Abstract
Objective: Flow-diverting devices treat intracranial aneurysms by modifying blood exchange between the parent artery and aneurysm. We quantified local flow changes after implantation using patient-specific Computational Fluid Dynamics(CFD).
Methods: Three patients were evaluated: two treated with a flow-diverting device alone and one with adjunctive coiling. Patient-specific vascular models were reconstructed from clinical images. Devices and coils were virtually deployed, and steady-state CFD with explicit device geometry compared untreated and treated conditions. Findings were descriptively compared with immediate and 12-month angiography.
Results: Device implantation redistributed lesion-region flow and reduced mean velocity by 11.9%, 29.4%, and 6.7% in Cases 1-3. In the giant aneurysm, adjunctive coiling produced additional reductions of 65.1% in sac-averaged velocity and 17.7% in neck-plane velocity. Immediate angiography showed contrast stasis in all cases. At 12 months, all lesions were completely occluded and parent arteries remained patent, consistent with the direction of flow suppression observed by CFD.
Conclusion: Across these representative patient-specific configurations, the device produced local flow redirection while preserving parent-artery patency. Adjunctive coiling provided additional intra-saccular flow suppression in the giant aneurysm case.
Keywords:
intracranial aneurysm
; flow-diverting device
; Computational Fluid Dynamics
; patient-specific modeling
; virtual deployment
1. Introduction
Intracranial aneurysms are common cerebrovascular lesions. Their rupture can cause subarachnoid hemorrhage and carries substantial risks of disability and death [1]. For wide-necked aneurysms, complex morphologies, or lesions unsuitable for conventional embolization, flow-diverting devices reconstruct the flow channel within the parent artery [2] and offer an alternative to intrasaccular filling alone [3]. After implantation, the braided mesh spans the aneurysm neck or diseased segment. It modifies blood exchange between the parent artery and lesion cavity, reduces high-velocity inflow [4], and progressively promotes aneurysm occlusion and parent-artery reconstruction [5].
The hemodynamic effect of a flow-diverting device depends not only on metal coverage and braid structure but also on parent-artery diameter, curvature, lesion morphology, and the deployed device configuration [6,7]. Devices of the same nominal size may develop different local pore patterns and strut distributions after deployment in different patient anatomies [8]. Consequently, nominal device parameters alone may not fully characterize local flow regulation under realistic anatomical conditions [9]. Hemodynamic analysis that incorporates patient-specific geometry and the deployed configuration is therefore important [10].
Computational Fluid Dynamics(CFD) can quantify local flow changes before and after implantation while retaining patient-specific vascular geometry [11]. Unlike porous-medium or equivalent-boundary representations, explicit stent-strut models resolve local flow near individual pores [12] and represent the spatial configuration of the device after deployment in a curved vessel [13,14]. For cases treated with adjunctive coiling, explicitly retaining the intra-saccular coil geometry enables assessment of the combined hemodynamic effects of the device and the space-occupying coil mass [15,16].
This study evaluated three patients treated with a flow-diverting device who had complete immediate postprocedural and 12-month imaging follow-up. The cases represented a small sidewall saccular aneurysm, a dissecting or locally dilated vertebral artery lesion, and a giant internal carotid artery aneurysm. Patient-specific vascular reconstruction, virtual device deployment, and CFD with explicit device geometry were used to assess local hemodynamic changes after treatment with the device alone or with adjunctive coiling. Computational observations were descriptively compared with the corresponding angiographic evolution.
2. Materials and Methods
2.1. Study Design and Cases
Three patients with intracranial aneurysms treated using a flow-diverting device manufactured by Skynor Medical (Shanghai) Co., Ltd. were included. All patients had complete preprocedural imaging, immediate postprocedural cerebral angiography, and 12-month imaging follow-up. Two patients received the flow-diverting device alone, whereas one patient with a giant aneurysm received the device with adjunctive coil embolization. Patient-specific vascular models were reconstructed from preprocedural clinical images. Virtual treatment models were then created according to the devices and treatment strategies used clinically for subsequent CFD analysis. The patient characteristics were as follows.
Case 1 was a 49-year-old man with a small saccular aneurysm in the C6 segment of the right internal carotid artery. The maximum aneurysm diameter was approximately 5.0 mm, and the parent artery measured approximately 3.5 mm. The patient was treated with a flow-diverting device alone.
Case 2 was a 53-year-old man with a lesion in the V4 segment of the right vertebral artery. The maximum aneurysm diameter was approximately 8.6 mm, and the parent artery measured approximately 3.5 mm. The lesion was clinically characterized as locally dilated and lacked a distinct neck typical of a saccular aneurysm. The patient was treated with a flow-diverting device alone.
Case 3 was a 60-year-old woman with a giant aneurysm in the ophthalmic segment of the left internal carotid artery. The maximum aneurysm diameter was approximately 23.8 mm. Clinical treatment comprised a flow-diverting device and coils. Untreated, device-only, and device-plus-coil models were constructed to evaluate sequential local hemodynamic changes after device implantation and adjunctive coiling.
2.2. Flow-Diverting Device
A braided flow-diverting device manufactured by Skynor Medical (Shanghai) Co., Ltd. was used. The device had a phosphorylcholine (PC)-modified surface. Phosphorylcholine groups were chemically grafted onto the device surface to improve blood compatibility and were designed to reduce platelet adhesion and device-related thrombogenicity. This study assessed the deployed geometry and its mechanical effects on local hemodynamics. The biological or antithrombotic effects of the PC surface modification were not evaluated separately. All devices had a nominal metal coverage of 40%. Device size was selected according to parent-artery diameter and the clinical treatment length. Device parameters are summarized in Table 1.
For each patient-specific deployment, the initial device model was constructed using the clinically implanted size and braid parameters. The device was then deployed according to the parent-artery geometry to obtain the patient-specific spatial configuration after implantation [17].
2.3. Patient-Specific Vascular Reconstruction and Virtual Device Deployment
The parent artery and aneurysm region were segmented and reconstructed in three dimensions from each patient's preprocedural clinical images. The patient-specific luminal surface was exported in STL format. The model was then imported into Geomagic Wrap 2017 (Geomagic, USA) for geometric repair. Processing included removal of isolated structures and surface noise unrelated to the target vessel, local smoothing, defect repair, and truncation of the inlet and outlet segments according to the computational domain [18].
The repaired vascular model was imported into the Vascular Modeling Toolkit (VMTK). The parent-artery centerline was extracted using Delaunay triangulation, the Voronoi diagram, and shortest-path searching. This centerline described the vessel course and provided a geometric reference for initial device positioning and expansion. An initial braided stent mesh was created in MATLAB (MathWorks, USA) and positioned along the centerline. The initial mesh radius was set to 0.01 mm solely for numerical initialization and did not represent the physical wire radius.
Device expansion was simulated using a virtual deployment method based on nodal dynamics and geometric constraints imposed by the vessel wall. Stent-mesh nodes were iteratively updated in MATLAB under internal forces, external forces, and damping:
where is the equivalent mass of a stent node, is the position vector of node , is the iteration time, is the damping coefficient, is the internal force that maintains mesh connectivity and local structure, and is the external force that drives expansion toward the vessel wall. After discretization, the nodal positions were updated as follows:
Here, and are weighting coefficients for the internal and external forces, respectively. During iteration, the internal force maintained connectivity and relative spacing among braided nodes. The external force drove radial expansion, whereas damping suppressed numerical oscillation. The patient-specific vessel wall was treated as an impenetrable geometric boundary. Iteration ended when the nodes reached stable apposition to the wall and further iterations produced no appreciable geometric change, yielding the patient-specific implanted configuration.
After virtual deployment, the nodal coordinates and connectivity were exported from MATLAB. Solid stent struts were then generated along the deployed braid paths in SolidWorks (Dassault Systèmes, France) and spatially registered to the patient-specific vascular model.
For the giant aneurysm in Case 3, a coil-assisted embolization model corresponding to the clinical treatment was also created. The coils were represented by serially connected three-dimensional Timoshenko beam elements. The material was simplified as an alloy of 92% platinum and 8% tungsten, with an elastic modulus of 7.5 GPa, density of 21.3 g/cm3, and Poisson ratio of 0.39. The coil wire diameter was 0.2 mm, and the element length was 1.5 times the wire diameter. A displacement boundary condition was applied to the proximal coil node to guide the coil into the sac. Self-contact, coil-to-coil contact, and contact with the aneurysm wall were included. Coil-microcatheter contact was frictionless, and the coil-to-coil friction coefficient was 0.2 [19].
The coil volume embolization ratio (VER) was defined as the total volume of coil material within the aneurysm divided by the aneurysm cavity volume:
where is the total solid coil volume within the sac and is the aneurysm cavity volume. The VER was set to 20% in the combined-treatment model for Case 3. After virtual deployment, solid coil geometry was generated in SolidWorks and registered to the patient-specific aneurysm and deployed device. Both the stent struts and coils were explicitly retained in the fluid domain. Figure 1
2.4. Computational Mesh
Fluid domains before and after device implantation were imported into ANSYS ICEM CFD 2022 R2 for meshing. Unstructured tetrahedral meshes were generated using the Tetra/Mixed type and Robust Octree method. The global maximum element size was 1.0 mm, and the minimum curvature/proximity size was 0.02 mm. Untreated models included prism layers along the vessel wall, with a first-layer height of 0.01 mm, growth rate of 1.25, 12 layers, and a total thickness of approximately 0.542 mm. Device-treated models did not include prism layers because the small wire diameter and complex local topology created narrow gaps between the device and vessel wall. Instead, curvature/proximity-based tetrahedral refinement resolved flow around the device wires and coils [20]. These parameters were selected from a prior mesh-independence assessment. Figure 2
2.5. CFD Model and Numerical Solution
Steady-state blood flow was simulated in ANSYS CFX. Blood was modeled as an incompressible Newtonian fluid under laminar flow, with a density of 1060 kg/m3 and dynamic viscosity of 0.0035 Pa·s. The vessel wall, stent struts, and coils were treated as rigid walls with no-slip boundary conditions. These assumptions enabled within-patient comparison of relative flow changes across device conditions under a common inlet perfusion condition [21].
Steady incompressible flow satisfied the continuity and Navier-Stokes momentum equations:
where is the velocity vector, is static pressure, is blood density, and is dynamic viscosity. A volumetric flow rate of 4.6 mL/s was prescribed at the inlet of each model and converted to a uniform normal velocity using the patient-specific inlet cross-sectional area:
where is the inlet volumetric flow rate, is the inlet cross-sectional area, and is the mean inlet velocity. Cases 1 and 3 each had one inlet and one outlet. The outlet was specified as an Opening boundary with a relative pressure of 0 Pa, and local backflow was permitted. Case 2 had one inlet and two outlets. The inlet flow was distributed between the outlets in proportion to their cross-sectional areas:
This formulation maintained mass conservation by matching the total outlet flow to the inlet flow. All treated conditions used the same inlet boundary and outlet allocation as the corresponding untreated model, ensuring consistent within-case comparisons.
The solver used Steady State and Laminar settings with the High Resolution advection scheme. Auto Timescale was used with a timescale factor of 1.0. The maximum number of iterations was 800. Convergence required RMS residuals below 1 × 10−5, and inlet-outlet mass-flow balance was also checked. A solution was considered converged when the residual criterion was met and mass conservation remained stable.
2.6. Hemodynamic and Angiographic Evaluation
Hemodynamic results were postprocessed in CFD-Post. Within each case, all conditions were compared using the same number of streamlines and similar viewing angles. Velocity isosurfaces showed regions reaching a specified threshold. Thresholds were 0.50, 0.40, and 0.29 m/s for Cases 1, 2, and 3, respectively. The same threshold was applied across conditions within each case to compare the spatial extent of high-velocity regions before and after treatment. Because thresholds differed across cases, isosurface volumes were not compared directly between patients.
Quantitative evaluation included mean velocity in an analysis region and on an analysis plane. For Cases 1 and 3, sac-averaged velocity and neck-plane mean velocity were calculated. Case 2 lacked a typical aneurysm neck; therefore, mean velocity was calculated in the dilated vascular region and on a representative analysis plane in the dilated segment. The relative reduction after device implantation was defined as:
where and are the target velocity metrics before and after implantation, respectively. For Case 3, reductions after the device alone and after combined treatment were calculated relative to the untreated condition. The additional reduction with adjunctive coiling was also calculated relative to the device-only condition.
Angiographic evaluation included preprocedural, immediate postprocedural, and 12-month follow-up cerebral angiography. The assessment focused on contrast filling or stasis within the lesion, aneurysm occlusion, and parent-artery patency. All hemodynamic comparisons were descriptive and performed within each patient-specific case. No between-case statistical inference was conducted, and numerical differences among the three cases were not used to infer a general relationship between aneurysm size and treatment response.
3. Results
3.1. Patient-Specific Models and Virtual Device Deployment
Preprocedural images from all three patients were reconstructed into continuous patient-specific luminal models. After geometric repair, the inlets, outlets, lesion regions, and principal vessel curvatures remained intact for device deployment and flow simulation. Each device expanded along the parent artery and covered the intended treatment region. In Case 1, the device followed the curved right internal carotid artery and spanned the aneurysm neck. In Case 2, it traversed the dilated vertebral artery lesion and connected the proximal and distal segments. In Case 3, it followed the parent artery adjacent to the neck of the giant aneurysm, with coils distributed within the sac in the combined-treatment model. Deployed struts and coils were explicitly retained for CFD analysis.
3.2. Case 1: Small Sidewall Saccular Aneurysm
Under the untreated condition, the main flow followed the curved parent artery. Some streamlines deviated near the neck, entered the sac, and formed recirculation. After device implantation, the main flow through the parent artery remained continuous. Fewer high-velocity streamlines entered the sac, intra-aneurysmal streamlines shifted toward lower velocities, and recirculation weakened. At the 0.50 m/s isosurface threshold, the untreated high-velocity region occupied the parent artery and adjacent neck, with local extension into the sac entrance. After implantation, it was more confined to the parent artery and contracted near the neck. The neck-plane map likewise showed less high-velocity area and more low-velocity area after treatment. Figure 3A shows these changes.
Sac-averaged velocity decreased from 0.378 m/s under the untreated condition to 0.333 m/s after device implantation, a reduction of 11.9%. Neck-plane mean velocity decreased from 0.437 to 0.359 m/s, a reduction of 17.8%. Velocity decreased in both regions while the parent-artery flow pathway remained continuous.
3.3. Case 2: Dissecting or Locally Dilated Vertebral Artery Lesion
Under the untreated condition, flow dispersed after entering the dilated vertebral artery segment. Some streamlines deviated from the main direction, formed complex recirculation within the dilated region, and then entered the two outlets. After device implantation, streamlines were redistributed into a more concentrated pathway along the covered segment. Low-velocity flow increased around the dilated region, and both outlets remained perfused. At the 0.40 m/s threshold, the untreated high-velocity region extended continuously from the inlet into much of the dilated segment. After treatment, its transverse extent contracted and was concentrated within the device and toward the outlets. On the representative analysis plane, the area of moderate-to-high velocity decreased and the peripheral low-velocity area increased. Figure 3B shows these changes.
Mean velocity in the dilated region decreased from 0.340 to 0.240 m/s, a reduction of 29.4%. Mean velocity on the representative analysis plane decreased from 0.422 to 0.138 m/s, a reduction of 67.3%. Thus, in this patient-specific vertebral artery lesion, device implantation altered both the flow pathway and velocity distribution within the dilated segment.
3.4. Case 3: Giant Aneurysm
Under the untreated condition, numerous streamlines entered the giant aneurysm through the neck and formed large-scale circulation within the sac. After implantation of the flow-diverting device alone, intra-saccular streamline velocity decreased and the proportion of low-velocity flow increased, although pronounced circulation remained. After coils were added, fewer streamlines penetrated deeply into the sac. Residual flow became more confined to the neck region, and low-velocity flow predominated in the sac. At the 0.29 m/s threshold, the untreated high-velocity region extended from the parent artery through the neck into the sac. It contracted after device implantation but remained near the neck. With adjunctive coiling, it contracted further and was concentrated near the neck and parent artery. Figure 3C shows these changes.
Sac-averaged velocities were 0.089, 0.083, and 0.029 m/s under the untreated, device-only, and device-plus-coil conditions, respectively. The device alone reduced sac-averaged velocity by 6.7% relative to untreated. The combined treatment reduced it by 67.4% relative to untreated and by an additional 65.1% relative to the device alone. Neck-plane mean velocities were 0.136, 0.113, and 0.093 m/s, respectively. The corresponding reductions were 16.9% for the device alone versus untreated, 31.6% for combined treatment versus untreated, and an additional 17.7% for combined treatment versus the device alone.
In this representative case, adjunctive coiling produced a substantially greater additional reduction in sac-averaged velocity than in neck-plane velocity (65.1% vs. 17.7%). This within-case difference suggests that the coils provided additional suppression of flow within the main sac, whereas flow redirection at the neck remained primarily associated with the device spanning the aneurysm interface.
Table 2 summarizes mean velocities and relative changes across device conditions for all three cases.
3.5. Immediate Postprocedural and 12-Month Angiographic Follow-Up
All three patients underwent immediate postprocedural cerebral angiography and 12-month imaging follow-up. In Case 1, immediate angiography showed a patent parent artery and marked contrast pooling and stasis within the aneurysm, without complete occlusion. At 12 months, the sac no longer opacified, no evident residual neck was present, and the parent artery remained patent. In Case 2, immediate angiography showed patent parent and distal vessels after the device covered the lesion, with residual contrast stasis in the dilated region. At 12 months, the lesion cavity no longer opacified, no evident residual lesion remained, and the parent artery was patent. Case 3 was treated with the device and coils. Immediate angiography showed a patent parent artery and residual contrast stasis within the sac, without complete occlusion. At 12 months, the giant aneurysm no longer opacified, no evident residual neck remained, and the parent artery was patent. Figure 4 shows the preprocedural, immediate postprocedural, and 12-month angiographic findings.
Across the three cases, CFD showed lower lesion-region velocity or redistribution of the flow pathway after device implantation. Immediate angiography showed varying degrees of contrast stasis, and 12-month follow-up showed complete occlusion in all cases. The angiographic evolution was consistent with the direction of the immediate hemodynamic changes observed by CFD.
4. Discussion
This study evaluated the deployed configurations and hemodynamic effects of a flow-diverting device in three patient-specific clinical cases. Computational observations were descriptively compared with immediate and 12-month angiography. Three main observations emerged. First, device implantation reduced lesion-region velocity or redistributed flow while maintaining continuity of the parent-artery pathway. Second, the magnitude and pattern of the response varied with anatomy. Third, in the giant aneurysm, adjunctive coiling produced a greater additional reduction in sac-averaged velocity than in neck-plane velocity. These findings describe local device effects across representative treatment scenarios without implying uniform responses across aneurysm classifications.
A principal function of a flow-diverting device is to reduce direct exchange between parent-artery flow and the aneurysm or dilated lesion while preserving the parent artery [22]. In Case 1, spanning the neck reduced high-velocity inflow. Neck-plane and sac-averaged velocities decreased by 17.8% and 11.9%, respectively. Case 2 lacked a typical saccular neck. Nevertheless, implantation redirected the diffuse flow in the dilated region toward a more concentrated pathway along the covered segment, and analysis-plane mean velocity decreased by 67.3%. Thus, the local response depended on lesion morphology, vessel course, and the spatial configuration achieved after deployment [23,24].
The giant aneurysm enabled within-case comparison of the device alone and adjunctive coiling. After device implantation, neck-plane mean velocity decreased by 16.9%, whereas sac-averaged velocity decreased by 6.7% and large-scale intra-saccular circulation remained. At a VER of 20%, adjunctive coiling reduced sac-averaged velocity from 0.083 to 0.029 m/s, an additional 65.1% reduction relative to the device alone. Neck-plane velocity decreased by an additional 17.7%. In this representative case, the coils occupied the sac, disrupted large-scale intra-aneurysmal circulation, increased resistance to intra-saccular flow, and limited deep penetration of residual inflow. The device and coils therefore exerted distinct but complementary local effects.
Aneurysm healing after flow-diverting device implantation generally involves immediate hemodynamic alteration, subsequent intra-aneurysmal thrombosis, and longer-term endothelialization and vascular remodeling. The present CFD analysis described only the immediate hemodynamic stage and did not simulate thrombosis or endothelialization. Immediate angiography showed contrast stasis in all three cases, consistent with reduced blood exchange within the lesion. All lesions were completely occluded at 12 months.
The study device also had a PC-modified surface designed to improve blood compatibility and reduce thrombogenicity at the blood-material interface. The computational model evaluated only the deployed geometry and its mechanical regulation of the flow field. Surface chemistry, platelet adhesion, and coagulation were not represented. From a device-evaluation perspective, patient-specific CFD retained the local pore structure after deployment in realistic anatomy and enabled direct comparison of flow before and after implantation. This provides local hemodynamic information that nominal metal coverage or idealized geometry alone cannot capture [25]. In the combined-treatment case, explicitly retaining the coils also distinguished neck-level redirection from flow suppression within the sac. The approach may therefore complement structural device evaluation and clinical imaging observations.
This study had three principal limitations. First, only three patient-specific cases were included, and lesion location, morphology, and treatment strategy differed among them. The findings therefore describe within-case flow changes in representative clinical scenarios and do not support meaningful statistical inference. Second, CFD assumed steady flow, Newtonian blood, and rigid vessel walls [26]. Pulsatile flow, non-Newtonian behavior, and wall compliance were not considered. The results primarily represent relative differences among device conditions under standardized computational assumptions. Third, all lesions were completely occluded at 12 months, with no non-occluded or treatment-failure case. A predictive relationship between computed hemodynamic changes and clinical outcome therefore could not be established.
5. Conclusions
Flow-diverting devices alter blood exchange between the parent artery and the aneurysm or dilated lesion. This study evaluated three patients using patient-specific vascular reconstruction, virtual device deployment, and CFD with explicit device geometry. Implantation reduced lesion-region velocity and contracted high-velocity regions while preserving parent-artery flow. In the representative giant aneurysm, adjunctive coiling provided additional suppression of intra-saccular flow. Overall, the observations support patient-specific computational-angiographic evaluation of local hemodynamic alterations across representative anatomical and treatment scenarios.
Author Contributions
Z.Z. and Z.Q. contributed equally to this work (co-first author). Conceptualization, methodology, clinical data collection, angiographic image interpretation, original draft writing, formal analysis, data validation, Z.Z.; Numerical simulation design manuscript revision and polishing, literature research, Z.Q.; Visualization, data curation and result sorting, L.X.; Supervision, project administration, manuscript review and revision, correspondence work, W.S. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding .
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Informed consent was obtained from all subjects involved in the study.
Data Availability Statement
The datasets generated and/or analyzed during the current study are not publicly available but are available from the corresponding author on reasonable request and with appropriate agreements in place.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| CFD | Computational Fluid Dynamics |
| PC | phosphorylcholine |
| VMTK | Vascular Modeling Toolkit |
References
- Brinjikji, W.; Murad, M.H.; Lanzino, G.; Cloft, H.J.; Kallmes, D.F. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke 2013, 44(2), 442–447. [Google Scholar] [CrossRef]
- Wakhloo, A.K.; Lylyk, P.; de Vries, J.; Taschner, C.; Lundquist, J.; Biondi, A.; Hartmann, M.; Szikora, I.; Pierot, L.; Sakai, N.; Imamura, H.; Sourour, N.; Rennie, I.; Skalej, M.; Beuing, O.; Bonafé, A.; Mery, F.; Turjman, F.; Brouwer, P.; Boccardi, E.; Valvassori, L.; Derakhshani, S.; Litzenberg, M.W.; Gounis, M.J. Group, for the S.S. Surpass Flow Diverter in the Treatment of Intracranial Aneurysms: A Prospective Multicenter Study. Am. J. Neuroradiol. 2015, 36, 98–107. [Google Scholar] [CrossRef]
- Fischer, S.; Vajda, Z.; Aguilar Perez, M.; Schmid, E.; Hopf, N.; Bäzner, H.; Henkes, H. Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections. Neuroradiology 2012, 54, 369–382. [Google Scholar] [CrossRef]
- Becske, T.; Kallmes, D.F.; Saatci, I.; McDougall, C.G.; Szikora, I.; Lanzino, G.; Moran, C.J.; Woo, H.H.; Lopes, D.K.; Berez, A.L.; Cher, D.J.; Siddiqui, A.H.; Levy, E.I.; Albuquerque, F.C.; Fiorella, D.J.; Berentei, Z.; Marősfoi, M.; Cekirge, S.H.; Nelson, P.K. Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical Trial. Radiology 2013, 267, 858–868. [Google Scholar] [CrossRef]
- Kallmes, D.F.; Hanel, R.; Lopes, D.; Boccardi, E.; Bonafé, A.; Cekirge, S.; Fiorella, D.; Jabbour, P.; Levy, E.; McDougall, C.; Siddiqui, A.; Szikora, I.; Woo, H.; Albuquerque, F.; Bozorgchami, H.; Dashti, S.R.; Almandoz, J.E.D.; Kelly, M.E.; Turner, R.; Woodward, B.K.; Brinjikji, W.; Lanzino, G.; Lylyk, P. International Retrospective Study of the Pipeline Embolization Device: A Multicenter Aneurysm Treatment Study. Am. J. Neuroradiol. 2015, 36, 108–115. [Google Scholar] [CrossRef]
- MacDonald, D.E.; Cancelliere, N.M.; Pereira, V.M.; Steinman, D.A. Sensitivity of hostile hemodynamics to aneurysm geometry via unsupervised shape interpolation. Comput. Methods Programs Biomed. 2023, 241, 107762. [Google Scholar] [CrossRef]
- Chen, S.; Bai, B.; Lv, N.; Cheng, Y.; Ji, B. Hemodynamic analysis and implantation strategies of delayed intracranial aneurysm rupture after flow diverter treatment. Ann. Transl. Med. 2021, 9, 1735–1735. [Google Scholar] [CrossRef]
- Abdehkakha, A.; Hammond, A.L.; Patel, T.R.; Siddiqui, A.H.; Dargush, G.F.; Meng, H. Cerebral aneurysm flow diverter modeled as a thin inhomogeneous porous medium in hemodynamic simulations. Comput. Biol. Med. 2021, 139, 104988. [Google Scholar] [CrossRef]
- Zhou, Y.; Yang, P.-F.; Fang, Y.-B.; Xu, Y.; Hong, B.; Zhao, W.-Y.; Li, Q.; Zhao, R.; Huang, Q.-H.; Liu, J.-M. A Novel Flow-Diverting Device (Tubridge) for the Treatment of 28 Large or Giant Intracranial Aneurysms: A Single-Center Experience. Am. J. Neuroradiol. 2014, 35, 2326–2333. [Google Scholar] [CrossRef]
- Velvaluri, P.; Pravdivtseva, M.S.; Berg, P.; Wodarg, F.; de Miranda, R.L.; Hövener, J.-B.; Jansen, O.; Quandt, E. Thin-Film Patient-Specific Flow Diverter Stents for the Treatment of Intracranial Aneurysms. Adv. Mater. Technol. 2021, 6, 2100384. [Google Scholar] [CrossRef]
- Bouillot, P.; Brina, O.; Ouared, R.; Yilmaz, H.; Farhat, M.; Erceg, G.; Lovblad, K.-O.; Vargas, M.I.; Kulcsar, Z.; Pereira, V.M. Geometrical deployment for braided stent. Med. Image Anal. 2016, 30, 85–94. [Google Scholar] [CrossRef]
- White, T.G.; Santhumayor, B.A.; Turpin, J.; Shah, K.; Toscano, D.; Teron, I.; Link, T.; Patsalides, A.; Woo, H.H. Flow diverter surface modifications for aneurysm treatment: A review of the mechanisms and data behind existing technologies. Interv. Neuroradiol. 2026, 32, 109–125. [Google Scholar] [CrossRef]
- Zhang, Y.; Chong, W.; Qian, Y. Investigation of intracranial aneurysm hemodynamics following flow diverter stent treatment. Med. Eng. Phys. 2012, 35, 608. [Google Scholar] [CrossRef]
- Janiga, G.; Daróczy, L.; Berg, P.; Thévenin, D.; Skalej, M.; Beuing, O. An automatic CFD-based flow diverter optimization principle for patient-specific intracranial aneurysms. J. Biomech. 2015, 48, 3846–3852. [Google Scholar] [CrossRef]
- Kasem, R.A.; Hubbard, Z.; Cunningham, C.; Almorawed, H.; Isidor, J.; Tahhan, I.S.; Sowlat, M.-M.; Babool, S.; Abodest, L.; Spiotta, A.M. Comparison of flow diverter alone versus flow diverter with coiling for large and giant intracranial aneurysms: systematic review and meta-analysis of observational studies. J. NeuroInterventional Surg. 2026, 18, 739–749. [Google Scholar] [CrossRef]
- Sein, M.E.; Ceciliano, A.; Navarro, F.; Molina, S.; Villasante, F.; Tejado, E.E.; Requena, M.; Hernandez, D.; de Dios Lascuevas, M.; Tomasello, A.; Diana, F. Flow diverter with or without adjunctive coils in the treatment of large and giant intracranial aneurysms: a meta-analysis. Neurosurg. Rev. 2025, 48, 519. [Google Scholar] [CrossRef]
- Jeken-Rico, P.; Chau, Y.; Goetz, A.; Lannelongue, V.; Sédat, J.; Hachem, E. Virtual flow diverter deployment and embedding for hemodynamic simulations. Comput. Biol. Med. 2024, 180, 109023. [Google Scholar] [CrossRef]
- Ding, A.; Braschkat, A.; Guber, A.; Cattaneo, G. New Concept of Patient-specific Flow Diversion Treatment of Intracranial Aneurysms. Clin. Neuroradiol. 2021, 31, 671–679. [Google Scholar] [CrossRef]
- Jiang, Y.; Xu, L.; Zhang, Q.; Xu, G.; Yu, L.; Li, X.; Zhang, X.; Wan, J.; Wang, S. Hemodynamic effect of coil packing density after tubridge flow diverter for anterior circulation small and medium-sized aneurysms. Sci. Rep. 2026, 16, 18892. [Google Scholar] [CrossRef]
- Li, X.; Geng, J.; Feng, Y.; Wang, S.; Zhang, H. Real-Time Surgical Planning for Cerebral Aneurysms Treated With Intrasaccular Flow Disruption Devices Based on Fast Virtual Deployment and Discrete Element Method. Int. J. Numer. Methods Biomed. Eng. 2024, 40, e3886. [Google Scholar] [CrossRef]
- Wan, J.; Jiang, Y.; Xu, L.; Zhang, Q.; Xu, G.; Yu, L.; Li, X.; Zhang, X.; Wang, S. Exploration of the effect of morphology and location on hemodynamics of small aneurysms: a variable-controlled study based on two cases with tandem aneurysms. BioMed Eng. OnLine 2025, 24, 42. [Google Scholar] [CrossRef]
- Xu, L.; Zhao, B.; Liu, X.; Liang, F. Computational methods applied to analyze the hemodynamic effects of flow-diverter devices in the treatment of cerebral aneurysms: Current status and future directions. Med. Nov. Technol. Devices 2019, 3, 100018. [Google Scholar] [CrossRef]
- Kim, S.; Yang, H.; Oh, J.H.; Kim, Y.B. Quantitative analysis of hemodynamic changes induced by the discrepancy between the sizes of the flow diverter and parent artery. Sci. Rep. 2024, 14, 10653. [Google Scholar] [CrossRef]
- Roloff, C.; Berg, P. Effect of flow diverter stent malposition on intracranial aneurysm hemodynamics—An experimental framework using stereoscopic particle image velocimetry. PLoS ONE 2022, 17, e0264688. [Google Scholar] [CrossRef]
- Sun, A.; Zhao, C.; Gao, Z.; Deng, X.; Qiu, H. A proposed design of flow diverter and it’s hemodynamic validation. Med. Nov. Technol. Devices 2021, 9, 100049. [Google Scholar] [CrossRef]
- Uchiyama, Y.; Fujimura, S.; Takao, H.; Suzuki, T.; Ishibashi, T.; Otani, K.; Karagiozov, K.; Fukudome, K.; Yamamoto, H.; Yamamoto, M.; Murayama, Y. Role of patient-specific blood properties in computational fluid dynamics simulation of flow diverter deployed cerebral aneurysms. Technol. Health Care 2022, 30, 839–850. [Google Scholar] [CrossRef]
Figure 1.
Patient-specific aneurysm models and virtual deployment of interventional devices.

Figure 2.
Computational mesh of a patient-specific aneurysm model.

Figure 3.
Comparison of simulated hemodynamics before and after treatment in the three cases.

Figure 4.
Angiographic evolution after flow-diverting device treatment over 12 months.

Table 1.
Device parameters and clinical treatment for the three cases.
| Case | Device Size | Nominal Metal Coverage | Clinical Treatment |
| Case 1 | 4.0 × 20 mm | 40% | FD |
| Case 2 | 4.0 × 30 mm | 40% | FD |
| Case 3 | 5.0 × 25 mm | 40% | FD + 20%VER |
Note: The coil volume embolization ratio (VER) was defined as the total volume of coil material within the aneurysm divided by the aneurysm cavity volume.
Table 2.
Mean velocities and relative changes across device conditions.
| Case | Condition | Lesion-Region Mean Velocity (m/s) |
Reduction from Untreated (%) |
Analysis-Plane Mean Velocity (m/s) |
Reduction from Untreated (%) |
| Case 1 | Untreated | 0.378 | — | 0.437 | — |
| FD | 0.333 | 11.9% | 0.359 | 17.8% | |
| Case 2 | Untreated | 0.340 | — | 0.422 | — |
| FD | 0.240 | 29.4% | 0.138 | 67.3% | |
| Case 3 | Untreated | 0.089 | — | 0.136 | — |
| FD | 0.083 | 6.7% | 0.113 | 16.9% | |
| FD + coils | 0.029 | 67.4% | 0.093 | 31.6% |
Note: For Cases 1 and 3, the lesion region and analysis plane corresponded to the aneurysm sac and aneurysm neck plane, respectively; for Case 2, they corresponded to the dilated vascular region and a representative analysis plane.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2026 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license.
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.