Abstract
Background. Bloodstream infection is a leading cause of death from antimicrobial resistance, and delays in effective therapy cost lives. Two classes of rapid test now report resistance infor-mation directly from positive blood cultures, hours to days before conventional subculture and susceptibility testing: multiplex molecular identification panels, which detect selected resistance genes, and rapid phenotypic antimicrobial susceptibility testing (AST) systems, which measure growth in the presence of antimicrobials. The two classes answer different questions. A detect-ed resistance gene is strong evidence of resistance, but a gene that is not detected does not es-tablish susceptibility, because resistance can arise through mechanisms a panel does not target. Phenotypic systems report susceptibility directly, but only for the antimicrobial and organism combinations their authorization covers. These boundaries are described in each device's FDA documentation, yet that documentation is spread across dozens of decision summaries written over nearly two decades in different formats. A laboratory choosing a rapid test, writing result comments, or verifying a new system under federal laboratory rules therefore has no single public source showing what each FDA-authorized test detects and what its FDA documenta-tion says a negative resistance result does and does not rule out. Methods. We conducted a cross-sectional analysis of public U.S. regulatory records. Device records were identified through the openFDA 510(k) interface (data current to 26 October 2026) in nine product codes covering positive blood culture identification panels, rapid AST systems and tests for S. aureus/methicillin-resistant S. aureus (MRSA), supplemented by a device-name search. Records were included if the test is performed directly on positive blood culture and reports at least one resistance gene, a resistance phenotype or antimicrobial susceptibility. Submissions for the same assay were grouped into device families, and one FDA document per family, normally the most recent decision summary restating the full indications for use, was abstracted. For each family we recorded organism and resistance targets as written, the organ-isms with which each resistance result is reported or attributed, every antimicrobial and organ-ism combination in the AST indications for use, specimen requirements, and labeled limitations transcribed verbatim and classified into 16 predefined types. Every value is traced to its sub-mission number and page. The primary outcome was whether each molecular or MRSA-screen document states that a negative resistance-marker result does not indicate susceptibility. As a pre-specified secondary comparison, documents dated before and from 2020 were compared with Fisher's exact test. Before release, all openFDA queries were re-run, every submission field was checked against live records, extracted data were reviewed against the source docu-ments, and all derived files were reproduced from a fresh copy. Results. Of 72 records screened, 40 submissions (38 510(k) clearances and 2 De Novo authoriza-tions, dated 2007-12-20 to 2026-09-24) from 24 device families were included: 12 multiplex mo-lecular panels, 5 molecular S. aureus/MRSA tests, 1 phenotypic MRSA screen and 6 rapid phe-notypic AST systems. Rapid phenotypic AST is recent: 10 of its 12 submissions were decided in 2024 to 2026. 16 documents tie use to a Gram-stain result. The documents named 217 organism targets and 90 resistance targets (83 genes and 7 qualitative phenotypes) in 28 marker families. Breadth ranged from 1 resistance target on single-organism S. aureus tests to 18 on the broadest Gram-negative panel. Mec genes were most widely represented (11 devices), followed by KPC and van genes (7 each) and CTX-M and NDM (6 each). Only 4 Gram-negative panels named all five carbapenemase families most often reported in the United States (KPC, NDM, VIM, IMP and OXA-48-like). 16 marker families appeared on a single device, 13 of them on two panels authorized in 2026. Attribution of resistance results varied: three documents report a gene only with an associated organism, four state that genes are not attributed in mixed growth, 9 warn that a detected gene may not belong to the causative organism, and for 6 panels the indications do not say which organisms a resistance result is reported with. Of 18 molecular and MRSA-screen documents, 9 stated that a negative resistance-marker result does not indicate suscepti-bility (one with hedged wording) and 9 did not (one only by indirect reference to the package insert). All 6 documents dated 2020 or later carried the statement, compared with 3 of 12 dated earlier (Fisher's exact p = 0.009). The documents without it, dated 2007-12-20 to 2019-06-28, in-clude all four tests limited to S. aureus/MRSA. 4 documents also warned that a negative result may reflect a pathogen the panel does not target. The 6 phenotypic AST systems listed 841 an-timicrobial and organism indications covering 32 antimicrobials. 7 agents were indicated on all six systems (amikacin, cefepime, ceftazidime, ertapenem, gentamicin, meropenem, piperacillin-tazobactam), whereas ceftazidime-avibactam was indicated on 5, meropenem-vaborbactam on 3 and ceftolozane-tazobactam on 2. Only two systems covered Gram-positive organisms. Of 292 normalized organism and antimicrobial pairs, 73 (25.0%) were indicated on only one system and 19 (6.5%) on all six. 5 systems required organism identification from a separate method, 5 stated that results are not valid for polymicrobial samples, 4 listed combinations requiring an al-ternate method before reporting, and 4 listed combinations whose ability to detect resistance is unknown because too few resistant isolates were studied. All 24 documents stated that subcul-ture of the positive blood culture is still required. Conclusions. FDA-authorized rapid blood culture resistance tests differ substantially in what they detect and in what their FDA documents say about a negative result. Documents dated be-fore 2020 often omit the warning that a negative resistance-marker result does not indicate sus-ceptibility, so laboratories using older tests should check their labeling and consider adding that warning to their own result comments. Resistance gene attribution is limited in mixed cultures, and rapid phenotypic AST coverage remains fragmented across systems, with a quarter of or-ganism and antimicrobial pairs available on only one platform. These findings describe FDA documents rather than current package inserts or test performance, and the era comparison is exploratory. The dataset, code, comparison tables and provenance log are openly released to support test selection, verification and result reporting, and will be refreshed as new devices are authorized.
Keywords:
rapid diagnostics
; blood culture identification panel
; rapid antimicrobial susceptibility testing
; FDA 510(k)
; De Novo
; test verification
; CLIA
; carbapenemase
; mecA
; diagnostic stewardship
I. Introduction
Bloodstream infection is a leading cause of death from antimicrobial resistance worldwide [1,2], and every hour of ineffective therapy in septic shock is associated with lower survival [3]. Conventional workup of a positive blood culture needs subculture, identification and isolate-based susceptibility testing, so definitive results usually arrive 48–72 hours after the bottle signals positive. Two classes of rapid test now shorten that interval. Multiplex molecular blood culture identification panels identify common organisms and selected resistance genes within about an hour, and rapid phenotypic AST systems measure growth directly from the positive bottle and report minimum inhibitory concentrations (MICs) within a working shift [4,5,6]. With antimicrobial stewardship, rapid testing shortens time to optimal therapy and, in meta-analysis, is associated with lower mortality [7,8].
The two classes answer different questions. A detected resistance gene is strong evidence that the organism carrying it is resistant. A gene that is not detected says only that the targeted sequence was absent; it does not establish susceptibility, because resistance can arise through mechanisms a panel does not target [9]. Phenotypic systems report susceptibility directly, but only for the antimicrobial–organism combinations their FDA authorization covers and usually only for monomicrobial cultures. Each of these boundaries is described in the device’s FDA documentation, but that documentation is spread across dozens of decision summaries written over nearly two decades in different formats.
This matters most where specialist support is thinnest. A community hospital laboratory selecting a rapid panel, writing result comments for a negative mecA or KPC call, or verifying a new phenotypic system under the Clinical Laboratory Improvement Amendments (CLIA) needs to know precisely what each FDA-authorized test detects, which organism each resistance result is attributed to, and what its FDA documentation says a negative result does and does not rule out [10,11]. Published comparisons of rapid tests report analytical or clinical performance from selected studies and date quickly as devices are added or modified [4,5]. To our knowledge, no open, machine-readable, source-traced dataset of FDA-authorized rapid blood culture resistance tests and their labeled limitations exists.
Here we describe the FDA-Cleared Rapid Resistance Test Landscape, version 1.0.0 [12] (Figure 1). It (i) enumerates FDA-cleared and De Novo-authorized tests that report resistance directly from positive blood culture; (ii) records, from FDA documents, the organism and resistance targets, antimicrobial–organism indications, specimen requirements and labeled limitations of each, with every value traced to its source page; and (iii) describes how the devices differ, with particular attention to what each FDA document says about negative resistance results. It complements the author’s earlier benchmark of gene–phenotype concordance for rapid panel targets [13] and crosswalk of reference isolates to those targets [14]. Section II describes the methods, Sec. III the results, Sec. IV their use and limits, and Sec. V our conclusions.
II. Materials and Methods
A. Data sources
This is a cross-sectional documentary analysis of public U.S. regulatory records (Table 1). The universe of devices was drawn from the openFDA 510(k) application programming interface, which also carries De Novo decisions, with data last updated 26 October 2026 [15]. We queried nine product codes chosen to capture every rapid blood culture resistance test: PAM (Gram-positive bacteria and their resistance markers in positive blood culture) and PEN (Gram-negative bacteria and associated resistance markers), both under 21 CFR 866.3365 [16]; PEO (yeast in positive blood culture); PRH (positive blood culture identification and AST kit), SAN (automated AST system for positive blood culture samples) and QZX (positive blood culture processor for AST inoculum), all under 21 CFR 866.1650 [17]; NQX (MRSA nucleic acid amplification tests); OUS (MRSA/MSSA blood culture test); and MYI (MRSA and mecA detection tests). Two further records were found by device-name search. The raw openFDA record for every included submission is archived with the dataset.
Device content was abstracted from FDA decision summaries [18], the reviewer’s account of each submission, published on accessdata.fda.gov. Decision summaries are U.S. government works and state the indications for use, specimen requirements, measurands and, in most cases, limitations relevant to result interpretation. Package inserts and published evaluations were not used, so every value traces to a single public regulatory source.
B. Eligibility
Records were included if the device is performed directly on positive blood culture (or on smears from it) and either reports at least one resistance gene or resistance phenotype, or reports antimicrobial susceptibility. Records were excluded if the specimen was not positive blood culture (nasal swabs, rectal swabs, isolated colonies), if the device reported identification only (yeast-only panels), or if it prepared samples without reporting results. Whole-blood, culture-independent assays were outside scope.
C. Device families and source documents
Submissions for the same assay were grouped into device families. For each family one source document was abstracted, normally the most recent decision summary that restated the full indications for use. Special 510(k)s that changed only software, instrument compatibility or labeling format were listed but not re-abstracted. Where no decision summary could be retrieved, the sponsor’s 510(k) summary was used and flagged (one family, Verigene BC-GP).
D. Data extraction
For each family we recorded device name, manufacturer, category, technology, organism coverage, whether the device itself identifies the organism, specimen wording and source document. Targets were recorded as written, with class (organism, pan target, resistance gene, resistance phenotype), page and, for resistance targets, the organisms with which the source says a result is reported or attributed. For phenotypic AST systems every antimicrobial–organism combination in the indications for use was recorded. Labeled limitations were transcribed verbatim and classified into 16 predefined types (codebook in the repository), among them statements that a negative resistance-marker result does not indicate susceptibility, that a negative result may reflect an untargeted pathogen, that polymicrobial samples are not valid, that an alternate method is needed before reporting certain combinations, and that the ability to detect resistance is unknown for certain combinations.
Text was read from the FDA documents with an automated document reader and transcribed with page references; submission metadata came from openFDA. Before release the author reviewed the extracted data against the source documents, including line-by-line checks of the antimicrobial–organism indications of three AST systems (375 rows). Corrections from that review are logged in the repository.
