Rubidium-82 Generator Yield and Efficiency for PET Perfusion Imaging: Comparison of Two Clinical Systems
Version 1
: Received: 21 April 2020 / Approved: 22 April 2020 / Online: 22 April 2020 (07:32:27 CEST)
How to cite: deKemp, R. Rubidium-82 Generator Yield and Efficiency for PET Perfusion Imaging: Comparison of Two Clinical Systems. Preprints.org 2020, 2020040393 deKemp, R. Rubidium-82 Generator Yield and Efficiency for PET Perfusion Imaging: Comparison of Two Clinical Systems. Preprints.org 2020, 2020040393
Abstract
Introduction. Strontium-82/Rubidium-82 (82Sr/82Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the 82Rb isotope yield and production efficiency of two FDA-approved 82Sr/82Rb generators were compared.Methods. N=515 sequential daily quality assurance (QA) reports from 9 CardioGen-82® and 9 RUBY-FILL® generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL® system to determine an empirical correction-factor used to convert CardioGen-82® daily QA values of 82Rb activity (dose-calibrator ‘maximum’ of 50 mL elution at 50 mL/min) to RUBY-FILL® equivalent values (integrated ‘total’ of 35 mL elution at 20 mL/min). The generator yield (82Rb) and production efficiency (82Rb yield / 82Sr parent activity) were measured and compared after this conversion to a common scale. Results. At the start of clinical use, the system reported 82Rb activity from daily QA was lower for CardioGen-82® vs RUBY-FILL® (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, p<0.001) despite having similar 82Sr activity. Dose-calibrator ‘maximum’ (CardioGen-82®) values were found to under-estimate the integrated ‘total’ (RUBY-FILL®) activity by ~24% at 50 mL/min. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82® efficiency remained slightly lower than the RUBY-FILL® system on average (88 ± 4% vs 95 ± 4%, p<0.001). The efficiency of 82Rb production improved for both systems over the respective periods of clinical use.Conclusions. 82Rb generator yield was significantly under-estimated using the CardioGen-82® vs RUBY-FILL® daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated 82Rb production efficiency of the CardioGen-82® system was ~7% lower than RUBY-FILL® over the full period of clinical use.
Keywords
myocardial perfusion imaging; positron emission tomography; Rubidium-82 generator; daily quality assurance
Subject
Medicine and Pharmacology, Other
Copyright: This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Comments (0)
We encourage comments and feedback from a broad range of readers. See criteria for comments and our Diversity statement.
Leave a public commentSend a private comment to the author(s)