An Exploratory In Vitro Pilot Study of Connector Failure and Censored Load Tolerance in Cast and Laser Melted Cobalt Chromium Frameworks

Submitted:

30 October 2026

Posted:

03 November 2026

You are already at the latest version

Abstract
A load ceiling can obscure manufacturing-related differences in cobalt–chromium framework testing. This exploratory in vitro pilot analyzed ten frameworks: five cast and five fabricated by selective laser melting (SLM). The documented protocol used a hemispherical indenter at 1 mm/min after a 100 N preload. Cast specimens fractured at 938.53–1134.63 N (mean 1012.47 ± 77.40 N); SLM specimens remained macroscopically intact at 4498.36–4501.34 N and were right-censored. At the common observed limit of 4498.36 N, the capped mean contrast was 3485.89 N, with a 4.44 ratio and a two-sided exact permutation p-value of 0.00794. Single-specimen deletion changed the contrast by at most 0.876%. Exact 95% failure-probability intervals were 47.8–100% for casting and 0–52.2% for SLM. Bayesian direction favored SLM across three symmetric priors, while magnitude remained prior-sensitive. With five reported displacements per workflow, terminal displacement was 1.459 ± 0.357 mm for casting and 2.048 ± 0.134 mm for SLM. The five calculated SLM terminal secants averaged 2155.16 ± 139.28 N/mm; different endpoint forces preclude a matched-force stiffness comparison. The findings support independent validation; censored outcomes, unavailable raw traces, and unknown batch structure limit population and causal interpretation.
Keywords: 
;  ;  ;  ;  ;  ;  ;  

1. Introduction

The performance of a cobalt–chromium (Co–Cr) fixed dental prostheses (FDP) is a structural outcome: alloy properties, connector geometry, processing history, and the supporting assembly jointly determine its response to load. Contemporary comparisons have documented manufacturing-dependent differences in hardness, strength, grain structure, and porosity among cast, milled, and additively produced dental Co–Cr alloys [1,2]. A load measured on a complete framework therefore cannot be interpreted as an intrinsic alloy strength without specifying the geometry and boundary conditions. Framework testing remains useful precisely because it interrogates the manufactured component rather than an idealized material coupon.
Selective laser melting (SLM), a laser powder bed fusion process, changes the thermal history through localized melting and rapid solidification. A controlled comparison of cast and SLM Co–Cr specimens linked differences in bend-fatigue behavior with their microstructures [3]. Residual stresses, phase constitution, strength, and ductility can subsequently respond to thermal treatment [4,5]. Such evidence establishes plausible pathways by which fabrication affects performance, but the direction and magnitude of those pathways depend on the actual alloy, processing parameters, and postprocessing sequence.
Geometry creates a second, interacting pathway. Experiments modifying the internal architecture of multiunit Co–Cr frameworks have shown that changes in mass need not map directly onto changes in measured flexural response [6]. Recent experimental and finite element work on metal-ceramic FDPs identified an association between pontic volume, connector stress, and fracture resistance [7]. These findings make connector failure a relevant outcome and show why a nominally common digital design must be distinguished from verified as-built dimensions. A study of one design can compare its recorded outcomes; it cannot identify a span–resistance relationship.
Evidence from ceramic CAD/CAM restorations also shows why geometry requires material-specific assessment. Hajaj et al. found that cervical margin design affected fracture performance in lithium disilicate crowns but not in zirconia crowns under their test conditions [8]. A posterior zirconia single-retainer case report illustrates a different approach to preparation and supplementary support [9]. These findings motivate explicit reporting of preparation and support geometry; they do not establish fracture-load thresholds for Co–Cr frameworks.
Manufacturing can affect dimensional adaptation and material response. Layer thickness and build orientation have affected the fit of laser-produced multiunit Co–Cr frameworks [10,11]. Orientation and thermal treatment can interact with fit and hardness, while manufacturing direction influences the deformation and tensile response of heat-treated specimens [12,13]. Studies of posterior and anterior FDPs report workflow-dependent adaptation [14,15], and clinical measurements show that marginal and internal fit may favor different workflows [16]. These distinctions matter mechanically because seating and support contact help define how a framework transmits the applied force.
Digital design transfer also requires a clear distinction between workflow feasibility and quantitative validation. Titihazan et al. described printed and milled preparation guides as a qualitative technical feasibility study, without a statistical comparison of dimensional accuracy [17]. For this manufacturing question, a common CAD design remains the intended geometry; equivalence between the manufactured connectors and their seating requires direct measurement.
The completed restoration introduces further processing and additional failure endpoints. Annealing and porcelain firing have altered adaptation in additive workflows [18], while firing studies of cast and milled implant-cemented frameworks document dimensional and microstructural changes [19,20]. A framework-only test therefore addresses a narrower question than veneer adhesion, fatigue durability, or clinical survival. That narrower question becomes statistically challenging when a framework remains intact at the maximum permitted test load: the terminal reading indicates a lower bound on fracture resistance rather than a measured fracture load.
The present exploratory in vitro pilot study analyzes endpoint data from ten Co–Cr FDP frameworks assigned to casting or SLM to characterize initial mechanical observations and inform independent validation. Its primary analytical objective was to compare load tolerance over a common, fully observed force range while retaining the distinction between fractures and intact specimens stopped at the test limit. Secondary objectives were to quantify specimen-level identification bounds, examine the influence of individual observations and the selected force limit, and assess uncertainty in failure probability using exact and explicitly prior-sensitive Bayesian analyses. The documented experiment used one nominal digital design. An analytical beam illustration identified geometric and compliance measurements needed for a subsequent validation experiment; it was not fitted to the specimens.

2. Materials and Methods

2.1. Pilot Study Design and Provenance of the Experimental Data

This exploratory in vitro pilot study reports a secondary endpoint analysis of ten Co–Cr FDP frameworks: five cast and five produced by SLM. The laboratory endpoint tables, experimental descriptions, and photographs document MELT_1–MELT_5 and SINT_1–SINT_5. The documented testing site was the Faculty of Mechanical Engineering, Politehnica University Timișoara, Romania. Original TestXpert exports, manufacturing logs, dimensional metrology, and independent-batch records were unavailable.
The experimental description concerned a posterior maxillary segment with an edentulous space bounded by two abutments – 14 and 17. Both workflows used a common nominal digital framework design; one design and one nominal span were documented. Each framework was the unit of observation. We did not count connectors, photographs, sampled curve points, or the 25 possible cross-workflow specimen pairs as independent replicates. The authors did not document allocation randomization, blinded assessment, or randomized testing order. The number of independently manufactured batches was also unknown. Table 1 distinguishes reported experimental conditions from unverified variables.
The pilot sample comprised all ten available force/event records, with five specimens per workflow. An a priori sample-size calculation was not documented. The pilot designation reflects the analysis's exploratory scope and its role in identifying endpoint, measurement, and reporting requirements for independent validation. Prospective power calculations for a future prespecified binary endpoint were performed separately and are reported first in Results (Section 3.1), with methods in Section 2.7 and Supplementary Methods S4. They do not establish the adequacy of the observed pilot sample for a confirmatory comparison.

2.2. Fabrication Workflows and Exposure Definition

For casting, the working models were scanned and digital files were prepared in a workflow reported to use hyperDENT. The source did not separate its CAD and CAM functions or provide software versions. Wax patterns were milled using an imes-icore CORiTEC 250i dry unit, checked on the model, invested in refractory material, and subjected to wax elimination. Co–Cr alloy was cast into the molds. After cooling and recovery, frameworks were finished, airborne-particle abraded, and checked for seating. The commercial alloy, investment product, temperature history, and finishing parameters were not recoverable from the supplied record.
The SLM workflow used the same nominal digital design and Co–Cr powder. Frameworks were produced layer by layer, removed from the build plate, finished, airborne-particle abraded, and checked for seating and interference. The record did not identify the printer, powder lot or reuse history, build orientation, layer thickness, laser power, scan velocity, hatch spacing, support strategy, or heat-treatment schedule. No porcelain veneer was described on the tested frameworks. Visual seating checks establish the intended assembly procedure but do not demonstrate equivalent connector dimensions or quantify adaptation.
The exposure was therefore the recorded fabrication workflow, casting versus SLM. It was not an isolated intervention on manufacturing technology with chemically identical alloys and verified equivalent final geometry. This definition prevents an alloy-composition difference, dimensional discrepancy, or postprocessing effect from being automatically attributed to laser melting. No undocumented parameter was filled in using typical settings from another laboratory or manufacturer.

2.3. Mechanical Loading and Stopping Rule

The documented protocol positioned frameworks on a common experimental support to reproduce the relationship between the two abutments. The protocol did not document the support material, dimensions, compliance, or any cementation or retention procedure. The photographs show the seating and indenter configuration for the photographed subset (Figure 1). The report described loading as perpendicular and applied at the same nominal contact location, but it did not provide coordinates or contact-position tolerances. The nominal design, fixture, and contact arrangement were reported as common, rather than verified through specimen-level metrology or complete logs.
The laboratory documentation specified a Zwick/Roell Z005 universal testing machine, a 5 kN load cell with accuracy class 0.5, and TestXpert software under displacement control at ambient temperature. The test applied a 100 N preload at 10 mm/min, then loaded at 1 mm/min using a 10 mm hemispherical indenter. The documented protocol recorded force and crosshead displacement. Cast records ended at structural fracture; SLM records ended near the nominal administrative limit of 4501.34 N without reported macroscopic fracture. The load-cell class was retained as an equipment specification and was not converted into specimen-specific uncertainty without calibration records.

2.4. Event Definition and Endpoint Extraction

The event of interest was macroscopic structural fracture. Connector location after deformation was documented for MELT_1–MELT_5. For specimen i, Tᵢ denotes its latent load at that event, Cᵢ the stopping load, Xᵢ the observed endpoint, and δᵢ the event indicator. Equation (1) expresses the observation mechanism. An intact framework provides Tᵢ > Cᵢ; it does not provide Tᵢ = Cᵢ. The five SLM readings were 4500.07, 4499.41, 4499.62, 4498.36, and 4501.34 N with δᵢ = 0. All five cast readings had δᵢ = 1. Every available force/event record was included.
X i = min ( T i ,   C i ) ,      δ i = 1 { T i   ≤   C i }
Endpoint crosshead displacement was available for all five specimens per workflow. All ten reported values were included without imputation. Cast displacement occurred at fracture, whereas SLM displacement occurred at the stopping limit. The three source-reported SLM values labeled stiffness matched force above the 100 N preload divided by endpoint displacement within rounding precision. Terminal secant ratios were therefore calculated uniformly from the force and displacement of every specimen, before rounding. These are calculated system-response ratios, distinct from independently recorded elastic slopes. Displacement zero, permanent set, local strain, flexural stress, yield load, and integrated energy absorption were not recoverable from endpoint records.
Figure 1. Available photographic documentation: (a) laboratory model; (b) loading arrangement; (c) lateral view of indenter contact; (d) grouped frameworks after testing in the photographed six-specimen subset. Panels are reproduced from the experimental record and are not scaled. They illustrate the documented setup and specimen appearance; photographs of MELT_4, MELT_5, SINT_4, and SINT_5 were unavailable. These panels do not constitute microscopic fracture analysis.
Figure 1. Available photographic documentation: (a) laboratory model; (b) loading arrangement; (c) lateral view of indenter contact; (d) grouped frameworks after testing in the photographed six-specimen subset. Panels are reproduced from the experimental record and are not scaled. They illustrate the documented setup and specimen appearance; photographs of MELT_4, MELT_5, SINT_4, and SINT_5 were unavailable. These panels do not constitute microscopic fracture analysis.
Preprints 241565 g001

2.5. Statistical Analysis

2.5.1. Descriptive Summaries and the Restricted Estimand

The pilot analysis was developed after inspection of the record and was not preregistered. Individual reported values are presented with means, sample standard deviations (denominator n − 1), medians, ranges, and sample coefficients of variation where the endpoint is observed. Force, event, displacement, and calculated terminal-secant summaries use five specimens per workflow; the three source-reported rounded SLM secants are retained separately for arithmetic verification. Means and variability of calculated secants use the unrounded specimen ratios. No measurement was imputed and no normality screen was used to justify a large-sample test. We did not treat attained SLM loads as a fracture-strength distribution; their proximity to 4500 N reflects the stopping rule. Displacement and secant summaries are descriptive because workflows ended at different forces and event states.
We defined the common truncation load τ as the smallest attained load in the pooled set of tests that ended without fracture: 4498.36 N. All five such tests belonged to SLM, but the selection rule used event records rather than reassigned workflow labels. Every specimen was observed through either fracture or τ, so no censoring occurred before this common limit. Capped load tolerance Yᵢ(τ) and its group means are defined in Equation (2). The primary contrast Δ̂τ is SLM minus cast; the capped mean ratio was also reported. This is an analysis over an empirically supported force range, not a claim that τ is a clinically validated limit.
Y i ( τ ) = min ( T i ,   τ )
Δ ^ τ = Y ¯ SLM ( τ ) − Y ¯ Cast ( τ ) ,      R ^ τ = Y ¯ SLM ( τ ) Y ¯ Cast ( τ )
The group mean capped load equals the area under the empirical fracture-free proportion curve from zero to τ, as expressed in Equation (3). This identity makes the mechanical interpretation explicit: the estimand measures load tolerance accumulated over the observed force range. Its axis is applied force in N, not follow-up time. Since there is no earlier censoring, the area can be calculated directly from the ten capped endpoints without estimating an unobserved tail. The SLM capped mean equals τ by construction; a zero sample variance of its capped values would not imply a zero variance of true fracture loads.
μ ^ g ( τ ) = 1 n g ∑ i = 1 n g Y gi ( τ ) = ∫ 0 τ S ^ g ( f )   df

2.5.2. Exact Inference and Frequentist Uncertainty

The two-sided permutation calculation enumerated all 252 ways to assign ten capped endpoints to groups of five. For each allocation, the difference in group means was calculated. Equation (4) gives the fraction whose absolute contrast was at least as large as the observed value. Enumeration removes reliance on an asymptotic reference distribution, but does not remove the need for independent, exchangeable specimens under the null. In the absence of documented randomization and batch replication, this is a conditional exploratory analysis of group separation within the pilot, not a randomization-based causal estimate.
p exact = 1 252 ∑ A :   | A | = 5 1 {   | Δ ^ A |   ≥   | Δ ^ obs |   }
Failure by τ was additionally treated as a binary outcome. Group-specific equal-tailed 95% Clopper–Pearson confidence intervals were reported, and Fisher’s exact test was used only as a consistency check on the same separation. For the failure-risk difference θCast − θSLM, a conservative interval with at least 95% joint coverage at a fixed threshold was constructed by combining two 97.5% Clopper–Pearson intervals through the Bonferroni inequality. Its limits are the cast lower limit minus the SLM upper limit and the cast upper limit minus the SLM lower limit. This construction is deliberately conservative and does not provide selection-adjusted coverage for choosing τ after examining the experiment. The failure-risk difference has the opposite orientation to the capped-load contrast because fewer fractures favor SLM.
Complete cross-group ordering and Cliff’s delta were retained as descriptive summaries. No degenerate bootstrap interval from the completely separated observations was used to suggest precision. No unrestricted SLM mean or variance, finite hazard ratio, Weibull modulus, or extrapolated failure percentile was fitted. The absence of SLM fractures supports lower bounds but leaves its distribution beyond the stopping loads unidentified.

2.5.3. Exploratory Bayesian Prior Sensitivity

A secondary beta-binomial analysis quantified uncertainty in failure probability by the already selected common force. This analysis followed the conjugate updating framework described in a recent methodological review [21]. Independent, identical, symmetric Beta(a,a) priors were assigned to the two workflow probabilities. Three choices were examined: a = 0.5 (Jeffreys), a = 1 (uniform), and a = 2 (central regularizing). None favors a workflow before the observations; their concentrations a + a are 1, 2, and 4, respectively. They are analytical sensitivity choices, not priors elicited from historical manufacturing data.
With kCast = 5 and kSLM = 0 among five specimens per workflow, the posterior distributions are Beta(a + 5,a) and Beta(a,a + 5), respectively (Equation (5)). For D = θCast − θSLM, the posterior median, equal-tailed 95% credible interval, probability D > 0, and probability D > 0.50 were computed. The 0.50 threshold is an illustrative absolute failure-probability contrast and is not a validated clinically important difference. Posterior distribution functions were integrated deterministically by adaptive quadrature, and quantiles were obtained by root finding with numerical tolerances specified in the supplement.
θ g   ∼   Beta ( a ,   a )
θ g   |   k g   ∼   Beta ( a + k g ,   a + n g − k g )
These posterior quantities condition on the fixed post hoc τ and independent Bernoulli sampling within each workflow. They do not adjust for selection of τ, undocumented batch dependence, alloy differences, or geometry confounding. They were not calibrated as a confirmatory decision rule and were not substituted for the primary two-sided p value. Their purpose was to expose the dependence of inferred magnitude on prior assumptions with only five specimens per workflow.

2.5.4. Identification Bounds and Influence Diagnostics

Deterministic lower bounds were calculated by retaining Tᵢ > Cᵢ for each intact SLM specimen. The minimum SLM attained load was compared with the largest measured cast fracture load; the true SLM sample mean and its contrast with the observed cast mean were bounded using all five individual stopping loads. These are statements about the tested specimens, not population confidence limits or clinical safety factors. Computed lower bounds involving means or ratios were rounded downward to preserve the inequalities. No finite upper bound was available.
Two descriptive diagnostics were evaluated. First, capped means were recomputed at lower common force limits, all within the available observed range. Second, each of the ten specimens was omitted once while retaining the original common τ = 4498.36 N. The resulting contrasts describe influence on the sample estimate; they are not independent replications or jackknife confidence intervals. No additional primary hypothesis was declared from these post hoc checks.

2.6. Analytical Framework for Geometry and Compliance

An idealized, centrally loaded, simply supported prismatic beam was used solely as a dimensional illustration. For a rectangular section with width b, depth h in the bending direction, span L, and elastic modulus E, its second moment I, linear stiffness K, and maximum nominal bending stress are given in Equation (6). These expressions follow from elementary beam mechanics under small, linear-elastic deflection. The actual FDP is not prismatic, its connectors have complex contours, and its support fixation is incompletely described. Consequently, no values of b, h, L, E, local stress, or safety factor were assigned to the experimental specimens, and no finite element analysis was claimed.
I = b h 3 12 ,      K = 48 EI L 3 ,      σ nom = 3 FL 2 b h 2
The illustration separates geometric scaling from intrinsic material strength. At a fixed force and unchanged width, nominal stress scales with L/h², whereas linear stiffness scales with Eh³/L³. A local stress concentration would require a geometry-dependent concentration factor and a stress field not supplied by the source record. The same model also motivates distinguishing framework displacement from system displacement. Over a verified linear interval, crosshead compliance can contain contributions from the machine, fixture, seating, and framework. Equation (7) is a series-compliance approximation for that interval, not a correction applied to the terminal ratios.
C cross = d δ cross d F   ≈   C machine + C fixture + C seating + C framework
K app = 1 C cross

2.7. Computational Reproducibility and Prospective Calculations

Calculations used Python 3.12.14, NumPy 2.3.5, and SciPy 1.17.0. The package provides ten force/event/displacement input records with record-level provenance, all calculated terminal secants, every permutation allocation, truncation and deletion diagnostics, posterior results and density coordinates, exact prospective power curves, numerical verification outputs, and figure-generation code. The 96 independent numerical checks include exact rational displacement and secant arithmetic, sample variability, and alternative posterior integrations. The uniform-prior directional probability was checked against 923/924. Reported values retain their supplied precision; full computational precision is retained in machine-readable output, and manuscript estimates are rounded for readability.
Prospective binary-endpoint planning used hypothetical workflow failure probabilities of 0.80 versus 0.20 and 0.70 versus 0.30. Exact two-sided Fisher-test power at α = 0.05 was obtained by summing joint binomial probabilities over the complete rejection region for each balanced sample size. A separate precision calculation identified the number of independent, zero-failure observations needed for a two-sided 95% Clopper–Pearson upper limit below 10% or 5%. These are future-design scenarios for a prespecified common force; they are not observed power, estimates of the pilot’s population failure probabilities, or sample size calculations for fatigue and span interactions.
Candidate balanced sample sizes were assessed for a future binary endpoint of macroscopic fracture by a prespecified common force. The calculation distinguished the minimum attainable two-sided p value under complete separation, obtained as two divided by the balanced allocation count, from power averaged over all future event tables. Exact power assumed failure probabilities of 0.80 versus 0.20 or 0.70 versus 0.30 and independent specimens. For zero failures, precision used the upper endpoint of a two-sided 95% Clopper–Pearson interval (Supplementary Equation (S4)). Supplementary Table S9 summarizes candidate designs. Planning for the primary capped-load mean contrast would instead require a prespecified mechanically meaningful difference and assumptions about future capped-load distributions and censoring. The pilot endpoint records do not establish these inputs, so the binary calculations are not a power calculation for the continuous primary contrast.
During manuscript preparation, ChatGPT (OpenAI, GPT-6) assisted with final editing. The authors are responsible for verifying the experimental record, computational files, references, and interpretation before submission.

3. Results

3.1. Pilot Sample and Prospective Power Assessment

The observed pilot analysis included ten frameworks, with five specimens per workflow. No documented a priori size calculation was available. For this balanced allocation, complete separation has a minimum attainable two-sided permutation p-value of 2/252 = 0.00794. This describes the test resolution at the observed sample size; it does not establish adequate prospective power. At n = 5 per workflow, the separate hypothetical binary-endpoint scenarios below give exact powers of 37.58% for failure probabilities of 0.80 versus 0.20 and 14.95% for 0.70 versus 0.30 (Supplementary Table S9). These are scenario calculations, not estimates of observed power.
Prospective exact planning for a future binary endpoint of fracture by a prespecified common force gave the following results at two-sided alpha = 0.05. For hypothetical failure probabilities of 0.80 versus 0.20, the first integer sample sizes attaining 80% and 90% power were 12 and 16 specimens per workflow, with achieved power of 81.15% and 91.05%, respectively. For 0.70 versus 0.30, the corresponding sizes were 29 and 37, with achieved power of 81.43% and 90.90% (Supplementary Table S7 and Figure S4). These failure probabilities are sensitivity assumptions, not estimates adopted from the observed complete separation.
The full candidate-size sensitivity analysis is provided in Supplementary Table S9. A separate zero-failure precision target below 10% or 5% would require 36 or 72 independent observations with zero failures, respectively (Supplementary Table S8). These prospective calculations exclude clustering, attrition, and multiplicity. They address failure probability at a future prespecified force; the required sample size for the pilot’s primary capped-load mean contrast cannot be determined without additional distributional and effect-size assumptions.

3.2. Recorded Specimen Outcomes

All ten pilot specimens had endpoint load, macroscopic fracture status, and terminal displacement available (Table 2). MELT_1–MELT_5 fractured at 964.26, 1134.63, 938.53, 987.46, and 1037.48 N, respectively. The mean measured fracture load was 1012.47 ± 77.40 N, median 987.46 N, range 938.53–1134.63 N, and a sample coefficient of variation of 7.65%. We documented connector fracture after deformation for MELT_1–MELT_5. Photographs cover the documented subset. No quantitative porosity measurement, metallographic section, or microscopic fracture-origin assessment was available.
SINT_1–SINT_5 remained macroscopically intact at attained loads of 4500.07, 4499.41, 4499.62, 4498.36, and 4501.34 N, respectively. Each contributed a censored endpoint and a lower bound on its true fracture load. These five terminal readings describe the experiment's stopping behavior, not the distribution of SLM fracture strength. The observations do not quantify residual deformation after unloading or exclude earlier subvisual damage.

3.3. Capped Load Tolerance and Exact Inference

At the common fully observed limit τ = 4498.36 N, the cast capped mean was 1012.47 N and the SLM capped mean was 4498.36 N. The primary observed contrast was 3485.89 N and the capped mean ratio was 4.44 (Table 3). Every SLM lower bound exceeded every measured cast fracture load. All 25 dependent cross-workflow specimen comparisons favored SLM, yielding a descriptive Cliff’s delta of 1.00. These comparisons reuse the same ten frameworks.
Two of the 252 equally sized allocations were at least as extreme as the observed absolute contrast, giving p = 0.00794 for the two-sided exact permutation calculation. Fisher’s exact consistency check gave the same value. For complete separation with five specimens per workflow, 2/252 is the smallest attainable two-sided value in this balanced allocation scheme. The conditional exploratory comparison is below 0.05, but neither that threshold crossing nor complete separation establishes prospective power, a causal manufacturing effect, or a confirmatory design.
Failure by τ occurred in 5/5 cast specimens and 0/5 SLM specimens. Their group-specific exact 95% confidence intervals were 47.8–100% and 0–52.2%. The observed cast-minus-SLM failure-risk difference was 100 percentage points; its conservative simultaneous 95% interval extended from −16.7 to 100 percentage points. This conservative construction can include zero despite the exact test result because it combines two simultaneous group intervals rather than inverting that test. Figure 2 retains event and censoring status alongside the force-dependent fracture-free proportions and group-specific intervals.

3.4. Lower Bounds on the Unobserved Fracture Loads

The weakest attained SLM lower bound exceeded the greatest measured cast fracture load by 3363.73 N. Their attained-load ratio was approximately 3.9646; every tested SLM fracture load was more than 3.9646 times the greatest cast fracture load observed in this set. This is a conservative ordering statement about these ten specimens; it is not a population strength ratio or an allowable clinical load.
Using all five censoring limits, the true mean SLM fracture load for these specimens was strictly greater than 4499.76 N. Its true sample mean contrast with the cast group was therefore greater than 3487.28 N, and its true sample mean ratio greater than 4.4443. There was no identifiable upper bound. The capped analysis uses one common limit, whereas these deterministic bounds retain each specimen’s own attained load. Neither calculation supplies an unrestricted population mean or its confidence interval.