E. Normalization
Source wording is preserved in every file. For comparison only, a normalized label merges trivial variants (for example S. agalactiae and S. agalactiae (GBS)), and resistance targets are grouped into marker families so that combined assays (mecA/mecC) and separate assays (mecA, mecC) count together. OXA assays were assigned to a group (OXA-23, OXA-24/40, OXA-48-like, OXA-58) only where the source named one. In AST indications, species and their “group” or “complex” designations were merged, and genus entries that listed their species were split into those species.
F. Primary outcome
For each molecular or MRSA-screen device the source document was classified by whether it states that a negative (“Not Detected”) resistance-marker result does not indicate susceptibility: States; States (hedged), where the wording was weaker (“do not always indicate”); Indirect only, where the FDA review referred to a package-insert limitation without a statement in the indications or limitations; or None found. Before a document was classed None found it was searched for “negative”, “not detected” and “susceptib”. Phenotypic AST systems were not classified because they report susceptibility directly.
G. Statistical analysis
Analyses are descriptive. As a pre-specified secondary comparison, the proportion of documents carrying the negative-marker statement (States or States (hedged)) was compared between documents dated before and from 1 January 2020 with two-sided Fisher’s exact test. Coverage of antimicrobial–organism pairs was summarized as the number of AST systems indicated per normalized pair. Analyses used Python 3 with SciPy [19]; figures used matplotlib [20] and seaborn [21].
H. Reproducibility and quality control
A validation script checks referential integrity, submission-number format, dates, one source document per family, page references, duplicates and required fields. A build script regenerates every comparison table and a public, filterable comparison page from the curated files, and the scripts in the manuscript folder regenerate every number, table and figure in this paper. Before release all openFDA queries were re-run (no new in-scope clearances), every submission field was confirmed against live records, and all derived files were reproduced identically from a fresh copy.
I. Ethics
The study used only public regulatory records. No human participants or patient data were involved.
III. Results
A. Included devices
The nine product-code queries returned 70 records and the name search 2 more (Figure 1). Of these, 32 were excluded: 18 NQX assays for specimens other than blood culture (among them BD MAX StaphSR, for nasal swabs), 9 MYI isolate or colony tests, 2 yeast-only panels, 1 inoculum-preparation device and 2 isolate-based AST records. The 40 included submissions (33 traditional and 5 special 510(k)s, 2 De Novo authorizations) formed 24 device families (Table 2). 23 families were abstracted from FDA decision summaries and one from the sponsor’s 510(k) summary.
Decision dates spanned 2007-12-20 to 2026-09-24 (Figure 2). The earliest submissions were S. aureus/MRSA tests and the first multiplex panels. Rapid phenotypic AST is recent: the first system was authorized by De Novo in 2017, and 10 of the 12 phenotypic AST submissions were decided in 2024–2026. Overall, 15 of the 40 submissions (37.5
B. Specimen requirements
All 24 devices are performed on positive blood culture, but their FDA documents set different preconditions. 16 documents tie use to a Gram-stain result: molecular panels restricted to cultures containing Gram-positive or Gram-negative organisms, S. aureus/MRSA tests restricted to Gram-positive cocci, and all five externally identified AST systems restricted to Gram-negative bacilli or to monomicrobial cultures confirmed by Gram stain. Three further documents (BioFire BCID and BCID2, Accelerate PhenoTest BC) state that results are read together with the Gram stain, and mecA XpressFISH applies only to smears from cultures already identified as S. aureus by a companion FISH assay. Four documents set no Gram-stain condition. Three name specific blood culture bottle types (Verigene BC-S, Xpert MRSA/SA BC, KeyPath), and both Icubate assays exclude cultures with mixed Gram-stain results. These conditions decide which positive bottles a laboratory may test.
C. Organism and resistance targets
Across the 24 families the documents named 217 organism targets and 7 pan targets (66 distinct after normalization) and 90 resistance targets (83 genes and 7 qualitative resistance phenotypes; 40 distinct normalized targets in 28 marker families; Figure 3). Breadth ranged from 1 resistance target on single-organism S. aureus tests to 18 on QIAstat-Dx BCID GN Plus AMR. The mec family was the most widely represented (11 devices), followed by KPC and the van genes (7 each), CTX-M and NDM (6 each), a methicillin-resistance phenotype call (6), and IMP and VIM (5 each) (Table 3). mecC was named on 4 devices, separately or within a combined mecA/mecC assay.
Carbapenemase coverage was uneven. All five carbapenemase families most often reported in the United States (KPC, NDM, VIM, IMP, OXA-48-like) were named on 4 Gram-negative panels (BioFire BCID2, ePlex BCID-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN). Verigene BC-GN named four of these plus an OXA target whose group its indications do not specify; iC-GN named KPC and NDM; the original BioFire BCID named KPC only. Colistin-resistance genes appeared on 2 panels (mcr-1 on BCID2, MCR on LIAISON PLEX BCN). 16 of the 28 marker families appeared on a single device, and 13 of these came from the two QIAstat-Dx panels authorized in August and September 2026, which added extended-spectrum β-lactamase and AmpC genes (SHV, TEM, PER, VEB, GES, CMY/LAT/DHA/FOX), further carbapenemases (OXA-24/40, OXA-58, SPM), the 16S rRNA methyltransferase armA, and Gram-positive markers for linezolid (cfr), macrolide–lincosamide–streptogramin B (erm(A), erm(C)) and tetracycline (tet(K), tet(M)) resistance.
