Submitted:
10 October 2026
Posted:
13 October 2026
You are already at the latest version
Abstract
Treatment with short implants is considered as predictable as conventional implants. Objective: This in silico experimental study aimed to determine the stress distribution in short zirconia dental implants using finite element analysis (FEA). Methods: Thirty-six implant-crown designs were modeled in ANSYS 2019 R2. Loading conditions included an oblique force of 150 N (30°, ISO 14801:2016), vertical forces of 300 N and 400 N, and a 32 Ncm screw tightening torque across Type I and Type III bone models using three implant diameters (4.1, 4.8, and 5.0 mm). Results: Under both oblique and vertical loads, zirconia generated lower bone deformation but higher overall stress concentrations than titanium. When stratified by bone density, zirconia concentrated higher peri-implant stress in low-density (Type III) bone. Discussion: In both loading scenarios, a progressive increase in implant diameter successfully and proportionally mitigated stress concentrations, which aligns with established biomechanical literature on titanium systems while matching findings from recent comparative studies showing elevated stress levels in zirconia structures. Conclusion: Stress concentrations are higher in zirconia than titanium, particularly in low-density bone; however, this biomechanical limitation can be effectively compensated for by increasing the implant diameter.
Keywords:
zirconia
; dental implant
; finite element analysis
; short implants
; titanium
1. Introduction
The loss or congenital absence of teeth drastically affects patients’ masticatory function and facial aesthetics. To resolve this, endosseous dental implants have consolidated as the standard surgical device to achieve oral rehabilitation through osseointegration, a biological process of direct bonding between bone and implant [1]. Classically, commercially pure titanium has been the ‘gold standard’ in the manufacturing of these devices due to its excellent biocompatibility and scientific validation [2]. However, titanium presents critical limitations: risk of galvanic corrosion, potential allergic reactions, and aesthetic disadvantages due to the shadowing of its grayish color through the oral mucosa [3].
As an aesthetically favorable alternative, zirconium dioxide or ZrO2 has gained ground due to the optical advantages it offers in crowns and abutments manufactured using CAD/CAM technology. When doped with 3 mol% yttria (Y2O3), zirconia stabilizes into a tetragonal crystalline phase that triggers a martensitic transformation hardening mechanism. This provides high fracture toughness and flexural strength, while releasing 50% fewer particles into surrounding tissues and reducing bacterial adhesion compared to titanium [4].
Historically, implants required lengths of 12 to 20 mm to ensure stability. However, tooth loss causes alveolar bone resorption, leaving limited bone height in posterior areas and an imminent risk of damaging anatomical structures such as the maxillary sinus or the inferior alveolar nerve. Although bone grafting techniques (such as sinus augmentation) exist, they increase morbidity and complications. This prompted the development of ‘short’ (7–10 mm) and ‘ultrashort’ (<7 mm) implants. Despite their clinical success, a latent biomechanical controversy exists within the scientific community regarding whether the drastic reduction in length under occlusal loads compromises force distribution in the surrounding bone [5].
Given that traditional analytical mathematical methods are incapable of evaluating stress in complex biological structures, Finite Element Analysis (FEA) has become the numerical tool of choice. FEA allows for the accurate simulation of the actual biomechanical interaction between the implant and oral tissues under different load vectors. Previous evaluations suggest that peak stresses are mainly concentrated at the bone crest (within the first 5 mm), showing similar load distributions among implants of different lengths and reducing compressive stresses in the prosthesis compared to cantilever structures [6].
Despite these findings, the exact interaction between length reduction and material change (titanium versus zirconia) remains not fully elucidated when supporting a single restoration. Therefore, the main objective of this study is to quantitatively determine the stress distribution in short dental implants manufactured from titanium and zirconium dioxide, restored with an individual prosthetic crown, using a Finite Element Analysis model under controlled occlusal loads in Newtons (N). As principal conclusion, this study demonstrates that under force vector magnitudes in Newtons (N), short zirconia implants with single crowns offer a stress dissipation profile at the bone crest comparable to that of titanium. Likewise, it is concluded that increasing the implant diameter plays a more critical and efficient biomechanical role in reducing and homogeneously distributing loads in the surrounding bone compared to length, showing a progressive optimization of stress behavior when transitioning from 4.1 mm platforms to wider diameters of 4.8 mm and 5.0 mm. This numerically validates that the use of reduced-length implants—regardless of the analyzed material—represents a biomechanically viable and safe alternative for clinical application in areas with deficient bone height, provided that anatomical conditions allow for maximizing the platform diameter.
2. Materials and Methods
This study obtained an exemption from the Bioethics Committee: CEISH-USGP-EXP-ODO-2025-00/ CEB IUNIR: 25-020. The research design was an in silico observational study involving 36 implant-supported, screw-retained zirconia crown designs for a maxillary second molar (tooth 27) developed using FEA software (ANSYS 2019 R2) with a fully assembled mesh. Inclined occlusive forces of 150 N at a 30° angle were applied in accordance with ISO 14801:2016 standards, along with vertical forces of 300 N and 400 N, and a screw torque force of 32 Ncm. These 36 simulations were created with screw-retained crowns divided into two groups, A1–A2: three short titanium implants (4.1 mm, 4.8 mm, and 5 mm x 8 mm) osseointegrated into type I and III bone; and two groups, B1–B2: three short zirconia implants (4.1 mm, 4.8 mm, and 5 mm x 8 mm) osseointegrated into type I and III bone.
Figure 1.
Assembly of the models.

Table 1.
Nodes, elements and mesh size of the FE models.
| Components | Nodes | Elements | Mesh |
|---|---|---|---|
| Mesh – Implant | 79376 | 50324 | |
| Mesh – Screw | 30433 | 20107 | 1435684 (N) |
| Mesh – abutment | 83762 | 53267 | 1007620 (E) |
| Bone | 271275 | 187339 | |
| Crown | 894306 | 650167 | |
| Cement | 29388 | 14491 | |
| Dental Resin | 47144 | 31925 |
Subsequently, the physical properties under loading conditions—which influenced stress and its distribution across the bone, crown, and abutment (Young’s modulus and Poisson’s ratio)—were established and compared. The research results were evaluated using Von Mises stress analysis.
Table 2.
Material Properties.
| Material | Elastic Modulus (MPa) | Poisson’s Ratio | Reference |
|---|---|---|---|
| Titanium – Implant | 110000 | 0.34 | [7,8,9] |
| Titanium – Screw | 110000 | 0.35 | [10] |
| Zirconia – Implant | 210000 | 0.30 | [10] |
| Zirconia – Crown | 210000 | 0.26 | [10] |
| Titanium – abutment | 110000 | 0.35 | [10] |
| Cortical Bone | 13700 | 0.30 | [8] |
| Cancellous Bone | 1370 | 0.30 | [8] |
| Cement | 10760 | 0.35 | [8] |
| Dental Resin | 7000 | 0.20 | [8] |
3. Results
The following sections present the deformation and stress results obtained for both titanium and zirconia. (see File S1)
3.1. Deformations in Titanium and Zirconia Within the Implant, Abutment-Screw, and Bone
Figure 2.
Deformation with a 150 N Load – 30° Incline. 4.1 mm diameter Titanium Implant and Type I Bone Density. Greater deformations at the coronal level 0.0407 mm.
Figure 2.
Deformation with a 150 N Load – 30° Incline. 4.1 mm diameter Titanium Implant and Type I Bone Density. Greater deformations at the coronal level 0.0407 mm.

Figure 3.
Deformation with a 150 N Load – 30° Incline. 4.1 mm diameter Zirconia Implant and Type I Bone Density. Greater deformations at the coronal level 0.028008 mm.
Figure 3.
Deformation with a 150 N Load – 30° Incline. 4.1 mm diameter Zirconia Implant and Type I Bone Density. Greater deformations at the coronal level 0.028008 mm.

Table 3.
Maximum Deformation in Titanium - Zirconia Structure / Component.
| Loading Condition | Structure / Component | Maximum Titanium Deformation (mm) | Maximum Zirconia Deformation (mm) | |
|---|---|---|---|---|
| Inclined Force (150 N / 30°) | Crown | 0.040700 | 0.002800 | |
| Inclined Force (150 N / 30°) | Implant-abutment-screw | 0.031100 | 0.021900 | |
| Inclined Force (150 N / 30°) | Surrounding Bone | 0.007233 | 0.006700 | |
| Vertical Force (300 N) | Crown | 0.012200 | 0.011000 | |
| Vertical Force (300 N) | Implant-abutment-screw | 0.004400 | 0.003200 | |
| Vertical Force (300 N) | Surrounding Bone | 0.002450 | 0.002000 | |
| Vertical Force (400 N) | Crown | 0.016200 | 0.014000 | |
| Vertical Force (400 N) | Implant-abutment-screw | 0.006000 | 0.004300 | |
| Vertical Force (400 N) | Surrounding Bone | 0.003250 | 0.002700 |
The deformations in zirconia are lower than those in titanium. Greater deformation occurs in the crown and within the implant-abutment-screw than in the surrounding bone. Deformation increases as forces increase.
Table 4.
Deformations in Type I-III Bone (Titanium- Zirconia) - 150 N Inclined Force.
| Loading Condition | Structure / Component | Maximum Titanium Deformation (mm) |
Maximum Zirconia Deformation (mm) |
|---|---|---|---|
| Inclined Force (150 N) | Type I Bone, 4.1 mm Implant Diameter | 0.040700 | 0.028000 |
| Inclined Force (150 N) | Type III Bone, 4.1 mm Implant Diameter | 0.059500 | 0.047200 |
| Inclined Force (150 N) | Type I Bone, 4.8 mm Implant Diameter | 0.037000 | 0.025900 |
| Inclined Force (150 N) | Type III Bone, 4.8 mm Implant Diameter | 0.054800 | 0.043100 |
| Inclined Force (150 N) | Type I Bone, 5 mm Implant Diameter-screw | 0.036000 | 0.025300 |
| Inclined Force (150 N) | Type III Bone, 5 mm Implant Diameter-screw | 0.054400 | 0.041800 |
With the application of 150 N inclined forces in type I and III bone, it is observed that deformations are lower in type I bone and higher in type III bone; likewise, they are lower in zirconia than in titanium, and decrease even further as the implant diameter increases.
Table 5.
Deformations in Type I-III Bone (Titanium- Zirconia) - 300 N Vertical Force.
| Loading Condition | Structure / Component | Maximum Titanium Deformation (mm) |
Maximum Zirconia Deformation (mm) |
|---|---|---|---|
| Vertical Force (300 N) | Type I Bone, 4.1 mm Implant Diameter | 0.012270 | 0.011200 |
| Vertical Force (300 N) | Type III Bone, 4.1 mm Implant Diameter | 0.019980 | 0.018390 |
| Vertical Force (300 N) | Type I Bone, 4.8 mm Implant Diameter | 0.012050 | 0.011010 |
| Vertical Force (300 N) | Type III Bone, 4.8 mm Implant Diameter | 0.019010 | 0.017970 |
| Vertical Force (300 N) | Type I Bone, 5 mm Implant Diameter-screw | 0.011980 | 0.010940 |
| Vertical Force (300 N) | Type III Bone, 5 mm Implant Diameter-screw | 0.018710 | 0.017620 |
With the application of 300 N vertical forces in type I and III bone, it is observed that deformations are lower in type I bone than in type III bone; likewise, they are lower in zirconia than in titanium and decrease even further as the implant diameter increases.
Table 6.
Deformations in Type I-III Bone (Titanium- Zirconia) - 400 N Vertical Force.
| Loading Condition | Structure / Component | Maximum Titanium Deformation (mm) |
Maximum Zirconia Deformation (mm) |
|---|---|---|---|
| Vertical Force (400 N) | Type I Bone, 4.1 mm Implant Diameter | 0.016330 | 0.014920 |
| Vertical Force (400 N) | Type III Bone, 4.1 mm Implant Diameter | 0.026600 | 0.025200 |
| Vertical Force (400 N) | Type I Bone, 4.8 mm Implant Diameter | 0.016040 | 0.014660 |
| Vertical Force (400 N) | Type III Bone, 4.8 mm Implant Diameter | 0.025300 | 0.023900 |
| Vertical Force (400 N) | Type I Bone, 5 mm Implant Diameter-screw | 0.015940 | 0.014570 |
| Vertical Force (400 N) | Type III Bone, 5 mm Implant Diameter-screw | 0.024900 | 0.023500 |
With the application of 400 N vertical forces in type I and III bone, it is observed that deformations are lower in type I bone than in type III bone; likewise, they are lower in zirconia than in titanium, and decrease even further as the implant diameter increases
3.2. Stresses in Titanium and Zirconia Within the Implant, Abutment-Screw, and Bone
Figure 4.
Tension with a 150 N Load – 30° Incline. 4.1 mm diameter Titanium Implant and Type I Bone Density. The highest tension is located at the alveolar crest 372.25 MPa.
Figure 4.
Tension with a 150 N Load – 30° Incline. 4.1 mm diameter Titanium Implant and Type I Bone Density. The highest tension is located at the alveolar crest 372.25 MPa.

Figure 5.
Tension with a 150 N Load – 30° Incline. 4.1 mm diameter Zirconia Implant and Type I Bone Density. The highest tension is located at the alveolar crest 407.88 MPa.
Figure 5.
Tension with a 150 N Load – 30° Incline. 4.1 mm diameter Zirconia Implant and Type I Bone Density. The highest tension is located at the alveolar crest 407.88 MPa.

Table 7.
Maximum Stresses in Titanium - Zirconia Structures / Component.
| Loading Condition | Structure / Component | Maximum Titanium Stresses (MPa) | Maximum Zirconia Stresses (MPa) |
|---|---|---|---|
| Inclined Force (150 N / 30°) | Crown | 372.25 | 404.11 |
| Inclined Force (150 N / 30°) | Implant-abutment-screw | 372.25 | 404.11 |
| Inclined Force (150 N / 30°) | Surrounding Bone | 26.192 | 14.349 |
| Vertical Force (300 N) | Crown | 340.03 | 363.24 |
| Vertical Force (300 N) | Implant-abutment-screw | 340.03 | 363.24 |
| Vertical Force (300 N) | Surrounding Bone | 7.9812 | 9.5938 |
| Vertical Force (400 N) | Crown | 355.02 | 378.05 |
| Vertical Force (400 N) | Implant-abutment-screw | 355.02 | 378.75 |
| Vertical Force (400 N) | Surrounding Bone | 10.558 | 12.805 |
Stresses in titanium are lower than those in zirconia, and they decrease even further as the implant diameter increases. Higher stress concentrations occur in the crown and within the implant-abutment-screw than in the surrounding bone.
Table 8.
Stresses in Type I-III Bone (Titanium- Zirconia)- 150 N Inclined Force.
| Loading Condition | Structure / Component | Maximum Titanium Stresses (MPa) |
Maximum Zirconia Stresses (MPa) |
|---|---|---|---|
| Inclined Force (150 N) | Type I Bone, 4.1 mm Implant Diameter | 372.25 | 407.88 |
| Inclined Force (150 N) | Type III Bone, 4.1 mm Implant Diameter | 376.11 | 409.03 |
| Inclined Force (150 N) | Type I Bone, 4.8 mm Implant Diameter | 370.11 | 404.11 |
| Inclined Force (150 N) | Type III Bone, 4.8 mm Implant Diameter | 372.20 | 405.61 |
| Inclined Force (150 N) | Type I Bone, 5 mm Implant Diameter-screw | 369.52 | 403.19 |
| Inclined Force (150 N) | Type III Bone, 5 mm Implant Diameter-screw | 372.11 | 404.78 |
With the application of 150 N inclined forces in type I and III bone, it is observed that stresses are lower in type I bone and higher in type III bone; likewise, they are lower in titanium than in zirconia and decrease even further as the implant diameter increases.
Table 9.
Stresses in Type I-III Bone (Titanium- Zirconia) - 300 Vertical Force.
| Loading Condition | Structure / Component | Maximum Titanium Stresses (MPa) |
Maximum Zirconia Stresses (MPa) |
|---|---|---|---|
| Vertical Force (300 N) | Type I Bone, 4.1 mm Implant Diameter | 340.03 | 367.47 |
| Vertical Force (300 N) | Type III Bone, 4.1 mm Implant Diameter | 345.32 | 370.55 |
| Vertical Force (300 N) | Type I Bone, 4.8 mm Implant Diameter | 337.00 | 363.24 |
| Vertical Force (300 N) | Type III Bone, 4.8 mm Implant Diameter | 341.20 | 366.33 |
| Vertical Force (300 N) | Type I Bone, 5 mm Implant Diameter-screw | 336.00 | 362.61 |
| Vertical Force (300 N) | Type III Bone, 5 mm Implant Diameter-screw | 339.62 | 364.69 |
With the application of 300 N vertical forces in type I and III bone, it is observed that stress is lower in type I bone than in type III bone; likewise, stresses are lower in titanium than in zirconia and decrease even further as the implant diameter increases.
Table 10.
Stresses in Type I-III Bone (Titanium- Zirconia) - 400 N Vertical Force.
| Loading Condition | Structure / Component | Maximum Titanium Stresses (MPa) |
Maximum Zirconia Stresses (MPa) |
|---|---|---|---|
| Vertical Force (400 N) | Type I Bone, 4.1 mm Implant Diameter | 355.02 | 384.17 |
| Vertical Force (400 N) | Type III Bone, 4.1 mm Implant Diameter | 361.92 | 388.17 |
| Vertical Force (400 N) | Type I Bone, 4.8 mm Implant Diameter | 351.14 | 378.75 |
| Vertical Force (400 N) | Type III Bone, 4.8 mm Implant Diameter | 356.64 | 382.80 |
| Vertical Force (400 N) | Type I Bone, 5 mm Implant Diameter-screw | 349.86 | 377.02 |
| Vertical Force (400 N) | Type III Bone, 5 mm Implant Diameter-screw | 340.90 | 380.68 |
With the application of 400 N vertical forces in type I and III bone, it is observed that stresses are lower in type I bone than in type III bone; likewise, they are lower in titanium than in zirconia and decrease even further as the implant diameter increases.
3.3. Zirconia Optimizes Deformations but Increases Stresses
An inversely proportional relationship exists due to the distinct elastic moduli (Young’s modulus) of the biomaterials. Yttria-stabilized zirconium dioxide exhibits significantly higher stiffness than commercially pure titanium. Consequently, zirconia components undergo lower structural deformation under all loading regimens (both vertical and inclined) compared to titanium. However, due to this inherent rigidity, zirconia absorbs less energy, resulting in higher stress concentrations (MPa) within its structure and across the prosthetic assembly (implant-abutment-screw) than those observed in titanium systems.
3.4. Increasing Implant Diameter Is the Most Critical Optimization Factor
Regardless of the analyzed biomaterial, both stresses and deformations decrease drastically as the implant diameter increases, transitioning progressively from 4.1 mm to 4.8 mm and 5.0 mm. A wider diameter increases the overall surface contact area at the bone-implant interface, enabling a more homogeneous load distribution. This mechanically demonstrates that to mitigate stress and deformation in scenarios with deficient bone height (short implants), maximizing the platform diameter is biomechanically more efficient than increasing implant length.
3.5. Bone Density Governs Structural Resistance
The mechanical behavior varies drastically depending on the quality and histomorphology of the receptor bone tissue. Type I bone (predominantly dense cortical bone) exhibits lower deformations and lower stresses, as it possesses optimal mechanical properties to dissipate load vectors. Conversely, Type III bone (more trabecular and medullary) undergoes higher deformations and peak stresses, making the selection of wider diameter platforms critically necessary in these low-density areas to prevent crestal bone resorption.
3.6. Inclined Forces Represent the Critical Biomechanical Scenario
While vertical forces (300 N and 400 N) transmit loads axially—resulting in controlled and predictable deformation profiles—inclined forces (150 N at 30°) trigger a severe bending moment (lever effect), despite their lower numerical magnitude. This phenomenon triggers the absolute maximum peaks of stress and deformation at the alveolar bone crest and within the abutment-screw junction, consolidating as the highest clinical risk scenario for the long-term integrity of single restorations.
3.7. Titanium vs. Zirconia
It is important to recognize that both deformation and stress values in the surrounding bone are lower than those observed within the implant-abutment-screw complex. Furthermore, titanium transfers higher stress and causes greater deformation to the peri-implant bone than zirconia under all loading conditions.
4. Discussion
The success of a treatment with dental implants is highly dependent on stresses transmission to the supporting bone. This load transfer depends on the factors such as type of loading, the dimensions of the implant and the quality of bone among other factors. Finite element analysis is a tool that are used to simulate stress in the implant systemand also the bone contacting the implant [11].
According to the results obtained in this study, structural deformations were higher in titanium implants than in zirconia implants. This finding is consistent with that reported by Che-Lun Chang (2024) [12], who, using a three-dimensional finite element analysis (FEA), simulated masticatory forces with 100 N perpendicular and oblique loads on implants integrated into cortical and cancellous bone. In that study, it was determined that zirconia offers the best biomechanical performance due to its low structural deformation and high efficiency in dissipating and transferring stress to the surrounding area, outperforming both titanium and PEEK; the latter, despite registering the lowest maximum and average internal stress, demonstrated poor performance in dissipating these mechanical forces. Likewise, Claudia Peter (2024) [13] demonstrated through mechanical testing that zirconia led the modulus of elasticity at 208 ± 8 GPa, followed by titanium (112 ± 6 GPa) and ceramic (72 ± 6 GPa), with a p-value of less than 0.001
The results of this study demonstrate that oblique loads cause greater damage than vertical loads in both evaluated materials. This finding is consistent with that reported by Bilgi-Ozyetim (2025) [14] , who points out that ‘implant angulation, connection type, and loading conditions affected stress distribution in the peri-implant bone, implants, and frameworks.’ Furthermore, that author observed higher stress values in posterior implants with increased angulation, while also reporting that screw-retained models register superior stress levels compared to cement-retained ones under both loading conditions. In alignment with these biomechanical behaviors, Kitamura (2005) [15] analyzed the impact of applying axial and bucco-lingual forces independently to the center of the superstructure; their results similarly indicate that bone stress concentrates primarily around the implant neck, registering higher levels under bucco-lingual (oblique) than axial loads in both the cortical and cancellous bone, across models with and without resorption
The results of the present investigation demonstrated a decreasing pattern in stress transfer, with the highest stress magnitudes located in the crown, followed by the abutment and the prosthetic screw, before finally dissipating into the alveolar bone. This biomechanical behavior is consistent with the findings of Epifania (2023) [16], who reported that the peak stress in zirconia crowns (64.4 MPa) was significantly higher than that recorded in composite resin crowns, regardless of the presence of the prosthetic screw. Likewise, in agreement with D. Milone (2022) [17], it was observed that the maximum stress values are very similar for both materials in both vertical and horizontal directions. The main difference lies in the stress distribution, which tends to decrease more rapidly in the zirconia implant compared to the titanium implant.
Regarding the behavior of the peri-implant bone tissue, the results revealed that the load vectors are critically concentrated at the alveolar crest, specifically within the cervical portion of the peri-implant cortical bone. This finding is in close alignment with the observations of Araki (2020) [18], who detected that the localization of peak stress depends on the three-dimensional design of the prosthesis: it is situated above the crest in tissue-level designs, whereas it shifts intraosseously when the connection is located below the cortical crest. From a clinical perspective, this accumulation of mechanical stress in the cervical area provides a biomechanical explanation for the phenomenon of marginal bone remodeling, which typically manifests as an initial crestal bone loss of approximately 2 mm during the first year of functional loading, followed by a constant resorption rate of 0.2 mm annually in subsequent years [19]. Biomechanical Finite Element Method (FEM) studies have demonstrated that occlusal forces are primarily distributed toward the alveolar crest, rather than being uniformly transferred across the entire surface of the bone–implant interface. These forces are normally well tolerated by the bone, which may explain why implant length is not linearly related to biomechanical stability [9].
The results of the present investigation demonstrated that stress levels were superior in zirconia implants compared to titanium ones. This biomechanical behavior is consistent with the findings reported by Saz (2026) [20], who noted that Von Mises stress values were higher in all evaluated groups for ceramic implants than for titanium systems. Along the same lines, a study developed by Krennmair (2025) [21] quantified this mechanical difference, reporting that the stress values measured within the dental implants showed a significantly higher maximum stress peak for the zirconia implant (1137 MPa) than for the titanium implant model (686.7 MPa)
When evaluating the behavior of zirconia and titanium implants, it was observed that both vertical and inclined loads generated greater stress in lower bone densities. These findings are consistent with those reported by Shinichiro Tada (2003) [22], who determined that, regardless of the load direction, the maximum equivalent stress and strain in bone tissue increase as bone density decreases. This biomechanical behavior is grounded in the classic theories of Misch (1990) [23], who, through finite element analysis (FEA), explained that excessive stress at the bone–implant interface causes crestal bone loss and the subsequent early failure of the implant after restoration; furthermore, he demonstrated that the rate of bone resorption varies significantly across different bone densities, even under the application of similar forces.
Within this research, we found that as the implant diameter increases, loads are better distributed in type I and type III bone, similar to what was reported by Anitua (2021) [24], who found that a Ø3.3 mm diameter implant caused 2.5 to 4.4 times greater stress compared to a Ø4.75 mm implant, depending on the bone type. That study demonstrated that implant diameter directly influences stress concentration in the peri-implant bone, which in turn affects the long-term survival rate. In accordance with these findings, Yen-Wen Shen (2024) [25] reported that wider implants demonstrate lower stress levels in both the implant and bone compared with standard-diameter implants. Furthermore, that author noted that the quality of cancellous bone has a minimal impact on the stress values of the implant, abutment screw, and cortical bone. Regardless of occlusal arrangement or cancellous bone quality, a consistent pattern emerged: larger abutment screw diameters lead to increased stress levels on the screws, while stress levels in both cortical and cancellous bone show comparatively minor fluctuations.
It is suggested to conduct further research incorporating higher vertical loads and oblique loads at angles of less than 30°.
5. Conclusions
When evaluating the numerical simulations of the biomechanical assembly, a decreasing distribution of structural stresses and strains is observed, whose maximum values are located in the crown and progressively decrease toward the abutment, the screw, and the bone tissue; thus, the bone records the lowest magnitude of mechanical strains and stresses for both materials. Likewise, it is evident that this stress and strain state increases critically as bone density decreases. Regarding boundary conditions, the analysis reveals that oblique loads of 150 N at 30° induce higher biomechanical responses compared to vertical axial forces of 300 N or 400 N. Finally, the material properties significantly altered the system behavior: the use of titanium components generated more pronounced strains in contrast to zirconia, although the latter exhibited higher internal stress peaks due to its higher stiffness.
Supplementary Materials
The following supporting information can be downloaded at the website of this paper posted on Preprints.org.
Author Contributions
Conceptualization, C.M.T.A. and J.L.E.L.; methodology, J.L.E.L. and H.E.C.I.; software, J.L.E.L.; validation, C.M.T.A., J.L.E.L., C.A.C. and H.E.C.I.; formal analysis, J.L.E.L. and C.A.C.; investigation, C.M.T.A., C.A.C. and H.E.C.I.; resources, C.M.T.A.; data curation, J.L.E.L.; writing—original draft preparation, C.M.T.A.; writing—review and editing, C.M.T.A., J.L.E.L. and C.A.C.; visualization, J.L.E.L.; supervision, C.M.T.A.; project administration, C.M.T.A. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding. The APC was funded by the authors.
Institutional Review Board Statement
The study was conducted in accordance with the Declaration of Helsinki. Ethical review and approval were waived for this study by the Ethics Committee of Universidad San Gregorio de Portoviejo (Ecuador, protocol code CEISH-USGP-EXP-ODO-2025-001) on 8 April 2025, and an exemption was granted by the Institutional Review Board and Bioethics Committee of the Instituto Universitario Italiano de Rosario (IUNIR, Argentina, protocol code CEB IUNIR: 25-020) on 12 September 2025, as this research involved dental implant simulations and did not require direct human clinical intervention.
Informed Consent Statement
Not applicable.
Data Availability Statement
The raw data supporting the conclusions of this article will be made available by the authors on request.
Acknowledgments
During the preparation of this manuscript, the authors used Gemini (Google) for the purposes of English translation, text formatting, and manuscript template alignment. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Conflicts of Interest
The authors declare no conflict of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| ANSYS | 3D design and engineering simulation software |
| FEA | Finite element analysis |
| N | Newtons |
| MPa | Megapascals |
| Ncm | Newton centimeter |
| CAD/CAM | Computer aided design |
References
- Lavorgna, L.; Cervino, G.; Fiorillo, L.; Di Leo, G.; Troiano, G.; Ortensi, M.; et al. Reliability of a virtual prosthodontic project realized through a 2d and 3d photographic acquisition: An experimental study on the accuracy of different digital systems. Int. J. Environ. Res. Public Health 2019, 16(24), 5139. [Google Scholar] [CrossRef]
- Cionca, N.; Hashim, D.; Mombelli, A. Zirconia dental implants: where are we now, and where are we heading? Periodontol. 2000 2017, 73(1), 241–58. [Google Scholar] [CrossRef]
- Siddiqi, A.; Payne, A.G.T.; De Silva, R.K.; Duncan, W.J. Titanium allergy: could it affect dental implant integration? Clin. Oral Implant. Res. 2011, 22(7), 673–80. [Google Scholar] [CrossRef]
- Keith, O.; Kusy, R.P.; Whitley, J.Q. Zirconia brackets: An evaluation of morphology and coefficients of friction. Am. J. Orthod. Dentofac. Orthop. 1994, 106(6), 605–14. [Google Scholar] [CrossRef]
- Shah, A. Short implants - When, where and how? J. Int. Clin. Dent. Res. Organ. 2015, 7(3), 132. [Google Scholar] [CrossRef]
- Angeles Maluscan, R. J. ANÁLISIS DE TENSIONES EN ELEMENTOS FINITOS DE UNA CONEXIÓN TRIANGULAR CÓNICA EN IMPLANTES UNA NUEVA PROPUESTA” (Tesis de titulación/Trabajo académico) [Internet]. Repositorio Institucional UPCH; Universidad Peruana Cayetano Heredia: Lima, Perú, 2021. Available online: https://repositorio.upch.edu.pe/.
- Figueroa-Hernández, C.; Pantaleón-Matamoros, E.; Méndez-González, S.; García-Fernández, C.; Gómez-González, R.; Carvajal-de la Osa, J. Análisis del fenómeno de fatiga en implantes dentales monocomponente. Ing. Mecánica [Internet];Ing. Mecánica Available from. 2020, 23(1), e595. [Google Scholar]
- Lee, H.; Park, S.; Kwon, K.R.; Noh, G. Effects of cementless fixation of implant prosthesis: A finite element study. J. Adv. Prosthodont. 2019, 11(6), 341–9. [Google Scholar] [CrossRef]
- Loyola-González, P.O.; Torassa, D.; Dominguez, A. Estudio comparativo sobre el comportamiento y la distribución de las tensiones en implantes dentales cortos e implantes dentales estándares en la región posterior del maxilar superior. Un estudio en elementos finitos. Rev. Clínica De Periodoncia Implantol. Y Rehabil. Oral 2016, 9(1), 36–41. [Google Scholar] [CrossRef]
- Choi, S.M.; Choi, H.; Lee, D.H.; Hong, M.H. Comparative finite element analysis of mandibular posterior single zirconia and titanium implants: a 3-dimensional finite element analysis. J. Adv. Prosthodont. 2021, 13(6), 396–407. [Google Scholar] [CrossRef]
- Geng, J.P.; Tan, K.B.C.; Liu, G.R. Application of finite element analysis in implant dentistry: A review of the literature. J. Prosthet. Dent. 2001, 85(6), 585–98. [Google Scholar] [CrossRef]
- Chang, C.L.; Karmakar, R.; Mukundan, A.; Lu, S.H.; Choomjinda, U.; Chen, M.M.; et al. Mechanical Integrity of All-on-Four Dental Implant Systems: Finite Element Simulation of Material Properties of Zirconia, Titanium, and PEEK. Open Dent. J. 2024, 18(1). [Google Scholar] [CrossRef]
- Peter, C.; Shah, K.; Simon, L.; Pm, S.; N, A.; El-Shamy, F.M. Comprehensive Evaluation of Titanium, Zirconia, and Ceramic Dental Implant Materials: A Comparative Analysis of Mechanical and Esthetic Properties. Cureus;PubMed 2024, 16(5), e60582. [Google Scholar] [CrossRef] [PubMed]
- Yang, T.C.; Maeda, Y.; Gonda, T. Biomechanical rationale for short implants in splinted restorations: an in vitro study. Int. J. Prosthodont.;PubMed 2011, 24(2), 130–2. [Google Scholar] [PubMed]
- KITAMURA, E.; STEGAROIU, R.; NOMURA, S.; MIYAKAWA, O. Influence of marginal bone resorption on stress around an implant – a three-dimensional finite element analysis. J. Oral Rehabil. 2005, 32(4), 279–86. [Google Scholar] [CrossRef]
- Epifania, E.; Di Lauro, A.E.; Ausiello, P.; Mancone, A.; Garcia-Godoy, F.; Mendes Tribst, J.P. Effect of crown stiffness and prosthetic screw absence on the stress distribution in implant-supported restoration: A 3D finite element analysis. PLoS ONE 2023, 18(5), e0285421. [Google Scholar] [CrossRef]
- Milone, D.; Fiorillo, L.; Alberti, F.; Cervino, G.; Filardi, V.; Pistone, A.; et al. Stress distribution and failure analysis comparison between Zirconia and Titanium dental implants. Procedia Struct. Integr. 2022, 41, 680–91. [Google Scholar] [CrossRef]
- Araki, H.; Nakano, T.; Ono, S.; Yatani, H. Three-dimensional finite element analysis of extra short implants focusing on implant designs and materials. Int. J. Implant Dent. 2020, 6(1), 5. [Google Scholar] [CrossRef]
- Schultze-Mosgau, S.; Schliephake, H.; Radespiel-Tröger, M.; Neukam, F.W. Osseointegration of endodontic endosseous conesZirconium oxide vs titanium. Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology 2000, 89(1), 91–8. [Google Scholar] [CrossRef]
- Saz, B.; Ozyilmaz, O.Y.; Doganay Ozyilmaz, O. Stress distribution in titanium and zirconia implants with full-arch prostheses: a finite element analysis. BMC Oral Health 2026, 26(1), 1518. [Google Scholar] [CrossRef]
- Krennmair, S.; Postl, L.; Staudigl, C.; Malek, M.; Otto, S.; Hof, M. Zirconia vs. titanium dental implants in maxillofacial trauma: a preliminary comparative biomechanical analysis in the maxilla using the finite element method. Head Face Med. 2025, 22(1), 4. [Google Scholar] [CrossRef]
- Tada, S.; Stegaroiu, R.; Kitamura, E.; Miyakawa, O.; Kusakari, H. Influence of implant design and bone quality on stress/strain distribution in bone around implants: a 3-dimensional finite element analysis. In Int J Oral Maxillofac Implants; PubMed, 2003; Volume 18, 3, pp. 357–68. [Google Scholar] [PubMed]
- Misch, C.E. Density of bone: effect on treatment plans, surgical approach, healing, and progressive boen loading. Int. J. Oral Implantol.;PubMed 1990, 6(2), 23–31. [Google Scholar] [PubMed]
- Anitua, E.; Larrazabal Saez de Ibarra, N.; Morales Martín, I.; Saracho Rotaeche, L. Influence of dental implant diameter and bone quality on the biomechanics of single-crown restoration. A finite element analysis. Dent. J. 2021, 9(9). [Google Scholar] [CrossRef]
- Shen, Y.W.; Huang, H.L.; Hsu, J.T.; Fuh, L.J. Effects of diameters of implant and abutment screw on stress distribution within dental implant and alveolar bone: A three-dimensional finite element analysis. J. Dent. Sci. 2024, 19(2), 1126–34. [Google Scholar] [CrossRef]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.