Design and Manufacture of a Test Device for Small Vacuum Pumps

A peer-reviewed article of this preprint also exists.

Submitted:

06 December 2022

Posted:

20 December 2022

You are already at the latest version

Abstract
Vacuum pump wear is the most prevalent failure mode of the IBA Synthera® automated radiochemistry system. Rebuilding or replacing the pump causes equipment downtime and increases the radiation exposure of the service personnel. We built a dedicated test device to assess new or rebuilt pumps prior to installation, thus reducing downtime and radiation exposure during repairs. The Testbed incorporates a microprocessor that actuates the pump, valves, pressure sensor, and communicates with the user through lights, buttons, and an alphanumeric screen. The Testbed increases productivity and safety in the radiochemistry laboratory.