1. Introduction
Cardiotocography (CTG) is widely used in clinical practice to continuously measure the fetal heart rate (FHR) and uterine contractions during delivery. In the Japanese guidelines, FHR patterns are classified into 82 types based on three basic patterns (baseline, variability, and deceleration) and their subdivisions. Combinations of FHR patterns are further classified into five levels to estimate the degree of risk for conditions such as fetal hypoxia and acidemia. However, because pattern classification is performed visually by obstetricians and gynecologists, discrepancies are a problem [
1,
2,
3]. The Trium CTG (Trium Analysis Online GmbH, Munich, Germany), which performs automatic pattern classifications with a computer, has been introduced recently to address this issue. The purpose of this study was to examine the reliability of assessments performed by obstetricians and the Trium CTG, as well as identify factors that contribute to differences in assessments.
2. Materials and Methods
Obstetricians at our hospital followed the FHR guidelines (
Table 1 and
Table 2) [
4], recommended by the Perinatal Committee of the Japanese Society of Obstetrics and Gynecology, to perform pattern classifications. The classification of FHR patterns refers to the baseline, variability, and deceleration of the FHR. The data analyzed were obtained from January 2011 to December 2017 and were read by CTG at the National Cerebral and Cardiovascular Center and Mie University Faculty of Medicine which have the same FHR monitors and were read by CTG. Informed consent was provided verbally and the data analyzed retrospectively. The study was approved by the Research Ethics Committees of the Graduate School of Applied Informatics, University of Hyogo and Aino University (approval number: Aino2012–010).
During the assessments, the clinical background of the patients was not provided, nor was there any opportunity to exchange this kind of information among the raters. We added “assessment not possible” for cases to which the guidelines did not apply. The following analyses were conducted to examine the reliability of the assessments and factors associated with differences: inter-rater reliability, Trium reliability, and intra-rater reliability.
The Kappa score was used to evaluate reliability. Kappa scores were categorized using Landis & Koch's assessment table (
Table 3) [
5].
2.1. Inter-Rater Reliability
Five raters from the same institution and five from different institutions, all of whom were obstetricians in leadership positions in Japan, performed FHR pattern classification on 188 CTG sections from 14 pregnant women. Data for which assessment was not possible were excluded and those with agreement of more than half of the raters (three) were included. The Kappa scores of raters from the same and different institutions, as well as Kappa scores by category, were analyzed.
2.2. Trium Reliability
Eleven obstetricians working at the same institution performed FHR pattern classification on 101 CTG sections from 12 pregnant women. Data for which assessment was not possible were excluded and those with agreement of more than half of the raters were included. The Kappa scores between raters and Trium, as well as Kappa scores by category, were analyzed.
2.3. Intra-Rater Reliability
One obstetrician performed a second assessment of the pattern classifications of the 101 sections from the 12 pregnant women after a month had passed since the first assessment. Agreement between the first and second assessments was analyzed.
3. Results
3.1. Inter-Rater Reliability
3.1.1. Kappa Scores of Raters from the Same and Different Institutions
Table 4 shows the Kappa scores of the five raters from the same institution and the five from different institutions. Both had a fair agreement on variability and level. The reference values in the table indicate the degree of agreement between the majority opinions of raters from the same and different institutions. Agreement between them was “substantial" for variability and level, as well as “almost perfect" for baseline and deceleration.
3.1.1. Kappa Score by Category
Table 5 shows the Kappa scores of each FHR category. Kappa scores tended to be higher for normal patterns in all categories and for raters from both the same and different institutions. Moderate agreement was seen with tachycardia, and a severe prolonged pattern showed substantial to almost perfect agreement, demonstrating high levels of agreement overall.
3.2. Trium Reliability
3.2.1. Kappa Score Between Raters, as Well as Between Trium and Raters
The majority of the 11 raters (six or more) agreed on the pattern classifications of the 101 sections in baseline (98 [97%]), variability (96 [95%]), deceleration (61 [60%]), and level (66 [65%]). Kappa scores between the 11 raters and Trium ranged from fair to moderate agreement for all categories. However, agreement with the reference values ranged from substantial to almost perfect in baseline and variability but ranged from moderate in deceleration and level (
Table 6).
3.2.2. Kappa Score by Category
Table 7 shows the Kappa scores of each category. Kappa scores tended to be relatively high for the normal results in all categories, as well as for both raters from the same institution and Trium. Kappa scores of Trium tended to be higher in variability and baseline, while those of the 11 raters from the same institution tended to be higher in deceleration and level.
3.2.3. Intra-Rater Reliability
Table 8 shows the agreement when one obstetrician performed a second reading sometime after a month reading. The agreement between the two FHR pattern classifications was very high at ≥90% in baseline and variability, in which the Kappa scores were between substantial and almost perfect agreement.
In contrast, the agreement rate in deceleration and level was approximately 60%, and the Kappa scores also showed moderate agreement.
4. Discussion
The reliability of CTG readings has been assessed using various methods. One example is to link the CTG results with actual birth outcomes (umbilical cord artery blood data) for analysis, assessment of CTG data by different readers, and checking the agreement rate.
The Japanese Society of Obstetrics and Gynecology has established five levels to estimate the degree of risk of conditions such as fetal hypoxia and acidemia. FHR pattern classification includes 82 categories, based on the results of baseline, variability, and deceleration. However, the number of categories means that it is difficult to perform assessments; therefore, a five-level classification is generally used. This is not seen as a problem because the medical treatment is the same. However, this also means that discrepancies may exist in level assessments. We analyzed inter-rater reliability, Trium reliability, and intra-rater reliability to examine the reliability of FHR pattern classification and factors contributing to discrepancies.
The results of Kappa scores within the same institution and between institutions showed fair agreement on both variability and level. However, the degree of agreement between readers was relatively good, and there were no differences between institutions. The factor contributing to differences in the interpretation between the automatic CTG assessment system (Trium) and obstetricians was identified as decelerations. The reason could be due to differences between the assessment of mild variable and mild late, as well as between severe variable and severe late, which caused the agreement on levels to decrease.
Intra-rater reliability for a single obstetrician is higher than the inter-rater agreement rate between obstetricians [
3]. Previous studies have shown that agreement rates among different raters are highly variable [
2], which was also the case in the present study. The reason for this was the assessment of deceleration, with the most common disagreements occurring between mild variable and severe variable, as well as between severe variable and severe late. However, we believed this was due, in part, to assessor factors.
In this study, we examined the factors of disagreement in CTG readings. Differences associated with the subjectivity of readers and interpretation of guidelines were some hypothesized factors, but the results of this study negate these hypotheses. Another factor was the quality of CTG, particularly with the automatic CTG assessment system (Trium), where unreadable data is not an option and all readings are classified into some category. Results are always classified into some category even when patterns consist of many indistinguishable complicated patterns in the 10 minutes. In this study as well, mild variable, mild late, severe variable, and severe late sometimes appeared as temporary changes, and these were likely important factors for disagreement between readings.
Kappa scores were used in this study, which are known to be lowered by data bias [
6]. Previous publications have shown that agreement rates decline with worse FHR patterns [
7,
8,
9], which, therefore, was avoided by utilizing CTG from cases with umbilical cord arterial blood pH <7.15 at birth in the current study. Nevertheless, the high percentage of normal waveforms suggested that some Kappa scores may have been lower than they would have normally been. Moreover, one of the limitations of the study was the discrepancy in the results of the automatic CTG assessment system (Trium) due to deceleration; it is suggested to improve this to enhance the agreement on the level.
5. Conclusions
This study found a high rate of agreement in FHR pattern readings between specialist obstetricians. Meanwhile, the assessment of deceleration by the same obstetrician showed only 60% agreement, highlighting the ongoing difficulty in interpreting decelerations.
Author Contributions
Conceptualization: T.I.; Data curation: T.I.; Formal analysis: H.K.; Investigation: N.M.; Methodology: H.K.; Project administration: T.I.; Resources: H.K.; Software: H.K.; Supervision: T.I.; Validation: N.M.; Visualization: H.K.; Writing – original draft: H.K.; Writing – review & editing: H.K.
Funding
This research received no external funding.
Institutional Review Board Statement
The study was approved by the Research Ethics Committees of the Graduate School of Applied Informatics, University of Hyogo and Aino University (approval number: Aino2012–010).
Informed Consent Statement
Informed consent was provided verbally and the data analyzed retrospectively.
Data Availability Statement
The data used to support the findings of this study are available from the corresponding author upon request.
Acknowledgments
We would like to thank the doctors of the Mie University Hospital Department of Obstetrics for their assistance in performing the CTG assessments.
Conflicts of Interest
The authors declare no conflicts of interest.
Abbreviations
The following abbreviations are used in this manuscript:
| FHR |
fetal heart rate |
| CTG |
Cardiotocography |
References
- Devane, D.; Lalor, J. Midwives’ visual interpretation of intrapartum cardiotocographs: Intra- and inter-observer agreement. J Adv Nurs 2005, 52, 133–141. [CrossRef]
- Blackwell, S.C.; Grobman, W.A.; Antoniewicz, L.; Hutchinson, M.; Gyamfi Bannerman, C. Interobserver and intraobserver reliability of the NICHD 3-Tier fetal heart rate Interpretation System. Am J Obstet Gynecol 2011, 205, 378.e1–378.e5. [CrossRef]
- Hernandez Engelhart, C.; Gundro Brurberg, K.; Aanstad, K.J.; Pay, A.S.D.; Kaasen, A.; Blix, E.; Vanbelle, S. Reliability and agreement in intrapartum fetal heart rate monitoring interpretation: A systematic review. Acta Obstet Gynecol Scand 2023, 102, 970–985. [CrossRef]
- Okai, T.; Ikeda, T.; Kawarabayashi, T.; Kozuma, S.; Sugawara, J.; Chisaka, H.; Yoneda, S.; Matsuoka, R.; Nakano, H.; Okamura, K.; et al. Intrapartum management guidelines based on fetal heart rate pattern classification. J Obstet Gynaecol Res 2010, 36, 925–928. [CrossRef]
- Landis, J.R.; Koch, G.G. The measurement of observer agreement for categorical data. Biometrics 1977, 33, 159–174. [CrossRef]
- Tsushima, E. Application of coefficient of reliability to the studies of physical therapy. Rigakuryoho Kagaku 2002, 17, 181–187. [CrossRef]
- Bernardes, J.; Costa-Pereira, A.; Ayres-de-Campos, D.; van Geijn, H.P.; Pereira-Leite, L. Evaluation of interobserver agreement of cardiotocograms. Int J Gynaecol Obstet 1997, 57, 33–37. [CrossRef]
- Ayres-de-Campos, D.; Bernardes, J.; Costa-Pereira, A.; Pereira-Leite, L. Inconsistencies in classification by experts of cardiotocograms and subsequent clinical decision. Br J Obstet Gynaecol 1999, 106, 1307–1310. [CrossRef]
- Epstein, A.J.; Twogood, S.; Lee, R.H.; Opper, N.; Beavis, A.; Miller, D.A. Interobserver reliability of fetal heart rate pattern interpretation using NICHD definitions. Am J Perinatol 2013, 30, 463–468. [CrossRef]
Table 1.
The 5-level classification of FHR patterns (82 categories) [
4].
Table 1.
The 5-level classification of FHR patterns (82 categories) [
4].