D. Attribution of resistance results
Documents differed in how they tie a resistance result to an organism. Three state that a resistance result is reported only when an associated organism is detected (BioFire BCID, ePlex BCID-GP, iC-GPC). Four state that in mixed growth the gene is not attributed to a specific organism (Verigene BC-GP and BC-S, iC-GN, LIAISON PLEX BCP). 9 state that the detected gene may not belong to the organism causing disease. For 6 panels, including all four with full carbapenemase coverage, the indications do not say which organisms a resistance result is reported with; three of these (Verigene BC-GN, ePlex BCID-GN, LIAISON PLEX BCN) instead say that subculture is needed to associate resistance genes with a specific organism.
E. What the FDA documents say about a negative result
Of the 18 molecular and MRSA-screen documents, 9 state that a negative resistance-marker result does not indicate susceptibility (Table 4, Figure 4). 8 use direct wording, typically that negative results “do not indicate susceptibility” because multiple mechanisms of resistance exist; Great Basin Staph ID/R says that negative mecA results “do not always indicate susceptibility”. The other 9 documents carry no such statement. For one of these (mecA XpressFISH) the FDA review notes that inability to detect methicillin resistance from mechanisms other than mecA is a limitation in the package insert, but the indications for use do not say so. For the remaining 8, no statement was found after a targeted search.
The statement tracked document date. All 6 documents dated 2020 or later carry it, compared with 3 of the 12 dated earlier (Fisher’s exact p = 0.009). The documents without it are dated 2007-12-20 to 2019-06-28 and include all four tests limited to S. aureus/MRSA (Xpert MRSA/SA BC, BD GeneOhm StaphSR, KeyPath, mecA XpressFISH) and all three Verigene assays. 4 documents, all dated 2018 or later, also say that a negative result may reflect a pathogen the panel does not target, and 4 say that a positive result does not exclude co-infection with an untargeted organism.
F. Rapid phenotypic AST systems
The 6 phenotypic AST systems listed 841 antimicrobial–organism indications covering 32 antimicrobials (Table 5, Figure 5). 7 agents were indicated on all six: amikacin, cefepime, ceftazidime, ertapenem, gentamicin, meropenem, piperacillin-tazobactam. Newer β-lactam/β-lactamase-inhibitor combinations were less consistently available: ceftazidime–avibactam on 5 systems, meropenem–vaborbactam on 3 and ceftolozane–tazobactam on 2. Only Accelerate PhenoTest BC and Selux with the PBC Separator included Gram-positive indications (vancomycin, daptomycin, linezolid, ceftaroline); the other four were Gram-negative only.
After normalization the indications covered 292 distinct organism–antimicrobial pairs. 73 (25.0 (Figure 6). Meropenem, for example, was indicated for E. coli, K. pneumoniae, P. mirabilis and P. aeruginosa on all six systems but for Morganella morganii on only two. Only Accelerate identifies the organism; the other 5 need identification from a separate FDA-cleared method. 5 systems state that results are not valid or should not be reported for polymicrobial samples; Accelerate instead directs subculture. 4 list combinations requiring an alternate method before reporting, and 4 list combinations whose ability to detect resistance is unknown because too few resistant isolates were studied.
G. Labeled limitations
Every one of the 24 documents states that subculture of the positive blood culture is still required, for susceptibility testing of untargeted organisms, polymicrobial samples, untargeted antimicrobials or epidemiological typing (Figure 7). Statement types otherwise split by technology: warnings about negative markers, attribution and untargeted pathogens appear only in molecular documents, and warnings about polymicrobial samples, alternate methods and unknown resistance detection only in AST documents.
IV. Discussion
A. Principal findings
This study assembled an open device-by-target dataset of FDA-authorized rapid blood culture resistance tests in which every target, indication and limitation traces to a page of an FDA document. Three findings stand out. Half of the FDA documents for molecular and MRSA-screen devices, all dated before 2020, do not state that a negative resistance-marker result does not indicate susceptibility, whereas every document from 2020 onward does. Resistance targets vary widely: only 4 Gram-negative panels name all five common carbapenemase families, and most of the newest gene targets appear on a single panel. And rapid phenotypic AST coverage is fragmented: a quarter of organism–antimicrobial pairs are indicated on only one of six systems, and most systems depend on separate identification and exclude polymicrobial samples.
B. Negative resistance results
The asymmetry between positive and negative gene results is well established: a detected mecA, van or carbapenemase gene reliably predicts resistance, but absence of a targeted gene does not establish susceptibility, because resistance can arise from untargeted genes, mutations, efflux, porin loss or altered expression [4,9]. The author’s reference-isolate crosswalk found carbapenem-resistant Enterobacterales and Acinetobacter that carry no panel carbapenemase, which is exactly the situation the warning covers [14]. The newer FDA documents say so plainly. The older ones, including those for single-target S. aureus/MRSA assays still in use, do not, at least in the decision summary. The pattern most likely reflects evolving review practice and the growth of multi-target panels rather than any difference in biology, and package inserts may carry the warning where the decision summary does not. A laboratory relying on an older document cannot assume the warning has reached clinicians and should consider adding it to its own result comments, for example for a mecA-not-detected S. aureus or a carbapenemase-negative Enterobacterales result.
C. Attribution and polymicrobial cultures
Genes detected on a multiplex panel are not physically linked to an organism. Several documents say so explicitly, and others say only that subculture is required to associate a gene with an organism. In a polymicrobial culture a detected carbapenemase or mecA may belong to any organism present, including one the panel does not identify. Similarly, five of six phenotypic AST systems are not valid for polymicrobial samples, and at least one FDA document (PhAST) notes that a small percentage of polymicrobial samples may escape detection by Gram stain and rapid identification. These boundaries argue for reporting rules that qualify resistance interpretations when more than one organism is detected.