3.5. Displacement and Endpoint Secant Ratios

Across all five cast specimens, mean displacement at fracture was 1.459 ± 0.357 mm, median 1.547 mm, range 0.871–1.804 mm, and sample coefficient of variation 24.46%. Across all five SLM specimens, mean displacement at administrative termination was 2.048 ± 0.134 mm, median 2.053 mm, range 1.897–2.225 mm, and sample coefficient of variation 6.52%. These summaries include the four investigator-supplied displacement records. The larger terminal SLM displacement occurred at a substantially higher force and in a different event state. It does not establish a matched-force difference in elastic deflection or clinical deformability.
Calculated SLM terminal secant ratios for SINT_1–SINT_5 were 1977.56, 2265.40, 2143.02, 2318.59, and 2071.22 N/mm. Calculating all five uniformly as (endpoint force − 100 N)/endpoint displacement gave a mean of 2155.16 ± 139.28 N/mm, median 2143.02 N/mm, range 1977.56–2318.59 N/mm, and sample coefficient of variation 6.46%. The first three calculated ratios agree with the three source-reported rounded values; SINT_4 and SINT_5 are derived from their reported endpoint pairs. Cast terminal ratios are also tabulated, with a mean of 662.66 ± 196.10 N/mm, solely to document the same arithmetic at a different force and event state. These are system-response summaries, not intrinsic framework elasticity, matched-force stiffness ratios, or between-batch reproducibility (Supplementary Table S2 and Figure S1).

3.6. Truncation and Specimen Influence Diagnostics

At τ = 500 N, all ten specimens reached the chosen force, so the capped contrast was zero. At 1000 N, the cast-capped mean was 978.05 N, the SLM mean was 1000 N, and the contrast was 21.95 N with an exact two-sided p-value of 0.16667. Once the limit exceeded the greatest cast fracture load of 1134.63 N, the between-workflow ordering was complete. At 1500, 2000, 3000, and 4000 N, capped contrasts were 487.53, 987.53, 1987.53, and 2987.53 N, respectively, with p = 0.00794 at each limit. Effect magnitude depends on the stated force range, whereas separation above all cast events persists (Figure 3a).
Omitting each of the ten specimens once while fixing τ at 4498.36 N produced capped contrasts from 3467.40 to 3516.43 N (Figure 3b). The greatest absolute relative change from the complete-sample contrast was 0.876%. Removing any single SLM specimen left its capped mean unchanged because the remaining capped observations also equaled the fixed τ. This diagnostic indicates that no individual specimen drives the large capped contrast within the dataset. It does not assess an unobserved SLM tail, compensate for missing batches, or increase population precision.

3.7. Posterior Sensitivity of Failure Probability

The Bayesian sensitivity analysis favored a lower failure probability for SLM across all three symmetric prior choices, while the inferred magnitude varied (Table 4). Under the uniform prior, posterior mean failure probabilities were 0.857 for casting and 0.143 for SLM. The median cast-minus-SLM probability difference was 0.745, its equal-tailed 95% credible interval was 0.301–0.960, and the directional posterior probability was 99.89%.
The Jeffreys prior yielded a median difference of 0.874 (95% credible interval 0.453–0.995) and a directional probability of 99.98%. The central regularizing prior yielded 0.574 (0.146–0.860) and 99.49%, respectively. All three equal-tailed 95% intervals excluded zero under their stated models. For the illustrative contrast D > 0.50, posterior probabilities ranged from 64.66% under central regularization to 96.27% under Jeffreys, with 87.82% under the uniform prior. Thus, direction was consistent across the examined priors, while the probability of a very large effect remained prior-sensitive. These are conditional summaries of the same ten observations, not independent evidence or proof of clinical superiority.

4. Discussion

4.1. Interpretation of the Pilot Mechanical Findings

The principal pilot observation is a separation in recorded failure behavior: all five cast frameworks fractured near 1 kN, while all five SLM frameworks remained macroscopically intact at attained loads near 4.5 kN. We specifically documented connector failure for MELT_1–MELT_5. Preserving censoring status distinguishes observed fractures from the five intact stopping records. The restricted contrast quantifies the demonstrated force-range difference, and the specimen-level lower bounds show how far the true sample fracture contrast must extend. The SLM tail remains unobserved, and the lack of specimen-specific traces and protocol logs limits verification of experimental comparability.
The direction is compatible with published differences in mechanical properties and microstructure among dental Co–Cr workflows [1,2] and with improved fatigue behavior in a controlled cast-versus-SLM coupon comparison [3]. Those studies provide plausible mechanisms, not measurements of the present frameworks. The present record cannot establish whether connector failure initiated at a casting pore, an inclusion, a locally thin section, a finishing defect, or a stress concentration altered by seating. A finer laser-produced microstructure is also a hypothesis until these specimens independently demonstrate it.
A mechanism-focused follow-up should connect the structural event with measurements at the fracture origin. Microcomputed tomography could identify internal discontinuities before testing; dimensional scans could quantify the finished connector sections; metallography and phase-sensitive characterization could document the substrate; and scanning electron microscopy with compositional analysis could examine fracture surfaces. These measurements would distinguish a geometry-related reduction in section modulus from a defect-associated reduction in local resistance. Their proposed use should not be confused with analyses already performed in this pilot.

4.2. Connector Geometry and the Meaning of Apparent Stiffness

The dimensional beam illustration explains why alloy strength and framework stiffness are separate targets. In the stated linear-elastic model, stiffness depends on E and the section's second moment, while nominal bending stress at a given force depends on span and section modulus. At constant width and span, increasing depth by 10% raises the model stiffness by 33.1% and lowers its nominal bending stress by 17.4%. Reducing depth by 10% lowers stiffness by 27.1% and raises stress by 23.5%. These values are analytical consequences of the cubic and squared scaling laws, not estimated differences between the cast and SLM connectors (Figure 4a).
The same illustration predicts that a 20% increase in span, with material and cross-section held constant, increases linear deflection under a fixed force by 72.8% and nominal bending stress by 20%. It is a reason to measure span and test several geometries, not evidence that the experiment has already observed a span effect (Figure 4b). The actual numerical coefficients depend on fixation, load position, cross-sectional shape, and the multiunit anatomy. Recent internal-design experiments and pontic-volume studies reinforce the need to analyze the entire architecture [6,7]. Connector depth and pontic volume are related design features but are not interchangeable geometric interventions.
Nominal section area alone is insufficient for describing bending. Two sections with the same area can have different second moments if material is redistributed away from the neutral axis. Connector curvature can also create a local stress field absent from a simple rectangular beam. The nominally common digital file therefore constrains design intent, while final metrology is required to establish comparable section modulus, contour, and seating. In a future geometry study, force values in N should be accompanied by measured dimensions and a defensible stress model if comparisons in MPa are intended.
Crosshead stiffness presents an additional identifiability problem. In a locally linear series system, machine, fixture, seating, and framework compliances add. The reciprocal total compliance is an apparent system stiffness, which cannot be assigned to the framework alone without separate displacement or compliance measurements. Local framework displacement measured optically or by an appropriately positioned sensor would reduce this ambiguity. Repeated loading within a defined elastic interval and an unloading segment would help distinguish reversible displacement from permanent set. Those procedures were not documented in the present source.
The terminal secant ratio combines an entire loading history into one number. Agreement with force above the preload divided by terminal displacement identifies its arithmetic, not an initial tangent slope, a secant over a common force interval, or an elastic modulus. All five SLM ratios are calculable from the reported endpoint pairs, giving a sample coefficient of variation of 6.46%. That spread describes these five terminal system responses; absent common-force curves and independently replicated batches, it cannot establish material stiffness or manufacturing reproducibility. Cast endpoints occur at lower forces and at fracture, so comparing the two groups’ terminal ratios would confound load level, event state, support compliance, and possible nonlinear deformation.

