopy procedure twin, and the five-layer system architecture it implies. Layer 1, airway-tree and target modeling; Layer 2, motion and CT-to-body divergence modeling; Layer 3, scope–twin coupling and confirmation; Layer 4, closed-loop recalibration. Arrows indicate the bidirectional data flow required for Tier 3 function.
Figure 2.
The four-layer technical stack of the bronchoscopy procedure twin, and the five-layer system architecture it implies. Layer 1, airway-tree and target modeling; Layer 2, motion and CT-to-body divergence modeling; Layer 3, scope–twin coupling and confirmation; Layer 4, closed-loop recalibration. Arrows indicate the bidirectional data flow required for Tier 3 function.

Figure 3.
Research agenda on an indicative timeline. Near term (achievable with existing technology): P1 motion-envelope overlay, P2 registration-error dashboard, P3 scaled simulator credentialing. Mid term: P4 predictive targeting and digital margin rehearsal. Long term: P5 the longitudinal twin.
Figure 3.
Research agenda on an indicative timeline. Near term (achievable with existing technology): P1 motion-envelope overlay, P2 registration-error dashboard, P3 scaled simulator credentialing. Mid term: P4 predictive targeting and digital margin rehearsal. Long term: P5 the longitudinal twin.

Figure 4.
End-to-end proof-of-concept workflow. Pre-procedural four-dimensional CT builds the patient-specific model (Layer 1); motion mathematics populates the envelope (Layer 2); tracking and confirmation streams couple the procedure to the replica (Layer 3); the twin recalibrates continuously and issues predictive guidance (Layer 4); post-procedural imaging and histology feed validation (Section 6).
Figure 4.
End-to-end proof-of-concept workflow. Pre-procedural four-dimensional CT builds the patient-specific model (Layer 1); motion mathematics populates the envelope (Layer 2); tracking and confirmation streams couple the procedure to the replica (Layer 3); the twin recalibrates continuously and issues predictive guidance (Layer 4); post-procedural imaging and histology feed validation (Section 6).

Table 1.
Publications explicitly adopting the digital-twin concept in bronchoscopy and airway medicine (PubMed, 22 November 2026; complete inventory, N=9).
Table 1.
Publications explicitly adopting the digital-twin concept in bronchoscopy and airway medicine (PubMed, 22 November 2026; complete inventory, N=9).
| Application | Twin object | Tier | Enters the procedure room? | Models motion? |
|---|---|---|---|---|
| Bronchoscopy simulator training, randomized [39] | Procedure (scope+airway+operator) | 1–2 | No (training) | No |
| Simulator-based risk disclosure, multicenter RCT [40] | Procedure (patient communication) | 1 | No (consent) | No |
| Simulator learning-curve trial, ongoing [41] | Procedure (training) | 1–2 | No (training) | No |
| Ventilator-protocol virtual patient [42] | Lung physiology | 1 | No | No |
| Aerosol delivery modeling [43,44] | Airway deposition | 1 | No | Partial (flow dynamics) |
| Acute respiratory failure decision rules [45] | ICU physiology | 1 | No | No |
| Airway remodeling prediction in asthma [46] | Airway biology | 1 | No | No |
| Inhalable biomaterial design [47] | Drug–airway interaction | 1 | No | No |
| Breathing-gas source logistics [48] | Device infrastructure | 1 | No | No |
Table 2.
The four-layer stack: validated components, bronchoscopy-specific barriers, gaps, and success criteria.
Table 2.
The four-layer stack: validated components, bronchoscopy-specific barriers, gaps, and success criteria.
| Layer | Validated components (examples) | Bronchoscopy-specific barrier | Concrete gap | Success criterion |
|---|---|---|---|---|
| L1 Airway-tree & target modeling | Airway segmentation [51]; pathway planning [2]; bronchus-sign mapping [13] | Ground-glass margin uncertainty; scope-geometry passability unencoded | No planning system exploits traversability | Complete, traversable path to target in ≥95% of cases |
| L2 Motion & divergence modeling | 4D CT/DIR mathematics [52,53]; motion-model biopsy prototype [37]; surface surrogates [54] | Target motion 1–2 cm with hysteresis; CBCT captures one phase [24] | No platform carries a motion model into the procedure | Predicted target position within 5 mm at sampling |
| L3 Scope–twin coupling & confirmation | ssRAB/ENB tracking [22,23]; pose optimization [58]; tool-in-lesion CBCT [24] | Confirmation terminates at the operator’s eyes | Confirmation events not assimilated as observations | Confirmation stream logged and machine-readable in 100% of procedures |
| L4 Closed-loop recalibration | Model-order-reduction surrogates [59]; virtual-heart precedent [17] | No state estimation in current stack | Correction episodic, operator-initiated | Continuous recalibration with divergence alerts during the procedure |
Table 3.
Five-rung validation ladder for the bronchoscopy procedure twin.
| Rung | Design | Endpoint | Anchor precedent |
|---|---|---|---|
| 1 Phantom/simulation | Benchtop airway models under programmed motion | Targeting error vs static map | Simulator RCT methodology [39,50] |
| 2 Retrospective rehearsal | Twin replays archived procedures | Reproduces rEBUS view, tool-in-lesion, yield; must beat regression | Planning-model precedents [60,61] |
| 3 Accuracy gate | Prospective, pre-specified | Targeting error below contemporary plateau | Navigation cohorts [13,24] |
| 4 Prospective utility | Randomized, noninferiority-then-superiority | Diagnostic yield, procedure time, complications | AR localization RCT [62] |
| 5 Therapeutic extension | Treat-and-resect with histopathology | Predicted vs achieved margin | NCT05281237 [34]; 3D margin assessment [63,64] |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.