D. Implications for laboratories
Under CLIA a laboratory introducing an FDA-cleared test must verify the manufacturer’s performance specifications for its own use and must report within the device’s indications [10,11,22]. The dataset supports this in four ways. It lists, for each device, the organisms and resistance targets in scope, so verification sets can be built from reference isolates covering each claimed target [14]. It shows which antimicrobial–organism combinations each AST system may report, and which need an alternate method or carry unknown resistance detection. It provides FDA wording that laboratories can adapt for result comments. And it records where FDA documentation is silent on negative results, so laboratories can supply the warning themselves. Because AST indications change with breakpoint updates and predetermined change control plans, the dataset should be read with the current package insert and the FDA Susceptibility Test Interpretive Criteria [23].
E. Practical recommendations
Laboratories using or selecting rapid blood culture resistance tests can: enumerate[leftmargin=1.4em,itemsep=1pt,topsep=2pt] check whether the FDA document and package insert of each test in use state that a negative resistance-marker result does not indicate susceptibility, and add that statement to result comments where they do not; report panel resistance genes with an explicit note when more than one organism is detected; confirm that each combination a rapid AST system reports is within its current indications, and route combinations flagged for an alternate method or with unknown resistance detection to reference testing; build Gram-stain, bottle-type and monomicrobial preconditions into reflex rules so that only eligible bottles are tested; and keep subculture and conventional AST for every positive culture, as all 24 FDA documents require. enumerate
F. Limitations
The dataset describes FDA documents, not current labeling: each family was abstracted from one document, other submissions were listed but not re-abstracted, and package inserts may contain statements absent from decision summaries. It records labeled claims and limitations, not analytical or clinical performance, and should not be read as a ranking. One family was abstracted from a sponsor 510(k) summary, and for two AST systems (ASTar, PhAST) the full lists of combinations needing an alternate method or with unknown resistance detection are summarized rather than fully transcribed. Text was read with an automated reader, and page numbers may be off by one in places. The era comparison is exploratory, rests on small numbers and is confounded by manufacturer and panel design. Commercial availability was not assessed, and several included devices are no longer marketed. Scope was limited to tests on positive blood culture; whole-blood and isolate-based tests were excluded.
G. Future work
Planned extensions are to add package-insert limitations alongside FDA wording, extend scope to culture-independent whole-blood assays, link each resistance target to gene–phenotype concordance in U.S. clinical isolates [13], and refresh the dataset as new devices are authorized.
V. Conclusions
FDA-authorized rapid blood culture resistance tests differ substantially in what they detect and in what their FDA documents say about a negative result. Documents dated before 2020 often omit the warning that a negative resistance-marker result does not indicate susceptibility, and rapid phenotypic AST coverage remains fragmented across systems. The dataset gives laboratories a single, source-traced reference for test selection, verification and result reporting, and will be refreshed as new devices are authorized.
Author Contributions
F.A.: conceptualization, methodology, data curation, formal analysis, validation, visualization, writing (original draft, review and editing).
Acknowledgments
The author thanks the U.S. Food and Drug Administration for publishing decision summaries and the openFDA data service.
Competing Interests: The author declares no competing interests and no relationship with FDA or any manufacturer of the devices studied. This work is not affiliated with or endorsed by FDA.
Use of AI Tools: Claude (Anthropic) assisted with reading FDA documents, code, data processing, figure generation and drafting of this manuscript. The author made the analytic decisions, reviewed the extracted data against the FDA documents, and takes responsibility for the content.
References
- Antimicrobial Resistance Collaborators. Global burden of bacterial antimicrobial resistance in 2019: a systematic analysis. The Lancet 2022, 399(10325), 629–655. [Google Scholar] [CrossRef] [PubMed]
- Centers for Disease Control and Prevention. Antibiotic Resistance Threats in the United States, 2019; U.S. Department of Health and Human Services, CDC: Atlanta, GA, 2019. Available online: https://www.cdc.gov/antimicrobial-resistance/data-research/threats/.
- Kumar, Anand; et al. Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit. Care Med. 2006, 34(6), 1589–1596. [Google Scholar] [CrossRef] [PubMed]
- Banerjee, Ritu; Humphries, Romney. Rapid Antimicrobial Susceptibility Testing Methods for Blood Cultures and Their Clinical Impact. Front. Med. 2021, 8, 635831. [Google Scholar] [CrossRef] [PubMed]
- Humphries, Romney M.; et al. CLSI Methods Development and Standardization Working Group Best Practices for Evaluation of Antimicrobial Susceptibility Tests. J. Clin. Microbiol. 2018, 56(4), e01934-17. [Google Scholar] [CrossRef] [PubMed]
- Pliakos, Elina Eleftheria; et al. The Cost-Effectiveness of Rapid Diagnostic Testing for the Diagnosis of Bloodstream Infections with or without Antimicrobial Stewardship. Clin. Microbiol. Rev. 2018, 31(3), e00095-17. [Google Scholar] [CrossRef] [PubMed]
- Timbrook, Tristan T.; et al. The Effect of Molecular Rapid Diagnostic Testing on Clinical Outcomes in Bloodstream Infections: A Systematic Review and Meta-analysis. Clin. Infect. Dis. 2017, 64(1), 15–23. [Google Scholar] [CrossRef] [PubMed]
- Banerjee, Ritu; et al. Randomized Trial of Rapid Multiplex Polymerase Chain Reaction–Based Blood Culture Identification and Susceptibility Testing. Clin. Infect. Dis. 2015, 61(7), 1071–1080. [Google Scholar] [CrossRef] [PubMed]
- Bush, Karen; Bradford, Patricia A. Epidemiology of β-Lactamase-Producing Pathogens. Clin. Microbiol. Rev. 2020, 33(2), e00047-19. [Google Scholar] [CrossRef] [PubMed]
- Code of Federal Regulations. Title 42, § 493.1253, Standard: Establishment and verification of performance specifications. Available online: https://www.ecfr.gov/current/title-42/section-493.1253.
- Clinical and Laboratory Standards Institute. Verification of Commercial Microbial Identification and Antimicrobial Susceptibility Testing Systems. CLSI guideline M52, 1st edition; Clinical and Laboratory Standards Institute: Wayne, PA, 2015. [Google Scholar]
- Abdulwahab, Fatima. FDA-Cleared Rapid Resistance Test Landscape, version 1.0.0. 2026. Available online: https://github.com/FatimaAbdulwahab/fda-rapid-resistance-test-landscape.
- Abdulwahab, Fatima. U.S. Rapid Panel Resistance Marker Concordance Benchmark, version 1.0.0. 2026. Available online: https://github.com/FatimaAbdulwahab/us-rapid-panel-amr-concordance.
- Abdulwahab, Fatima. CDC & FDA Antimicrobial Resistance Isolate Bank: Rapid Panel Target Crosswalk, version 1.0.0. 2026. Available online: https://github.com/FatimaAbdulwahab/ar-isolate-bank-panel-crosswalk.
- U.S. Food and Drug Administration. openFDA device 510(k) API. 26 October 2026. Available online: https://open.fda.gov/apis/device/510k/ (accessed on 6 November 2026).
- Code of Federal Regulations. Title 21, § 866.3365, Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures. Available online: https://www.ecfr.gov/current/title-21/section-866.3365.
- Code of Federal Regulations. Title 21, § 866.1650, A cellular analysis system for multiplexed antimicrobial susceptibility testing. Available online: https://www.ecfr.gov/current/title-21/section-866.1650.
- U.S. Food and Drug Administration. Premarket Notification 510(k) database and decision summaries Document URLs for all 24 device families are listed in the dataset provenance log. Available online: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm (accessed on 6 November 2026).
- Virtanen, Pauli; et al. SciPy 1.0: fundamental algorithms for scientific computing in Python. Nat. Methods 2020, 17(3), 261–272. [Google Scholar] [CrossRef] [PubMed]
- Hunter, John D. Matplotlib: A 2D Graphics Environment. Comput. Sci. Eng. 2007, 9(3), 90–95. [Google Scholar] [CrossRef]
- Waskom, Michael. seaborn: statistical data visualization. JOSS 2021, 6(60), 3021. [Google Scholar] [CrossRef]
- Clinical and Laboratory Standards Institute. Performance Standards for Antimicrobial Susceptibility Testing. CLSI supplement M100, 35th edition; Clinical and Laboratory Standards Institute: Wayne, PA, 2025. [Google Scholar]
- U.S. Food and Drug Administration. Antibacterial Susceptibility Test Interpretive Criteria. Available online: https://www.fda.gov/drugs/development-resources/antibacterial-susceptibility-test-interpretive-criteria (accessed on 6 November 2026).
Figure 1.
Study design. Records from nine openFDA product codes were screened against the scope criteria; included submissions were grouped into device families, each abstracted from one FDA document into an open, source-traced dataset.
Figure 1.
Study design. Records from nine openFDA product codes were screened against the scope criteria; included submissions were grouped into device families, each abstracted from one FDA document into an open, source-traced dataset.

Figure 2.
FDA submissions per device family, 2007–2026. Filled markers show the submission abstracted for each family; open markers, other submissions; diamonds, De Novo authorizations. Lines join submissions of the same family.
Figure 2.
FDA submissions per device family, 2007–2026. Filled markers show the submission abstracted for each family; open markers, other submissions; diamonds, De Novo authorizations. Lines join submissions of the same family.

Figure 3.
Resistance targets named in FDA documents, by marker family. Shaded cells mean the device’s source document names at least one target in the family; combined and separate assays are grouped (mecA, mecC and mecA/mecC under mec). Numbers in parentheses give devices per family. Phenotypic AST systems appear only where they report a qualitative resistance phenotype.
Figure 3.
Resistance targets named in FDA documents, by marker family. Shaded cells mean the device’s source document names at least one target in the family; combined and separate assays are grouped (mecA, mecC and mecA/mecC under mec). Numbers in parentheses give devices per family. Phenotypic AST systems appear only where they report a qualitative resistance phenotype.

Figure 4.
Negative resistance-marker language by date of the FDA document. Each point is one molecular or MRSA-screen device at the decision date of the document abstracted; the dashed line marks 1 January 2020.
Figure 4.
Negative resistance-marker language by date of the FDA document. Each point is one molecular or MRSA-screen device at the decision date of the document abstracted; the dashed line marks 1 January 2020.

Figure 5.
Antimicrobials on rapid phenotypic AST systems. Cells give the number of organism entries for which each agent is indicated; blank, not indicated. Rows are ordered by the number of systems carrying the agent.
Figure 5.
Antimicrobials on rapid phenotypic AST systems. Cells give the number of organism entries for which each agent is indicated; blank, not indicated. Rows are ordered by the number of systems carrying the agent.

Figure 6.
Coverage of the 292 normalized organism–antimicrobial pairs across the six rapid phenotypic AST systems.
Figure 6.
Coverage of the 292 normalized organism–antimicrobial pairs across the six rapid phenotypic AST systems.

Figure 7.
Labeled-limitation statements in each FDA document. Shaded cells show that the statement type was transcribed from the device’s source document. The negative-marker row was confirmed by a targeted search of each document; for other rows an unshaded cell means only that the statement was not among those transcribed.
Figure 7.
Labeled-limitation statements in each FDA document. Shaded cells show that the statement type was transcribed from the device’s source document. The negative-marker row was confirmed by a targeted search of each document; for other rows an unshaded cell means only that the statement was not among those transcribed.