4.3. From the Framework to the Finished Restoration

A high framework load tolerance does not resolve the failure pathways introduced by veneering. Recent interfacial studies document alloy- and workflow-dependent ceramic bonding [22,23]. Annealing temperature can affect metal-ceramic bond characteristics [24], and surface treatments may act differently across Co–Cr substrates [25]. SLM hatch distance and laser power have altered properties relevant to porcelain bonding [26,27], while comparisons of laser texturing and conventional conditioning show that the chosen surface process matters [28]. The metal substrate, oxide formation, firing sequence, and interface therefore constitute an additional system requiring its own validation.
A recent pilot comparison reported higher flexural interface performance for the tested cast Co–Cr ceramic system than for the printed system [29]. That finding need not conflict with the present framework outcome: the events, assemblies, and loading definitions are different. Studies of screw-retained metal-ceramic restorations and aged tooth-to-implant FDPs have likewise identified veneer or connection damage that a framework-only monotonic test does not measure [30,31]. Work varying cantilever length and material combinations further demonstrates the importance of assembly and geometry [32]. Manufacturing decisions should therefore be evaluated against the intended restoration endpoint rather than a single large terminal force.
Failure detection also defines the measured tolerance. An acoustic-emission pilot on implant-supported hybrid bars detected microdamage before a conspicuous terminal event [33]. The present event was limited to source-reported macroscopic fracture, so we cannot exclude earlier microdamage. Standardized bending studies of titanium and Co–Cr [34] address a material question under controlled dimensions, whereas the present experiment measures the force tolerance of the supported framework. Coupon strength in MPa, framework force in N, and survival under fatigue are distinct outcomes and require separate translation models.
Other restorative studies make the need for separate performance endpoints explicit. Titihazan et al. evaluated fracture resistance and color change as distinct outcomes for CAD/CAM and chairside provisional crowns [35]. Rotar et al. investigated mouthwash-related color changes in hybrid ceramics and reinforced CAD/CAM composites [36]. These studies support evaluating mechanical and optical durability separately; their nonmetallic materials and endpoints do not supply failure-load estimates for the present Co–Cr frameworks.

4.4. Processing Quality and Biological Performance

SLM is a process family rather than one reproducible set of material properties. Evidence that fit depends on layer thickness, orientation, thermal treatment, and firing [10,11,12,18] supports recording the entire fabrication sequence. Clinical adaptation can also differ by the measured region [16]. Adjacent removable-prosthesis evidence is useful for illustrating this dependence: a randomized crossover trial found the overall adaptation of direct digital SLM manufacture consistent with conventional manufacture, while clasp experiments linked retentive response with deformation and processing [37,38]. Optimized powder-bed-fused clasps and studies of presintered Co–Cr discs further show that the relevant performance criterion is component-specific [39,40]. These findings inform process control but do not provide survival estimates for the present FDPs.
Powder history is part of that process definition. Reuse studies have reported preserved metal-ceramic bonding while also documenting changes in powder morphology, framework microstructure, or ion release [41,42,43]. Reuse, stress relief, surface finishing, and ceramic firing can therefore produce different outcomes even when a single monotonic test suffices. Because these histories were unknown here, the observed SLM tolerance should be attributed to the recorded workflow, not generalized to every powder reuse or thermal protocol.
Mechanical and biological validation are also separate. Comparative surface work has sometimes favored digitally manufactured Co–Cr over casting [44], but composition and manufacturing route both affect electrochemical behavior [45]. One investigation measured greater release of several ions from laser-sintered than milled specimens [46], whereas a more recent immersion experiment found lower release from additive than cast specimens under its conditions [47]. Proteins and build orientation can modify release and tribocorrosion [48], and multiroute studies emphasize interactions with the chemical environment [49]. These context-dependent findings preclude inferring corrosion resistance, ion release, or cytocompatibility from the load-tolerance contrast.
Finishing can influence both the surface environment and the final geometry. Recent polishing studies of laser-produced frameworks reported effects on roughness, corrosion-related behavior, and dimensional features [50,51]. Therefore, the abrasive medium, pressure, exposure, polishing sequence, and dimensional change should be reported and assessed alongside fabrication. The present experiment records finishing and airborne-particle abrasion in general terms but provides no surface characterization or biological endpoint.

4.5. Statistical Interpretation and Prior Dependence

With five specimens per workflow, the balanced permutation reference contains 252 allocations; complete ordering gives the observed two-sided p = 0.00794. This conditional exploratory evidence of separation does not establish prospective power or resolve the undocumented sampling and batch structure. The binary planning scenarios give only 37.58% or 14.95% power at the same n under their respective hypothetical alternatives. The separate exact 95% intervals still permit SLM failure probabilities up to 52.2%. The conservative simultaneous risk-difference interval spans zero because it combines two group intervals rather than inverting the permutation or Fisher test. It therefore addresses uncertainty differently from the reported hypothesis test.
The restricted mean and the deterministic lower bounds solve different problems. The former is fully observed within τ and can be compared conditionally without assigning unobserved failures; the latter proves an ordering for these particular intact specimens. Neither identifies the SLM population tail. Raising τ while no SLM fracture is observed necessarily increases the capped contrast once all cast specimens have failed. That dependence is an intrinsic feature of the estimand and should be reported, as it is here, rather than hidden behind one headline ratio.
The posterior sensitivity analysis makes prior assumptions explicit [21]. All three identical symmetric prior choices produce positive equal-tailed 95% intervals for the cast-minus-SLM failure-probability difference. The central regularizing prior nevertheless gives a smaller median difference (0.574) and a 64.66% probability of D > 0.50, compared with 0.874 and 96.27% under Jeffreys. Five binary outcomes per workflow thus support a consistent direction under these models while leaving the probability of a very large effect prior-sensitive. Posterior probabilities, exact-test p values, and conservative confidence intervals answer different conditional questions. None adjusts for threshold selection, manufacturing confounding, or undocumented batches; agreement in direction is not independent replication or confirmatory validation.
Deletion analysis supports stability of the observed capped contrast to any one of the ten recorded specimens, while leaving the structural limitations intact. Undocumented batches could create shared variation within a workflow. If all five specimens per workflow came from one batch, fabrication method and batch could not be separated by specimen-level testing alone. Neither exact enumeration, Bayesian updating, nor an influence diagnostic supplies missing batch replication. All ten terminal displacements are reported, but unavailable traces prevent verification of displacement origin, a common elastic interval, and specimen-specific protocol implementation. Future inference should specify whether its target is a laboratory workflow, an independently produced batch, or a broader manufacturing process.

4.6. Implications of the Pilot for Independent Validation

The next study should first recover the alloy identities, chemistry, powder and casting histories, actual connector dimensions, support properties, fixation protocol, load coordinates, calibration records, and original traces. A common digital file should include as-built metrology. Manufacturing should span independently replicated batches, with specimens randomized across batch positions and testing order where feasible. Specify event classification before testing, ideally separating first detectable damage, permanent set, connector fracture, and veneer damage when those outcomes are relevant. These reporting and bias-control priorities are consistent with the domains addressed by the QUIN tool for in vitro dental studies [52]; a formal QUIN score was not retrospectively assigned from incomplete records.
For monotonic testing, a force range that produces observable failures in both workflows would improve estimation of the full fracture distribution, provided fixture and load-cell capacity remain appropriate. When censoring is expected, the primary range and stopping rule should be set in advance and censored observations retained in the analysis. Local displacement measurement, machine and fixture compliance checks, and common loading intervals would support a defensible stiffness endpoint. An independent cyclic experiment with defined aging, loading, and run-out criteria is needed for fatigue resistance; monotonic force tolerance does not supply that endpoint.
A span-related question requires a factorial design with both workflows represented at multiple prespecified spans and with connector geometry documented at each span. The workflow-by-span interaction should be an explicit target rather than inferred from separate significance tests at different geometries. Batch should enter the sampling and analysis structure. Once sufficient failures are observed, a fracture-load or censoring-aware model could evaluate this interaction while accounting for measured geometry. Its sample size and assumptions must match that multivariable endpoint rather than the present ten-record pilot.
The prospective binary-endpoint calculations illustrate the scale of independent validation informed by the pilot. For hypothetical failure probabilities of 0.80 versus 0.20 at a prespecified common force, exact two-sided Fisher-test power first exceeded 80% at 12 specimens per workflow and 90% at 16. For 0.70 versus 0.30, the corresponding counts were 29 and 37. These counts exclude clustering, attrition, multiple spans, and additional primary endpoints. With zero failures, 36 independent observations are needed for a two-sided 95% Clopper–Pearson upper failure limit below 10%, and 72 for a limit below 5%. Five intact SLM specimens yield an upper limit of 52.2%, so the pilot does not establish a low population failure probability despite the high attained forces.

5. Conclusions

In this exploratory in vitro pilot study of ten frameworks, five cast Co–Cr FDPs fractured at 938.53–1134.63 N, while five SLM frameworks remained macroscopically intact at 4498.36–4501.34 N. Connector fracture was documented for the first three cast specimens; location was unavailable for the other two. At τ = 4498.36 N, the capped mean contrast was 3485.89 N, the ratio 4.44, and the two-sided exact permutation p-value 0.00794. Single-specimen deletion changed the contrast by at most 0.876%. Terminal displacement averaged 1.459 ± 0.357 mm for casting and 2.048 ± 0.134 mm for SLM, with five records per workflow; the five calculated SLM terminal secants averaged 2155.16 ± 139.28 N/mm. Different endpoint forces and event states preclude a matched-force stiffness interpretation. Bayesian direction favored fewer SLM fractures, with prior-sensitive magnitude. These preliminary findings support independent validation; the prospective scenarios do not establish adequate power for this pilot or its continuous primary contrast. The SLM population fracture distribution, material stiffness, fatigue performance, clinical benefit, and span effect remain unestablished.

Supplementary Materials

Supplementary Results S1; Supplementary Methods S1–S4; Table S1–S9; Figure S1–S4; ten specimen-level force/event/displacement records and calculated secants; all 252 permutation allocations; truncation, deletion, Bayesian, precision, and exact power outputs; candidate future sample-size assessments; executable Python analysis, independent verification, and figure code; and verified reference metadata are supplied separately.

Author Contributions

Conceptualization, F.T., T.H., C.S. and M.R.; methodology, F.T., T.H., M.L.N. and I.-A.T.; software, I.-A.T.; validation, C.S., M.L.N., I.V. and M.R.; formal analysis, I.-A.T.; investigation, F.T., T.H., I.V., C.M. and D.S.; resources, C.S., M.L.N., D.S. and M.R.; data curation, F.T., I.V., I.-A.T., C.M. and D.S.; writing—original draft preparation, F.T., T.H. and I.-A.T.; writing—review and editing, F.T., T.H., C.S., M.L.N., I.V., I.-A.T., C.M., D.S. and M.R.; visualization, F.T., I.-A.T. and C.M.; supervision, C.S., M.L.N. and M.R.; project administration, F.T., T.H. and C.S. All authors have read and agreed to the published version of the manuscript.

Funding

The Victor Babes University of Medicine and Pharmacy, Timisoara, supported publication costs. The findings, interpretations, and conclusions drawn in this study remain entirely independent of the financial support provided.

Institutional Review Board Statement

Not applicable to the laboratory experiment as documented, which involved manufactured frameworks and laboratory models without human participants, identifiable patient data, or animals. The responsible institution should confirm this classification before submission.

Data Availability Statement

All ten specimen-level force/event/displacement records, calculated terminal secants, three source-reported rounded SLM secants, record-level provenance, and complete analysis and figure-generation code are included in the Supplementary Materials. Fracture locations for MELT_4 and MELT_5 were not supplied. Original machine exports, CAD files, manufacturing logs, and dimensional metrology were unavailable for this reanalysis; their availability should be confirmed with the investigators before submission.

Acknowledgments

ChatGPT (OpenAI, GPT-6) assisted with minor typographical corrections and formatting adjustments to the manuscript. All authors reviewed and approved the final version and take full responsibility for its content.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

Abbreviation Definition
CAD/CAM Computer-aided design and computer-aided manufacturing
Co–Cr Cobalt–chromium
FDP Fixed dental prosthesis
SLM Selective laser melting
CI Confidence interval
CrI Credible interval
QUIN Quality Assessment Tool for In Vitro Studies

