The VEMoS AXL system is a new optical biometer based on spectral domain optical coherence tomography (SD-OCT) that has been tested in terms of intrasession repeatability and compared with a swept-source optical coherence tomography biometer (SS-OCT), which is considered the gold standard, for an agreement analysis. A biometric analysis was performed three consecutive times in 120 healthy eyes of 120 patients aged between 18 and 40 years with the SD-OCT system and afterwards a single measurement was obtained with the SS-OCT system. Within-subject standard deviations were 0.004 mm, 4.394 µm and 0.017 mm for axial length (AL), central corneal thickness (CCT), and anterior chamber depth (ACD) measures obtained with the SD-OCT biometer, re-spectively. The agreement between devices was good for AL (Limits of agreement, LoA: -0.04 to 0.03 mm) and CCT (LoA: -4.36 to 14.38 µm), whereas differences between devices were clinically relevant for ACD (LoA: 0.03 to 0.21 mm). In conclusion, the VEMoS AXL system provides consistent measures of anatomical parameters, being most of them interchangeable with those provided by the SS-OCT-based gold standard.