Table 1.
Data sources.
| Source | Content used |
| openFDA 510(k) API | Submission metadata for 9 product codes (data last updated 26 October 2026); raw record archived for every included submission |
| FDA decision summaries | 23 decision summaries: targets, indications, specimen wording, limitations, with page numbers |
| FDA 510(k) summary | 1 sponsor-prepared summary (Verigene BC-GP) where the decision summary was not retrievable |
| Dataset v1.0.0 | Curated CSV files, codebook, provenance log and validation scripts (CC BY 4.0 / MIT) |
Table 2.
FDA-authorized rapid blood culture resistance tests in the dataset. Source: FDA document abstracted, with its decision date. Subs: FDA submissions in the family. ID: organism and pan targets. Res.: resistance genes and phenotypes. Drugs: antimicrobials in the AST indications for use. Manufacturer as named in the source, later owner in parentheses. Commercial availability was not assessed.
Table 2.
FDA-authorized rapid blood culture resistance tests in the dataset. Source: FDA document abstracted, with its decision date. Subs: FDA submissions in the family. ID: organism and pan targets. Res.: resistance genes and phenotypes. Drugs: antimicrobials in the AST indications for use. Manufacturer as named in the source, later owner in parentheses. Commercial availability was not assessed.
| Device | Manufacturer | Category | Source | Decision | Subs | ID | Res. | Drugs |
| Verigene BC-GP | Nanosphere (now Luminex) | Molecular panel | K122514 | 2012-09-19 | 2 | 12 | 3 | – |
| BioFire BCID | BioFire Diagnostics | Molecular panel | K130914 | 2013-06-21 | 4 | 24 | 4 | – |
| Verigene BC-GN | Nanosphere (now Luminex) | Molecular panel | K132843 | 2014-01-08 | 1 | 8 | 6 | – |
| iC-GPC | Icubate | Molecular panel | K163390 | 2017-08-08 | 1 | 5 | 3 | – |
| ePlex BCID-GP | GenMark Diagnostics (now Roche) | Molecular panel | K181663 | 2018-12-20 | 1 | 22 | 4 | – |
| iC-GN | Icubate | Molecular panel | K190341 | 2019-06-28 | 1 | 8 | 3 | – |
| BioFire BCID2 | BioFire Diagnostics | Molecular panel | K193519 | 2020-03-18 | 2 | 33 | 10 | – |
| ePlex BCID-GN | GenMark Diagnostics (now Roche) | Molecular panel | K213236 | 2022-04-27 | 2 | 23 | 7 | – |
| LIAISON PLEX BCN | Luminex | Molecular panel | K243013 | 2025-04-18 | 1 | 18 | 8 | – |
| LIAISON PLEX BCP | Luminex | Molecular panel | K243490 | 2025-06-06 | 1 | 13 | 3 | – |
| QIAstat-Dx BCID GPF | QIAGEN | Molecular panel | K254194 | 2026-08-06 | 1 | 20 | 10 | – |
| QIAstat-Dx BCID GN | QIAGEN | Molecular panel | K262168 | 2026-09-24 | 1 | 13 | 18 | – |
| BD GeneOhm StaphSR | BD Diagnostics (GeneOhm Sciences) | Molecular S. aureus/MRSA | K071026 | 2007-12-20 | 1 | 1 | 1 | – |
| Verigene BC-S | Nanosphere (now Luminex) | Molecular S. aureus/MRSA | K112424 | 2011-12-16 | 1 | 2 | 1 | – |
| mecA XpressFISH | AdvanDx | Molecular S. aureus/MRSA | K140619 | 2014-10-10 | 1 | 1 | 2 | – |
| Great Basin Staph ID/R | Great Basin Scientific | Molecular S. aureus/MRSA | K152470 | 2016-03-25 | 1 | 3 | 1 | – |
| Xpert MRSA/SA BC | Cepheid | Molecular S. aureus/MRSA | K190771 | 2019-04-25 | 4 | 1 | 2 | – |
| KeyPath MRSA/MSSA | MicroPhage | Phenotypic MRSA screen | K120563 | 2012-03-30 | 2 | 1 | 1 | – |
| Accelerate PhenoTest BC | Accelerate Diagnostics | Phenotypic AST + ID | K192665 | 2020-09-15 | 2 | 16 | 1 | 17 |
| Selux (PBC Separator) | Selux Diagnostics | Phenotypic AST | K244044 | 2025-03-28 | 2 | – | 1 | 22 |
| LifeScale LSGN | Affinity Biosensors | Phenotypic AST | K251875 | 2025-10-27 | 3 | – | – | 14 |
| ASTar BC G- | Q-linea | Phenotypic AST | K253573 | 2026-04-15 | 2 | – | – | 23 |
| PhAST BC GN | Phast Corp. | Phenotypic AST | K253316 | 2026-06-26 | 1 | – | – | 15 |
| VITEK REVEAL GN | Specific Diagnostics / bioMerieux | Phenotypic AST | K260160 | 2026-07-08 | 2 | – | 1 | 22 |
Table 3.