References

  1. Kassapidou, M.; Stenport, V.F.; Johansson, C.B.; Syverud, M.; Hammarström Johansson, P.; Börjesson, J.; Hjalmarsson, L. Cobalt chromium alloys in fixed prosthodontics: Investigations of mechanical properties and microstructure. J. Prosthet. Dent. 2023, 130, 255.e1–255.e10. [Google Scholar] [CrossRef]
  2. Schettini, A.; Pesun, I.J.; França, R. Assessment of mechanical properties and microstructure of Co-Cr dental alloys manufactured by casting, milling, and 3D printing. J. Prosthet. Dent. 2025, 133, 906.e1–906.e7. [Google Scholar] [CrossRef]
  3. Wu, M.; Dong, X.; Qu, Y.; Yan, J.; Li, N. Analysis of microstructure and fatigue of cast versus selective laser-melted dental Co-Cr alloy. J. Prosthet. Dent. 2022, 128, 218.e1–218.e7. [Google Scholar] [CrossRef]
  4. Ko, K.H.; Kang, H.G.; Huh, Y.H.; Park, C.J.; Cho, L.R. Effects of heat treatment on the microstructure, residual stress, and mechanical properties of Co-Cr alloy fabricated by selective laser melting. J. Mech. Behav. Biomed. Mater. 2022, 126, 105051. [Google Scholar] [CrossRef]
  5. Konieczny, B.; Szczesio-Wlodarczyk, A.; Andrearczyk, A.; Januszewicz, B.; Lipa, S.; Zieliński, R.; Sokolowski, J. Enhancing the Mechanical Properties of Co-Cr Dental Alloys Fabricated by Laser Powder Bed Fusion: Evaluation of Quenching and Annealing as Heat Treatment Methods. Materials 2024, 17, 5313. [Google Scholar] [CrossRef]
  6. Kocak, E.F.; Ekren, O.; Ucar, Y. Effect of Internal Design Modification on the Mechanical Properties of Laser Sintered Cobalt-Chromium Multi-Unit Metal-Ceramic Frameworks. J. Prosthodont. 2022, 31, 766–770. [Google Scholar] [CrossRef]
  7. Cebi Gul, B.; Bahce, E.; Atmaca, A.; Kuzudisli, F.; Demirci, F. Effect of pontic volume on fracture resistance of three-unit metal-ceramic fixed dental prostheses: an in vitro and finite element analysis study. BMC Oral Health 2026, 26, 1933. [Google Scholar] [CrossRef]
  8. Hajaj, T.; Marian, D.; Zaharia, C.; Niculescu, S.T.; Negru, R.M.; Titihazan, F.; Rominu, M.; Sinescu, C.; Novac, A.C.; Dobrota, G.; Veja, I. Fracture Resistance of CAD/CAM-Fabricated Zirconia and Lithium Disilicate Crowns with Different Margin Designs: Implications for Digital Dentistry. J. Funct. Biomater. 2025, 16, 205. [Google Scholar] [CrossRef]
  9. Andrei, A.D.; Constantin, C.A.; Tabitha, A.L.; Anca, J. Zirconia single retainer fixed dental prostheses for the posterior region—A novel preparation technique and literature review. Clin. Case Rep. 2024, 12, e9460. [Google Scholar] [CrossRef]
  10. Daou, E.E. Effect of lamination layer thickness and abutment preparation on the fit of Co-Cr multi-unit prostheses fabricated by additive manufacturing: An in vitro study. J. Prosthet. Dent. 2024, 131, 652–657. [Google Scholar] [CrossRef]
  11. Daou, E.E. Does the Build Orientation Parameter Affect the Fit of Fixed Partial Dentures Fabricated by Additive Manufacturing? Int. J. Prosthodont. 2024, 37, S41–S47. [Google Scholar] [CrossRef]
  12. Htat, H.L.; Limpuangthip, N.; Lee, J.H.; Lee, S.J. Printing orientation and heat treatment effects on fit and hardness of selective laser melting-manufactured cobalt-chromium crowns. J. Dent. 2025, 160, 105856. [Google Scholar] [CrossRef]
  13. Zhou, Y.; Liu, Z.; Du, H.; Jiang, J.; Li, J.; Ma, X.; Yi, Y. Effects of rapid infrared radiation heating on the warping deformation and mechanical properties of selective laser melted Co-Cr dental alloy. J. Prosthet. Dent. 2025, 133, 907.e1–907.e9. [Google Scholar] [CrossRef]
  14. Pekkan, G.; Degirmenci, K.; Tuna, S.H.; Hekimoğlu, C.; Saridag, S. Comparison of the overall fit of three-unit posterior fixed dental prostheses fabricated with laser sintering and conventional casting methods. Clin. Oral Investig. 2025, 29, 153. [Google Scholar] [CrossRef]
  15. Pekkan, G.; Gülgezen Aydın, H.; Tuna, S.H.; Hekimoğlu, C. Fit of three-unit anterior fixed dental prosthesis frameworks fabricated by different manufacturing techniques. BMC Oral Health 2026, 26, 1466. [Google Scholar] [CrossRef]
  16. Bani-Younes, S.A.; Al Fodeh, R.S.; Khasawneh, L.; Tabnjh, A. Clinical internal and marginal fit of metal-ceramic fixed dental prostheses fabricated with selective laser melting and 3D-printed pattern casting using cobalt-chromium metal alloy. J. Prosthodont. 2024, 33, 861–868. [Google Scholar] [CrossRef]
  17. Titihazan, F.; Hajaj, T.; Novac, A.C.; Pop, D.M.; Sinescu, C.; Negruțiu, M.L.; Romînu, M.; Zaharia, C. Dental Preparation Guides—From CAD to PRINT and CAM. Oral 2026, 6, 12. [Google Scholar] [CrossRef]
  18. Daou, E.E.; Baba, N.Z. The Effect of Firing Cycles on the Fit of Cobalt-Chromium Multiunit Prostheses Fabricated by Additive Manufacturing. J. Prosthodont. 2023, 32, 71–76. [Google Scholar] [CrossRef]
  19. Nikolopoulos, M.; Dimitriadis, K.; Yannikakis, S.; Galiatsatos, A.; Tsolka, P. Influence of repeated porcelain firing cycles on the microstructure, marginal and internal fit of implant-cemented Co-Cr frameworks. Eur. J. Oral Sci. 2025, 133, e70009. [Google Scholar] [CrossRef]
  20. Nikolopoulos, M.; Dimitriadis, K.; Yannikakis, S.; Sfikas, A.K.; Galiatsatos, A.; Tsolka, P. Implant-cemented Co-Cr frameworks produced by hard and soft milling: The impact of porcelain firing cycles on microstructure and fit accuracy. Eur. J. Oral Sci. 2025, 133, e70031. [Google Scholar] [CrossRef]
  21. Lee, S.Y. Using Bayesian statistics in confirmatory clinical trials in the regulatory setting: a tutorial review. BMC Med. Res. Methodol. 2024, 24, 110. [Google Scholar] [CrossRef]
  22. Al Jabbari, Y.S.; Dimitriadis, K.; Sufyan, A.; Zinelis, S. Bond strength and interfacial analysis of six CoCr alloys made by conventional casting and selective laser melting. J. Prosthodont. 2026, 35, 936–944. [Google Scholar] [CrossRef]
  23. Daou, E.E.; Özcan, M. Evaluation of ceramic adherence to cobalt-chromium alloys fabricated by different manufacturing techniques. J. Prosthet. Dent. 2022, 128, 1364.e1–1364.e8. [Google Scholar] [CrossRef]
  24. Zhou, Y.; Dong, X.; Li, N.; Yan, J. Effects of post-treatment on metal-ceramic bond properties of selective laser melted Co-Cr dental alloy. Part 1: Annealing temperature. J. Prosthet. Dent. 2023, 129, 657.e1–657.e9. [Google Scholar] [CrossRef]
  25. AboElhassan, R.G.; Morsy, N. Effect of laser etching on surface characteristics and porcelain bond to soft milled and direct metal laser sintered cobalt chromium alloys. J. Prosthet. Dent. 2024, 132, 269.e1–269.e6. [Google Scholar] [CrossRef]
  26. Sezer, T.; Albayrak, H.; Ayata, M.; Yavuz, H.İ.; Yamanoğlu, R. Effect of hatch distance on porcelain bond strength of metal frameworks produced by selective laser melting. J. Prosthet. Dent. 2025, 134, 469.e1–469.e8. [Google Scholar] [CrossRef]
  27. Karahasan, B.; Dogan, D.O.; Babacan, N. Investigation of Metal-Ceramic Bond Strength in Co-Cr Alloys Produced by Selective Laser Melting at Different Laser Powers. Int. J. Prosthodont. 2026, 39, 509–515. [Google Scholar] [CrossRef]
  28. Ayata, M.; Albayrak, H.; Sezer, T. Effect of femtosecond and nanosecond laser and tribochemical silica coating on metal-ceramic bond strength: An in vitro study. J. Prosthet. Dent. 2026, 135, e166–e175. [Google Scholar] [CrossRef]
  29. Biculescu, A.E.; Popescu, A.I.; Ionescu, T.P.; Ciorniciuc, I.A.M.; Referendaru, D.A.; Coman, C.; Constantinovici, A.; Chirsanov-Capanu, S.E.; Stoian-Albulescu, M.; Comaneanu, R.M. Evaluation of Flexural Resistance in Co-Cr Ceramic Systems: Conventional Casting Versus 3D Printing-A Pilot Study. Dent. J. 2025, 13, 583. [Google Scholar] [CrossRef]
  30. Mohebbi, H.; Habibzadeh, S.; Alikhasi, M.; Mosaddad, S.A.; Ghodsi, S. The effect of fabrication procedures and thermomechanical loading on the structural properties of screw-retained metal-ceramic implant restorations: An in vitro study. J. Prosthodont. advance online publication. 2024. [Google Scholar] [CrossRef]
  31. Waldecker, M.; Custodio, L.J.; Ohlmann, B.; Rammelsberg, P.; Rues, S. Influence of artificial aging on the in vitro fracture load of tooth-to-implant-supported fixed partial dentures with different retainer designs. J. Prosthet. Dent. 2025, 134, 435.e1–435.e8. [Google Scholar] [CrossRef]
  32. Rubiños, N.M.; Mosaddad, S.A.; Díaz, P.; Revilla-León, M.; Gómez-Polo, M. Mechanical behavior of complete arch implant-supported prostheses with distal cantilevers fabricated from different digital material combinations. J. Prosthet. Dent. 2026, 136, 239–249. [Google Scholar] [CrossRef]
  33. Arteaga-Losada, L.; Laura-Fernadez, H.; Puerta-Dominguez, M.A.; Ascaso-Terren, C.; Vives, E.; Escuin-Henar, T.; Torné-Durán, S. In-vitro mechanical performance of three CAD-CAM bar designs for implant-supported metal-resin hybrid prostheses: a preliminary pilot study. J. Oral Biol. Craniofac. Res. 2026, 16, 101407. [Google Scholar] [CrossRef]
  34. Audi, L.; Almustafa, A. Bending strength and hardness comparison of titanium and cobalt-chromium alloys for dentures made by selective laser melting. Sci. Rep. 2025, 15, 36463. [Google Scholar] [CrossRef]
  35. Titihazan, F.; Veja, I.; Zaharia, C.; Hajaj, T.; Sinescu, C.; Constantin, G.D.; Rominu, M. Comparative Evaluation of Color Stability and Fracture Resistance of CAD/CAM and Chairside Provisional Restorations: An In Vitro Study. J. Funct. Biomater. 2025, 16, 426. [Google Scholar] [CrossRef]
  36. Rotar, R.; Cândea, A.; Măroiu, A.; Faur, A.B.; Cuzic, C.; Rotar, R.E.; Jivănescu, A. Impact of Various Mouthwashes on the Color Stability of Hybrid Ceramic and Reinforced Composite CAD/CAM Restorative Materials: An In Vitro Study. Materials 2026, 19, 758. [Google Scholar] [CrossRef]
  37. Gan, N.; Yao, M.; Ruan, Y.; Wang, C.; Xiong, Y.; Jiao, T. Adaptation of maxillary removable partial denture frameworks fabricated with a direct digital workflow: A randomized crossover clinical trial. J. Dent. 2025, 154, 105588. [Google Scholar] [CrossRef]
