Background: Few studies have reported electrocardiogram data collected during the neonatal period. The aim of this study was to evaluate electrocardiogram variations in neonates during the early postnatal period. Methods: Electrocardiogram samples taken during the first hour of life from newborns born at our hospital were analyzed in this prospective observational study. Demographic data and possible electrocardiogram changes were studied. The results were statistically analyzed. Results: A total of 260 patients were included during the study period. Among these, 50% were male (n=130), the mean gestational age was 38.1±1.4 weeks, and the mean birth weight was 3.2±1.4 kg. In the electrocardiograms obtained, low atrial rhythm was detected in 0.3% of the patients (n=1). Right axis deviation was observed in 1.5% of the patients (n=4), and left axis deviation was observed in 1.2% of the patients (n=3). An abnormal P-axis was found in one patient (0.3%), and an abnormal QRS-T angle was found in one patient (0.3%). According to the normogram of Davignon and colleagues, T-wave changes were significantly greater in lead V1 (p=0.02). No statistically significant differences were observed in the other parameters. Conclusion: Different electrocardiogram changes can be observed in the early neonatal period. Further studies are needed to clarify the interpretation of the electrocardiogram findings.

Keywords:
newborn
; electrocardiogram
; ECG variations
; early postnatal period
1. Introduction
Immediately after birth, with the transition from the fetal period to the neonatal period, certain anatomical and hemodynamic changes occur in the cardiovascular system 1,2]. In the fetus, the placenta acts as a low-resistance vascular bed, and the right ventricle (RV) is the dominant ventricle, generating 60% of the cardiac output. After the umbilical cord is cut and breathing begins, there is a decrease in pulmonary vascular resistance and an increase in systemic vascular resistance as the placenta is removed from the circulation. This results in a shift of the dominant ventricle from the RV to the left ventricle (LV).
Hemodynamic and anatomical changes can be assessed in practice via clinical parameters (heart rate, oxygen saturation, respiratory pattern, and cardiac auscultation) and complementary tests (e.g., echocardiogram, serum lactate, and serum sodium bicarbonate). However, the exact effects of these circulatory changes on electrocardiography (ECG) during the first few days are not well understood. In addition, there are fewer electrocardiographic studies in the neonatal period than in other stages of life [3,4,5,6].
The aim of this study was to evaluate ECG changes in neonates during the first hour after birth.
2. Materials and Methods
This prospective observational study was conducted on neonates born at our hospital between 1 November 2023 and 1 February 2024. Newborns who received resuscitation after birth, those with congenital anomalies, advanced preterm infants, those admitted to the neonatal intensive care unit, and those whose families did not give consent were excluded from the study.
The study was designed in accordance with the Declaration of Helsinki after approval by the local ethics committee. A study form was completed for each case, including maternal demographics, gestational age at birth, mode of delivery, sex, birth weight, physical examination findings, and ECG data.
ECGs were recorded via the Philips Page Writer Trim II device at a paper speed of 25 mm/second, an amplitude of 10 mm/mV and a bandwidth filter of F 60 ~ 0.15--150 Hz with 12 leads. The electrodes were placed with solid adhesive gel on the right and left shoulders, the iliac region and the areas recommended for the V1–V6 leads. The shoulders and iliac regions were chosen instead of the arms and legs to reduce artifacts caused by the natural movements of the newborn and to improve the quality of the ECG signal. The following parameters were assessed:
- Heart rate (bpm, measured automatically by the device)
- Frontal plane QRS axis (°)
- P-wave amplitude (mm), duration (ms) and PR interval in lead II (ms)
- Q-wave amplitude in leads III, aVF and V5-V6
- S-wave amplitude in leads aVR, V1--V2 and V4--V6
- R/S ratio in leads V1 and V6
- QRS duration, QT and QTc intervals (using Bazett's formula) in lead II (ms)
- T-wave duration (ms) and orientation (+/-) in leads V1 and V6
The P axis, T axis and QRS axis were calculated as the mean vector angle on the Einthoven plane, taking into account the amplitudes of these waves in the DI and aVF leads. A P-axis between 0° and 90° was considered normal sinus rhythm, whereas values outside this range were considered abnormal sinus rhythm. For the QRS axis, a range of 70° (10–125°) was considered normal, and values outside this minimum and maximum range were considered abnormal. A QRS axis between 125° and 180° was classified as right axis deviation (RAD), that between 0° and -90° was classified as left axis deviation (LAD), and that between -90° and -180° was classified as superior axis deviation. The QRS-T angle was defined as the absolute difference between the QRS axis and the T axis, with a difference >90° considered pathological [6]. The results obtained were compared with the ECG scales of Davignon and colleagues [4].
ECG was performed in the delivery room immediately after birth after stabilization according to routine care suggested by the neonatal resuscitation program; however, the baby was still placed under the open bed.
Echocardiography was performed via a Philips Affiniti 50 ultrasound machine (Philips Affiniti 50 Cardiac Ultrasound, Bothell, WA, USA) equipped with a transducer operating at a carrier frequency of 4.2 or 8.3 MHz. Standard pediatric imaging windows were used, including parasternal (long- and short-axis), apical (four- and five-chamber), subcostal and suprasternal views. Morphological assessment was performed via the segmental approach, which assesses the direction of blood flow. The main components of this approach include the atrial situs, venoatrial connections (systemic and pulmonary venous return), atrioventricular (AV) connections, ventricles, ventriculo-arterial (VA) connections, spatial positions of the great arteries relative to each other, intracardiac defects, and extracardiac vascular anomalies. Congenital heart disease was classified according to definitions found in the literature [5]. Echocardiography was performed by the same pediatric cardiologist according to the outpatient clinic patient number. Echocardiographic examination was performed either on the same day or before discharge (on the 1st and 3rd days).
Statistical Methods:
The distribution of variables in the study was categorized via computer software, and descriptive results were obtained via SPSS version 23 (Statistical Package for the Social Sciences for Windows). Descriptive results are expressed as the mean ± standard deviation (Std), median (interquartile range), and percentage percentile. Chi-square tests and Fisher's exact tests were used to compare categorical variables with normal normograms. p<0.05 was considered statistically significant.
3. Results
There were 490 births during the study period, and 260 cases were included in the study. Among the patients, 50% were male (n=130), with a mean gestational age of 38.1 ± 1.4 weeks and a mean birth weight of 3.2 ± 1.4 kg. The demographic data of the patients are shown in Table 1.
A grade 1/6 systolic murmur was detected at the lower left sternal border in two neonates during cardiac examination.
The ECG findings and variations of the cases are shown in Table 2. In the ECG recordings, low atrial rhythm was detected in 0.3% of the patients (n=1). Right axis deviation was observed in 1.5% of the patients (n=4), and left axis deviation was observed in 1.2% of the patients (n=3). A total of 0.3% of the patients (n=1) had an abnormal P-axis, and 0.3% of the patients (n=1) had an abnormal QRS-T-angle.
According to the normogram of Davignon and colleagues, T wave changes were significantly more common in lead V1 (p=0.02). No statistically significant differences were found between the other parameters.
Echocardiography was performed during the first three days of life before discharge. Sixteen patients (6.1%) had patent foramen ovale in the interatrial septum, four patients (1.5%) had patent ductus atreiosus (hemodynamically insignificant), two patients (0.6%) had atrial septal defects (moderate, with 6 mm color Doppler flow), and one patient (0.3%) had ventricular septal defects (small midline, 2 mm width).
4. Discussion
In this prospective observational study, we investigated ECG variations recorded during the first postnatal hour in healthy newborns. We found that during this transitional period, ECG variations may differ from those in other age groups, with a significantly greater incidence of T-wave variability in lead V1. This study is valuable because it provides cross-sectional data from the very early postnatal period.
ECG is a useful medical test for assessing the electrical activity of the heart and, to some extent, the structures of the heart. The primary parameters assessed via ECG include rhythm, rate, P axis, PR interval, QRS duration and axis, T wave characteristics, the QTc interval and ST-T changes, which are important in the diagnosis of various structural or congenital heart diseases [4,6,7]. Both term and preterm infants undergo cardiovascular changes at birth, coinciding with the transition from placental circulation to the lung as the respiratory organ. Changes in fetal cardiac physiology during late pregnancy and perinatal transition result in variability in neonatal ECG parameters [8]. In addition, studies focusing on neonatal ECG data are limited.
The most comprehensive pediatric study of ECG variability was conducted by Davignon et al. [4] and was based on measurements from 2,141 children. Normal ranges were established for several parameters, and the percentile values of the ECG parameters in our study are quite comparable to those in Davignon's study.
In our study, heart rate, mean QRS axis, QRS duration, and R-wave and S-wave amplitudes in leads V1 and V6 measured in the first hour after birth were similar. The results were also consistent with those of recent studies [8,9].
A recent study revealed significant differences in the orientation of T waves in leads V1, V2, V3 and V4 across different age groups. The large proportion of positive factors in the neonatal group can be explained by the rapid pressure applied to the lungs, which directly affects the early phase of ventricular repolarization. As pulmonary pressure decreases in the first years of life, it has been suggested that infants may rely more on T-wave derivations (V1 and V4) to maintain repolarization [10]. Although the precise mechanism underlying this phenomenon remains unclear, it may be influenced by negative T V1 and various factors, including congenital heart disease, pulmonary arterial hypertension, myocardial ischemia, oxygen deficiency, and adrenaline levels. [11]
In their study, Pimenta et al. [3] reported a significant difference in T wave positivity in leads V1--4 between neonates studied within the first 24 hours and those studied between days 1--3 and 3--7. Similarly, in our study, we found a significantly greater incidence of T wave positivity in lead V1, which is consistent with Pimenta's findings. Additionally, Meng et al. [11] reported that T waves change 15 hours after birth and that there is a relationship between PDA and pulmonary hypertension. In their cohort, the biphasic and positive T wave incidence was 85%, and the negative T wave incidence was 15%. Although they reported that a negative T wave after 15 hr of birth was associated with a higher incidence of PDA, our findings did not support this finding. These findings may be related to the reduction in pulmonary vascular resistance during the early postnatal period.
Limitations: The main limitation of this study is that it was conducted in a single center with a limited number of patients. Owing to the challenges of obtaining ECG recordings in neonates, all recordings were performed by the same team. The inability to perform echocardiography immediately in the early period may have affected the results of the measurements, particularly the presence of patent ductus arteriosus and patent foramen ovale in the interatrial septum. Our number of Cesarean section patients is relatively high because our hospital is a referral center for gynecology and obstetrics, and the drugs that are used for anesthesia might affect ECGs. In addition, as the results were compared with those of Davignon and colleagues, racial and ethnic factors may have influenced the results.
5. Conclusion
In conclusion, different ECG changes can be observed in neonates during the early period. In particular, T wave positivity was significantly greater in the first hour after birth. Therefore, studies with larger numbers of cases are needed for a clearer interpretation of ECG findings in this age group.
Author Contributions
Duygu Acar, MD: conceptualization, data curation, investigation, methodology, validation, writing - original draft, and writing - review and editing; Erkut Ozturk, MD: conceptualization, formal analysis, investigation, methodology, validation, writing - original draft, and writing - review and editing.
Funding Sources
This study was not supported by any sponsor or funder.
Institutional Review Board Statement
The study was designed in accordance with the Declaration of Helsinki after approval by the local ethics committee. Informed consent was obtained from eligible participants in accordance with the Declaration of Helsinki. This study protocol was reviewed and approved by the ethics committee of Gaziosmanpasa Research and Training Hospital, 10.25.2023, approval number ‘142’.
Informed Consent Statement
The patient’s family provided informed consent for evaluation of their child and presentation of his or her clinical course.
Data availability statement
All the data generated or analyzed during this study are included in this article. Further inquiries can be directed to the corresponding author.
Acknowledgments
Thanks to our delivery room nurses for their helpful.
Competing Interest declaration
there are no competing interests.
Abbreviation
RV: right ventricle.
LV: left ventricle.
ECG: electrocardiography.
RAD: right axis deviation.
LAD: left axis deviation.
AV: atrioventricular.
VA: ventriculo-arterial.
PDA: patent ductus arteriosus.
References
- Bensley, J.G.; De Matteo, R.; Harding, R.; Black, M.J. The effects of preterm birth and its antecedents on the cardiovascular system. Acta Obstet. et Gynecol. Scand. 2016, 95, 652–663. [Google Scholar] [CrossRef] [PubMed]
- Rijnbeek, P.; Witsenburg, M.; Schrama, E.; Hess, J.; Kors, J. New normal limits for the paediatric electrocardiogram. Eur. Hear. J. 2001, 22, 702–711. [Google Scholar] [CrossRef] [PubMed]
- Pimenta, M.S.; Samesima, N.; Pastore, C.A.; Krebs, V.L.J.; Leal, G.N.; Carvalho, W.B. Electrocardiographic Evaluation of Normal Newborns in the First Week of Life - Observational Study. Arq Bras Cardiol. 2022, 119, 587–592. [Google Scholar] [PubMed]
- Davignon, A.; Rautaharju, P.; Boisselle, E.; Soumis, F.; Mégélas, M.; Choquette, A. Normal ECG standards for infants and children. Pediatr. Cardiol. 1980, 1, 123–131. [Google Scholar] [CrossRef]
- Frommelt, P. Update on pediatric echocardiography. Curr Opin Pediatr 2005, 17, 579–85. [Google Scholar] [CrossRef] [PubMed]
- Khan, G.M. A new electrode placement method for obtaining 12-lead ECGs. Open Hear. 2015, 2, e000226. [Google Scholar] [CrossRef] [PubMed]
- Lue, H.-C.; Wu, M.-H.; Wang, J.-K.; Lin, M.-T.; Lu, C.-W.; Chiu, S.-N.; Chen, C.-A.; Wu, E.-T.; Wang, C.-C.; Fu, C.-M.; et al. Normal ECG standards and percentile charts for infants, children and adolescents. Pediatr. Neonatol. 2023, 64, 256–273. [Google Scholar] [CrossRef] [PubMed]
- Habibulla, S.K.M.; Bhowmik, A.; Saha, J.; Hazra, A.; Halder, S.K.; Mondal, R. Electrocardiographic Parameters in Indian Newborns. Indian Pediatr. 2019, 56, 917–922. [Google Scholar] [CrossRef]
- Pærregaard, M.M.; Kock, J.; Pihl, C.; Pietersen, A.; Iversen, K.K.; Bundgaard, H.; Christensen, A.H. The Evolution of the Neonatal QRS Axis during the First Four Weeks of Life. Neonatology 2021, 118, 155–162. [Google Scholar] [CrossRef] [PubMed]
- Pastore, C.A.; Pinho, J.A.; Pinho, C.; et al. III Diretrizes da Sociedade Brasileira de Cardiologia Sobre Análise e Emissão de Laudos Eletrocardiográficos. Arq Bras Cardiol. 2016, 106, 1–23. [Google Scholar] [CrossRef] [PubMed]
- Meng, H.; Wang, J.; Jiang, X.; Tian, L.; Liu, T.; Ye, Y. Association between negative T waves in lead V1 and delayed patent ductus arteriosus closure in preterm neonates. J. Cardiothorac. Surg. 2025, 20, 1–9. [Google Scholar] [CrossRef] [PubMed]
Table 1.
General Characteristics of the Cases.
| Variable | n =260 |
|---|---|
| Mother's Age, years | 27±3 |
| Gestational Age, weeks | 38.1±1.4 |
| Delivery Type, NSD (Normal Spontaneous Delivery) CS (Caesarean Section) |
100 (38) 160 (62) |
| Birth Weight, kg | 3.2±1.4 |
| Classification SGA (Small for Gestational Age) AGA (Appropriate for Gestational Age) LGA (Large for Gestational Age) |
16 (6) 218 (84) 26 (10) |
| Gender, Male | 130 (50) |
| Etnicity, (Turkish,%) | 214 (82) |
| Maternal Drug Use Levothrone Alpha methyl dopa Low dose asprin Novorapid Methimazole |
2(0.7) 1(0.3) 2(0.7) 1(0.3) 1(0.3) |
N(%) or mean±standart derivation.
Table 2.
Electrocardiography Variability in the Cases.
| Variable | |
|---|---|
| Heart rate, bpm | 120 (110-130) |
| Amplitude P Lead II (mm) | 1 (0.7-1.4) |
| PR LeadII (msec) | 90 (80-100) |
| QT Lead II (msec) | 310 (290-340) |
| QTc (msec) (Bazzett) | 400 (360-440) |
| QRS axis (°) | 130 (100-160) |
| Amplitude Q DIII (mm) | 4 (2-6) |
| Amplitude Q V6 (mm) | 1.0 (0.6-1.5) |
| Amplitude R V1 (mm) | 12.5 (11-15) |
| Amplitude R V2 (mm) | 14 (10-18) |
| Amplitude R V5 (mm) | 14 (10-16) |
| Amplitude R V6 (mm) | 13 (9-17) |
| Amplitude S V1 (mm) | 10 (6-14) |
| Amplitude S V2 (mm) | 15.5 (12-19) |
| Amplitude S V5 (mm) | 9 (5-15) |
| Amplitude S V6 (mm) | 6 (4-8) |
| R/S V1 | 1.4 (1-1.8) |
| R/S V6 | 5.2 (3-8) |
| Duration QRS Lead II (msec) | 45 (40-50) |
| V1 T Positive Negative Biphasic |
130 (50) 65 (25) 65 (25) |
| V6 T Positive Negative Biphasic |
234 (90) 26 (10) - |
| QRS axis Normal Right Axis Left Axis |
253 (97.3) 4 (1.5) 3 (1.2) |
| Abnormal QRS-T angle | 1 (0.3) |
| Rhythm Normal Abnormal |
259 (99.7) 1 (0.3) |
n(%).
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |