Preclinical Safety and Efficacy Evaluation of A Novel Phosphorylcholine-Modified Flow Diverter

Submitted:

19 November 2026

Posted:

19 November 2026

You are already at the latest version

Abstract
Background: Flow diverters are widely used to treat intracranial aneurysms, but thrombus formation triggered by foreign-body irritation remains a major complication. SkyVega is a new-generation flow diverter with phosphorylcholine surface modification designed to improve aneurysm occlusion and hemocompatibility. We evaluated aneurysm occlusion, branch vessel preservation, and long-term vascular tissue responses to the phosphorylcholine-modified SkyVega flow diverter in an elastase-induced rabbit aneurysm model. Methods: Twenty-six New Zealand white rabbits were assigned to 30-day, 90-day and 180-day follow-up cohorts. The modified flow diverters were implanted in the parent arteries and abdominal aortas in the experimental group. Unmodified flow diverters implanted in parent arteries served as the 180-day control. Aneurysm occlusion and vessel patency were assessed by digital subtraction angiography (DSA), and biocompatibility was evaluated by histopathology. Results: DSA showed progressive aneurysm occlusion in the experimental group. Parent arteries and covered branches remained patent for up to 180 days, and local tissue reactions were mild in all groups. Phosphorylcholine-modified flow diverters were associated with thinner, concentric, and uniform neointimal growth, whereas controls showed eccentric neointimal hyperplasia and one case of in-stent thrombosis. Conclusion: SkyVega achieves stable, progressive aneurysm occlusion with favorable branch vessel preservation. The phosphorylcholine coating promotes orderly concentric endothelialization and may reduce the risk of thrombosis.
Keywords: 
;  ;  ;  ;  

1. Introduction

Flow diverters are now well established for the treatment of intracranial aneurysms [1,2]. As permanent foreign implants, however, they may irritate or injure the vessel wall, triggering platelet activation and aggregation and thereby promoting thrombus formation [3].Phosphorylcholine-modified flow diverters are grafted with phosphorylcholine groups on the surface to mimic the biological properties of cell membranes.This modification is intended to improve hemocompatibility, reduce platelet adhesion, lower the risk of thrombosis, and promote early neointimal coverage [4,5,6].
After a flow diverter is delivered to the parent artery and deployed across the aneurysm neck, its dense meshwork redirects blood flow away from the aneurysm sac and toward the distal vessel through the aneurysm neck. This markedly reduces flow impact and shear stress on the aneurysm wall, slows and stagnates the intra-aneurysmal flow, promotes thrombus formation, and induces spontaneous aneurysm occlusion by remodeling intra-aneurysmal hemodynamics [2]. Nevertheless, when branch vessels exist at the target lesion, the flow diverter may compromise branch perfusion [7,8,9].
SkyVega (Skynor Medical (Shanghai) Co., Ltd.) is a novel phosphorylcholine-modified flow diverter. Its long-term vascular tissue responses, neointimal growth after implantation, and the relationship between metal coverage and branch vessel patency in aneurysm models had not been systematically evaluated in vivo. Accordingly, we implanted SkyVega flow diverters in the parent arteries of elastase-induced aneurysms in New Zealand White rabbits and used angiography to assess occlusion and histopathology to evaluate tissue responses and the long-term effects of the phosphorylcholine coating on neointimal growth and thrombus inhibition. Flow diverters were also implanted in the abdominal aorta to examine whether metal coverage affects branch vessel perfusion.

2. Materials and Methods

2.1. Test Device

SkyVega is a new-generation flow diverter braided from 52 nitinol wires and 12 platinum-tungsten radiopaque wires, with a metal coverage rate of approximately 40%. A 3-5nm phosphorylcholine coating is covalently grafted onto the device surface, comparable to the 3 nm coating of the Medtronic Pipeline Shield [10]. The commercially available Tubridge flow diverter (MicroPort NeuroTech Co., Ltd.), an unmodified device, served as the control at 180 days to compare long-term in vivo performance.

2.2. Aneurysm Creation and Study Design

Elastase-induced aneurysms in New Zealand White rabbits reproduce the hemodynamic features, geometry, and histology of human intracranial aneurysms [11,12,13,14] and are therefore a well-established model for testing flow diverters. Aneurysms were created as previously described and were allowed to mature for 1 month before flow diverter implantation.
The study was approved by the institutional review board and conducted at PharmaLegacy Diagnostic Laboratories (Shanghai) Co., Ltd. in accordance with institutional animal care guidelines. Aneurysms were created at the origin of the right common carotid artery in 26 New Zealand White rabbits (age 180–210 days, body weight approximately 4 kg, either sex). According to the predefined study protocol, 22 rabbits were allocated to the SkyVega test group and followed up at 30 days (n = 7), 90 days (n = 7), or 180 days (n = 8), while an additional four rabbits received Tubridge flow diverters as controls for the 180-day evaluation. Animals were euthanized upon reaching their designated follow-up time points. Each rabbit in the SkyVega test group received one SkyVega flow diverter in the parent artery to assess aneurysm occlusion and histopathological responses, and a second SkyVega in the abdominal aorta to evaluate branch vessel preservation. All 26 animals enrolled in the study were included in the final analysis, and no animals or data points were excluded. Group allocation was based on the predefined study design rather than randomization. No randomization or blinding procedures were implemented, which are acknowledged as limitations of the study design.
All rabbits received dual antiplatelet therapy with aspirin (5 mg/kg) and clopidogrel (5 mg/kg) orally, beginning 3 days before implantation and continuing until euthanasia.

2.3. Flow Diverter Implantation and DSA Imaging

All procedures were performed under general anesthesia. Atropine (0.04 mg/kg) was given subcutaneously, followed by intramuscular Zoletil 50 (6 mg/kg) for sedation and induction; anesthesia was maintained with 1-5% isoflurane under mechanical ventilation.
The right common femoral artery was surgically exposed, and a 5F introducer sheath was placed. Microcatheters were used to deliver flow diverters to the right subclavian artery, and the flow diverter was deployed across the aneurysm neck, or through the right vertebral artery to cover the aneurysm orifice. For abdominal aortic implantation, the procedure was identical except that the device was positioned to cover the ostia of at least one pair of lumbar arteries. Flow diverter diameters ranged from 3.0 to 4.5 mm and lengths from 15 to 20 mm. A total of 44 SkyVega devices were implanted.
Digital subtraction angiography (DSA) was performed before the procedure, immediately after implantation, and at each scheduled follow-up. Preprocedural DSA was used to assess aneurysm and parent artery anatomy for flow diverter sizing. Immediate postprocedural DSA confirmed device position, apposition, filling of the aneurysm sac, and branch vessel patency. At each follow-up, aneurysm occlusion was graded with the Raymond-Roy classification [15], and parent artery stenosis and in-stent thrombosis were recorded.

2.4. Histopathology

At each follow-up, the stented parent artery segments containing the aneurysm neck and surrounding tissues were harvested for histopathological evaluation. The following parameters were scored: inflammatory cell infiltration (0, normal; 1, mild; 2, moderate; 3, moderate to severe); injury to the intima and internal elastic lamina (IEL) (0, normal; 1, injury confined between the intima and the IEL; 2, injury extending between the IEL and the adventitia; 3, injury involving the adventitia); fibrosis (0, normal; 1, <25% of the vessel wall area; 2, 25-50% of the vessel wall area; 3, >50% of the vessel wall area); granulation tissue (0, normal; 1, small foci in the adventitia; 2, moderate adventitial granulation tissue; 3, abundant granulation tissue or replacement of the adventitia by granulation tissue); endothelialization (0, no endothelial coverage; 1, <25% of the device covered; 2, 25-50% covered; 3, >50% covered); and thrombus formation (0, normal; 1, thrombus occupying <25% of the lumen area; 2, thrombus occupying 25-50% of the lumen area; 3, thrombus occupying >50% of the lumen area). The percentage of neointimal coverage of the aneurysm neck was also measured.

2.5. Statistical Method

Continuous variables were expressed as mea±standard deviation (SD). Histopathological findings were evaluated descriptively based on representative tissue sections.