Resistance marker families named on two or more devices.
| Marker family | n | Devices |
| mec (mecA/mecC) | 11 | BioFire BCID, BioFire BCID2, ePlex BCID-GP, Great Basin Staph ID/R, iC-GPC, LIAISON PLEX BCP, mecA XpressFISH, QIAstat-Dx BCID GPF, Verigene BC-GP, Verigene BC-S, Xpert MRSA/SA BC |
| blaKPC | 7 | BioFire BCID, BioFire BCID2, ePlex BCID-GN, iC-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN, Verigene BC-GN |
| van (vanA/vanB) | 7 | BioFire BCID, BioFire BCID2, ePlex BCID-GP, iC-GPC, LIAISON PLEX BCP, QIAstat-Dx BCID GPF, Verigene BC-GP |
| blaCTX-M | 6 | BioFire BCID2, ePlex BCID-GN, iC-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN, Verigene BC-GN |
| methicillin resistance (phenotype) | 6 | Accelerate PhenoTest BC, BD GeneOhm StaphSR, KeyPath MRSA/MSSA, mecA XpressFISH, Selux (PBC Separator), Xpert MRSA/SA BC |
| blaNDM | 6 | BioFire BCID2, ePlex BCID-GN, iC-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN, Verigene BC-GN |
| blaIMP | 5 | BioFire BCID2, ePlex BCID-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN, Verigene BC-GN |
| blaVIM | 5 | BioFire BCID2, ePlex BCID-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN, Verigene BC-GN |
| blaOXA-48-like | 4 | BioFire BCID2, ePlex BCID-GN, LIAISON PLEX BCN, QIAstat-Dx BCID GN |
| aac(6’) | 2 | QIAstat-Dx BCID GN, QIAstat-Dx BCID GPF |
| mcr | 2 | BioFire BCID2, LIAISON PLEX BCN |
| blaOXA-23 | 2 | ePlex BCID-GN, QIAstat-Dx BCID GN |
Table 4.
Negative resistance-marker language in the FDA document of each molecular and MRSA-screen device. Status categories are defined in Sec. II. Wording is reproduced from the FDA document; “...” marks omitted text and square brackets curator additions. A dash means no statement was found in the document reviewed; current package inserts may differ.
Table 4.
Negative resistance-marker language in the FDA document of each molecular and MRSA-screen device. Status categories are defined in Sec. II. Wording is reproduced from the FDA document; “...” marks omitted text and square brackets curator additions. A dash means no statement was found in the document reviewed; current package inserts may differ.
| Device | Document | Status | FDA wording (page) |
| BD GeneOhm StaphSR | 2007-12-20 | None found | – |
| Verigene BC-S | 2011-12-16 | None found | – |
| KeyPath MRSA/MSSA | 2012-03-30 | None found | – |
| Verigene BC-GP | 2012-09-19 | None found | – |
| BioFire BCID | 2013-06-21 | States | “Negative results for these select antimicrobial resistance gene assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, and carbapenems exist.” (IFU (section H)) |
| Verigene BC-GN | 2014-01-08 | None found | – |
| mecA XpressFISH | 2014-10-10 | Indirect only | “... to detect methicillin resistance conferred by mechanisms other than the presence of the mecA gene ... [is] reflected as a Limitation in the Package Insert.” (p.14) |
| Great Basin Staph ID/R | 2016-03-25 | States (hedged) | “Negative results for mecA antimicrobial resistance gene assays do not always indicate susceptibility, as other mechanisms of resistance to methicillin exist.” (p.2) |
| iC-GPC | 2017-08-08 | None found | – |
| ePlex BCID-GP | 2018-12-20 | States | “Antimicrobial resistance can occur via multiple mechanisms. A “Not Detected” result for the BCID-GP antimicrobial resistance gene assays does not indicate antimicrobial susceptibility.” (pp.3-4) |
| Xpert MRSA/SA BC | 2019-04-25 | None found | – |
| iC-GN | 2019-06-28 | None found | – |
| BioFire BCID2 | 2020-03-18 | States | “Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of resistance to methicillin, vancomycin, βlactams, and colistin exist.” (p.4) |
| ePlex BCID-GN | 2022-04-27 | States | “Negative results for these select antimicrobial resistance assays do not indicate susceptibility, as there are multiple mechanisms of resistance in gram-negative bacteria.” (p.2) |
| LIAISON PLEX BCN | 2025-04-18 | States | “Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of resistance to βlactams and colistin exist.” (p.4) |
| LIAISON PLEX BCP | 2025-06-06 | States | “Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.” (p.4) |
| QIAstat-Dx BCID GPF | 2026-08-06 | States | “Negative results for certain antimicrobial resistance assays do not indicate susceptibility, as multiple mechanisms of resistance exist.” (p.5) |
| QIAstat-Dx BCID GN | 2026-09-24 | States | “Negative results for certain antimicrobial resistance assays do not indicate susceptibility as multiple mechanisms of resistance exist.” (p.4) |
Table 5.
Rapid phenotypic AST systems. ID: system identifies the organism. Rows: antimicrobial × organism entries in the indications for use. Poly. invalid: results not valid for polymicrobial samples (“Subculture”: subculture directed, invalidity not stated). Alt. method: combinations requiring an alternate method before reporting. Res.\ unknown: combinations whose ability to detect resistance is unknown.
Table 5.
Rapid phenotypic AST systems. ID: system identifies the organism. Rows: antimicrobial × organism entries in the indications for use. Poly. invalid: results not valid for polymicrobial samples (“Subculture”: subculture directed, invalidity not stated). Alt. method: combinations requiring an alternate method before reporting. Res.\ unknown: combinations whose ability to detect resistance is unknown.
| System | ID | Scope | Drugs | Rows | Poly. invalid | Alt. method | Res. unknown |
| Accelerate PhenoTest BC | Yes | GP+GN | 17 | 95 | Subculture | Yes | – |
| Selux (PBC Separator) | No | GP+GN | 22 | 131 | Yes | – | Yes |
| LifeScale LSGN | No | GN | 14 | 134 | Yes | Yes | Yes |
| ASTar BC G- | No | GN | 23 | 215 | Yes | Yes | Yes |
| PhAST BC GN | No | GN | 15 | 117 | Yes | Yes | Yes |
| VITEK REVEAL GN | No | GN | 22 | 149 | Yes | – | – |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |