Paired-View Sensitivity and Threshold Transfer in Frozen Vision Encoders for Fracture Detection

  † These authors contributed equally to this work.

Submitted:

23 November 2026

Posted:

25 November 2026

You are already at the latest version

Abstract
Background: A second view can change which fractures a model detects. We quantified this change and the transfer of a GRAZ threshold across cohorts. Methods: We trained frozen-encoder classifiers on 3,917 GRAZPEDWRI-DX examinations (7,834 radiographs). At GRAZ-derived thresholds we compared paired- and single-view sensitivity in 1,052 fracture-positive PediURF cases (DINOv2-504, SKELEX-224), and scored DINOv2-224 and SKELEX-224 on 491 MFXR patient-ID groups, 85 OrthoFrac-XR images, and 425 BoneFract Wrist images. Results: Adding the lateral (LAT) view to anteroposterior (AP) input raised sensitivity by 2.66 percentage points (pp; 95% CI +1.62 to +3.80) for DINOv2-504 and 0.38 pp (−0.29 to +1.05) for SKELEX-224; adding AP to LAT raised it by 1.62 pp (+0.86 to +2.47) and 4.37 pp (+3.14 to +5.70). Paired input missed five DINOv2-504 and four SKELEX-224 cases found by single views. In MFXR, the area under the curve (AUROC) was 0.599, 0.803, and 0.754 for DINOv2-224, SKELEX-224 and their average; the average reached 0.705 in OrthoFrac-XR and 0.711 in BoneFract, with specificity 2.9% and 6.5% at the GRAZ threshold (sensitivity 100% and 98.0%). Conclusions: The change depended on the model and starting projection. The fixed average showed moderate discrimination but classified most controls as positive at the GRAZ threshold.
Keywords: 
;  ;  ;  ;  ;  

1. Introduction

Frontal and lateral radiographs provide complementary information about a fracture. A classifier given both views can miss a fracture it detects on one view alone. Previous pediatric studies compared single- and dual-view models [1,2,3]. At a source-derived threshold, we counted detections gained and lost when a second view was added to the same fracture-positive cases.
External testing asks whether scores rank fractures above controls and how a fixed threshold performs in a new cohort. A 2026 meta-analysis found high pooled diagnostic performance for pediatric fracture artificial intelligence (AI), with substantial variation across studies [4]. Performance also changed with test-case difficulty within GRAZPEDWRI-DX [5]. We used that dataset’s annotated wrist examinations for development [6] and fitted the same head architecture to features from each pretrained encoder, DINOv2 and SKELEX [7,8]. We then assessed paired-view sensitivity, external discrimination, threshold transfer, and how spatial feature replacement affected model scores.
We used the PediURF (pediatric ulna and radius fracture) dataset to compare paired and single views [3] and the MFXR (Monteggia fracture) dataset to assess patient-level discrimination of Monteggia fractures from mixed controls in a cross-anatomic task [9]. In OrthoFrac-XR and BoneFract, we assessed image-level discrimination and performance at the GRAZ-derived threshold [10,11].

2. Materials and Methods

2.1. Study Design and Development Cohort

This retrospective study used publicly released, de-identified radiographs. From GRAZPEDWRI-DX v2 (hereafter GRAZ), we selected one frontal and one lateral view of the same wrist per patient. The final cohort contained 3,917 patients and 7,834 radiographs. Source labels identified 1,928 positive and 1,989 negative examinations. We excluded 278 candidates with unresolved perceptual hash matches and one positive examination whose source tag could not be linked to either selected view of the same side.
For model input, we use anteroposterior (AP) as the label for the frontal view and lateral (LAT) for the side view. The GRAZ descriptor calls its typical frontal projection posteroanterior. PediURF specifies anteroposterior [3,6]. Within a positive GRAZ examination, we treated a view without a fracture box in the Extensible Markup Language (XML) annotations as unannotated.

2.2. Frozen Classifiers

We initialized the vision transformers with the official DINOv2 ViT-B/14 and public SKELEX ViT-L/16 masked-autoencoder weights [7,8]. For the held-out GRAZ, spatial, MFXR, OrthoFrac-XR and BoneFract analyses, both encoders received a 224 × 224 image resized from the normalized 504 × 504 canvas. Each view was represented by a classification token and 16 spatial tokens obtained by pooling the patch map to a 4 × 4 grid. We refer to these configurations as DINOv2-224 and SKELEX-224. PediURF used SKELEX-224 and an earlier DINOv2 configuration, DINOv2-504. The latter used the classification token of the 504 × 504 full image and the classification tokens of sixteen 252 × 252 tiles from a 1008 × 1008 canvas. It used its own fitted heads and GRAZ calibration parameters. The classification head projected tokens to 256 dimensions, added view, position, coordinate and padding features, and applied two four-head within-view transformer blocks (feed-forward width 1,024 and dropout 0.1). It mean-pooled tokens within each view, averaged available views (F0), and passed the pooled representation and view-availability indicator to a binary output head. Encoder weights remained frozen.
We evaluated the S0-F0 configuration: examination-level binary supervision without auxiliary view labels, and independent within-view encoding followed by mean pooling. During training, each example used AP-only, LAT-only, or paired input with probabilities 0.25, 0.25, and 0.50, respectively. We trained the head with binary cross-entropy and a positive-class weight calculated from the training partition and clipped to 0.5–5. Optimization used AdamW (learning rate 3 × 10−4; weight decay 10−4), batches of 32, gradient-norm clipping at 1.0, and at most 60 epochs. The learning rate was warmed linearly over five epochs and then cosine-decayed to 10−6. We retained the checkpoint with the highest area under the receiver operating characteristic curve (AUROC) on the selection partition and stopped after ten epochs without an improvement greater than 0.0001, once at least ten epochs had run.
We used five outer patient folds and three training seeds (2026–2028), yielding 15 fits per feature configuration. Within each fold, patients belonged to one of four disjoint partitions for training, selection, calibration, and testing. The five test folds covered all 3,917 patients once. Across folds, the partitions included 2,140–2,161 training, 384–404 selection, 581–591 calibration, and 778–791 test patients. For each fit, we estimated a scalar temperature from paired-view logits in its GRAZ calibration partition. We scored each external case with the encoder and heads held fixed, then averaged the 15 temperature-scaled probabilities for the applicable configuration. Figure 1 shows the workflow and cohorts.

2.3. External Cohorts and Endpoints

PediURF. We linked source test entries to their paired frontal and lateral image folders and verified that both images were readable. One entry lacked an exact match, leaving 1,052 fracture-positive cases. Fractures were distal in 676 cases, midshaft in 265, and proximal in 111 [3]. Because the cohort contained no negative cases, AUROC and specificity were not estimable; we report sensitivity comparisons only. For each fitted model, a class-conditional conformal risk-control rule selected the largest threshold t satisfying 1 + c n + 1 ≤ 0.10 [12], where c is the number of positive calibration scores below t and n is the total number of positive calibration scores. We used the median of the 15 fit-specific GRAZ thresholds as one operating point for the 15-fit mean probability: 0.35279235 for DINOv2-504 and 0.31790158 for SKELEX-224. The 10% target applies to each fit’s calibration procedure under exchangeability; the median-threshold ensemble was not separately calibrated and carries no formal risk guarantee on the external data.
We subtracted single-view sensitivity from paired-view sensitivity and counted cases whose detection changed in either direction. Across all three inputs, we counted the eight possible combinations of decisions. We plotted score distributions and summarized detection by fracture location. For the joint score maps and case plots, we used 100 × (score minus the encoder threshold). The joint maps used bins 2.5 score points wide. In a supplementary analysis, we moved the threshold from −0.15 to +0.15 relative to its frozen value and applied each offset to all inputs.
We scored each MFXR, OrthoFrac-XR, and BoneFract image in two single-view passes, once in the frontal input position and once in the lateral position. For each fit, we averaged the two temperature-scaled probabilities. We then averaged the scores across 15 fits per encoder.
MFXR. The official patient-held-out test partition contained 491 patient IDs: 321 source-labeled Monteggia cases and 170 controls with normal findings or other fractures [9]. For each ID, we averaged calibrated scores across all released rows, which may include original radiographs, different views, and regional crop derivatives. MFXR JPEG/RGBA images were converted with Pillow convert(“L”), expanded from 8-bit grayscale to the 16-bit range, and passed through the unchanged GRAZ preprocessing. We calculated AUROC and average precision by patient ID for each encoder and their fixed 1:1 probability average, with image-row point estimates as a secondary description. Because this endpoint differs from wrist-fracture detection, we did not transfer the GRAZ threshold.
OrthoFrac-XR and BoneFract. OrthoFrac-XR images were selected from release metadata before scoring: age under 18 years, bone type radius or ulna, and either a distal-fracture label whose observation text identified the distal radius/ulna end or a non-fracture label. Proximal-fracture and post-fracture records were excluded. The subset contained 85 images (50 distal fractures and 35 non-fractures); patient, view, and site identifiers were unavailable. BoneFract’s official Wrist test contained 602 rows (193 positive, 409 negative). We collapsed exact within-test duplicates (602 rows to 599 unique image hashes) and excluded every test image with an exact SHA-256 match in the training or validation partitions, leaving 425 images (149 positive, 276 negative). Among the overlapping test images, 36 exact-hash groups had a source-label conflict with training or validation; these test entries were excluded. The local BoneFract release did not provide age, projection, or a verified patient/examination key. Source image formats were converted to grayscale and passed through the shared GRAZ preprocessing; 8-bit BoneFract values were expanded to the 16-bit range before normalization. We evaluated images individually. For each fold and seed, we averaged the two models’ calibrated probabilities for the same GRAZ-positive calibration examinations. We took the empirical 10th percentile using the lower order statistic and used the median across the 15 fits as the threshold for the fixed 1:1 average (0.32759076). Each individual encoder used the median of its 15 GRAZ thresholds. We calculated AUROC from continuous scores and sensitivity and specificity at the fixed operating points. No external labels were used to select a threshold or model.
The cohort sizes, analysis units, and endpoints are summarized in Table 1.

2.4. Spatial Feature Replacement in GRAZ

We included 1,928 positive examinations with paired images and at least one source fracture box. After mapping each Pascal VOC box through the original padding and resize, we identified the spatial tokens whose contributing patches overlapped it. With SKELEX adaptive pooling, adjacent tokens can share patches. We replaced each selected token with the nearest available non-target vector from the same view.
The control replaced an equal number of tokens per view outside the box-aligned cells. Donor vectors came from outside both the control mask and the box-aligned cells. We averaged control scores across up to 20 masks and kept the global feature, view identity, and coordinates fixed.
For each case, we compared the score drop after box-aligned replacement with the drop under the matched control. We calculated AUROC across all 3,917 cases after each intervention, with feature replacement in positive cases and the original scores in negative cases. For the illustration, we chose the DINOv2 case with boxes in both views whose difference in score drop was closest to the median of the positive cohort.

2.5. Statistical Analysis and Reporting

For GRAZ, we averaged the three calibrated paired-input scores from each patient’s held-out fold and pooled patient averages for ROC and precision–recall curves. We assessed calibration with ten equal-frequency reliability bins, Brier score, log loss, a calibration-in-the-large intercept with slope fixed at 1, and a joint logistic recalibration intercept and slope. We calculated 2,000 patient bootstrap intervals stratified by fold and outcome for these calibration summaries. For OrthoFrac-XR and BoneFract, AUROC bootstrap intervals were stratified by source label because patient identifiers were unavailable; they do not account for possible within-patient dependence. We calculated 95% Wilson score intervals for sensitivity and specificity at fixed external thresholds from the image-level counts.
We used 5,000 paired patient bootstrap resamples for 95% confidence intervals (CIs) on PediURF sensitivity differences. We applied Holm’s correction to the four exact McNemar tests. For MFXR, we resampled patient IDs within each outcome class. For image-level AUROC, we resampled images. The feature replacement analysis used 5,000 patient resamples stratified by fold and, where applicable, by outcome. The intervals are conditional on the fitted models and evaluated cohorts. Each external estimate used data from its own source. We used relevant CLAIM 2024 and STARD-AI items as reporting guides where applicable [13,14]. The reproducibility environment was Python 3.13.17 with PyTorch 2.11.0+cu128, Transformers 5.22.2, NumPy 2.5.4, SciPy 1.18.1, Pillow 13.0.0, Pandas 3.1.1, Matplotlib 3.11.2, and scikit-learn 1.9.1. Exact source files, model configurations, and external replay commands are listed in Supplementary Information.

2.6. Ethics and Data Governance

This secondary analysis used only publicly released, de-identified radiographs, metadata, and source labels. No images, reports, or records from the authors’ hospital were used; no participants were recruited or contacted, no intervention was performed, and no identifiable information was accessed. Available source documentation reports the following ethics and consent information for the original datasets: GRAZPEDWRI-DX was approved by the Medical University of Graz (EK 31-108 ex 18/19), with consent waived [6]; PediURF reports approval by Shenzhen Children’s Hospital (202505902) and parent/guardian consent for participation and data sharing [3]; MFXR reports approval by the same hospital committee, with consent waived [9]; and OrthoFrac-XR reports approval by Daffodil International University (REC-FSIT-2025/No. 12639), while its release documentation reviewed here does not specify consent procedures [10]. The BoneFract repository reports de-identification and names Peoples University of Medical & Health Sciences for Women as the source institution, but provides no associated publication or explicit ethics/consent statement [11]. These source-level statements concern original data collection and do not determine the ethics classification of this secondary analysis. The institutional determination for this secondary analysis is reported in the Institutional Review Board Statement.

2.7. Use of Generative Artificial Intelligence

During the preparation of this work, the authors used DeepSeek V4.1FLASH to assist with editing of the analysis code and with language polishing of author-written text. Instructions given to the tool were limited to modifying existing analysis scripts and to editing existing author-written sentences. The tool was not used to generate research data, to design the analyses, or to produce the reported results. Every code change was executed, inspected and re-run by the authors, and each reported value was recomputed from the released code and the archived outputs. The authors reviewed and edited all text and take full responsibility for the content of the publication.

3. Results

3.1. GRAZ Held-Out Performance

Across 3,917 held-out patient examinations, pooled AUROC was 0.885 For DINOv2 and 0.908 for SKELEX. Average precision was 0.905 and 0.922, respectively. Calibration summaries for the fixed average showed a Brier score of 0.120 (95% bootstrap CI, 0.115 to 0.125) and log loss of 0.384 (0.373 to 0.395). Its calibration-in-the-large intercept was −0.254 (−0.299 to −0.211), with slope fixed at 1; the jointly estimated recalibration slope was 1.594 (1.502 to 1.695) (Supplementary Table S6). In the reliability plot, fractures occurred less often than predicted in several low and intermediate score bins (Figure 2).

3.2. Paired-View Sensitivity in PediURF

DINOv2-504 detected 1,037 of 1,052 cases with paired input (98.6%), compared with 1,009 with AP only (95.9%) and 1,020 with LAT only (97.0%). Paired input increased sensitivity by 2.66 pp versus AP only (95% CI, +1.62 to +3.80) and by 1.62 pp versus LAT only (+0.86 to +2.47).
SKELEX-224 detected 1,046 cases with paired input (99.4%), 1,042 with AP only (99.0%), and 1,000 with LAT only (95.1%). Paired minus single-view sensitivity was 0.38 pp versus AP only (95% CI, −0.29 to +1.05) and 4.37 pp versus LAT only (+3.14 to +5.70). The largest net gains came from adding LAT for DINOv2-504 and adding AP for SKELEX-224 (Figure 3, Table 2).
For DINOv2-504, adding LAT to AP recovered 31 detections and lost 3 AP-only detections. The model missed 10 cases under every input condition, and paired input missed 5 cases detected by at least one single view. SKELEX-224 missed 2 cases under every input condition and missed 4 cases with paired input that AP alone detected. For each encoder, every paired-view detection was also made with at least one single view. Figure 4 shows the score changes, and Figure 5 illustrates three cases. Supplementary Figure S1, Figure S2 and Figure S4, and S5 report subgroup results, score distributions, joint decisions, and changes across threshold offsets. Supplementary Table S2 lists every paired-view difference, and Supplementary Table S3 gives the subgroup estimates.

3.3. Discrimination in the MFXR Monteggia Task

Across 491 patient IDs, AUROC for Monteggia fracture versus mixed controls was 0.599 for DINOv2-224 (95% CI, 0.547 to 0.649), 0.803 for SKELEX-224 (0.764 to 0.841), and 0.754 for the fixed model average (0.710 to 0.796). Average precision was 0.747, 0.896, and 0.863, respectively. AUROC at the image-row level was 0.541, 0.650, and 0.617 for the same models (Figure 6, Table 3; Supplementary Table S4). In paired patient-ID bootstrap comparisons, the fixed average differed from DINOv2 by +0.155 AUROC (95% CI, +0.123 to +0.189) and from SKELEX by −0.049 (−0.082 to −0.018) (Supplementary Table S5).

3.4. Threshold transfer in OrthoFrac-XR and BoneFract

The fixed average yielded an AUROC of 0.705 in OrthoFrac-XR distal-fracture versus non-fracture images (95% CI, 0.585 to 0.812) and 0.711 in the BoneFract Wrist subset (0.658 to 0.761). At the transferred threshold, sensitivity was 100.0% (50/50; 95% Wilson CI, 92.9 to 100.0%) and 98.0% (146/149; 94.2 to 99.3%), respectively. Specificity was 2.9% (1/35; 0.5 to 14.5%) and 6.5% (18/276; 4.2 to 10.1%), corresponding to 34 and 258 false positives (Figure 7, Table 4; Supplementary Table S1). Across all three classifiers, false-positive fractions ranged from 91.4% to 97.1% in OrthoFrac-XR and from 89.1% to 93.5% in BoneFract (Supplementary Figure S6). Supplementary Figure S3 documents the BoneFract test-split overlap screen.

3.5. Spatial Feature Replacement in GRAZ

Replacing box-aligned features reduced scores more than matched non-box replacement. The mean excess reduction was 0.302 for DINOv2 (95% CI, 0.291 to 0.313) and 0.471 for SKELEX (0.458 to 0.484). For DINOv2, AUROC was 0.885 before replacement, 0.891 after control replacement, and 0.593 after box-aligned replacement in positive cases. The corresponding SKELEX values were 0.908, 0.914, and 0.537 (Figure 8).

4. Discussion

In PediURF, adding LAT to AP with DINOv2-504 recovered 31 cases and lost 3 detections made on AP alone. With SKELEX-224, adding AP to LAT recovered 46 cases and retained every LAT detection. Reporting gains and losses alongside the net change in sensitivity distinguishes these patterns [1,2,3].
In MFXR, the fixed average reached a patient-level AUROC of 0.754, below the 0.803 for SKELEX alone. In OrthoFrac-XR and BoneFract, discrimination was moderate (AUROC 0.705 and 0.711). At the GRAZ threshold, it classified 34 of 35 OrthoFrac-XR controls and 258 of 276 BoneFract controls as positive. Because radiograph-model performance can change across populations [15], external validation should report the false-positive burden at the intended operating point alongside AUROC.
The fixed average showed miscalibration on held-out GRAZ cases. Its calibration-in-the-large intercept was −0.254, and the joint recalibration slope was 1.594. The contribution of calibration to the low specificity after threshold transfer remains uncertain. Between-study heterogeneity and changes in performance with case difficulty within GRAZ support reporting the endpoints and case mix of each external test cohort [4,5].
In both encoders, replacing pooled cells that overlapped fracture boxes lowered scores more than replacing matched non-box cells. This score difference measures sensitivity to the perturbation. Assessing localization accuracy would require a spatial endpoint and an independent reference [16].

4.1. Limitations

All classifiers used the S0-F0 head, and operating thresholds came from GRAZ. DINOv2 used different input sizes and token extraction in PediURF and the other cohorts, so the paired-view results apply to the model and its input pipeline together. The datasets also differ in endpoint and analysis unit. PediURF contains only positive cases. MFXR aggregates released image and crop rows by source-defined patient ID. OrthoFrac-XR and BoneFract contribute selected images without verified patient linkage. OrthoFrac-XR included 85 images, and we used source labels without independent adjudication. These differences limit direct comparisons.
We did not assess calibration in the external cohorts. Case mix, label definitions, acquisition, and preprocessing may have contributed to low specificity. The external releases had incomplete patient, projection, and site metadata. Exact-hash screening in BoneFract could miss transformed copies and near-duplicates from the same patient. We could not verify pixel-for-pixel agreement with the original GRAZ preprocessing. Bootstrap intervals reflect uncertainty conditional on the cases and fitted models. The spatial intervention perturbs pooled features, which limits its use for judging localization. We did not measure reader performance or patient outcomes.

4.2. Future Research

Estimating paired-view specificity requires fracture-positive and fracture-negative examinations with patient-linked projections. Age, projection, and site metadata would help describe case mix. Weights, endpoints, preprocessing, and thresholds should be set before testing. Fit any recalibration on a separate subset and evaluate it on untouched patients, with site-level separation where feasible. Reporting discrimination, calibration, sensitivity, and specificity across cohorts with different case difficulty would show how threshold performance changes. Predictive values can be estimated where prevalence is known.
A multicenter, multireader, multicase crossover study with ethics oversight could test whether AI assistance changes clinical decisions in representative cases. Randomize the order of unaided and AI-assisted readings, and specify the clinical action and washout period in advance. Record sensitivity, specificity, false-positive burden, reading time, confidence, and errors after AI suggestions. Pediatric studies have documented gains in reader accuracy and automation-related errors. Multivendor evaluations found differences by software and fracture type [17,18,19]. Decision-curve analysis could estimate the net benefit of acting at the tested threshold [20].
Compare view-aware fusion with mean pooling in independent cohorts that include several fracture locations. For the spatial analysis, repeat feature replacement with finer spatial representations and distribution-matched perturbations. Where independent expert annotations are available, use them to assess localization.

5. Conclusions

A second view changed which fractures the models detected, with gains and losses that depended on the classifier and starting projection. The fixed average showed moderate discrimination in OrthoFrac-XR and BoneFract. At the GRAZ threshold, it detected most fractures and classified most controls as positive. Paired-view evaluations should report case-level changes alongside net sensitivity. External evaluations should report sensitivity and specificity at the intended operating point alongside discrimination.

Supplementary Materials

The following supporting information can be downloaded at the website of this paper posted on Preprints.org. Figure S1. Paired-view sensitivity increment by fracture location in PediURF; Figure S2. Model-score distributions among fracture-positive PediURF cases; Figure S3. BoneFract image provenance and sensitivity analysis; Figure S4. Overlap of missed cases across the three PediURF input conditions; Figure S5. Positive-case sensitivity differences across threshold offsets in PediURF; Figure S6. Frozen operating points on external image-level sets; Table S1. Image-level threshold transfer in OrthoFrac-XR and BoneFract; Table S2. Paired-view sensitivity differences in PediURF; Table S3. PediURF paired-view sensitivity differences by fracture location; Table S4. Patient-ID-aggregated AUROC in the MFXR Monteggia task; Table S5. Paired patient-ID AUROC contrasts for the MFXR fixed average; Table S6. Calibration metrics for held-out GRAZ patients; Supplementary File S1. Analysis scripts, protocols, calibration parameters, and runtime instructions; Table S7 and Table S8. Public source files and external reproduction checks (in Supplementary Information).

Author Contributions

Conceptualization, J.D. and L.W.; methodology, J.D., J.G. and L.W.; software, J.D. and Y.Y.; validation, Z.D. and P.Z.; formal analysis, J.D.; investigation, Z.D.; resources, J.G.; data curation, Z.D.; writing—original draft preparation, J.D.; writing—review and editing, Z.D., Y.Y., J.G., P.Z. and L.W.; visualization, Y.Y. and P.Z.; supervision, L.W.; project administration, L.W.; funding acquisition, L.W. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the Biological Agricultural Joint Fund of the Natural Science Foundation of Hebei Province (Grant No. C2023402063) and the Science Research Project of Hebei Education Department (Grant No. QN2022002).

Institutional Review Board Statement

Ethical review and approval were waived for this study, which was conducted in accordance with the Declaration of Helsinki. All analyses used publicly released, de-identified datasets (GRAZPEDWRI-DX v2, PediURF v2, MFXR, OrthoFrac-XR v2 and BoneFract v1), each distributed under a CC BY 4.0 license; no new data were collected and no human participants were recruited.

Data Availability Statement

The public datasets analyzed in this study were GRAZPEDWRI-DX v2 [21], PediURF v2 [22], MFXR [23], OrthoFrac-XR v2 [10], and BoneFract v1 [11]. Supplementary File S1 provides the code, software requirements, model revisions, input checksums, fit-level calibration and source audits. The 45 fitted classification heads required for frozen inference in the four external cohorts are available from the corresponding author on reasonable request. The authors retain the original GRAZ cohort identifiers, fold assignments, feature caches and individual predictions. Detailed access instructions and the scope of reproduction are given in Supplementary Information. The code archive is deposited in a Zenodo record (DOI 10.5281/zenodo.24442840), which is available to reviewers through a private preview link and becomes publicly accessible upon acceptance.

Acknowledgments

The authors thank the providers of the public datasets (GRAZPEDWRI-DX, PediURF, MFXR, OrthoFrac-XR and BoneFract) for releasing their data. The use of generative artificial intelligence is disclosed in Section 2.7.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Janisch, M.; Apfaltrer, G.; Hržić, F.; Castellani, C.; Mittl, B.; Singer, G.; Lindbichler, F.; Pilhatsch, A.; Sorantin, E.; Tschauner, S. Pediatric radius torus fractures in x-rays—how computer vision could render lateral projections obsolete. Front. Pediatr. 2022, 10, 1005099. [Google Scholar] [CrossRef] [PubMed]
  2. Suzuki, H.; Teramoto, A.; Honmoto, T.; Niki, A.; Kono, T.; Fujita, H. Automated detection of pediatric forearm fractures in X-ray images using deep learning. Radiol. Phys. Technol. 2026, 19, 830–841. [Google Scholar] [CrossRef] [PubMed]
  3. Tang, S.; Ou, L.; Li, W.; Xiong, Z.; Li, N.; Liu, H.; Liang, Y.; Zhao, Z. A Comprehensive X-ray Dataset for Pediatric Ulna and Radius Fractures Analysis. Sci. Data 2026, 13, 308. [Google Scholar] [CrossRef] [PubMed]
  4. Pahlevan-Fallahy, M.-T.; Salmanian, E.; Asgari, A.-M.; Kian, N.; Malek, M.; Shaker, F. Evaluating the diagnostic performance of artificial intelligence for pediatric appendicular fractures: a systematic review and meta-analysis. Pediatr. Radiol. advance online publication. 2026. [Google Scholar] [CrossRef] [PubMed]
  5. Till, T.; Scherkl, M.; Stranger, N.; Singer, G.; Hankel, S.; Flucher, C.; Hržić, F.; Štajduhar, I.; Tschauner, S. Impact of test set composition on AI performance in pediatric wrist fracture detection in X-rays. Eur. Radiol. 2025, 35, 6853–6864. [Google Scholar] [CrossRef] [PubMed]
  6. Nagy, E.; Janisch, M.; Hržić, F.; Sorantin, E.; Tschauner, S. A pediatric wrist trauma X-ray dataset (GRAZPEDWRI-DX) for machine learning. Sci. Data 2022, 9, 222. [Google Scholar] [CrossRef] [PubMed]
  7. Oquab, M.; Darcet, T.; Moutakanni, T.; Vo, H.V.; Szafraniec, M.; Khalidov, V.; Fernandez, P.; Haziza, D.; Massa, F.; El-Nouby, A.; et al. DINOv2: learning robust visual features without supervision. Trans. Mach. Learn. Res. 2024. Available online: https://openreview.net/forum?id=a68SUt6zFt.
  8. Kim, S.; Lee, S.; Shin, K.; Kim, H.-S.; Kim, Y.; Kim, M.; Nam, J.; Ko, S.; Kwon, D.; Huh, W.; et al. A large-scale vision foundation model for musculoskeletal radiographs. npj Digit. Med. 2026, 9, 668. [Google Scholar] [CrossRef] [PubMed]
  9. Li, N.; Xiong, Z.; Tang, S.; Zhao, Z.; Xue, S.; Jiang, C.; Liu, A.; Liang, Y. An X-ray Dataset and Benchmark for AI-Based Diagnosis of Monteggia Fractures. Sci. Data 2026, 13, 1494. [Google Scholar] [CrossRef] [PubMed]
  10. Tabib, M.A.; Liza, S.S.; Bijoy, M.H.I.; Hasan, M.Z.; Khan, M.R.I. OrthoFrac-XR: A Clinically Validated Multimodal High-Resolution X-ray Imaging Dataset for Bone Fracture Detection and Localization. Figshare 2026, version 2. [Google Scholar] [CrossRef]
  11. Noor, R.; Ali, A. BoneFract: A Bone Fracture Dataset. Mendeley Data 2026, version 1. [Google Scholar] [CrossRef]
  12. Angelopoulos, A.N.; Bates, S.; Fisch, A.; Lei, L.; Schuster, T. Conformal Risk Control. In Proceedings of the International Conference on Learning Representations, 2024; Available online: https://arxiv.org/abs/2208.02814.
  13. Tejani, A.S.; Klontzas, M.E.; Gatti, A.A.; Mongan, J.T.; Moy, L.; Park, S.H.; Kahn, C.E., Jr. CLAIM 2024 Update Panel. Checklist for Artificial Intelligence in Medical Imaging (CLAIM): 2024 Update. Radiol. Artif. Intell. 2024, 6, e240300. [Google Scholar] [CrossRef] [PubMed]
  14. Sounderajah, V.; Guni, A.; Liu, X.; Collins, G.S.; Karthikesalingam, A.; Markar, S.R.; Golub, R.M.; Denniston, A.K.; Shetty, S.; Moher, D.; et al. The STARD-AI reporting guideline for diagnostic accuracy studies using artificial intelligence. Nat. Med. 2025, 31, 3283–3289. [Google Scholar] [CrossRef] [PubMed]
  15. Zech, J.R.; Badgeley, M.A.; Liu, M.; Costa, A.B.; Titano, J.J.; Oermann, E.K. Variable generalization performance of a deep learning model to detect pneumonia in chest radiographs: A cross-sectional study. PLoS Med. 2018, 15, e1002683. [Google Scholar] [CrossRef] [PubMed]
  16. Hooker, S.; Erhan, D.; Kindermans, P.-J.; Kim, B. A Benchmark for Interpretability Methods in Deep Neural Networks. Adv. Neural Inf. Process. Syst. 2019, 32. Available online: https://arxiv.org/abs/1806.10758.
  17. Zech, J.R.; Carotenuto, G.; Igbinoba, Z.; Tran, C.V.; Insley, E.; Baccarella, A.; Wong, T.T. Detecting pediatric wrist fractures using deep-learning-based object detection. Pediatr. Radiol. 2023, 53, 1125–1134. [Google Scholar] [CrossRef] [PubMed]
  18. Ziegner, M.; Pape, J.; Lacher, M.; Brandau, A.; Kelety, T.; Mayer, S.; Hirsch, F.W.; Rosolowski, M.; Gräfe, D. Real-life benefit of artificial intelligence-based fracture detection in a pediatric emergency department. Eur. Radiol. 2025, 35, 5881–5890. [Google Scholar] [CrossRef] [PubMed]
  19. Altmann-Schneider, I.; Geiger, L.S.; Kellenberger, C.J.; Callaghan, F.; Seiler, M. Multivendor comparison study of artificial intelligence software for automated fracture detection in paediatric patients. Pediatr. Radiol. 2026, 56, 2067–2077. [Google Scholar] [CrossRef] [PubMed]
  20. Vickers, A.J.; Elkin, E.B. Decision curve analysis: a novel method for evaluating prediction models. Med. Decis. Mak. 2006, 26, 565–574. [Google Scholar] [CrossRef] [PubMed]
  21. Nagy, E.; Janisch, M.; Hržić, F.; Sorantin, E.; Tschauner, S. GRAZPEDWRI-DX. Figshare, version 2. 2022. [Google Scholar] [CrossRef]
  22. Tang, S.; Ou, L.; Li, W.; Xiong, Z.; Li, N.; Liu, H.; Liang, Y.; Zhao, Z. AComprehensive X-ray Dataset for the Pediatric Ulna and Radius Fractures Analysis. Figshare 2025, version 2. [Google Scholar] [CrossRef]
  23. Li, N.; Xiong, Z.; Tang, S.; Zhao, Z.; Xue, S.; Jiang, C.; Liu, A.; Liang, Y. An X-ray Dataset for AI-Based Diagnosis of Monteggia Fractures. Figshare 2026. [Google Scholar] [CrossRef]
Figure 1. Study design and evaluation endpoints. (A) An illustrative PediURF anteroposterior (AP) and lateral (LAT) radiograph pair. (B) GRAZ model development, with patient-level partitions for training, selection, calibration, and testing. (C) External scoring with frozen weights and thresholds. (D) Analysis units and endpoints across external datasets. Radiographs from Tang et al., PediURF v2, CC BY 4.0, resized and padded for display.
Figure 1. Study design and evaluation endpoints. (A) An illustrative PediURF anteroposterior (AP) and lateral (LAT) radiograph pair. (B) GRAZ model development, with patient-level partitions for training, selection, calibration, and testing. (C) External scoring with frozen weights and thresholds. (D) Analysis units and endpoints across external datasets. Radiographs from Tang et al., PediURF v2, CC BY 4.0, resized and padded for display.
Preprints 245968 g001
Figure 2. Discrimination and calibration for held-out GRAZ patients. (A) Receiver operating characteristic (ROC) curves and the area under the ROC curve (AUROC). (B) Precision–recall curves and average precision. The dashed line in B shows the source-cohort positive fraction. (C) Mean predicted probability and observed positive fraction within ten equal-frequency bins per model, with patient-bootstrap 95% intervals on the observed fractions. Curves pool predictions from 3,917 held-out patients after averaging their three seed-specific paired-input scores within each outer fold.
Figure 2. Discrimination and calibration for held-out GRAZ patients. (A) Receiver operating characteristic (ROC) curves and the area under the ROC curve (AUROC). (B) Precision–recall curves and average precision. The dashed line in B shows the source-cohort positive fraction. (C) Mean predicted probability and observed positive fraction within ten equal-frequency bins per model, with patient-bootstrap 95% intervals on the observed fractions. Curves pool predictions from 3,917 held-out patients after averaging their three seed-specific paired-input scores within each outer fold.
Preprints 245968 g002
Figure 3. Case-level detection transitions with paired views in PediURF. (A–D) Within-patient transition matrices for each encoder and single-view comparator. Rows show the single-view decision and columns show the paired decision. Within each panel’s comparison, teal marks detections unique to paired input and rose marks detections unique to the single-view input. Each panel contains the same 1,052 fracture-positive cases. (E) Paired minus single-view sensitivity differences with 95% paired patient bootstrap intervals. DINOv2-504 and SKELEX-224 each use their GRAZ threshold across all input conditions.
Figure 3. Case-level detection transitions with paired views in PediURF. (A–D) Within-patient transition matrices for each encoder and single-view comparator. Rows show the single-view decision and columns show the paired decision. Within each panel’s comparison, teal marks detections unique to paired input and rose marks detections unique to the single-view input. Each panel contains the same 1,052 fracture-positive cases. (E) Paired minus single-view sensitivity differences with 95% paired patient bootstrap intervals. DINOv2-504 and SKELEX-224 each use their GRAZ threshold across all input conditions.
Preprints 245968 g003
Figure 4. Joint single-view and paired-view score margins in PediURF. Panels A and B show DINOv2-504 for frontal and lateral comparators, respectively. Panels C and D show the same comparisons for SKELEX-224. Each panel contains 1,052 observations. The color scale shows the number of cases in each two-dimensional bin. Threshold-centered margins on both axes are 100 × (score minus that model’s frozen threshold). Markers identify cases detected by only one input condition. The upper-left quadrant contains paired-only detections and the lower-right quadrant contains single-view-only detections.
Figure 4. Joint single-view and paired-view score margins in PediURF. Panels A and B show DINOv2-504 for frontal and lateral comparators, respectively. Panels C and D show the same comparisons for SKELEX-224. Each panel contains 1,052 observations. The color scale shows the number of cases in each two-dimensional bin. Threshold-centered margins on both axes are 100 × (score minus that model’s frozen threshold). Markers identify cases detected by only one input condition. The upper-left quadrant contains paired-only detections and the lower-right quadrant contains single-view-only detections.
Preprints 245968 g004
Figure 5. Paired radiographs and model decisions in three PediURF cases. Each row shows the patient’s full frontal and lateral radiographs beside the model scores. The score axis gives 100 × (score minus the encoder threshold). (A) SKELEX-224 detected the case with paired input after missing it with lateral-only input. (B) DINOv2-504 missed the case with paired input after detecting it with frontal-only input. (C) DINOv2-504 missed the case in all three conditions. The complete test set included 46, 3, and 10 cases meeting these criteria, respectively. We selected the first eligible case for each pattern in the archived order. All three cases had distal fracture labels. Filled markers indicate detection and open markers indicate a miss. The dashed zero line marks each encoder’s threshold. Labels beside the points show the scores that define each illustrated decision pattern. Radiographs from Tang et al., PediURF v2, CC BY 4.0, resized and padded for display.
Figure 5. Paired radiographs and model decisions in three PediURF cases. Each row shows the patient’s full frontal and lateral radiographs beside the model scores. The score axis gives 100 × (score minus the encoder threshold). (A) SKELEX-224 detected the case with paired input after missing it with lateral-only input. (B) DINOv2-504 missed the case with paired input after detecting it with frontal-only input. (C) DINOv2-504 missed the case in all three conditions. The complete test set included 46, 3, and 10 cases meeting these criteria, respectively. We selected the first eligible case for each pattern in the archived order. All three cases had distal fracture labels. Filled markers indicate detection and open markers indicate a miss. The dashed zero line marks each encoder’s threshold. Labels beside the points show the scores that define each illustrated decision pattern. Radiographs from Tang et al., PediURF v2, CC BY 4.0, resized and padded for display.
Preprints 245968 g005
Figure 6. Cross-anatomic ranking on the MFXR Monteggia challenge. (A) Class composition of the image rows and patient IDs. (B) AUROC. (C) Average precision. Circles show patient-ID-level estimates with class-stratified patient bootstrap 95% intervals, and open diamonds show image-row point estimates. Dashed lines represent chance ranking in B and the patient-ID Monteggia fraction in C. Patient scores average all released rows for the ID, which can include cropped derivatives. The mixed control group includes non-Monteggia fractures.
Figure 6. Cross-anatomic ranking on the MFXR Monteggia challenge. (A) Class composition of the image rows and patient IDs. (B) AUROC. (C) Average precision. Circles show patient-ID-level estimates with class-stratified patient bootstrap 95% intervals, and open diamonds show image-row point estimates. Dashed lines represent chance ranking in B and the patient-ID Monteggia fraction in C. Patient scores average all released rows for the ID, which can include cropped derivatives. The mixed control group includes non-Monteggia fractures.
Preprints 245968 g006
Figure 7. Classification outcomes and discrimination in the image-level external tests. (A, B) Confusion matrices for the fixed 1:1 average at the GRAZ-derived threshold of 0.32759076. Cells show counts and percentages within each source-label row. (C) AUROC with 95% image-bootstrap intervals.
Figure 7. Classification outcomes and discrimination in the image-level external tests. (A, B) Confusion matrices for the fixed 1:1 average at the GRAZ-derived threshold of 0.32759076. Cells show counts and percentages within each source-label row. (C) AUROC with 95% image-bootstrap intervals.
Preprints 245968 g007
Figure 8. GRAZ fracture-box feature replacement and matched controls. (A) The upper images show the source radiographs from one held-out DINOv2 case. The lower images show the corresponding source fracture boxes, the pooled cells overlapping them, and one of the matched control masks. The control score is the mean across valid control masks for that case. (B) Mean paired difference in score drop from replacing box-aligned cells versus same-count non-box cells among 1,928 source-positive patients. (C) Source-label AUROC among all 3,917 held-out patients. Each replacement was applied to the positive cases, with negative scores held fixed. Intervals are 95% patient bootstrap CIs. The overlay marks regions associated with replaced feature vectors. Radiographs and source boxes from Nagy et al., GRAZPEDWRI-DX v2, CC BY 4.0, display-normalized, resized, and overlaid.
Figure 8. GRAZ fracture-box feature replacement and matched controls. (A) The upper images show the source radiographs from one held-out DINOv2 case. The lower images show the corresponding source fracture boxes, the pooled cells overlapping them, and one of the matched control masks. The control score is the mean across valid control masks for that case. (B) Mean paired difference in score drop from replacing box-aligned cells versus same-count non-box cells among 1,928 source-positive patients. (C) Source-label AUROC among all 3,917 held-out patients. Each replacement was applied to the positive cases, with negative scores held fixed. Intervals are 95% patient bootstrap CIs. The overlay marks regions associated with replaced feature vectors. Radiographs and source boxes from Nagy et al., GRAZPEDWRI-DX v2, CC BY 4.0, display-normalized, resized, and overlaid.
Preprints 245968 g008
Table 1. Cohorts, analysis units, and endpoints.
Table 1. Cohorts, analysis units, and endpoints.
Cohort Analysis set Unit Endpoint
GRAZPEDWRI-DX 3,917 patients, 7,834 paired images (1,928 positive / 1,989 negative examinations) Patient examination Model development, calibration, and held-out evaluation
PediURF 1,052 fracture-positive cases with paired AP/LAT Patient case Paired minus single-view sensitivity
MFXR 491 IDs (321 Monteggia / 170 mixed controls) Patient ID, mean over released rows Monteggia versus mixed-control AUROC
OrthoFrac-XR 85 images (50 distal-fracture / 35 non-fracture) Image Distal radius/ulna fracture vs non-fracture AUROC and frozen-threshold sensitivity and specificity
BoneFract 425 images (149 positive / 276 negative) Image Wrist-fracture vs non-fracture AUROC and frozen-threshold sensitivity and specificity
AP, anteroposterior projection in PediURF. LAT, lateral projection. AUROC, area under the receiver operating characteristic curve. ID, identifier.
Table 2. Paired-view incremental detection in PediURF. The same encoder-specific threshold from GRAZ calibration was applied to paired and single-view inputs. CIs are patient-paired bootstrap intervals. All 1,052 cases were fracture-positive. AP, anteroposterior; LAT, lateral.
Table 2. Paired-view incremental detection in PediURF. The same encoder-specific threshold from GRAZ calibration was applied to paired and single-view inputs. CIs are patient-paired bootstrap intervals. All 1,052 cases were fracture-positive. AP, anteroposterior; LAT, lateral.
Encoder Contrast Paired sensitivity Single-view sensitivity Difference, percentage points (95% CI) Paired-only / single-only detections Holm-adjusted McNemar p
DINOv2-504 Paired vs AP-only 1,037/1,052 (98.6%) 1,009/1,052 (95.9%) +2.66 (+1.62 to +3.80) 31 / 3 2.30 × 10−6
DINOv2-504 Paired vs LAT-only 1,037/1,052 (98.6%) 1,020/1,052 (97.0%) +1.62 (+0.86 to +2.47) 19 / 2 4.43 × 10−4
SKELEX-224 Paired vs AP-only 1,046/1,052 (99.4%) 1,042/1,052 (99.0%) +0.38 (−0.29 to +1.05) 8 / 4 0.388
SKELEX-224 Paired vs LAT-only 1,046/1,052 (99.4%) 1,000/1,052 (95.1%) +4.37 (+3.14 to +5.70) 46 / 0 1.14 × 10−13
Table 3. MFXR patient-ID-aggregated AUROC. Scores are averages across all released rows per source-defined ID and may include original images and crop derivatives. CIs use patient-ID bootstrap resampling.
Table 3. MFXR patient-ID-aggregated AUROC. Scores are averages across all released rows per source-defined ID and may include original images and crop derivatives. CIs use patient-ID bootstrap resampling.
Frozen model Source IDs Monteggia / mixed control AUROC (95% CI)
DINOv2-224 491 321 / 170 0.599 (0.547 to 0.649)
SKELEX-224 491 321 / 170 0.803 (0.764 to 0.841)
Fixed 1:1 probability average 491 321 / 170 0.754 (0.710 to 0.796)
Table 4. Frozen-threshold image-level stress tests. Threshold 0.32759076 and fixed 1:1 DINOv2/SKELEX average were derived from GRAZ calibration only. CIs for AUROC use image bootstrap.
Table 4. Frozen-threshold image-level stress tests. Threshold 0.32759076 and fixed 1:1 DINOv2/SKELEX average were derived from GRAZ calibration only. CIs for AUROC use image bootstrap.
Source and subset Images (positive / control) AUROC (95% CI) Sensitivity (95% Wilson CI) Specificity (95% Wilson CI)
OrthoFrac-XR v2; age <18, radius/ulna; distal fracture vs non-fracture 85 (50 / 35) 0.705 (0.585 to 0.812) 50/50 (100.0%; 92.9 to 100.0%) 1/35 (2.9%; 0.5 to 14.5%)
BoneFract v1; clean Wrist test subset 425 (149 / 276) 0.711 (0.658 to 0.761) 146/149 (98.0%; 94.2 to 99.3%) 18/276 (6.5%; 4.2 to 10.1%)
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Copyright: This open access article is published under a Creative Commons CC BY 4.0 license, which permit the free download, distribution, and reuse, provided that the author and preprint are cited in any reuse.