II.1. Reference electrode
The
standard electrocardiogram (ECG) employs 12 leads and includes a reference electrode positioned on the Right Leg (RL), resulting in a total of 9 electrodes [
2] responsible for its generation:
The three limb electrodes give rise to the three vertices of Einthoven’s Triangle [
2]. These vertices are separated from each other by an angle of 120º, considering the lines originating from the centroid.
Based on this, we can define:
where
represents the cardiac electrical potential measured by an
electrode on the limb or, equivalently, at a vertex point of Einthoven’s Triangle.
Eq.1 signifies that the values of various potentials are inherently linked to the potential of the right leg,
. Assuming that
is measured concerning a neutral potential (e.g., earth ground)[
2], any potential difference measurement should inherently refer to
. For instance:
.
Similarly, the augmented leads introduced by Goldberger are calculated as the difference between one of the limb potentials (
,
, and
) and the average of the remaining two [
5]. For instance:
This action amplifies the signal by 50 % if the reference were established using the WCT (
non-augmented leads: VR, VL, and VF):
where
is the potential of the WCT, which historically has been defined as [
2,
5,
6]:
II.2. Wilson’s Central Terminal
As evident in Eq.2, the calculation of WCT involves taking the average of potentials recorded by the cardinal leads of Einthoven’s Triangle. However, an additional aspect needs consideration: a high-value resistance (≥ 5 kΩ) is added to those connected to each of the electrodes. These three corresponding electrodes are positioned to converge at a common point, where the theoretical placement of the WTC is posited: at the centroid of Einthoven’s Triangle [
2,
6] (
Figure 1).
Clearly represented in Eq.2, WCT is calculated as the average of the potentials registered by the cardinal leads of the Einthoven’s Triangle; but we must add one aspect: from those that are connected to each of the electrodes a resistance of high value (≥ 5 kΩ). The three corresponding electrodes are made to coincide at a common point, which is where the theoretical location of the WTC is assumed: the centroid of the Einthoven’s Triangle [
2,
6] (
Figure 1).
The WCT serves as a neutral reference point for the precordial leads within the 12-lead ECG. As demonstrated in Eq.2, its influence is rooted in the three cardinal electrodes, implying that its value hinges on the positioning of the limb electrodes. However, recent observations have unveiled fluctuations in the WCT throughout the cardiac cycle, resembling behaviour akin to a lead [
2,
7,
8]. There has even been suggesting its potential existence beyond the geometric plane defined by the cardinal leads [
2], in essence, beyond the plane encompassing Einthoven’s Triangle.
Despite a few investigations [
2,
7,
8], research on the WCT has been relatively limited in recent years, perhaps attributed to the arduous methods necessitated by experimental studies attempting its measurement, such as the immersion of study subjects in water after their confinement within a metallic structure [
7].
In the 1950s,
Ernest Frank [
9], in his influential work
“General theory of heat-vector projection”, was the pioneer in raising concerns about fluctuations in the WCT during the cardiac cycle, an observation that has since been validated. This validation is exemplified by recent measurements, such as the quantification of the WCT amplitude relative to the DII amplitude, yielding a value of 51.2 % (with a reported standard deviation of 27.4 %) and even reaching peaks exceeding 100 %[
8].
The comprehensive investigation conducted by
Gargiulo et al. in 2016[
8] substantiates considering the WCT as an additional reference. However, the persistence of a theoretical discrepancy between Wilson’s original work and present-day observations raises questions. Could there be an opportunity to refine the definition of the WCT grounded in its physical-mathematical structures?
As of today, ongoing experiments continue to probe the significance of the WCT and its spatial location. Nonetheless, it remains imperative to synthesize the historical theoretical groundwork with the findings of contemporary empirical studies.
II.3. Bipolar Lead Vector Definition
The generation of an ECG recording is predicated on a model wherein the limb (or cardinal) leads constitute Einthoven’s Triangle. Within this triangle’s centroid, a dipole moment vector is postulated to be anchored at its origin, functioning as an analogy of cardiac electrical activity. Throughout the cardiac cycle, it rotates as a temporal function [
2]. We establish the concept of an arbitrary electric dipole consistent with theoretical principles [
6], represented as
, relative to the canonical basis
of the Cartesian Coordinate System. This dipole is affixed at the origin and is defined as follows:
where
,
, and
denote the components of
along the
,
, and
axes, respectively.
Consider an arbitrary point
located on the surface of any conductor. We establish the potential
due to the presence of the dipole
at point
as follows:
where
is a vector whose components,
,
, and
, correspond to the values of the measurement
under the assumption that the electric dipole is represented by the basis vectors themselves. The operation
denotes the dot product. This vector,
, is identified as the
lead vector. In this context, the reference point for potential measurements at point
is regarded as arbitrary [
6].
When we take into account an infinite number of points,
, distributed across the surface of the conductor and assume a constant dipole, we can express:
This enables us to define:
where
denotes the
electric potential difference between points
and
. We refer to these points as
“leads”, and they are the locations on the conductor’s surface where the electrodes are positioned [
6] (
Figure 2).
Following this, we have:
where
, and this difference is referred to as the
“bipolar lead vector” since it involves two measurement points (leads).
II.4. Lead Vectors of Einthoven’s Triangle
Within Einthoven’s Triangle, the lead vectors
,
and
are characterized as vectors originating from the centroid of the triangle and extending to the vertices R, L, and F (signifying the right arm, left arm, and left leg electrodes, respectively) (
Figure 3). Consequently, the potentials at the vertices can be established for an arbitrary constant dipole
as follows:
, and within Einthoven’s Triangle.
Einthoven himself provided the definition [
6]:
where:
These vectors (, and ) originate and terminate at the vertices of Einthoven’s Triangle. They are the vectors recognized in clinical practice, as they define the orientations of the axes DI, DII, and DIII within the Bailey Hexaxial System.
The notations DI, DII, and DIII essentially represent the “standard leads of the members”. In other words, they denote the notations employed for the scalar magnitudes of potential differences , and .
The magnitudes of the potential differences
,
and
in the directions
,
and
, respectively, they are not independent; rather, they adhere to Kirchhoff’s Law [
6]:
II.5. The Front Plane and Appropriate Coordinate System
To accurately define the spatial orientation of Einthoven’s Triangle, it is imperative to adopt a coordinate system that eliminates any ambiguity or confusion in the interpretation of magnitudes involved in our analysis.
For the purposes of this paper, we will establish a reference coordinate system in which the positive directions of the
,
, and
axes correspond to the front, right (from the observer’s perspective), and top directions, respectively. With this setup, assuming α represents the angle between the vector
, fixed at the centroid of Einthoven’s Triangle, and the horizontal axis, we can express this relationship as follows [
6]:
This relationship leads to the following conclusion:
It's noteworthy that these three vector equations involve the component
. In other words, the plane’s description disregards the
-coordinate of our chosen coordinate system (
Figure 4).
, and within Einthoven’s Triangle.
II.6. Vector Analysis of Central Terminal Structures
Based on the accumulated theory, let’s delve into the workings “inside” the definition of the WCT. According to Eq.2,
The sum invariably results in the additive identity of vector summation due to the additive inverse nature of vector operations. Additionally, the outcome of Eq.3 yields a scalar value of zero, stemming from the dot product of a vector , where its components , and are all null.
If Eq.3 were valid, the introduction of resistors to establish a neutral reference would not have been essential; indeed, the reference potential of WCT would have been precisely zero.
Conversely, in augmented leads, a similar strategy with resistors is adopted. Here, two resistances of
are integrated from the leads designated as reference points. These leads are connected and converge at a common point known as
Goldberger’s Central Terminal (GCT). Intriguingly, the orientation of the augmented lead vectors, namely aVR, aVL, and aVF, corresponds to that of the vectors
,
and
, respectively. This alignment signifies a clear and distinctive functional relationship between Einthoven’s and Goldberger’s leads:
As per Eq.4, it becomes necessary to suggest the positioning of three GCT at the midpoint of each side of Einthoven’s Triangle. Consequently, the orientation of augmented leads aligns with the medians of Einthoven’s Triangle, consistent with the respective vertices they are denoted by.
Interestingly, the parallels between the behaviour of GCT and WCT appear to diverge. The author’s literature review did not yield significant data that reproduces the phenomena observed with WCT.
II.7. Proposal for a New Central Terminal
Building on Goldberger’s concept of enhancing the signal from leads VR, VL, and VF by altering WCT to GCT, a notion arises to extract the potential from the specific electrode in question.
Given the intrinsic relationship between augmented leads and the cardinal leads, we can infer the subsequent equations [
5]:
While Goldberger’s concept was indeed ingenious, it’s noteworthy that the precordial leads continue to reference the WCT. Notably, no inherent issue arises from this arrangement, as none of the precordial leads are encompassed within the WCT framework.
The rationale behind augmenting the signal was particularly pertinent during an era when ECG traces were characterized by their high thickness and low amplitude [
10].
Could it be possible to formulate a Central Terminal for the precordial leads using reasoning akin to Goldberger’s approach? Goldberger extracted the electrode potentials from an existing Central Terminal (Wilson’s), while keeping the remaining two as references. However, in the case of precordial leads, they refer to a Central Terminal that doesn’t inherently include the precordial electrodes.
It raises the question: can we potentially devise a Central Terminal solely using precordial leads that could function as a reference point? If such a formulation were attainable, the terminal’s fluctuations might parallel the behaviour observed in GCT fluctuations within augmented leads, which demonstrate consistent comportment.
A study conducted by
Medias et al. [
10] presents intriguing insights. They suggested a shift from the 12-lead ECG to a 9-lead ECG comprising the 3 non-augmented leads (VR, VL, VF) and the 6 precordial leads. They proposed using VR, VL, and VF leads by placing the initial three precordial electrodes on the right arm, left arm, and left leg, respectively [
10]. This generated three signals that, while slightly reduced in amplitude compared to their magnified counterparts, maintained sufficient visual quality for clinical diagnosis.
Moreover, an alternate 9-lead ECG model has been employed, maintaining the concept of a reliable reference. For example, a model retained the limb leads (both Goldberger’s and Einthoven’s) but excluded V2, V4, and V5. Although this model still used 12 electrodes, it incorporated 6 distinct reference electrodes positioned on the right leg [
11].
After thoroughly examining prior research and aligning it with the theoretical framework developed in this paper, we were initially inclined to propose the following definition:
This concept aimed to establish a precordial lead in relation to the average of the other five leads, distributing electrodes equally around the thorax. However, upon more in-depth analysis, we realized that this formulation would not harmonize with the vector theory we have developed, as the lead vectors would lack a coherent arrangement.
Upon further scrutiny, and consistent with the current theory, we arrive at the following proposition as a potential solution to the issue of precordial reference electrodes:
In this context, indicates that the other two leads being considered must be separated by an angle of 120º.
To achieve the desired spatial distribution of the 6 precordial electrodes, it is necessary to position the electrodes around the thoracic circumference at intervals of 60º, as 360º/6 = 60º (
Figure 5).
As illustrated in
Figure 5, we can create a configuration resembling two “Einthoven’s Triangles” arranged horizontally, forming what we can refer to as a “six pointed star”. Drawing from the earlier vector analysis, the distinct Central Terminals will be situated at the midpoint of the opposing side of each precordial lead, with both triangles sharing the same centroids. Furthermore, the arrangement of the bipolar lead vectors reproduces that of Einthoven’s Triangle with a frontal orientation. This signifies that we can establish lead vectors within our horizontal triangles much like we described for Einthoven’s Triangle.
By placing the origin of our coordinate system at the intersection point of all the medians from both triangles, we acquire a coordinate framework in which we can accommodate the frontal Einthoven’s Triangle and the two Einthoven’s Triangles.
In light of this, we can conceive a “novel electrocardiograph” that is entirely standardized in terms of its Central Terminals and leads, exclusively utilizing “precordial” electrodes. Consequently, “non-augmented” frontal plane leads could be derived from the same precordial leads, as proposed by
Medias et al. [
10]. This approach, considering the insights presented by
Cho et al. [
11], suggests that even the reference to the right leg might not be dispensable.
Traditional electrocardiographs are structured with three main components: Einthoven’s leads, Goldberger’s leads, and precordial leads. The first two exhibit clear functional relationships, while the third, seemingly isolated, holds inherent potential to generate three frontal plane leads of adequate quality. In the contemporary context, it appears essential to standardize conventional electrocardiography, despite the historical concepts tailored to our clinical convenience.