  38. Husseiny, E.; Youssef, R.A.F.; Eltamaly, N.; Rokaya, D.; Mohamed, S.Z. Comparison of retentive force and deformation of selective laser melting and conventional cobalt-chromium clasps of removable partial denture. BMC Oral Health 2025, 26, 361. [Google Scholar] [CrossRef]
  39. Ma, K.; Chen, H.; Shen, Y.; Guo, Y.; Li, W.; Wang, Y.; Zhang, Y.; Sun, Y. Feasibility study and material selection for powder-bed fusion process in printing of denture clasps. Comput. Biol. Med. 2023, 157, 106772. [Google Scholar] [CrossRef]
  40. Kato, Y.; Tasaka, A.; Okano, H.; Uzawa, S.; Yamashita, S. Accuracy and mechanical properties of denture clasps fabricated utilizing pre-sintered Co-Cr alloy discs. J. Prosthodont. Res. 2025, 69, 578–587. [Google Scholar] [CrossRef]
  41. Albayrak, H.; Ayata, M.; Demirel, B. Recycling selective laser melting alloy powder on cobalt chromium-to-ceramic bond strength. J. Prosthet. Dent. 2023, 130, 786.e1–786.e7. [Google Scholar] [CrossRef]
  42. Guluzade, I.; Albayrak, H.; Yamanoğlu, R.; İsmail Yavuz, H.; Ayata, M. Effect of Recycling Cobalt-Chromium Alloy Powder on Metal-Ceramic Bond Strength and Microstructure. Int. J. Prosthodont. 2026, 39, 251–259. [Google Scholar] [CrossRef]
  43. Berk, T.Y.; Çelik Öge, S.; Uçar, Y.; Ekren, O. Effect of reusing cobalt-chromium alloy powders in direct metal laser sintering on powder characteristics and metal-ceramic bond strength. J. Mech. Behav. Biomed. Mater. 2026, 182, 107559. [Google Scholar] [CrossRef]
  44. Xing, X.; Hu, Q.; Liu, Y.; Wang, Y.; Cheng, H. Comparative analysis of the surface properties and corrosion resistance of Co-Cr dental alloys fabricated by different methods. J. Prosthet. Dent. 2022, 127, 497.e1–497.e11. [Google Scholar] [CrossRef]
  45. Al Jabbari, Y.S.; Karavoltsos, S.; Kokkinos, C.; Polychronis, G.; Economou, A.; Zinelis, S. Electrochemical characterization of three types of Co-Cr based alloys manufactured by casting and selective laser melting according to ISO 10271. Dent. Mater. 2022, 38, 1162–1172. [Google Scholar] [CrossRef]
  46. Şahin, M.; Ünalan, F.; Mutlu, İ. Corrosion, ion release, and surface hardness of Ti-6Al-4V and cobalt-chromium alloys produced by CAD-CAM milling and laser sintering. J. Prosthet. Dent. 2022, 128, 529.e1–529.e10. [Google Scholar] [CrossRef]
  47. Sakurai, N.; Sawada, T.; Kuwajima, Y.; Yamanaka, K.; Nomura, N.; Kasahara, M.; Chiba, A.; Satoh, K.; Takemoto, S. Metallic Ion Release Behaviors from Cobalt-Chromium Alloys Fabricated by Additive Manufacturing with Mechanical Grinding in an Acidic Saline Solution. Materials 2025, 18, 432. [Google Scholar] [CrossRef]
  48. Atapour, M.; Standish, T.E.; Henderson, J.D.; Wei, Z.; Dehnavi, V.; Hedberg, Y.S. Influence of Proteins and Building Direction on the Corrosion and Tribocorrosion of CoCrMo Fabricated by Laser Powder Bed Fusion. ACS Biomater. Sci. Eng. 2024, 10, 2880–2893. [Google Scholar] [CrossRef]
  49. Reclaru, L.; Buciu, G.; Petrescu, S.M.; Gheorghe, R.I.; Grecu, D.F.; Grecu, A.F. Corrosion Behavior and Ion Release of Co-Cr Dental Alloys Fabricated by Casting, CAD/CAM, SLM and DMLS: Influence of Manufacturing Route and Microstructure. Bioengineering 2026, 13, 406. [Google Scholar] [CrossRef]
  50. Gan, J.; Zhang, H.; Sun, J.; Zhu, C.; Jiao, T. Effects of various polishing processes on surface morphology, corrosion resistance, and RPD framework adaptation of SLM-printed dental Co-Cr alloys. J. Prosthodont. 2026, 35, 1358–1368. [Google Scholar] [CrossRef]
  51. Li, X.; Wang, Q.; Guan, Y. Laser polishing for integrated enhancement of thin-walled LPBFed dental Co-Cr-Mo alloy frameworks. J. Mech. Behav. Biomed. Mater. 2026, 180, 107463. [Google Scholar] [CrossRef]
  52. Sheth, V.H.; Shah, N.P.; Jain, R.; Bhanushali, N.; Bhatnagar, V. Development and validation of a risk-of-bias tool for assessing in vitro studies conducted in dentistry: The QUIN. J. Prosthet. Dent. 2024, 131, 1038–1042. [Google Scholar] [CrossRef]
Figure 2. Load tolerance and uncertainty in the ten-specimen pilot. (a) Five cast fracture loads (crosses) and five attained SLM loads without fracture (circles and right-censoring arrows). The dotted line marks τ = 4498.36 N. (b) Observed fracture-free proportions across force, truncated at τ without extrapolation. (c) Failure proportions by τ with exact 95% Clopper–Pearson confidence intervals, n = 5 per workflow. No SLM fracture load is imputed.
Figure 2. Load tolerance and uncertainty in the ten-specimen pilot. (a) Five cast fracture loads (crosses) and five attained SLM loads without fracture (circles and right-censoring arrows). The dotted line marks τ = 4498.36 N. (b) Observed fracture-free proportions across force, truncated at τ without extrapolation. (c) Failure proportions by τ with exact 95% Clopper–Pearson confidence intervals, n = 5 per workflow. No SLM fracture load is imputed.
Preprints 241565 g002
Figure 3. Descriptive sensitivity of the capped-load contrast. (a) Contrast versus common force limit, calculated from all ten endpoint records. The vertical line marks the greatest cast fracture load. (b) All ten single-specimen omissions, keeping τ = 4498.36 N fixed; the dashed line denotes the complete-sample contrast. These diagnostics do not supply additional specimens or confidence intervals.
Figure 3. Descriptive sensitivity of the capped-load contrast. (a) Contrast versus common force limit, calculated from all ten endpoint records. The vertical line marks the greatest cast fracture load. (b) All ten single-specimen omissions, keeping τ = 4498.36 N fixed; the dashed line denotes the complete-sample contrast. These diagnostics do not supply additional specimens or confidence intervals.
Preprints 241565 g003
Figure 4. Analytical dimensional illustration for a centrally loaded, simply supported rectangular beam. (a) Relative linear stiffness and nominal stress versus relative section depth at fixed width, span, material, and force. (b) Relative stiffness and nominal stress versus relative span at fixed section, material, and force. Each curve is normalized to the reference geometry. These are elementary scaling laws, not measurements or a finite element model of the tested FDPs; the actual specimen dimensions were unavailable.
Figure 4. Analytical dimensional illustration for a centrally loaded, simply supported rectangular beam. (a) Relative linear stiffness and nominal stress versus relative section depth at fixed width, span, material, and force. (b) Relative stiffness and nominal stress versus relative span at fixed section, material, and force. Each curve is normalized to the reference geometry. These are elementary scaling laws, not measurements or a finite element model of the tested FDPs; the actual specimen dimensions were unavailable.
Preprints 241565 g004
Table 1. Reported experimental conditions and documentation limits.
Table 1. Reported experimental conditions and documentation limits.
Variable Available record
Experimental units Ten frameworks: five cast and five SLM; force/event data for all ten
Nominal geometry One posterior maxillary design with two abutments; one edentulous span
Material class Co–Cr in both workflows; commercial identities and compositions unverified
Cast pattern manufacture Digital workflow reported to use hyperDENT; milled wax, CORiTEC 250i dry
SLM processing Layerwise manufacture from Co–Cr powder; printer and process settings undocumented
Finishing and seating Finishing, airborne-particle abrasion, and visual seating checks
Loading system Zwick/Roell Z005; 5 kN load cell; 10 mm hemispherical indenter
Loading protocol 100 N preload at 10 mm/min, then 1 mm/min; ambient temperature
Test termination Fracture for cast specimens; nominal ceiling of 4501.34 N for SLM
Support and actual dimensions Common support reported; material, compliance, fixation, and metrology undocumented
SLM, selective laser melting; Co–Cr, cobalt–chromium. Conditions describe the available experimental documentation. Common design denotes intended geometry, not verified equivalence of finished connectors or specimen-specific protocol compliance.
Table 2. Individual specimen measurements and fracture status.
Table 2. Individual specimen measurements and fracture status.
Specimen Workflow Observed load (N) Event status Endpoint displacement (mm)
MELT_1 Cast 964.26 Fracture 0.871
MELT_2 Cast 1134.63 Fracture 1.649
MELT_3 Cast 938.53 Fracture 1.804
MELT_4 Cast 987.46 Fracture 1.425
MELT_5 Cast 1037.48 Fracture 1.547
SINT_1 SLM 4500.07 Right censored 2.225
SINT_2 SLM 4499.41 Right censored 1.942
SINT_3 SLM 4499.62 Right censored 2.053
SINT_4 SLM 4498.36 Right censored 1.897
SINT_5 SLM 4501.34 Right censored 2.125
Right censored denotes no macroscopic fracture at the attained load. Displacement refers to fracture for casting and test termination for SLM; n = 5 per workflow, with no displacement imputation.
Table 3. Primary comparison at the common load limit.
Table 3. Primary comparison at the common load limit.
Measure Cast SLM
Measured fracture load, mean ± SD (N) 1012.47 ± 77.40 Not estimable
Capped load at τ, mean (N) 1012.47 4498.36
Failures by τ, n/N (%) 5/5 (100) 0/5 (0)
Exact 95% CI for failure (%) 47.8–100 0–52.2
τ = 4498.36 N. Between-workflow capped mean contrast: 3485.89 N; ratio: 4.44; exact two-sided permutation p = 0.00794 (2/252 allocations). Each workflow contains five specimens. Failure confidence intervals are Clopper–Pearson intervals. SD, sample standard deviation; CI, confidence interval.
Table 4. Exploratory posterior sensitivity of failure probability by τ.
Table 4. Exploratory posterior sensitivity of failure probability by τ.
Common prior Mean
θCast
Mean
θSLM
Median D (95% CrI) P(D > 0 | data)
Beta(0.5, 0.5) 0.917 0.083 0.874 (0.453–0.995) 99.98%
Beta(1, 1) 0.857 0.143 0.745 (0.301–0.960) 99.89%
Beta(2, 2) 0.778 0.222 0.574 (0.146–0.860) 99.49%
D = θCast − θSLM; positive values favor SLM. τ = 4498.36 N, selected post hoc. Each group contains five binary outcomes. CrI, equal-tailed credible interval. Model assumptions are detailed in Section 2.5.3.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.