Influence of Corneal Hysteresis on Agreement Between Corneal-Compensated, Non-Contact, and Rebound Tonometry

Submitted:

31 October 2026

Posted:

03 November 2026

You are already at the latest version

Abstract
Background/Objectives: Intraocular pressure measurements are influenced by corneal biomechanical properties, but their contribution to discrepancies between different tonometric methods remains poorly understood. This study evaluated the influence of corneal hysteresis (CH) on differences between corneal-compensated IOP (IOPcc), non-contact (NCT), and rebound tonometry, while adjusting for age and central corneal thickness (CCT). Methods: This cross-sectional study included 134 eyes. IOP was measured using NCT, iCare Pro rebound tonometry, and the Ocular Response Analyzer, which provided IOPcc and CH. Central corneal thickness was measured by spectral-domain optical coherence tomography. Agreement was assessed using regression-based Bland–Altman analysis. Multivariable linear regression evaluated CH, CCT, and age as determinants of the tonometric differences. Inter-method differences were also compared across CH tertiles. Results: Mean biases were +0.15 mmHg for IOPcc–NCT, +1.68 mmHg for IOPcc–iCare, and +1.18 mmHg for NCT–iCare. After adjustment for CCT and age, each 1-mmHg decrease in CH was associated with a 1.56-mmHg increase in the IOPcc–NCT difference and a 2.23-mmHg increase in the IOPcc–iCare difference (both p<0.001). Differences involving IOPcc progressively increased across decreasing CH tertiles. In the lowest CH tertile, the mean IOPcc–NCT difference was 2.50 ± 2.05 mmHg, whereas the mean IOPcc–iCare difference reached 5.51 ± 3.11 mmHg, indicating clinically relevant discrepancies in eyes with low CH. Conclusions: Lower corneal hysteresis was associated with greater discrepancies between biomechanically compensated and the other tonometric methods. While mean differences were small in the overall cohort, they became clinically relevant in eyes with low CH. These findings suggest that considering CH in addition to CCT may support a more accurate interpretation of IOP measurements in biomechanically vulnerable eyes.
Keywords: 
;  ;  ;  ;  ;  

1. Introduction

Accurate intraocular pressure (IOP) measurement is central to glaucoma diagnosis, risk assessment, and therapeutic monitoring. Decisions regarding treatment initiation, escalation, and evaluation of disease control are frequently based on relatively small changes in measured IOP. Consequently, even modest systematic differences between tonometric methods may influence clinical interpretation, especially when measurements obtained with different devices are compared over time.
Although Goldmann applanation tonometry remains the clinical reference standard, non-contact tonometry (NCT) and rebound tonometry are widely used in routine practice because of their practicality and suitability for different clinical settings and patient populations [1]. In routine practice, values obtained with these instruments may be interpreted as broadly comparable. However, agreement at the population level does not necessarily imply interchangeability in individual eyes, and clinically meaningful discrepancies may occur particularly in eyes with altered corneal properties [2,3,4,5,6].
Tonometric techniques estimate IOP indirectly by analysing the corneal response to an externally applied force or transient mechanical interaction. The measured value therefore depends not only on the pressure within the eye, but also on how the cornea responds to the specific measurement principle of each device. This response is determined by the interaction among IOP, corneal geometry, intrinsic biomechanical properties, and the magnitude and temporal characteristics of the applied force [3]. Because different tonometers interact with the cornea in different ways, they may not be equally affected by variations in corneal behaviour. Thus, two eyes with the same true IOP may yield different measured values if their corneas differ in thickness, rigidity and viscoelasticity. Similarly, the same eye may produce different IOP estimates when examined with devices based on different measurement principles.
The influence of the cornea on tonometry has been interpreted primarily in relation to central corneal thickness (CCT). Nevertheless, thickness represents only one aspect of corneal behaviour. Corneas with similar CCT may have substantially different deformation responses because of differences in tissue composition, collagen architecture, viscoelasticity and age-related or ocular pathology changes. Despite increasing recognition of these factors, corneal biomechanics remain less integrated into clinical assessment. In particular, limited evidence is available regarding the extent to which biomechanical variability explains disagreement among tonometric methods commonly used.
Corneal hysteresis (CH), measured with the Ocular Response Analyzer (ORA), reflects the viscoelastic capacity of the cornea to absorb and dissipate mechanical energy. The ORA also provides corneal-compensated intraocular pressure (IOPcc), an estimate designed to reduce the influence of corneal biomechanical characteristics on IOP measurement [1,2]. CH has increasingly been recognized as a clinically relevant biomarker, both for interpreting IOP and for assessing susceptibility to glaucomatous damage [2,3,6,7]. However, the extent to which CH accounts for discrepancies between IOPcc and measurements obtained using conventional tonometric methods remains insufficiently understood.
Previous studies comparing ORA-derived measurements, non-contact tonometry, rebound tonometry, and Goldmann applanation tonometry have reported variable levels of agreement, with both overestimation and underestimation depending on the device, population, and ocular characteristics [7,8,9,10,11]. Most investigations have focused on average inter-device agreement across entire study populations. This approach may obscure clinically important differences in subgroups with reduced corneal biomechanical resilience.
The aim of this study was to evaluate the influence of corneal hysteresis on the differences between IOPcc, NCT, and rebound tonometry; to determine the independent contribution of CH after adjustment for CCT and age; and to assess inter-method discrepancies across CH strata, with particular emphasis on eyes within the lowest CH tertile.

2. Materials and Methods

2.1. Study Design and Participants

This prospective cross-sectional study included 134 eyes of patients evaluated at an ophthalmology clinic in Targu-Mures, Romania, between March and July 2026. Exclusion criteria included ocular infection, central corneal opacities, keratoconus, previous corneal or refractive surgery, and current topical ocular or systemic corticosteroid treatment. Demographic data, intraocular pressure (IOP) measurements, corneal biomechanical parameters, and ocular biometric characteristics were collected for all participants.

2.2. Tonometric and Biomechanical Measurements

IOP measurements were obtained using three different tonometric methods: NCT was performed with an air puff tonometer (Topcon CT-1P, Topcon Corporation, Tokyo, Japan); rebound tonometry using iCare Pro (iCare Finland Oy, Helsinki, Finland); and the Ocular Response Analyzer (ORA; Reichert Technologies, Depew, NY, USA). For each instrument, three consecutive measurements were obtained. If the difference between measurements exceeded 2 mmHg, additional measurements were performed until three consecutive values within 2 mmHg were achieved. All measurements were obtained at approximately the same time of day to minimize the potential influence of diurnal variation in intraocular pressure.
Central corneal thickness was obtained using anterior pachymetry with spectral-domain optical coherence tomography (RTVue XR Avanti, Optovue Inc., Fremont, CA, USA). The average of three final measurements was taken for analysis.

2.3. Statistical Analysis

Continuous variables were expressed as mean ± standard deviation (SD). Ranges were additionally reported for tonometric parameters.

2.3.1. Agreement Analyses

Agreement between tonometric methods was assessed using Bland–Altman analysis. Mean bias and 95% limits of agreement (LoA) were calculated for each comparison. To evaluate proportional bias, regression-based Bland–Altman analyses were performed by regressing the inter-method difference against the mean IOP of the compared methods. Regression-based Bland–Altman analysis was selected because preliminary analyses indicated proportional bias, suggesting that constant limits of agreement might not adequately describe agreement across the full range of IOP values.

2.3.2. Multivariable Regression Analysis

Multivariable linear regression analysis was performed to identify independent determinants of inter-method differences. Separate regression models were fitted for each inter-tonometer difference (IOPcc–NCT, IOPcc–iCare, and NCT–iCare), with CH, CCT, and age included as independent variables. Regression coefficients (β), 95% confidence intervals (CI), and coefficients of determination (R²) were reported.

2.3.3. Corneal Biomechanical Stratification

To further evaluate the influence of corneal biomechanics on inter-tonometer agreement, eyes were stratified into tertiles according to CH based on the empirical distribution of CH values. Mean inter-method differences (IOPcc–NCT, IOPcc–iCare, and NCT–iCare) were compared across CH tertiles.
The normality of inter-method differences within each tertile was assessed using the Shapiro–Wilk test, while homogeneity of variances was evaluated using the Brown–Forsythe test. As the assumptions for parametric analysis were satisfied, one-way analysis of variance (ANOVA) was used to compare mean inter-method differences across CH tertiles. Pairwise comparisons were performed using the Bonferroni post-hoc test. Statistical significance was defined as a two-sided p value < 0.05.
All statistical analyses were performed using complete-case data. Consequently, the number of observations varied across analyses according to data availability for the variables included in each model.
All analyses were performed using Stata Statistical Software 17 (StataCorp LLC, College Station, TX, USA).

2.4. Ethics Approval

The study was conducted in accordance with the Declaration of Helsinki and approved by the ethics committee of the hospital where the study took place (2374/18.03.2026). Written informed consent was obtained from all subjects before procedures.

3. Results

A total of 134 eyes were included in the analysis. The mean age was 73.2 ± 5.4 years; baseline demographic, ocular biometric, biomechanical, and tonometric characteristics are summarized in Table 1.

3.1. Bland–Altman Agreement Analysis

Bland–Altman analysis demonstrated small mean biases but relatively wide limits of agreement between the tonometric methods. Regression-based Bland–Altman analysis revealed different patterns of proportional bias (Table 2, Figure 1). The mean bias between IOPcc and NCT was minimal (+0.15 mmHg), whereas IOPcc measured higher IOP values than iCare by an average of 1.68 mmHg. NCT also yielded slightly higher IOP measurements than iCare (mean bias +1.18 mmHg). Significant proportional bias was observed for the IOPcc–NCT (β = 0.341, p < 0.001) and NCT–iCare (β = −0.263, p = 0.004) comparisons. In contrast, no significant proportional bias was detected between IOPcc and iCare (β = 0.142, p = 0.155).

3.2. Multivariable Linear Regression

Multivariable linear regression analysis (Table 3) showed that corneal hysteresis was an independent determinant of the differences between IOPcc and both NCT and iCare measurements. After adjustment for central corneal thickness and age, each 1-mmHg decrease in CH was associated with a 1.56-mmHg increase in the IOPcc–NCT difference and a 2.23-mmHg increase in the IOPcc–iCare difference.
Central corneal thickness remained independently associated with the IOPcc–NCT and NCT–iCare differences, whereas age was not significantly associated with any inter-tonometer comparison. The models explained 44.0%, 52.3%, and 14.5% of the variance in the IOPcc–NCT, IOPcc–iCare, and NCT–iCare differences, respectively. Additional interaction analyses showed no significant interaction between CH and mean IOP for any inter-tonometer comparison (all interaction p > 0.05), indicating that the association between CH and inter-method differences was consistent across the observed IOP range.

3.3. Inter-Tonometer Differences According to Corneal Hysteresis

Inter-method differences varied according to corneal hysteresis (Figure 2). Differences involving IOPcc progressively increased with decreasing CH, whereas the difference between NCT and iCare remained relatively stable across tertiles. In eyes within the lowest CH tertile, the mean IOPcc–NCT difference was 2.50 ± 2.05 mmHg, whereas the mean IOPcc–iCare difference reached 5.51 ± 3.11 mmHg, indicating clinically relevant discrepancies between biomechanically compensated and the other tonometric methods.

