Submitted:
21 May 2025
Posted:
22 May 2025
You are already at the latest version
Abstract
Keywords:
1. Introduction
2. Materials and Methods
Teeth Selection and Preparation
Length Measurement
Electronic Length Measurement
- 0.9% saline (Pliva Hrvatska, Zagreb, Croatia),
- Ultradent FileEZE EDTA 19%,
- Glyde EDTA 17% (Densply products, York, PA, USA),
- Ultradent Consepsis V Chlorhexidine 2% CHX,
- Ultradent Chlorcid 3%,
- Ultradent Citric Acid 20%,
- Ultradent EDTA 18%.
Statistical Analysis
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
| ACL | Actual Canal Length |
| EALs | Electronic Apex Locators |
| AF | Apical Foramen |
| ECL | Electronic Canal Length |
| WL | Working Length |
| GS | Gold standard |
References
- Sjögren, U.; Hägglund, B.; Sundqvist, G.; Wing, K. Factors affecting the long-term results of endodontic treatment. J. Endod. 1990, 16, 498–504. [CrossRef]
- Dummer, P.M.H.; McGINN, J.H.; Rees, D.G. The position and topography of the apical canal constriction and apical foramen. Int. Endod. J. 1984, 17, 192–198. [CrossRef]
- Farzaneh, M.; Abitbol, S.; Lawrence, H.P.; Friedman, S. Treatment Outcome in Endodontics—The Toronto Study. Phase II: Initial Treatment. J. Endod. 2004, 30, 302–309. [CrossRef]
- Martos, J.; Ferrer-Luque, C.M.; González-Rodríguez, M.P.; Castro, L.A.S. Topographical evaluation of the major apical foramen in permanent human teeth. Int. Endod. J. 2009, 42, 329–334. [CrossRef]
- West JD, Roane JB, Goerig AC. Cleaning and shaping the root canal length system. In: Cohen S, Burns RC, eds. Pathways of the pulp, 6th ed; Cohen S, Burns RC,. Publisher: Mosby, St. Louis, USA, 1994; pp. 179-218.
- Suzuki K . Study on iontophoresis. J Jap Stomato 1942.;16:411-417.
- Ebrahim, A.; Wadachi, R.; Suda, H. An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals. Aust. Dent. J. 2007, 52, 193–197. [CrossRef]
- Huang, L. An experimental study of the principle of electronic root canal measurement. J. Endod. 1987, 13, 60–64. [CrossRef]
- Nasiri, K.; Wrbas, K.-T. Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis. Saudi Dent. J. 2021, 34, 11–20. [CrossRef]
- Meredith, N.; Gulabivala, K. Electrical impedance measurements of root canal length. Dent. Traumatol. 1997, 13, 126–131. [CrossRef]
- Jenkins JA, Walker WA 3rd, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the root ZX in the presence of various irrigants. J Endod 2001;27(3):209-211.
- Križaj, D.; Jan, J.; Valenčič, V. Modeling AC current conduction through a human tooth. Bioelectromagnetics 2004, 25, 185–195. [CrossRef]
- Venturi, M.; Breschi, L. A comparison between two electronic apex locators: an ex vivo investigation. Int. Endod. J. 2007, 40, 362–373. [CrossRef]
- Özsezer, E.; İnan, U.; Aydın, U. In Vivo Evaluation of ProPex Electronic Apex Locator. J. Endod. 2007, 33, 974–977. [CrossRef]
- Gomes, S.; Oliver, R.; Macouzet, C.; Mercadé, M.; Roig, M.; Duran-Sindreu, F. In Vivo Evaluation of the Raypex 5 by Using Different Irrigants. J. Endod. 2012, 38, 1075–1077. [CrossRef]
- Altenburger MJ, Tchorz JP, Somma F. In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II. Int Endod J 2012;45(11):1053-1054.
- Duran-Sindreu, F.; Gomes, S.; Stöber, E.; Mercadé, M.; Jané, L.; Roig, M. In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants. Int. Endod. J. 2012, 46, 769–774. [CrossRef]
- Kaufman, A.Y.; Keila, S.; Yoshpe, M. Accuracy of a new apex locator: an in vitro study. Int. Endod. J. 2002, 35, 186–192. [CrossRef]
- Meares, W.A.; Steiman, H.R. The Influence of Sodium Hypochlorite Irrigation on the Accuracy of the Root ZX Electronic Apex Locator. J. Endod. 2002, 28, 595–598. [CrossRef]
- Kang, J.-A.; Kim, S.K. Accuracies of seven different apex locators under various conditions. Oral Surgery, Oral Med. Oral Pathol. Oral Radiol. Endodontology 2008, 106, e57–e62. [CrossRef]
- Jain, S.; Kapur, R. Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study. Contemp. Clin. Dent. 2012, 3, 140–5. [CrossRef]
- Sinha, N.; Baruah, Q.; Singh, B.; Reddy, P.N.; Baruah, K.; Augustine, V. Comparative evaluation of accuracy of two electronic apex locators in the presence of contemporary irrigants: An in vitro study. J. Int. Soc. Prev. Community Dent. 2018, 8, 349–353. [CrossRef]
- Altunbaş, D.; Kuştarcı, A.; Toyoğlu, M. The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations. J. Endod. 2017, 43, 439–442. [CrossRef]
- Yahya, E.M.; Alchalabi, A.S.; Alkhalidi, E.F. Effect of Different Electroconductive Root Canal Irrigations on the Accuracy of Different Apex Locators: An In Vitro Comparative Study. J. Int. Soc. Prev. Community Dent. 2024, 14, 211–217. [CrossRef]
- Karunakar, P.; Reddy, M.S.R.; Faizuddin, U.; Priya, P.S.; Karteek, B.S.; Saideep, C.V. Comparative evaluation of working length determined using integrated apex locator and root ZX mini under various irrigating solutions: An in vivo study.. 2023, 26, 590–594.
- Cîmpean, S.I.; Chisnoiu, R.M.; Burtea, A.L.C.; Rotaru, R.; Bud, M.G.; Delean, A.G.; Pop-Ciutrilă, I.-S. In Vitro Evaluation of the Accuracy of Three Electronic Apex Locators Using Different Sodium Hypochlorite Concentrations. Medicina 2023, 59, 918. [CrossRef]
- Chukka, R.R.; Bellam, M.D.; Marukala, N.R.; Dinapadu, S.; Konda, N.K.; Nagilla, J. Efficiency of an integrated apex locator in determining working length in various irrigating solutions: An in vivo study. J. Pharm. Bioallied Sci. 2020, 12, 410–414. [CrossRef]
- Prasad AB, Harshit S, Aastha SA, Deepak R. An In vitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants. Ethiop J Health Sci 2016;26(5):457-462.
- Tinaz, A.C.; Maden, M.; Aydin, C.; Türköz, E. The accuracy of three different electronic root canal measuring devices: An in vitro evaluation.. J. Oral Sci. 2002, 44, 91–95. [CrossRef]
- Bernardes, R.A.; Duarte, M.A.; Vasconcelos, B.C.; Moraes, I.G.; Bernardineli, N.; Garcia, R.B.; Baldi, J.V.; Victorino, F.R.; Bramante, C.M. Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30. Oral Surgery, Oral Med. Oral Pathol. Oral Radiol. Endodontology 2007, 104, e91–e94. [CrossRef]
- D́assunção, F.L.C.; de Albuquerque, D.S.; Ferreira, L.C.d.Q. The Ability of Two Apex Locators to Locate the Apical Foramen: An In Vitro Study. J. Endod. 2006, 32, 560–562. [CrossRef]
- de Camargo, É.J.; Zapata, R.O.; Medeiros, P.L.; Bramante, C.M.; Bernardineli, N.; Garcia, R.B.; de Moraes, I.G.; Duarte, M.A.H. Influence of Preflaring on the Accuracy of Length Determination With Four Electronic Apex Locators. J. Endod. 2009, 35, 1300–1302. [CrossRef]
- Kim E, Lee SJ. Electronic apex locator. Dent Clin North Am 2004;48(1):35-54.
- Briseño-Marroquín, B.; Frajlich, S.; Goldberg, F.; Willershausen, B. Influence of Instrument Size on the Accuracy of Different Apex Locators: An In Vitro Study. J. Endod. 2008, 34, 698–702. [CrossRef]
- Marek, E.; Łagocka, R.; Kot, K.; Woźniak, K.; Lipski, M. The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators. BMC Oral Heal. 2019, 20, 1–8. [CrossRef]
- Herrera, M.; Ábalos, C.; Lucena, C.; Jiménez-Planas, A.; Llamas, R. Critical Diameter of Apical Foramen and of File Size Using the Root ZX Apex Locator: An In Vitro Study. J. Endod. 2011, 37, 1306–1309. [CrossRef]
- Keskin, I.S.Ç.; Yalçın, T.Y. Evaluation of heated sodium hypochlorite’s effect on the accuracy of contemporary electronic apex locators: an in vitro study. BMC Oral Heal. 2025, 25, 634. [CrossRef]
- Golvankar, K.; Kader, M.A.; Latheef, A.A.; Ali, A.B.M.; Abullais, S.S.; Sibagathullah, M. Comparison of Accuracy in Determining the Root Canal Working Length by Using Two Generations of Apex Locators – An In Vitro Study. Open Access Maced. J. Med Sci. 2019, 7, 3276–3280. [CrossRef]
- Rifaat S, Aljami A, Alshehri T, T Alameer S, Aldossary A, Almutairi W, N Almaliki M, A Farooqi F, Taymour N. The Effect of Coronal Pre-flaring and Type of Root Canal Irrigation on Working Length Accuracy Using Electronic Apex Locators. F1000Res 2023;12:533.
- Gürel, M.A.; Kıvanç, B.H.; Ekici, A. A COMPARATIVE ASSESSMENT OF THE ACCURACIES OF RAYPEX 5, RAYPEX 6, IPEX AND IPEX II ELECTRONIC APEX LOCATORS: AN IN VITRO STUDY. J. Istanb. Univ. Fac. Dent. 2016, 51, 28–33. [CrossRef]
- Nasiri, K.; Wrbas, K.-T. Comparing the accuracy of two electronic apex locators in the determination of working length and the detection of root perforations: An in vitro study. Dent. Oral Maxillofac. Res. 2019, 5. [CrossRef]
- Tufenkci, P.; Kalaycı, A. Evaluation of the accuracy of different apex locators in determiningthe working length during root canal retreatment. J. Dent. Res. Dent. Clin. Dent. Prospect. 2020, 14, 125–129. [CrossRef]
- Bilaiya, S.; Patni, P.; Jain, P.; Pandey, S.H.; Raghuwanshi, S.; Bagulkar, B. Comparative evaluation of accuracy of ipex, root zx mini, and epex pro apex locators in teeth with artificially created root perforations in presence of various intracanal irrigants. Eur. Endod. J. 2020, 5, 6–9. [CrossRef]
- Shekarbaghani, S.; Bolhari, B.; Khalilak, Z. The effect of different root canal irrigations on the accuracy of apex locators: A systematic review. J. Clin. Exp. Dent. 2024, 16, e1538–e1546. [CrossRef]
- Mull, J.P.; Manjunath, V.; Manjunath, M. Comparison of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study. J. Conserv. Dent. 2012, 15, 178–82. [CrossRef]
- Fan, W.; Fan, B.; Gutmann, J.L.; Bian, Z.; Fan, M.W. Evaluation of the accuracy of three electronic apex locators using glass tubules. Int. Endod. J. 2006, 39, 127–135. [CrossRef]


| Groups (mm) | Raypex 5 | ||||||
|
0.9% saline |
File EZE EDTA 19% | Glyde EDTA 17% | Consepsis V 2% CHX | ChlorCid 3% |
Citric Acid 20% |
EDTA 18% | |
| <-1 (A) | 0(0%) | 1(2.5%) | 9(22.5%) | 1(2.5%) | 0(0%) | 4(10%) | 1(2.5 %) |
| -1.00 to -0.5 (B) | 2(5%) | 10(25%) | 1(2.5%) | 4(10%) | 0(0%) | 9(22.5%) | 10(25%) |
| -0.5 to 0.0 (C) | 15(37.5%) | 14(35%) | 10(25%) | 12(30%) | 17(42.5%) | 11(27.5%) | 18(45%) |
| >0.0 to 0.5 (D) | 20(50%) | 11(27.5%) | 12(30%) | 18(45%) | 16(40%) | 11(27.5%) | 7(17.5%) |
| 0.5 to 1 (E) | 3(7.5%) | 4(10%) | 5(12.5%) | 4(10%) | 7(17.5%) | 4(10%) | 4(10%) |
| > 1 (F) | 0 (0%) | 0(0%) | 3(7.5%) | 1(2.5%) | 0(0%) | 1(2.5%) | 0(0%) |
| Total | 40 | 40 | 40 | 40 | 40 | 40 | 40 |
| Groups (mm) | Ipex | ||||||
|
0.9% saline |
FileEZE EDTA 19% | Glyde EDTA 17% | Consepsis V 2% CHX | ChlorCid 3% | Citric Acid 20% | EDTA 18% | |
| <-1 (A) | 0(0%) | 0(0%) | 0(0%) | 2(5%) | 0(0%) | 3(7.5%) | 2(5%) |
| -1.00 to -0.5 (B) | 3(7.5%) | 4(10%) | 2(5%) | 3(7.5%) | 4(10%) | 5(12.5%) | 10(25%) |
| -0.5 to 0.0 (C) | 18(45 %) | 5(12.5%) | 10(25%) | 11(27.5%) | 19(47.5%) | 17(42.5%) | 15(37.5%) |
| >0.0 to 0.5 (D) | 15(37.5 %) | 17(42.5%) | 13(32.5%) | 17(42.5%) | 13(32.5%) | 10(25%) | 11(27.5%) |
| 0.5 to 1 (E) | 4(10%) | 12(30%) | 12(30%) | 6(15%) | 4(10%) | 5(12.5%) | 1(2.5%) |
| > 1 (F) | 0(0 %) | 2(5%) | 3(7.5%) | 1(2.5%) | 0(0%) | 0(0%) | 1(2.5%) |
| Total | 40 | 40 | 40 | 40 | 40 | 40 | 40 |
| Mean value | SD | N | |
| Raypex 0.9% saline1 | 0.038 | 0.362 | 40 |
| Raypex File EZE EDTA 19%2 | -0.165 | 0.503 | 40 |
| Raypex Glyde EDTA 17% | -0.083 | 0.837 | 40 |
| Raypex Consepsis V 2% CHX | 0.047 | 0.484 | 40 |
| Raypex ChlorCid 3%2,3 | 0.087 | 0.381 | 40 |
| Raypex Citric Acid 20% | -0.183 | 0.649 | 40 |
| Raypex EDTA 18%1,3 | -0.239 | 0.488 | 40 |
| Ipex 0.9% saline4 | 0.017 | 0.351 | 40 |
| Ipex File EZE EDTA 19%5,6 | 0.266 | 0.513 | 40 |
| Ipex Glyde EDTA 17%7,8,9,10 | 0.299 | 0.549 | 40 |
| Ipex Consepsis V 2% CHX4 | 0.048 | 0.519 | 40 |
| Ipex ChlorCid 3%8 | -0.017 | 0.433 | 40 |
| Ipex Citric Acid 20%5,9 | -0.141 | 0.533 | 40 |
| Ipex EDTA 18%4,6,10 | -0.253 | 0.520 | 40 |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).