| Decelerations |
None |
Early |
Variable |
Late |
Prolonged |
| Mild |
Severe |
Mild |
Severe |
Mild |
Severe |
| Normal baseline variability |
Normocardia |
1 |
2 |
2 |
3 |
3 |
3 |
3 |
4 |
| Tachycardia |
2 |
2 |
3 |
3 |
3 |
4 |
3 |
4 |
| Mild bradycardia |
3 |
3 |
3 |
4 |
4 |
4 |
4 |
4 |
| Severe bradycardia |
4 |
4 |
|
4 |
4 |
4 |
|
|
Decreased baseline variability |
Normocardia |
2 |
3 |
3 |
4 |
3 |
4 |
4 |
5 |
| Tachycardia |
3 |
3 |
4 |
4 |
4 |
5 |
4 |
5 |
| Mild bradycardia |
4 |
4 |
4 |
5 |
5 |
5 |
5 |
5 |
| Severe bradycardia |
5 |
5 |
|
5 |
5 |
5 |
|
|
| Undetectable baseline variability |
4 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
| Marked baseline variability |
2 |
2 |
3 |
3 |
3 |
4 |
3 |
4 |
| Sinusoidal pattern |
4 |
4 |
4 |
4 |
5 |
5 |
5 |
5 |
Table 2.
Levels of FHR patterns [
4].
Table 2.
Levels of FHR patterns [
4].
| Level 1 |
Normal pattern |
| Level 2 |
Benign variant pattern |
| Level 3 |
Mild variant pattern |
| Level 4 |
Moderate variant pattern |
| Level 5 |
Severe variant pattern |
Table 3.
Kappa score (Landis & Koch) [
5].
Table 3.
Kappa score (Landis & Koch) [
5].
| <0.00 |
Poor agreement |
| 0.00–0.20 |
Slight agreement |
| 0.21–0.40 |
Fair agreement |
| 0.41–0.60 |
Moderate agreement |
| 0.61–0.80 |
Substantial agreement |
| 0.81–1.00 |
Almost perfect agreement |
Table 4.
Kappa scores of raters from the same and different institutions.
Table 4.
Kappa scores of raters from the same and different institutions.
| |
Same institution |
Different institutions |
Reference value |
| Baseline |
0.60 |
0.57 |
0.87 |
| Variability |
0.45 |
0.31 |
0.79 |
| Deceleration |
0.48 |
0.56 |
0.94 |
| Level |
0.39 |
0.38 |
0.76 |
Table 5.
Kappa score categories.
Table 5.
Kappa score categories.
Table 6.
Kappa score between raters, as well as between Trium and raters.
Table 6.
Kappa score between raters, as well as between Trium and raters.
| |
Assessments of 11 raters |
Trium |
Reference value |
| Baseline (n=98) |
0.34 |
0.49 |
0.94 |
| Variability (n=96) |
0.43 |
0.45 |
0.71 |
| Deceleration (n=61) |
0.55 |
0.36 |
0.54 |
| Level (n=66) |
0.45 |
0.38 |
0.46 |
Table 7.
Kappa scores by category.
Table 7.
Kappa scores by category.
| |
Category |
11 raters |
Trium |
| Variability |
Moderate |
0.32 |
0.59 |
| Minimal |
0.3 |
0.68 |
| Marked |
0.29 |
0 |
| Sinusoidal |
0 |
0 |
| Absent |
0 |
0 |
| Baseline |
Normocardia |
0.43 |
0.80 |
| Mild bradycardia |
0 |
0 |
| Severe bradycardia |
0 |
0 |
| Tachycardia |
0.44 |
0.80 |
| Deceleration |
None |
0.77 |
0.74 |
| Early |
0.54 |
0 |
| Mild variable |
0.54 |
0.46 |
| Mild late |
0.33 |
0 |
| Mild prolonged |
0.25 |
0 |
| Severe variable |
0.56 |
0.40 |
| Severe late |
0.33 |
0 |
| Severe prolonged |
0.53 |
0.37 |
| Level |
Level 1 |
0.67 |
0.67 |
| Level 2 |
0.40 |
0.31 |
| Level 3 |
0.38 |
0.28 |
| Level 4 |
0.24 |
0.65 |
| Level 5 |
0.13 |
0 |
Table 8.
Agreement for the same obstetrician (n=101).
Table 8.
Agreement for the same obstetrician (n=101).
| |
Baseline |
Variability |
Deceleration |
Level |
| Kappa coefficient |
0.93 |
0.71 |
0.54 |
0.45 |
| Agreement rate (%) |
99% |
93% |
60% |
60% |
|
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |