Abstract
Objective: This study aimed to evaluate the effects of coffee temperature and immersion duration on the percentage weight change, force degradation, mechanical properties (300% modulus, and strain at break), and surface morphology of orthodontic elastomeric chains. We also aimed to examine the relationship between weight change and force degradation. Materials and Methods: Latex-free closed-type elastomeric chains were stretched to generate a force of 300 gf and assigned to four groups: control (artificial saliva, 37 °C) and three coffee-immersion groups (cold, 4 °C; room temperature, 25 °C; hot, 65 °C). Coffee groups were immersed for 15 min/day and stored in artificial saliva at 37 °C between immersion cycles. Weight change, force degradation, 300% modulus, strain at break, and surface morphology (scanning electron microscopy [SEM]) were evaluated on days 1, 3, 7, 14, and 21 (N = 10/group/time point). Two-way ANOVA with Tukey’s HSD post hoc test was used to analyze the effects of temperature and duration, and Pearson’s correlation was used to assess the relationship between weight change and force degradation. Results: Weight change was significantly affected by immersion temperature and duration (both p < 0.001), with the hot group showing significantly greater overall weight change than the control group (1.61% vs. 0.87%). Immersion temperature and duration significantly affected 300% modulus and strain at break, with significant temperature-duration interactions (p < 0.001). The hot group showed a consistently higher modulus and lower strain at break than the control group throughout the 21-day period, accompanied by progressively increasing surface irregularity on SEM. Force degradation was significantly greater in the hot group than in all other groups at every time point (p < 0.001) and increased continuously over 21 days, whereas the control, cold-, and room-temperature groups plateaued after days 7–14. Weight change was significantly and positively correlated with force degradation overall (r = 0.332, p < 0.001) and within each temperature group (r = 0.388–0.633), with the strongest correlation observed in the hot group. Conclusions: Coffee temperature and immersion duration significantly affected the physicochemical and mechanical properties of orthodontic elastomeric chains with hot-coffee exposure producing the greatest weight change, mechanical stiffening, surface roughening, and force degradation. Weight change was positively associated with force degradation, which suggests a link between physicochemical alteration and mechanical force loss, particularly under hot-temperature exposure.
Keywords:
1. Introduction
2. Material and Methods
2.1. Study Design
2.2. Specimen Preparation
2.3. Immersion Protocol
2.4. Weight Change Measurement
2.5. Force Degradation Measurement
2.6. Strain at Break and Modulus at Strain 300% Measurement
2.7. Scanning Electron Microscopy
2.8. Statistical Analysis
3. Results
3.1. Effect of Different Coffee Temperatures and Immersion Times on Weight Change of Elastomeric Chains
3.2. Effect of Coffee Temperature and Immersion Duration on Mechanical Properties
3.3. Effects of Coffee Temperature and Immersion Duration on Force Degradation of Elastomeric Chains
3.4. Correlation Between Weight Change and Force Degradation
4. Discussion
Author Contributions
Data Availability Statement
Acknowledgments
Declaration of generative AI and AI-assisted technologies in the writing process
Declaration of interest
References
- Storey, E. Force in orthodontics and its relation to tooth movement. Aust. Dent. J. 1952, 56, 11–18. [Google Scholar]
- Oshagh, M.; Khajeh, F.; Heidari, S.; et al. The effect of different environmental factors on force degradation of three common systems of orthodontic space closure. Dent. Res. J. 2015, 12, 50–56. [Google Scholar] [CrossRef]
- Elsayed, H.S.; El-Beialy, A.R.; Alshazly, R.; Almohammad, A.; Elazab, K.; El-Badawy, R.; Palomo, J.M.; Mostafa, Y.A. Implant-supported canine retraction using different reactivation intervals of elastomeric chains: A CBCT-based split-mouth randomized controlled trial. Dent. Press J. Orthod. 2023, 28, e2321166. [Google Scholar] [CrossRef]
- Teixeira, L.; Pereira, B.R.; Bortoly, T.G.; et al. The environmental influence of Light Coke, phosphoric acid, and citric acid on elastomeric chains. J. Contemp. Dent. Pract. 2008, 9, 17–24. [Google Scholar] [CrossRef]
- Issa, A.R.; Kadhum, A.S. Force degradation of orthodontic elastomeric chains: A literature review. J. Baghdad Coll. Dent. 2022, 34, 51–58. [Google Scholar] [CrossRef]
- Virdi, G.K.; Prashar, A.; Kaur, G.; Jabbal, R.S.; Aggarwal, P.; Budh, S.; Singh, D. Force decay behavior of orthodontic elastomeric chains in simulated oral conditions. Cureus 2024, 16, e69908. [Google Scholar] [CrossRef]
- Samoggia, A.; Riedel, B. Coffee consumption and purchasing behavior review: Insights for further research. Appetite 2018, 129, 70–81. [Google Scholar] [CrossRef]
- Van Dam, R.M.; Hu, F.B.; Willett, W.C. Coffee, caffeine, and health. N. Engl. J. Med. 2020, 383, 369–378. [Google Scholar] [CrossRef]
- Boubakri, A.; Haddar, N.; Elleuch, K.; Bienvenu, Y. Impact of aging conditions on mechanical properties of thermoplastic polyurethane. Mater. Des. 2010, 31, 4194–4201. [Google Scholar] [CrossRef]
- Andhare, P.; Datana, S.; Agarwal, S.S.; Chopra, S.S. Comparison of in vivo and in vitro force decay of elastomeric chains/modules: A systematic review and meta-analysis. J. World Fed. Orthod. 2021, 10, 155–162. [Google Scholar] [CrossRef]
- Eyri, B.; Gul, O.; Karsli, N.G.; Yilmaz, T. Study of the thermal and hydrothermal aging effects on the properties of 3D-printed thermoplastic polyurethane. J. Vinyl Addit. Technol. 2025, 31, 1–14. [Google Scholar] [CrossRef]
- Salazar, M.R.; Thompson, S.L.; Laintz, K.E.; Pack, R.T. Degradation of a poly(ester urethane) elastomer. I. Absorption and diffusion of water in Estane 5703 and related polymers. J. Polym. Sci. B Polym. Phys. 2002, 40, 2896–2916. [Google Scholar] [CrossRef]
- Boubakri, A.; Haddar, N.; Elleuch, K.; Bienvenu, Y. Influence of thermal aging on tensile and creep behavior of thermoplastic polyurethane. C. R. Méc. 2011, 339, 666–673. [Google Scholar] [CrossRef]
- Albilali, A.T.; Baras, B.H.; Aldosari, M.A. Evaluation of water sorption and solubility and FTIR spectroscopy of thermoplastic orthodontic retainer materials subjected to thermoforming and thermocycling. Appl. Sci. 2023, 13, 5165. [Google Scholar] [CrossRef]
- Wang, Y.; Song, J.; Tian, Q.; Song, N.; Liang, S.; Tian, C.; et al. Understanding water absorption effect on molecular dynamics, microstructures and relaxation behavior of segmented polyurethane elastomers. Polym. Degrad. Stab. 2023, 214, 110415. [Google Scholar] [CrossRef]
- Khaleghi, A.; Ahmadvand, A.; Sadeghian, S. Effect of citric acid on force decay of orthodontic elastomeric chains. Dent. Res. J. 2021, 18, 31. [Google Scholar] [CrossRef]
- Blagec, T.; Šimunović, L.; Gjumlić, S.P.; Šutej, I.; Meštrović, S. Influence of pH levels and beverage exposure on force decay and color stability of orthodontic elastomeric chains: An experimental study. Saudi Dent. J. 2024, 36, 308–314. [Google Scholar] [CrossRef]
- Šimunović, L.; Blagec, T.; Šutej, I.; Meštrović, S. Do oral antiseptics affect the force degradation of elastomeric chains? Appl. Sci. 2023, 13, 7290. [Google Scholar] [CrossRef]
- Torres-Rosas, R.; Torres-Gómez, N.; Camero-Leal, J.A.; Jurado, C.; López-Ravelo, H.; Argueta-Figueroa, L. Force decay and elongation of orthodontic elastomeric chains exposed to different beverages common in the diet: An in vitro study. Dent. Med. Probl. 2023, 60, 413–420. [Google Scholar] [CrossRef]
- Nobahari, M.; Safari, F.; Geramy, A.; Hooshmand, T.; Kharazifard, M.J.; Arab, S. Effect of three common hot beverages on the force decay of orthodontic elastomeric chain within a 28-day period: An in vitro study. Korean J. Orthod. 2024, 54, 153–159. [Google Scholar] [CrossRef]
- Quezada-Calle, F.E.; Lima-Illescas, M.; Cuenca-León, K.; Pacheco-Quito, E.M. Degradation of the strength of elastomeric chains exposed to energy drinks in a simulated oral environment. Clin. Cosmet. Investig. Dent. 2026, 18, 604910. [Google Scholar] [CrossRef]
- Binlateh, T.; Jitprasertwong, P.; Sangharn, K.; Cheinglew, N.; Pootrakull, T.; Thavorn, C.; Theerasopon, P.; Puttaravuttiporn, P.; Leethanakul, C.; Thammanichanon, P. Influence of coffee temperature and exposure parameters on colour change, surface roughness, and force remaining of orthodontic elastomeric chains: An in vitro comparison of cold, room temperature, and hot coffee. Int. Orthod. 2026, 24, 101186. [Google Scholar] [CrossRef]
- Dimić, I.; Cvijović-Alagić, I.; Kostić, I.T.; Perić-Grujić, A.A.; Rakin, M.P.; Putić, S.S.; Bugarski, B.M. Metallic ion release from biocompatible cobalt-based alloy. Chem. Ind. Chem. Eng. Q. 2014, 20, 571–577. [Google Scholar] [CrossRef]
- Braga, E.; Souza, G.; Barretto, P.; Ferraz, C.; Pithon, M. Experimental evaluation of strength degradation of orthodontic chain elastics immersed in hot beverages. J. Indian Orthod. Soc. 2019, 53, 244–248. [Google Scholar] [CrossRef]




| Source | df | Mean Square | F | p-value | Partial η2 |
|---|---|---|---|---|---|
| Group | 3 | 4.688 | 5.698 | <.001 | 0.087 |
| Day (Time) | 4 | 10.248 | 12.456 | <.001 | 0.217 |
| Group × Day | 12 | 0.452 | 0.549 | 0.880 | 0.035 |
| Residual | 180 | 0.823 |
| Day | Control | Cold | Room | Hot | |
|---|---|---|---|---|---|
| Day 1 | Mean | 0.457 aA | 0.726 aA | 0.732 aA | 0.733 aA |
| SD | 0.457 | 0.511 | 0.435 | 0.667 | |
| 95% CI | (0.112, 0.802) | (0.341, 1.112) | (0.404, 1.061) | (0.230, 1.236) | |
| Day 3 | Mean | 0.610 aA | 1.006 aA | 0.783 aA | 1.036 aA |
| SD | 0.434 | 0.640 | 0.508 | 0.863 | |
| 95% CI | (0.283, 0.938) | (0.523, 1.489) | (0.400, 1.166) | (0.385, 1.687) | |
| Day 7 | Mean | 1.004 aA | 1.153 aA | 1.146 aA | 1.487 abA |
| SD | 0.627 | 0.837 | 0.689 | 1.023 | |
| 95% CI | (0.531, 1.477) | (0.523, 1.784) | (0.627, 1.666) | (0.716, 2.259) | |
| Day 14 | Mean | 1.056 aA | 1.814 aA | 1.564 aA | 2.457 bA |
| SD | 0.995 | 1.603 | 0.992 | 1.010 | |
| 95% CI | (0.306, 1.806) | (0.605, 3.023) | (0.816, 2.312) | (1.695, 3.219) | |
| Day 21 | Mean | 1.241 aA | 1.939 aA | 1.678 aA | 2.332 bA |
| SD | 0.518 | 1.291 | 0.790 | 1.167 | |
| 95% CI | (0.851, 1.632) | (0.965, 2.912) | (1.082, 2.274) | (1.452, 3.212) | |
| Overall (EMM) | EMM | 0.874 a | 1.328 a | 1.181 a | 1.609 b |
| SE | 0.128 | 0.128 | 0.128 | 0.128 | |
| 95% CI | (0.621, 1.127) | (1.075, 1.581) | (0.928, 1.434) | (1.356, 1.862) |
| Source | Strain at Break | 300% Modulus | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| df | Mean Square | F | p-value | Partial η2 | df | Mean Square | F | p-value | Partial η2 | |
| Group | 3 | 39176.850 | 301.670 | <.001 | 0.836 | 3 | 33.576 | 311.258 | <.001 | 0.840 |
| Day | 4 | 12750.126 | 98.179 | <.001 | 0.688 | 4 | 8.908 | 82.578 | <.001 | 0.650 |
| Group × Day | 12 | 1747.695 | 13.458 | <.001 | 0.476 | 12 | 1.551 | 14.377 | <.001 | 0.492 |
| Residual | 180 | 129.867 | 180 | 0.108 | ||||||
| Source | df | Mean Square | F | p-value | Partial η2 |
|---|---|---|---|---|---|
| Group | 3 | 17312.020 | 10380.292 | <.001 | 0.994 |
| Day | 4 | 940.266 | 563.784 | <.001 | 0.926 |
| Group × Day | 12 | 15.625 | 9.369 | <.001 | 0.384 |
| Residual | 180 | 1.668 |
| Day | Control | Cold | Room | Hot | |
|---|---|---|---|---|---|
| Day 1 | Mean | 8.30 aA | 12.83 aB | 13.53 aB | 47.87 aC |
| SD | 1.260 | 1.128 | 1.368 | 1.067 | |
| 95% CI | (7.350, 9.250) | (11.983, 13.684) | (12.502, 14.565) | (47.062, 48.671) | |
| Day 3 | Mean | 15.70 bA | 16.77 bAB | 18.33 bB | 52.17 bC |
| SD | 1.224 | 1.221 | 1.535 | 2.162 | |
| 95% CI | (14.777, 16.623) | (15.846, 17.687) | (17.176, 19.491) | (50.537, 53.797) | |
| Day 7 | Mean | 17.60 cA | 21.47 cB | 21.47 cB | 56.23 cC |
| SD | 1.114 | 1.213 | 1.013 | 1.065 | |
| 95% CI | (16.760, 18.440) | (20.552, 22.381) | (20.703, 22.231) | (55.430, 57.036) | |
| Day 14 | Mean | 20.90 dA | 22.37 dB | 21.93 cdAB | 60.27 dC |
| SD | 0.746 | 0.900 | 0.786 | 1.254 | |
| 95% CI | (20.337, 21.463) | (21.688, 23.045) | (21.341, 22.526) | (59.321, 61.212) | |
| Day 21 | Mean | 22.03 dA | 23.30 dA | 23.07 cA | 62.33 eC |
| SD | 1.149 | 0.912 | 1.340 | 1.397 | |
| 95% CI | (21.167, 22.900) | (22.612, 23.988) | (22.056, 24.077) | (61.334, 63.332) |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |