Cross Sectional Analysis of Impact of Seasonal Changes on Excimer Laser Ablation Performance on Polymethyl Methacrylate (PMMA)

Version 1 : Received: 4 May 2023 / Approved: 9 May 2023 / Online: 9 May 2023 (14:39:59 CEST)

A peer-reviewed article of this Preprint also exists.

Verma, S.; Hesser, J.; Mosquera, S.A. Cross Sectional Analysis of Impact of Seasonal Changes on Excimer Laser Ablation Performance on Polymethyl Methacrylate (PMMA). Vision 2023, 7, 50. Verma, S.; Hesser, J.; Mosquera, S.A. Cross Sectional Analysis of Impact of Seasonal Changes on Excimer Laser Ablation Performance on Polymethyl Methacrylate (PMMA). Vision 2023, 7, 50.

Abstract

Seasonal changes and varying degree of corneal hydration has been linked to excimer laser corneal ablation rates. Use of PMMA as calibration material in refractive lasers is well established. However, PMMA ablation may be equally affected by seasonal variations in temperature and humidity, intern affecting the calibration process. Aim of this work is to analyze the effect of seasonal changes in PMMA Performance using SCHWIND AMARIS laser system. PET and PMMA ablations conducted in climate-controlled environment with 826 consecutive AMARIS systems manufactured over 6 years were retrospectively analyzed. Lasers were stratified depending on seasons and months of the year. Metrics like Single laser pulse fluence, nominal number of laser pulses, mean performance, standard deviation, and technical performance of system were compared to global average values. Cyclic winter-summer variation was confirmed with seasons winter and summer showing statistically significant variations with respect to global values. Metric technical performance showed deeper PMMA ablation performance in summertime, with maximum seasonal deviation of 6%. Results were consistently confirmed in seasonal as well as monthly analyses. These findings could help minimize variance among laser systems by implementing compensation factors depending on seasons such that laser systems installed worldwide follow the same trend line of variation.

Keywords

Impact of seasonal variations; Humidity; Temperature; Excimer Laser Ablation; PMMA; Cornea; PET; Refractive surgery

Subject

Medicine and Pharmacology, Ophthalmology

Comments (0)

We encourage comments and feedback from a broad range of readers. See criteria for comments and our Diversity statement.

Leave a public comment
Send a private comment to the author(s)
* All users must log in before leaving a comment
Views 0
Downloads 0
Comments 0
Metrics 0


×
Alerts
Notify me about updates to this article or when a peer-reviewed version is published.
We use cookies on our website to ensure you get the best experience.
Read more about our cookies here.