4. Discussion

The present study indicates a clear biomechanical gradient in tonometric agreement, with eyes exhibiting low corneal hysteresis showing the greatest discrepancies between biomechanically compensated and conventional tonometric measurements. Although overall agreement between IOPcc and conventional tonometric methods was within acceptable limits, regression-based Bland–Altman analysis revealed significant proportional bias. Consistent with this finding, CH emerged as the strongest independent determinant of the differences between IOPcc and both NCT and iCare, whereas the contribution of CCT was comparatively modest. Stratification according to CH further demonstrated that these discrepancies progressively increased as CH decreased. In eyes with low CH, IOPcc exceeded iCare by an average of 5.5 mmHg and NCT by 2.5 mmHg. These mean differences are clinically relevant, and the observed variability indicates that substantially larger discrepancies may occur in individual eyes.
Whether IOPcc more closely approximates the true intracameral pressure cannot be established by the present study because no direct manometric reference was available. However, existing evidence indicates that IOPcc is less influenced by corneal biomechanical properties than conventional tonometric methods, supporting its use as a biomechanically compensated estimate of IOP [12,13,14,15,16].
Previous comparative studies have yielded mixed results regarding the agreement between ORA, rebound tonometry, and non-contact tonometry [7,8,9,10,11]. Although systematic differences between devices have been described, the mechanisms underlying these discrepancies remain poorly understood. Our findings suggest that these differences are largely explained by corneal biomechanics, with CH accounting for substantially more variability than CCT after adjustment for age. This indicates that corneal biomechanical behaviour, rather than corneal thickness alone, is the principal factor governing agreement between biomechanically compensated and conventional tonometric methods. More specifically, low CH amplifies the discrepancy between biomechanically compensated and conventional tonometric methods. This distinction is clinically relevant because it implies that the apparent agreement between tonometers observed at the population level may conceal clinically meaningful underestimation of IOP in biomechanically vulnerable eyes.
The clinical implications of these findings are particularly important in glaucoma. Several studies confirmed low CH as an independent predictor of glaucoma development and visual field progression, often exceeding the prognostic value of CCT [17,18,19,20,21].
Because the cornea and peripapillary sclera share similar extracellular matrix composition, CH is proposed as a surrogate marker of the plasticity of the peripapillary sclera [21,22]. Eyes with low CH are therefore thought to exhibit reduced capacity to dissipate mechanical stress generated by intraocular pressure fluctuations, rendering the optic nerve head, particularly the lamina cribrosa, more susceptible to deformation and glaucomatous injury [22,23,24]. This concept is particularly relevant to the current biomechanical theory of glaucoma pathogenesis. While elevated intraocular pressure remains the principal modifiable risk factor, growing evidence indicates that the biological response of the optic nerve head to a given pressure load depends on the mechanical properties of the supporting connective tissues. In primary open-angle glaucoma, elevated IOP induces posterior deformation of the lamina cribrosa and subsequent axonal injury. In contrast, in normal-tension glaucoma, glaucomatous damage occurs despite IOP values within the statistically normal range. This observation has led to the hypothesis that biomechanical properties of the optic nerve head may influence individual susceptibility to pressure-related axonal injury [22].
Our findings provide an additional clinical perspective on this biomechanical model. The largest discrepancies between IOPcc and conventional tonometric methods occurred in eyes with low CH, indicating that biomechanical vulnerability is accompanied by greater divergence between biomechanically compensated and conventional IOP measurements. Consequently, conventional tonometry may yield lower IOP estimates than biomechanically compensated measurements in eyes with reduced CH. Whether such differences contribute to the apparent mismatch between measured IOP and glaucoma progression observed in some patients remains to be determined in longitudinal studies.
These findings suggest that incorporating CH into routine clinical assessment may improve the interpretation of IOP measurements and provide complementary information when assessing patients with ocular hypertension, normal-tension glaucoma, or apparent disease progression despite satisfactory IOP measured by conventional tonometry.
An important strength of this study is its comprehensive evaluation of the relationship between corneal biomechanics and tonometric measurements. Rather than simply comparing agreement between devices, the study demonstrates that inter-tonometer differences are systematically influenced by corneal hysteresis, indicating that ocular biomechanics are a major determinant of how different tonometric principles estimate intraocular pressure.
Several limitations should be acknowledged. First, the cross-sectional design precludes conclusions regarding causality or the impact of CH-related tonometric discrepancies on long-term glaucoma progression. Second, the study included a relatively modest sample which may limit generalizability. Third, Goldmann applanation tonometry was not included as a reference method because the primary objective was to compare routinely used and biomechanically compensated techniques rather than evaluate agreement with the clinical gold standard. Finally, although IOPcc is designed to reduce the influence of corneal biomechanics, it cannot be regarded as a direct measure of true intracameral pressure. Prospective longitudinal studies are needed to determine whether incorporating CH into routine interpretation of IOP measurements translates into improved glaucoma risk stratification and clinical outcome.