3. Results

3.1. DSA Findings

3.1.1. Parent Artery and Aneurysm Occlusion

Immediately after SkyVega implantation in the parent artery, marked flow stasis was observed in every aneurysm. At 30 days, 4 of 7 animals had complete occlusion (Raymond-Roy grade 1) and 3 of 7 had residual filling (Raymond-Roy grade 2). At 90 days, 5 of 7 aneurysms were completely occluded (Raymond-Roy grade 1) and 2 of 7 showed minor residual filling (Raymond-Roy grade 2). At 180 days, 6 of 8 aneurysms were completely occluded (Raymond-Roy grade 1) and 2 of 8 showed slight residual filling (Raymond-Roy grade 2). All parent arteries remained patent, with no significant stenosis, at every time point.
DSA images of a representative animal in the experimental group before implantation and at each follow-up time point are shown in Figure 1. In this case, the aneurysm was clearly visualized before treatment, with rapid contrast transit, a short dwell time, and complete opacification of the sac. Immediately after deployment, the device fully covered the aneurysm neck without migration, and a distinct flow-diversion effect was evident: reduced inflow into the sac, prolonged filling time, prolonged contrast stasis within the sac, and slower flow. The parent artery remained patent, and no in-stent thrombosis was identified. At 30 days, no contrast retention was seen within the aneurysm, indicating complete occlusion, and the parent artery remained patent and thrombosis-free. The 90-day and 180-day images were similar, confirming sustained occlusion, a patent parent artery, and no visible thrombus.
In the control group, DSA findings paralleled those in the experimental group. Before treatment, contrast passed rapidly through the sac with a short dwell time and opacified the entire aneurysm. Immediately after implantation, the device fully covered the neck without migration and produced a clear flow-diversion effect, with markedly reduced inflow and prolonged filling. All parent arteries were patent, with no in-stent thrombosis. At 180 days, no contrast retention was seen in the aneurysm, consistent with complete occlusion (Raymond-Roy grade 1) .

3.1.2. Abdominal Aorta Implantation

Figure 2 shows DSA after implantation of SkyVega flow diverters in the rabbit abdominal aorta. All branch vessels were patent before deployment and remained patent afterward, with good opacification of the distal vasculature. At 180 days, the covered branches were still patent and the distal vessels remained fully opacified. No acute or delayed branch occlusion was observed on any immediate or follow-up angiogram. Thus, even with relatively high metal coverage, the device remodels blood flow without materially compromising branch perfusion.

3.2. Histopathology

Histological findings in the experimental group and control group are shown in Figure 3. At 30, 90, and 180 days, the stented parent artery at the aneurysm neck showed no prominent inflammatory infiltration and no evident injury between the intima and the IEL, with only minimal fibrosis and granulation tissue; most of the circumference was endothelialized, and no thrombus was seen. At 30 days, the sac contained mainly fresh, unorganized thrombus: loose, fragile clots composed of a fibrin mesh trapping numerous erythrocytes, platelets, and leukocytes, with no ingrowth or replacement by granulation tissue. Approximately 75.84±37.69% of the aneurysm neck was covered by neointima. By 90 days, organized thrombus predominated: endothelial cells and fibroblasts had grown into the clot to form granulation tissue, progressively replacing it with a more stable structure, and neointimal coverage of the neck was approximately 85.87±15.97%. By 180 days, the sac was filled with collagenized connective tissue, indicating completion of organization; the thrombus had been remodeled into dense connective tissue containing fibroblasts, abundant collagen and elastic fibers, and extracellular matrix. Approximately 91.83±11.50% of the neck was covered by a thin, concentric, uniform neointima. Inflammatory infiltration, vascular injury, fibrosis, and granulation tissue did not increase over time and remained minimal at all time points.
At 30 days, the flow diverter at the aneurysm neck was largely covered by neointima, which formed a biological barrier preventing blood from entering the aneurysm sac. This finding corresponded to the angiographic result: no contrast retention was seen in most aneurysm sacs at 30 days. At 90 days, the flow diverter at the aneurysm neck was further covered by neointima, which continued to reduce hemodynamic impact on the aneurysm sac. By 180 days, the device at the neck was almost completely covered by neointima, the aneurysm was further occluded, and the intra-aneurysmal thrombus had been fully organized, evolving from unorganized clot into stable connective tissue.
In the 180-day control group (Figure 3), the stented segment and surrounding tissues showed mild inflammatory infiltration and mild injury between the intima and the IEL, with minimal fibrosis and granulation tissue. The vessel was fully endothelialized, the sac contained organized thrombus, and about 92.70±12.70% of the neck was covered by neointima. However, the control neointima was markedly thicker and distributed eccentrically compared with the experimental group. No significant between-group differences were seen in inflammation, vascular injury, or fibrosis.
One of the four control specimens showed in-stent thrombosis (Figure 4). The device was well apposed to the vessel wall and retained its configuration, but a large thrombus occupied the lumen, resulting in significant in-stent restenosis.

4. Discussion

Flow diverters have become one of the main therapeutic modalities for intracranial aneurysms. Their therapeutic effect relies primarily on remodeling hemodynamics within the parent artery and aneurysm, and the complete occlusion rate increases over time [16,17]. This animal study confirmed the efficacy of the novel phosphorylcholine-modified SkyVega flow diverter. The hemodynamic effect was observed immediately after deployment, with reduced intra-aneurysmal flow velocity and prolonged contrast dwell time. With longer follow-up, all aneurysms achieved complete or near-complete occlusion by 180 days, while only a few aneurysms exhibited residual contrast opacification. No substantial in-stent restenosis or angiographically visible thrombus was detected at any time point in the experimental group. And angiographic outcomes were comparable to those in the control group. This progressive occlusion reflects the hemodynamic consequences of the low-porosity mesh: a marked reduction in inflow velocity and suppression of intra-aneurysmal vortices. In parallel, endothelial cells migrate along the device struts, endothelialize the implant, and seal the aneurysm neck, further blocking inflow. Stagnant blood within the sac promotes platelet and fibrin deposition and thrombus formation, which gradually organizes until the sac is filled with dense fibrous tissue [17].
Flow diverters are densely braided, high-metal-coverage meshes deployed across the aneurysm neck to prevent blood from entering the sac. Intracranial aneurysms, however, often arise at bifurcations within complex cerebrovascular anatomy, and that same high metal coverage may reduce branch perfusion or even occlude branches, raising the risk of ischemic stroke [18,19,20,21]. Many studies report that most covered branches remain patent after flow diversion, although a considerable minority develop reduced flow or occlusion and therefore warrant close attention [22,23,24]. Assessing branch patency after flow diverter placement is thus essential. A meta-analysis by Liu et al., covering 57 studies and 3,789 patients whose aneurysms were treated with branch-covered flow diverters, found a low risk of branch occlusion-related complications (<5%) [22]; Zhu et al. reported similar findings in an animal study [25]. In the present study, distal perfusion of the parent artery and of the abdominal aortic branches was preserved after implantation, and all covered branches remained patent throughout follow-up, including at 180 days. Within the limits of this experiment, the metal coverage of SkyVega did not impair branch blood flow.
The experimental group showed only mild histopathological responses-inflammation, vascular injury, fibrosis, and granulation tissue-at 30, 90, and 180 days, with no marked foreign-body reaction in the parent artery. At 180 days, neointimal hyperplasia was thin, the media was preserved, and the device was almost completely covered by endothelial tissue. Compared with controls, the neointima was thinner and circumferentially uniform, whereas controls showed uneven thickness around the circumference, with more pronounced hyperplasia at the aneurysm neck, and one control animal developed overt thrombus formation. Because the experimental group included 7-8 animals per time point and the 180-day control group only 4, the groups were unbalanced. Between-group histopathological differences are therefore described morphologically, without statistical testing.
Because of its limited spatial resolution and two-dimensional nature, DSA cannot detect microthrombi adherent to the device struts, which may therefore be underdiagnosed. Optical coherence tomography (OCT) is well established for identifying acute microthrombi in coronary interventions [26]. Marosfoi et al. [27]used OCT to detect thrombi 0.4-1.2 mm in diameter on flow diverter surfaces, appearing as irregular, layered thrombi across multiple slices without substantial luminal stenosis -findings consistent with the histology of our control group.
Matsuda et al. [28] compared neointimal growth after implantation of phosphorylcholine-coated Pipeline Shield and uncoated Pipeline Flex devices in a porcine carotid model, the rerults showed that neointima was asymmetric with the unmodified Pipeline Flex and Solitaire stents but concentric and uniform with modified Pipeline Shield. They concluded that neointimal hyperplasia is closely linked to early thrombus deposition on the struts, because neointima typically grows on thrombus already adherent to the struts-an observation that explains the eccentric, uneven neointima seen in our unmodified controls. By suppressing early thrombus adhesion, the phosphorylcholine coating appears to promote concentric, uniform neointimal growth, thereby limiting excessive late neointimal proliferation and reducing the risk of in-stent restenosis. Our results are consistent with these findings and further confirm that phosphorylcholine surface modification attenuates thrombus formation on flow diverters.
Although this study relied solely on DSA for macroscopic follow-up and may therefore have missed microthrombi on the struts, the combination of DSA and terminal histopathology allows macroscopic imaging and microscopic tissue findings to be cross-validated. DSA provides a global assessment of parent artery patency, aneurysm occlusion grade, and long-term hemodynamic remodeling, but cannot resolve micrometer-scale thrombi. Histopathology directly demonstrates thrombus distribution, endothelial coverage, neointimal thickness, and perivascular inflammation, thereby detecting lesions invisible to DSA and strengthening the reliability of the results.

5. Conclusions and Limitations

The SkyVega phosphorylcholine-modified flow diverter rapidly remodels parent artery hemodynamics and promotes progressive aneurysm occlusion while maintaining parent artery patency. It disturbs branch perfusion only minimally and elicits mild local vascular tissue reactions. The phosphorylcholine coating promotes orderly, uniform neointimal growth, inhibits early thrombus adhesion, limits excessive neointimal proliferation, and may reduce the risk of thrombosis.
This study has several limitations. Firstly, the sample size was small, precluding statistical comparison between the phosphorylcholine-modified and control devices. Secondly, follow-up imaging relied solely on DSA, which cannot detect early microthrombi or quantify the advantages of surface modification in thrombus suppression and neointimal regulation. Larger studies incorporating intravascular OCT are needed to address these questions.

Author Contributions

Conceptualization, methodology, data collection, angiographic image interpretation, original draft writing, formal analysis, data validation, Z.Z; Supervision, project administration, manuscript review and revision, correspondence work, S.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding .

Institutional Review Board Statement

The animal study was reviewed and approved by the Institutional Animal Care and Use Committee of [PharmaLegacy Diagnostic Laboratories (Shanghai) Co., Ltd] (Approval No. [PLJC-0092-1]). All procedures were conducted in accordance with the approved protocol and relevant animal welfare regulations.

Data Availability Statement

The datasets generated and/or analyzed during the current study are not publicly available but are available from the corresponding author on reasonable request and with appropriate agreements in place.

Conflicts of Interest

Zhenxin Zhao is the legal representative and a beneficial owner of Skynor Medical (Shanghai) Co., Ltd., the manufacturer of the flow-diverting device evaluated in this study. The remaining authors declare no conflicts of interest.

References

  1. Gaub, M.; Murtha, G.; Lafuente, M.; et al. Flow diversion for endovascular treatment of intracranial aneurysms: past, present, and future directions. J. Clin. Med. 2024, 13, 4167. [Google Scholar] [CrossRef] [PubMed]
  2. Al-Mufti, F.; Amuluru, K.; Gandhi, C.D.; et al. Flow Diversion for Intracranial Aneurysm Management: A New Standard of Care. Neurotherapeutics 2016, 13, 582–589. [Google Scholar] [CrossRef] [PubMed]
  3. Toma, A.; Essibayi, M.A.; Osama, M.; et al. Managing thrombosis risk in flow diversion: A review of antiplatelet approaches. Neuroradiol. J. 2025, 38, 588–598. [Google Scholar] [CrossRef] [PubMed]
  4. EINaamani, K.; Mastorakos, P.; Yudkoff, C.J.; et al. Comparison of PED/PED Flex and PED Shield in the treatment of unruptured intracerebral aneurysms. J. Neurosurg. 2023, 140, 436–440. [Google Scholar] [CrossRef] [PubMed]
  5. Ma, L.; Hoz, S.S.; Al-Bayati, A.R.; et al. Flow diverters with surface modification in patients with intracranial aneurysms: a systematic review and meta-analysis. WorldNeurosurg 2024, 185, 320–326.e17. [Google Scholar] [CrossRef] [PubMed]
  6. Starke, R.M.; Thompson, J.; Pagani, A.; et al. Preclinical safety and efficacy evaluation of the Pipeline Vantage Embolization Device with Shield Technology. J. NeuroInterventional Surg. 2020, 12, 981–986. [Google Scholar] [CrossRef] [PubMed]
  7. Hohenstatt, S.; Arrichiello, A.; Gonte, C.; et al. Branch vessel occlusion in aneurysm treatment with flow diverter stent. Acta Biomed. 2020, 91, e2020003. [Google Scholar] [CrossRef] [PubMed]
  8. Raymond, S.B.; Koch, M.J.; Rabinov, J.D.; et al. The Role of Collateral Circulation in Branch Vessel Occlusion After Flow Diversion. World Neurosurg. 2019, 124, e182–e187. [Google Scholar] [CrossRef] [PubMed]
  9. Wu, X.Z.; Tian, Z.B.; Li, W.Q.; et al. Patency of Branch Vessels After Pipeline Embolization: Comparison of Various Branches. Front Neurol. 2019, 10, 838. [Google Scholar] [CrossRef] [PubMed]
  10. Girdhar, G.; Andersen, A.; Pangerl, E.; et al. Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model. J. BioMed Mater. Res. A 2018, 106, 3195–3202. [Google Scholar] [CrossRef] [PubMed]
  11. Brinjikji, W.; Ding, Y.H.; Kallmes, D.F.; et al. From bench to bedside: utility of the rabbit elastase aneurysm model in preclinical studies of intracranial aneurysm treatment. J. NeuroInterventional Surg. 2016, 8, 521–525. [Google Scholar] [CrossRef] [PubMed]
  12. Altes, T.A.; Cloft, H.J.; Short, J.G.; et al. Creation of saccular aneurysms in the rabbit: a model suitable for testing endovascular devices. Am. J. Roentgenol. 2000, 174, 349–354. [Google Scholar] [CrossRef] [PubMed]
  13. Ding, Y.H.; Danielson, M.A.; Kadirvel, R.; et al. Modified technique to create morphologically reproducible elastase-induced aneurysms in rabbits. Neuroradiology 2006, 48, 528–532. [Google Scholar] [CrossRef] [PubMed]
  14. Zeng, Z.; Kallmes, D.F.; Durka, M.J.; et al. Hemodynamics and anatomy of elastase-induced rabbit aneurysm models: similarity to human cerebral aneurysms? Am. Neuroradiol. 2011, 32, 595–601. [Google Scholar] [CrossRef] [PubMed]
  15. Hanel, R.A.; Cortez, G.M.; Lopes, D.K.; et al. Prospective study on embolization of intracranial aneurysms with the pipeline device (PREMIER study): 3-year results with the application of a flow diverter specific occlusion classification. J. NeuroIntervent Surg. 2023, 15, 248–254. [Google Scholar] [CrossRef] [PubMed]
  16. Korkmazer, B.; Kocak, B.; Islak, C.; et al. Long-term results of flow diversion in the treatment of intracranial aneurysms: a retrospective data analysis of a single center. Acta Neurochir. 2019, 161, 1165–1173. [Google Scholar] [CrossRef] [PubMed]
  17. Lee, J.Y.; Cho, Y.D.; Kang, H.S.; et al. Healing of Aneurysm after Treatment Using Flow Diverter Stent: Histopathological Study in Experimental Canine Carotid Side Wall Aneurysm. J. Korean Neurosurg. Soc. 2020, 63, 34–44. [Google Scholar] [CrossRef] [PubMed]
  18. Brinjikji, W.; Lanzino, G.; Cloft, H.J.; et al. Patency of the posterior communicating artery after flow diversion treatment of internal carotid artery aneurysms. Clin. Neurol. Neurosurg. 2014, 120, 84–8. [Google Scholar] [CrossRef]
  19. Cagnazzo, F.; Lefevre, P.H.; Mantilla, D.; et al. Patency of the supraclinoid internal carotid artery branches after flow diversion treatment: a meta-analysis. J. Neuroradiol. 2019, 46, 9–14. [Google Scholar] [CrossRef] [PubMed]
  20. Rangel-Castilla, L.; Munich, S.A.; Jaleel, N.; et al. Patency of anterior circulation branch vessels after Pipeline embolization: longer-term results from 82 aneurysm cases. J. Neurosurg. 2017, 126, 1064–1069. [Google Scholar] [CrossRef] [PubMed]
  21. Adeeb, N.; Griessenauer, C.J.; Dmytriw, A.A.; et al. Risk of branch occlusion and ischemic complications with the Pipeline embolization device in the treatment of posterior circulation aneurysms. AJNR Am. J. Neuroradiol. 2018, 39, 1303–1309. [Google Scholar] [CrossRef] [PubMed]
  22. Liu, J.Y.; Cao, F.; ZhenMei, N.B.; et al. Flow-diverter stents in intracranial aneurysm treatment: impact on covered cerebral artery branches. Int. J. Surg. 2024, 110, 53–65. [Google Scholar] [CrossRef] [PubMed]
  23. Nelson, P.K.; Lylyk, P.; Szikora, I.; et al. The Pipeline Embolization Device for the intracranial treatment of aneurysms trial. AJNR Am. J. Neuroradiol. 2011, 32, 34–40. [Google Scholar] [CrossRef] [PubMed]
  24. Chalouhi, N.; Daou, B.; Kung, D.; et al. Fate of the ophthalmic artery after treatment with the Pipeline embolization device. Neurosurgery 2015, 77, 581–584. [Google Scholar] [CrossRef] [PubMed]
  25. Zhu, Y.N.; Zeng, F.Y.; Liu, J.; et al. Evaluation of the EMBOPIPE flow diverter device: in vivo and in vitro experiments. Chin. Neurosurg. J. 2024, 10, 8. [Google Scholar] [CrossRef] [PubMed]
  26. Kume, T.; Akasaka, T.; Kawamoto, T.; et al. Assessment of coronary arterial thrombus by optical coherence tomography. Am. J. Cardiol. 2006, 97, 1713–1717. [Google Scholar] [CrossRef] [PubMed]
  27. Marosfoi, M.; Clarencon, F.; Langan, E.T.; et al. Acute thrombus formation on phosphorilcholine surface modified flow diverters. J. Neurointerv Surg. 2017, 10, 406–411. [Google Scholar] [CrossRef] [PubMed]
  28. Matsuda, Y.; Chung, J.; Lopes, D.K. Analysis of neointima development in flow diverters using optical coherence tomography imaging. J. Neurointerv Surg. 2017, 10, 162–167. [Google Scholar] [CrossRef] [PubMed]
Figure 1. DSA images of SkyVega implanted in the parent artery.
Figure 1. DSA images of SkyVega implanted in the parent artery.
Preprints 245086 g001
Figure 2. Angiograms of the SkyVega at 180-day after abdominal aorta implantation.
Figure 2. Angiograms of the SkyVega at 180-day after abdominal aorta implantation.
Preprints 245086 g002
Figure 3. Histological specimens (HE staining) from the experimental and control groups.
Figure 3. Histological specimens (HE staining) from the experimental and control groups.
Preprints 245086 g003
Figure 4. In-stent thrombosis in the control group.
Figure 4. In-stent thrombosis in the control group.
Preprints 245086 g004
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.