Author Contributions

F.V., E.A.M., and K.U.H. contributed to the conceptualization and study design. E.A.M. performed data collection, data analysis, and preparation of the original manuscript draft. F.V. and K.U.H. contributed to manuscript review and editing and provided supervision. All authors reviewed and approved the final version of the manuscript.

Funding

This work was supported by the George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, Research Grant number 795/3/22.01.2025.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the ethics committee of the hospital where the study took place (protocol code 2374; date of approval: 18 March 2026). Written informed consent was obtained from all subjects before procedures.

Data Availability Statement

Anonymized individual-level data and analytic materials underlying the findings of this study will be made available upon reasonable request to the corresponding author, subject to institutional approval and applicable ethical and data protection requirements.

Acknowledgments

This work was supported by the George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, Research Grant number 795/3/22.01.2025.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

The following abbreviations are used in this manuscript:
CCT – central corneal thickness
CH – Corneal hysteresis
CRF – Corneal resistance factor
IOP – Intraocular pressure
IOPcc – Corneal compensated intraocular pressure
IOPg – Goldmann-correlated intraocular pressure
NCT – Non-contact tonometry
ORA – Ocular response analyzer

References

  1. Silva, F.; Lira, M. Intraocular pressure measurement: a review. Surv. Ophthalmol. 2022, 67, 1319–1331. [Google Scholar] [CrossRef] [PubMed]
  2. Komninou, M.A.; Seiler, T.G.; Enzmann, V. Corneal biomechanics and diagnostics: a review. Int. Ophthalmol. 2024, 44, 132. [Google Scholar] [CrossRef] [PubMed]
  3. Ariza-Gracia, M.Á.; Zurita, J.F.; Piñero, D.P. Coupled biomechanical response of the cornea assessed by non-contact tonometry: a simulation study. PLoS ONE 2015, 10. [Google Scholar] [CrossRef] [PubMed]
  4. Greene, P.R.; Sergienko, N.M.; Wang, S.K. Review: measurement techniques for intraocular pressure. Optom. Open Access 2016, 1, 118. [Google Scholar] [CrossRef]
  5. Aoki, S.; Asaoka, R.; Fujino, Y. Comparing corneal biomechanic changes among solo cataract surgery, microhook ab interno trabeculotomy and iStent implantation. Sci. Rep. 2023, 13, 19148. [Google Scholar] [CrossRef] [PubMed]
  6. Kaushik, S.; Pandav, S.S.; Banger, A. Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma. Am. J. Ophthalmol. 2012, 153, 840–849.e2. [Google Scholar] [CrossRef] [PubMed]
  7. Brown, L.; Foulsham, W.; Pronin, S. The influence of corneal biomechanical properties on intraocular pressure measurements using a rebound self-tonometer. J. Glaucoma 2018, 27, 511–518. [Google Scholar] [CrossRef] [PubMed]
  8. Vandewalle, E.; Vandenbroeck, S.; Stalmans, I. Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma. Eur. J. Ophthalmol. 2009, 19, 783–789. [Google Scholar] [CrossRef] [PubMed]
  9. Gillan, W.D.H. Intra-ocular pressure measurements using the Ocular Response Analyser and ICare tonometer: a comparison. Afr. Vis. Eye Health 2015, 74. [Google Scholar] [CrossRef]
  10. Pluháček, F.; Unzeitigová, A.; Marešová, K.; Rybář, J. Influence of cornea on intraocular pressure measurement by ICARE PRO and ORA. Cesk Slov. Oftalmol. 2019, 75, 111–118. [Google Scholar] [CrossRef] [PubMed]
  11. Wang, D.; Huang, W.; Li, Y. Intraocular pressure, central corneal thickness, and glaucoma in chinese adults: the liwan eye study. Am. J. Ophthalmol. 2011, 152, 454–62. [Google Scholar] [CrossRef] [PubMed]
  12. Medeiros, F.A.; Weinreb, R.N. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J. Glaucoma 2006, 15, 364–370. [Google Scholar] [CrossRef] [PubMed]
  13. Hager, A.; Loge, K.; Schroeder, B. Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes. J. Glaucoma 2008, 17, 361–365. [Google Scholar] [CrossRef] [PubMed]
  14. Oncel, B.; Dinc, U.A.; Orge, F. Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry. Eur. J. Ophthalmol. 2009, 19, 936–941. [Google Scholar] [CrossRef] [PubMed]
  15. Chui, W.S.; Lam, A.; Chen, D. The influence of corneal properties on rebound tonometry. Ophthalmology 2008, 115, 80–84. [Google Scholar] [CrossRef] [PubMed]
  16. Carbonaro, F.; Andrew, T.; Mackey, D.A. Comparison of three methods of intraocular pressure measurement and their relation to central corneal thickness. Eye 2010, 24, 1165–1170. [Google Scholar] [CrossRef] [PubMed]
  17. Medeiros, F.A.; Meira-Freitas, D.; Lisboa, R. Corneal hysteresis as a risk factor for glaucoma progression: a prospective longitudinal study. Ophthalmology 2013, 120, 1533–1540. [Google Scholar] [CrossRef] [PubMed]
  18. Zhang, C.; Tatham, A.J.; Abe, R.Y. Corneal hysteresis and progressive retinal nerve fiber layer loss in glaucoma. Am. J. Ophthalmol. 2016, 166, 29–36. [Google Scholar] [CrossRef] [PubMed]
  19. Sit, A.J.; Chen, T.C.; Takusagawa, H.L. Corneal hysteresis for the diagnosis of glaucoma and assessment of progression risk: a report by the American Academy of Ophthalmology. Ophthalmology 2023, 130, 433–442. [Google Scholar] [CrossRef] [PubMed]
  20. Susanna, C.N.; Diniz-Filho, A.; Daga, F.B. A prospective longitudinal study to investigate corneal hysteresis as a risk factor for predicting development of glaucoma. Am. J. Ophthalmol. 2018, 187, 148–152. [Google Scholar] [CrossRef] [PubMed]
  21. Murtagh, P.; O’Brien, C. Corneal hysteresis, intraocular pressure, and progression of glaucoma: time for a “hyst-oric” change in clinical practice? J. Clin. Med. 2022, 11, 2895. [Google Scholar] [CrossRef] [PubMed]
  22. Takeda, Y.; Hanyuda, A.; Takahashi, N. Relationship between corneal hysteresis and the site of damage to peripapillary retinal nerve fibre layer thickness in open-angle glaucoma. Sci. Rep. 2024, 14, 26329. [Google Scholar] [CrossRef] [PubMed]
  23. Wong, B.J.; Moghimi, S.; Zangwill, L.M.; et al. Relationship of corneal hysteresis and anterior lamina cribrosa displacement in glaucoma. Am. J. Ophthalmol. 2020, 212, 134–143. [Google Scholar] [CrossRef] [PubMed]
  24. Lee, K.M.; Kim, T.W.; Lee, E.J. Association of corneal hysteresis with lamina cribrosa curvature in primary open-angle glaucoma. Invest Ophthalmol. Vis. Sci. 2019, 60, 4171–4177. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Regression-based Bland–Altman analysis of agreement between IOPcc, non-contact tonometry (NCT), and rebound tonometry (iCare).
Figure 1. Regression-based Bland–Altman analysis of agreement between IOPcc, non-contact tonometry (NCT), and rebound tonometry (iCare).
Preprints 241717 g001
Figure 2. Inter-tonometer differences across corneal hysteresis tertiles.* Points represent mean inter-tonometer differences; error bars indicate 95% confidence intervals.
Figure 2. Inter-tonometer differences across corneal hysteresis tertiles.* Points represent mean inter-tonometer differences; error bars indicate 95% confidence intervals.
Preprints 241717 g002
Table 1. Descriptive characteristics of the study population. Data are presented as mean ± standard deviation; ranges are shown in parentheses.
Table 1. Descriptive characteristics of the study population. Data are presented as mean ± standard deviation; ranges are shown in parentheses.
Variable Total
(N=134)
Male
(N=42)
Female
(N=92)
Age (years) 73.2 ± 5.4
(61–86)
73.5 ± 4.5 73 ± 5.8
NCT (mmHg) 16.4 ± 2.8
(9.5–30)
15.8 ± 2.5 16.6 ± 2.9
iCare Pro (mmHg) 15.1 ± 3.5
(8–28)
15.2 ± 3.9 15.1 ± 3.2
IOPg (mmHg) 16.9 ± 3.8
(9.8–31.9)
16.9 ± 3.9 16.8 ± 3.8
IOPcc (mmHg) 16.7 ± 3.8
(7.3–31.0)
16.7 ± 3.9 16.7 ± 3.9
CH (mmHg) 10.7 ± 1.2
(8.3–15.9)
10.7 ± 1 10.7 ± 1.3
CCT (µm) 527.3 ± 33
(408-602)
531.3 ± 20.6 525.4 ± 37.5
Table 2. Regression-based Bland–Altman analysis of agreement between tonometric methods.
Table 2. Regression-based Bland–Altman analysis of agreement between tonometric methods.
Comparison Mean bias ± SD (mmHg) Slope β (95% CI) p R²
IOPcc − NCT 0.15 ± 2.70 0.341
(0.194–0.487)
<0.001 0.152
IOPcc − iCare 1.68 ± 3.62 0.142
(−0.055–0.339)
0.155 0.016
NCT − iCare 1.18 ± 2.86 −0.263
(−0.443 to −0.083)
0.004 0.068
Table 3. Multivariable linear regression analysis of factors associated with inter-tonometer differences. Values represent unstandardized regression coefficients (β) with 95% confidence intervals.
Table 3. Multivariable linear regression analysis of factors associated with inter-tonometer differences. Values represent unstandardized regression coefficients (β) with 95% confidence intervals.
Outcome Predictor β 95% CI p Model R²
IOPcc − NCT CH −1.56 −1.90 to −1.22 <0.001 0.440
CCT +0.023 0.011 to 0.036 <0.001
Age +0.043 −0.031 to 0.116 0.256
IOPcc − iCare CH −2.23 −2.64 to −1.82 <0.001 0.523
CCT +0.002 −0.013 to 0.017 0.769
Age −0.047 −0.137 to 0.042 0.295
NCT − iCare CH −0.390 −0.840 to 0.061 0.089 0.145
CCT −0.026 −0.043 to −0.010 0.002
Age −0.067 −0.163 to 0.029 0.167
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.